A multicenter retrieval study of the taper interfaces of modular hip prostheses

A multicenter retrieval analysis of 231 modular hip implants was done to investigate the effects of material combination, metallurgic condition, flexural rigidity, head and neck moment arm, neck length, and implantation time on corrosion and fretting of modular taper surfaces. Scores for corrosion a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Clinical orthopaedics and related research 2002-08, Vol.401 (401), p.149-161
Hauptverfasser: GOLDBERG, Jay R, GILBERT, Jeremy L, JACOBS, Joshua J, BAUER, Thomas W, PAPROSKY, Wayne, LEURGANS, Sue
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 161
container_issue 401
container_start_page 149
container_title Clinical orthopaedics and related research
container_volume 401
creator GOLDBERG, Jay R
GILBERT, Jeremy L
JACOBS, Joshua J
BAUER, Thomas W
PAPROSKY, Wayne
LEURGANS, Sue
description A multicenter retrieval analysis of 231 modular hip implants was done to investigate the effects of material combination, metallurgic condition, flexural rigidity, head and neck moment arm, neck length, and implantation time on corrosion and fretting of modular taper surfaces. Scores for corrosion and fretting were assigned to medial, lateral, anterior, and posterior quadrants of the necks, and proximal and distal regions of the heads. Neck and head corrosion and fretting scores were found to be significantly higher for mixed alloy versus similar alloy couples. Moderate to severe corrosion was observed in 28% of the heads of similar alloy couples and 42% of the heads of mixed alloy couples. Differences in corrosion scores were observed between components made from the same base alloy, but of different metallurgic conditions. Corrosion and fretting scores tended to be higher for heads than necks. Implantation time and flexural rigidity of the neck were predictors of head and neck corrosion and head fretting. The results of this study suggest that in vivo corrosion of modular hip taper interfaces is attributable to a mechanically-assisted crevice corrosion process. Larger diameter necks will increase neck stiffness and may reduce fretting and subsequent corrosion of the taper interface regardless of the alloy used. Increasing neck diameter must be balanced, however, with the resulting loss of range of motion and joint stability.
doi_str_mv 10.1097/00003086-200208000-00018
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_71992428</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>71992428</sourcerecordid><originalsourceid>FETCH-LOGICAL-c407t-9edfcc064c783a36e276eea11439264e0ee717d032836fa9b66f3c8f418574ec3</originalsourceid><addsrcrecordid>eNpFkF1LwzAUhnOhuDn9C5Ibvavmq01yOYZfMNiNgnchS09YpV1r0gr796auukA4JHnec8KDEKbknhItH0hanKgiY4QwotIpS5uqMzRPVWea0Y8ZuozxcwRFzi7QjDKaU6XZHG2WuBnqvnKw7yHgAH2o4NvWOPZDecCtx_0OcG-79FiNiLcO4njftOVQ24B3VYe70MbERYhX6NzbOsL1VBfo_enxbfWSrTfPr6vlOnOCyD7TUHrnSCGcVNzyApgsACylgmtWCCAAksqScKZ44a3eFoXnTnlBVS4FOL5Ad8e-afTXALE3TRUd1LXdQztEI6nWTKT0Aqkj6NIfYwBvulA1NhwMJWYUaP4Emn-B5ldgit5MM4ZtA-UpONlLwO0E2Ohs7YPduyqeOK6EoiznPzqTeqs</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>71992428</pqid></control><display><type>article</type><title>A multicenter retrieval study of the taper interfaces of modular hip prostheses</title><source>MEDLINE</source><source>Journals@Ovid Complete</source><creator>GOLDBERG, Jay R ; GILBERT, Jeremy L ; JACOBS, Joshua J ; BAUER, Thomas W ; PAPROSKY, Wayne ; LEURGANS, Sue</creator><creatorcontrib>GOLDBERG, Jay R ; GILBERT, Jeremy L ; JACOBS, Joshua J ; BAUER, Thomas W ; PAPROSKY, Wayne ; LEURGANS, Sue</creatorcontrib><description>A multicenter retrieval analysis of 231 modular hip implants was done to investigate the effects of material combination, metallurgic