Automated flow cytometric analysis of blood cells in cerebrospinal fluid: Analytic performance

We compared the performance of an automated method for obtaining RBC and WBC counts and WBC differential counts in cerebrospinal fluid (CSF) samples with the reference manual method. Results from 325 samples from 10 worldwide clinical sites were used to demonstrate the accuracy, precision, and linea...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of clinical pathology 2004-05, Vol.121 (5), p.690-700
Hauptverfasser: AUNE, Marthe W, BECKER, Joanne L, KUNICKA, Jolanta E, LACKNER, Karl J, MINCHELLO, Elaine, MULLENIX, Patricia A, MYERS, Michael, PETERSEN, Annette, TERNSTROM, Wanda, WILSON, Sandra J, BRUGNARA, Carlo, CANFIELD, William, DORFMAN, David M, FIEHN, W, FISCHER, Gena, FITZPATRICK, Patricia, FLAMING, Timothy H, HENRIKSEN, Hilde-Kristin
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 700
container_issue 5
container_start_page 690
container_title American journal of clinical pathology
container_volume 121
creator AUNE, Marthe W
BECKER, Joanne L
KUNICKA, Jolanta E
LACKNER, Karl J
MINCHELLO, Elaine
MULLENIX, Patricia A
MYERS, Michael
PETERSEN, Annette
TERNSTROM, Wanda
WILSON, Sandra J
BRUGNARA, Carlo
CANFIELD, William
DORFMAN, David M
FIEHN, W
FISCHER, Gena
FITZPATRICK, Patricia
FLAMING, Timothy H
HENRIKSEN, Hilde-Kristin
description We compared the performance of an automated method for obtaining RBC and WBC counts and WBC differential counts in cerebrospinal fluid (CSF) samples with the reference manual method. Results from 325 samples from 10 worldwide clinical sites were used to demonstrate the accuracy, precision, and linearity of the method. Accuracy statistics for absolute cell counts showed a high correlation between methods, with correlation coefficients for all reportable absolute counts greater than 0.9. Linearity results demonstrated that the method provides accurate results throughout the reportable ranges, including clinical decision points for WBCs of 0 to 10/microL. Interassay precision and intra-assay precision for the ADVIA 120 (Bayer HealthCare, Tarrytown, NY) method were acceptable at all levels. The ADVIA 120 CSF Assay enumerates and differentiates cells via flow cytometry in a minimally diluted sample, improving the analysis of typically hypocellular CSF samples. Study results demonstrate that the automated ADVIA 120 CSF Assay is an acceptable alternative to the labor-intensive manual method.
doi_str_mv 10.1309/EKFW9E3LLFXE15X9
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_71945570</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>71945570</sourcerecordid><originalsourceid>FETCH-LOGICAL-c195t-89b88ab5b5bce1afc31e18f81924949a23c8cad8eff2c202949a95822abca2db3</originalsourceid><addsrcrecordid>eNpdkE1LxDAQhoMo7rp69yS56K2aj2abeFuWrooFL4p7sqRpApG2WZMW2X9vyhYUmcN88LzDzAvAJUa3mCJxlz9v3kVOi2KzzTHbiiMwxyKlSZYRcgzmCCGSCJzRGTgL4RMhTDhKT8EMsxgEiTn4WA29a2Wva2ga9w3VPra691ZB2clmH2yAzsCqca6GSjdNgLaLhdeVd2FnIxOFg63v4Wrk-yjcaW-cb2Wn9Dk4MbIJ-mLKC_C2yV_Xj0nx8vC0XhWJwoL1CRcV57JiMZTG0iiKNeaGY0FSkQpJqOJK1lwbQxRBZJwJxgmRlZKkrugC3Bz27rz7GnToy9aG8VzZaTeEMou2MJahCKIDqOL5wWtT7rxtpd-XGJWjp-V_T6Pkato9VK2ufwWTiRG4ngAZlGyMj5_b8IdbpilDS_oDo06B0w</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>71945570</pqid></control><display><type>article</type><title>Automated flow cytometric analysis of blood cells in cerebrospinal fluid: Analytic performance</title><source>MEDLINE</source><source>EZB Electronic Journals Library</source><creator>AUNE, Marthe W ; BECKER, Joanne L ; KUNICKA, Jolanta E ; LACKNER, Karl J ; MINCHELLO, Elaine ; MULLENIX, Patricia A ; MYERS, Michael ; PETERSEN, Annette ; TERNSTROM, Wanda ; WILSON, Sandra J ; BRUGNARA, Carlo ; CANFIELD, William ; DORFMAN, David M ; FIEHN, W ; FISCHER, Gena ; FITZPATRICK, Patricia ; FLAMING, Timothy H ; HENRIKSEN, Hilde-Kristin</creator><creatorcontrib>AUNE, Marthe W ; BECKER, Joanne L ; KUNICKA, Jolanta E ; LACKNER, Karl J ; MINCHELLO, Elaine ; MULLENIX, Patricia A ; MYERS, Michael ; PETERSEN, Annette ; TERNSTROM, Wanda ; WILSON, Sandra J ; BRUGNARA, Carlo ; CANFIELD, William ; DORFMAN, David M ; FIEHN, W ; FISCHER, Gena ; FITZPATRICK, Patricia ; FLAMING, Timothy H ; HENRIKSEN, Hilde-Kristin</creatorcontrib><description>We compared the performance of an automated method for obtaining RBC and WBC counts and WBC differential counts in cerebrospinal fluid (CSF) samples with the reference manual method. Results from 325 samples from 10 worldwide clinical sites were used to demonstrate the accuracy, precision, and linearity of the method. Accuracy statistics for absolute cell counts showed a high correlation between methods, with correlation coefficients for all reportable absolute counts greater than 0.9. Linearity results demonstrated that the method provides accurate results throughout the reportable ranges, including clinical decision points for WBCs of 0 to 10/microL. Interassay precision and intra-assay precision for the ADVIA 120 (Bayer HealthCare, Tarrytown, NY) method were acceptable at all levels. The ADVIA 120 CSF Assay enumerates and differentiates cells via flow cytometry in a minimally diluted sample, improving the analysis of typically hypocellular CSF samples. Study results demonstrate that the automated ADVIA 120 CSF Assay is an acceptable alternative to the labor-intensive manual method.</description><identifier>ISSN: 0002-9173</identifier><identifier>EISSN: 1943-7722</identifier><identifier>DOI: 10.1309/EKFW9E3LLFXE15X9</identifier><identifier>PMID: 15151209</identifier><identifier>CODEN: AJCPAI</identifier><language>eng</language><publisher>Chicago, IL: American Society of Clinical Pathologists</publisher><subject>Autoanalysis - instrumentation ; Biological and medical sciences ; Cerebrospinal Fluid - cytology ; Erythrocyte Count - methods ; Flow Cytometry - methods ; General aspects ; Hematologic Tests - instrumentation ; Hematologic Tests - methods ; Humans ; Leukocyte Count - methods ; Medical sciences ; Reference Standards ; Reproducibility of Results ; Tumors</subject><ispartof>American journal of clinical pathology, 2004-05, Vol.121 (5), p.690-700</ispartof><rights>2004 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>315,782,786,27933,27934</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&amp;idt=15644506$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/15151209$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>AUNE, Marthe W</creatorcontrib><creatorcontrib>BECKER, Joanne L</creatorcontrib><creatorcontrib>KUNICKA, Jolanta E</creatorcontrib><creatorcontrib>LACKNER, Karl J</creatorcontrib><creatorcontrib>MINCHELLO, Elaine</creatorcontrib><creatorcontrib>MULLENIX, Patricia A</creatorcontrib><creatorcontrib>MYERS, Michael</creatorcontrib><creatorcontrib>PETERSEN, Annette</creatorcontrib><creatorcontrib>TERNSTROM, Wanda</creatorcontrib><creatorcontrib>WILSON, Sandra J</creatorcontrib><creatorcontrib>BRUGNARA, Carlo</creatorcontrib><creatorcontrib>CANFIELD, William</creatorcontrib><creatorcontrib>DORFMAN, David M</creatorcontrib><creatorcontrib>FIEHN, W</creatorcontrib><creatorcontrib>FISCHER, Gena</creatorcontrib><creatorcontrib>FITZPATRICK, Patricia</creatorcontrib><creatorcontrib>FLAMING, Timothy H</creatorcontrib><creatorcontrib>HENRIKSEN, Hilde-Kristin</creatorcontrib><title>Automated flow cytometric analysis of blood cells in cerebrospinal fluid: Analytic performance</title><title>American journal of clinical pathology</title><addtitle>Am J Clin Pathol</addtitle><description>We compared the performance of an automated method for obtaining RBC and WBC counts and WBC differential counts in cerebrospinal fluid (CSF) samples with the reference manual method. Results from 325 samples from 10 worldwide clinical sites were used to demonstrate the accuracy, precision, and linearity of the method. Accuracy statistics for absolute cell counts showed a high correlation between methods