condition, flexural rigidity, head and neck moment arm, neck length, and implantation time on corrosion and fretting of modular taper surfaces. Scores for corrosion and fretting were assigned to medial, lateral, anterior, and posterior quadrants of the necks, and proximal and distal regions of the heads. Neck and head corrosion and fretting scores were found to be significantly higher for mixed alloy versus similar alloy couples. Moderate to severe corrosion was observed in 28% of the heads of similar alloy couples and 42% of the heads of mixed alloy couples. Differences in corrosion scores were observed between components made from the same base alloy, but of different metallurgic conditions. Corrosion and fretting scores tended to be higher for heads than necks. Implantation time and flexural rigidity of the neck were predictors of head and neck corrosion and head fretting. The results of this study suggest that in vivo corrosion of modular hip taper interfaces is attributable to a mechanically-assisted crevice corrosion process. Larger diameter necks will increase neck stiffness and may reduce fretting and subsequent corrosion of the taper interface regardless of the alloy used. Increasing neck diameter must be balanced, however, with the resulting loss of range of motion and joint stability.</description><identifier>ISSN: 0009-921X</identifier><identifier>DOI: 10.1097/00003086-200208000-00018</identifier><identifier>PMID: 12151892</identifier><identifier>CODEN: CORTBR</identifier><language>eng</language><publisher>Heidelberg: Springer</publisher><subject>Adult ; Aged ; Aged, 80 and over ; Alloys ; Arthroplasty, Replacement, Hip ; Biological and medical sciences ; Corrosion ; Device Removal ; Equipment Failure Analysis ; Female ; Hip Prosthesis ; Humans ; Investigative techniques, diagnostic techniques (general aspects) ; Male ; Medical sciences ; Middle Aged ; Osteoarticular system. Muscles ; Pathology. Cytology. Biochemistry. Spectrometry. Miscellaneous investigative techniques ; Prosthesis Design ; Prosthesis Failure</subject><ispartof>Clinical orthopaedics and related research, 2002-08, Vol.401 (401), p.149-161</ispartof><rights>2002 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c407t-9edfcc064c783a36e276eea11439264e0ee717d032836fa9b66f3c8f418574ec3</citedby><cites>FETCH-LOGICAL-c407t-9edfcc064c783a36e276eea11439264e0ee717d032836fa9b66f3c8f418574ec3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&amp;idt=13848125$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/12151892$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>GOLDBERG, Jay R</creatorcontrib><creatorcontrib>GILBERT, Jeremy L</creatorcontrib><creatorcontrib>JACOBS, Joshua J</creatorcontrib><creatorcontrib>BAUER, Thomas W</creatorcontrib><creatorcontrib>PAPROSKY, Wayne</creatorcontrib><creatorcontrib>LEURGANS, Sue</creatorcontrib><title>A multicenter retrieval study of the taper interfaces of modular hip prostheses</title><title>Clinical orthopaedics and related research</title><addtitle>Clin Orthop Relat Res</addtitle><description>A multicenter retrieval analysis of 231 modular hip implants was done to investigate the effects of material combination, metallurgic condition, flexural rigidity, head and neck moment arm, neck length, and implantation time on corrosion and fretting of modular taper surfaces. Scores for corrosion and fretting were assigned to medial, lateral, anterior, and posterior quadrants of the necks, and proximal and distal regions of the heads. Neck and head corrosion and fretting scores were found to be significantly higher for mixed alloy versus similar alloy couples. Moderate to severe corrosion was observed in 28% of the heads of similar alloy couples and 42% of the heads of mixed alloy couples. Differences in corrosion scores were observed between components made from the same base alloy, but of different metallurgic conditions. Corrosion and fretting scores tended to be higher for heads than necks. Implantation time and flexural rigidity of the neck were predictors of head and neck corrosion and head fretting. The results of this study suggest that in vivo corrosion of modular hip taper interfaces is attributable to a mechanically-assisted crevice corrosion process. Larger diameter necks will increase neck stiffness and may reduce fretting and subsequent corrosion of the taper interface regardless of the alloy used. Increasing neck diameter must be balanced, however, with the resulting loss of range of motion and joint stability.</description><subject>Adult</subject><subject>Aged</subject><subject>Aged, 80 and over</subject><subject>Alloys</subject><subject>Arthroplasty, Replacement, Hip</subject><subject>Biological and medical sciences</subject><subject>Corrosion</subject><subject>Device Removal</subject><subject>Equipment Failure Analysis</subject><subject>Female</subject><subject>Hip Prosthesis</subject><subject>Humans</subject><subject>Investigative techniques, diagnostic techniques (general aspects)</subject><subject>Male</subject><subject>Medical sciences</subject><subject>Middle Aged</subject><subject>Osteoarticular system. Muscles</subject><subject>Pathology. Cytology. Biochemistry. Spectrometry. Miscellaneous investigative techniques</subject><subject>Prosthesis Design</subject><subject>Prosthesis Failure</subject><issn>0009-921X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2002</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpFkF1LwzAUhnOhuDn9C5Ibvavmq01yOYZfMNiNgnchS09YpV1r0gr796auukA4JHnec8KDEKbknhItH0hanKgiY4QwotIpS5uqMzRPVWea0Y8ZuozxcwRFzi7QjDKaU6XZHG2WuBnqvnKw7yHgAH2o4NvWOPZDecCtx_0OcG-79FiNiLcO4njftOVQ24B3VYe70MbERYhX6NzbOsL1VBfo_enxbfWSrTfPr6vlOnOCyD7TUHrnSCGcVNzyApgsACylgmtWCCAAksqScKZ44a3eFoXnTnlBVS4FOL5Ad8e-afTXALE3TRUd1LXdQztEI6nWTKT0Aqkj6NIfYwBvulA1NhwMJWYUaP4Emn-B5ldgit5MM4ZtA-UpONlLwO0E2Ohs7YPduyqeOK6EoiznPzqTeqs</recordid><startdate>20020801</startdate><enddate>20020801</enddate><creator>GOLDBERG, Jay R</creator><creator>GILBERT, Jeremy L</creator><creator>JACOBS, Joshua J</creator><creator>BAUER, Thomas W</creator><creator>PAPROSKY, Wayne</creator><creator>LEURGANS, Sue</creator><general>Springer</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20020801</creationdate><title>A multicenter retrieval study of the taper interfaces of modular hip prostheses</title><author>GOLDBERG, Jay R ; GILBERT, Jeremy L ; JACOBS, Joshua J ; BAUER, Thomas W ; PAPROSKY, Wayne ; LEURGANS, Sue</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c407t-9edfcc064c783a36e276eea11439264e0ee717d032836fa9b66f3c8f418574ec3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2002</creationdate><topic>Adult</topic><topic>Aged</topic><topic>Aged, 80 and over</topic><topic>Alloys</topic><topic>Arthroplasty, Replacement, Hip</topic><topic>Biological and medical sciences</topic><topic>Corrosion</topic><topic>Device Removal</topic><topic>Equipment Failure Analysis</topic><topic>Female</topic><topic>Hip Prosthesis</topic><topic>Humans</topic><topic>Investigative techniques, diagnostic techniques (general aspects)</topic><topic>Male</topic><topic>Medical sciences</topic><topic>Middle Aged</topic><topic>Osteoarticular system. Muscles</topic><topic>Pathology. Cytology. Biochemistry. Spectrometry. Miscellaneous investigative techniques</topic><topic>Prosthesis Design</topic><topic>Prosthesis Failure</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>GOLDBERG, Jay R</creatorcontrib><creatorcontrib>GILBERT, Jeremy L</creatorcontrib><creatorcontrib>JACOBS, Joshua J</creatorcontrib><creatorcontrib>BAUER, Thomas W</creatorcontrib><creatorcontrib>PAPROSKY, Wayne</creatorcontrib><creatorcontrib>LEURGANS, Sue</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Clinical orthopaedics and related research</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>GOLDBERG, Jay R</au><au>GILBERT, Jeremy L</au><au>JACOBS, Joshua J</au><au>BAUER, Thomas W</au><au>PAPROSKY, Wayne</au><au>LEURGANS, Sue</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>A multicenter retrieval study of the taper interfaces of modular hip prostheses</atitle><jtitle>Clinical orthopaedics and related research</jtitle><addtitle>Clin Orthop Relat Res</addtitle><date>2002-08-01</date><risdate>2002</risdate><volume>401</volume><issue>401</issue><spage>149</spage><epage>161</epage><pages>149-161</pages><issn>0009-921X</issn><coden>CORTBR</coden><abstract>A multicenter retrieval analysis of 231 modular hip implants was done to investigate the effects of material combination, metallurgic condition, flexural rigidity, head and neck moment arm, neck length, and implantation time on corrosion and fretting of modular taper surfaces. Scores for corrosion and fretting were assigned to medial, lateral, anterior, and posterior quadrants of the necks, and proximal and distal regions of the heads. Neck and head corrosion and fretting scores were found to be significantly higher for mixed alloy versus similar alloy couples. Moderate to severe corrosion was observed in 28% of the heads of similar alloy couples and 42% of the heads of mixed alloy couples. Differences in corrosion scores were observed between components made from the same base alloy, but of different metallurgic conditions. Corrosion and fretting scores tended to be higher for heads than necks. Implantation time and flexural rigidity of the neck were predictors of head and neck corrosion and head fretting. The results of this study suggest that in vivo corrosion of modular hip taper interfaces is attributable to a mechanically-assisted crevice corrosion process. Larger diameter necks will increase neck stiffness and may reduce fretting and subsequent corrosion of the taper interface regardless of the alloy used. Increasing neck diameter must be balanced, however, with the resulting loss of range of motion and joint stability.</abstract><cop>Heidelberg</cop><pub>Springer</pub><pmid>12151892</pmid><doi>10.1097/00003086-200208000-00018</doi><tpages>13</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0009-921X
ispartof Clinical orthopaedics and related research, 2002-08, Vol.401 (401), p.149-161
issn 0009-921X
language eng
recordid cdi_proquest_miscellaneous_71992428
source MEDLINE; Journals@Ovid Complete
subjects Adult
Aged
Aged, 80 and over
Alloys
Arthroplasty, Replacement, Hip
Biological and medical sciences
Corrosion
Device Removal
Equipment Failure Analysis
Female
Hip Prosthesis
Humans
Investigative techniques, diagnostic techniques (general aspects)
Male
Medical sciences
Middle Aged
Osteoarticular system. Muscles
Pathology. Cytology. Biochemistry. Spectrometry. Miscellaneous investigative techniques
Prosthesis Design
Prosthesis Failure
title A multicenter retrieval study of the taper interfaces of modular hip prostheses
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-26T19%3A41%3A29IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=A%20multicenter%20retrieval%20study%20of%20the%20taper%20interfaces%20of%20modular%20hip%20prostheses&rft.jtitle=Clinical%20orthopaedics%20and%20related%20research&rft.au=GOLDBERG,%20Jay%20R&rft.date=2002-08-01&rft.volume=401&rft.issue=401&rft.spage=149&rft.epage=161&rft.pages=149-161&rft.issn=0009-921X&rft.coden=CORTBR&rft_id=info:doi/10.1097/00003086-200208000-00018&rft_dat=%3Cproquest_cross%3E71992428%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=71992428&rft_id=info:pmid/12151892&rfr_iscdi=true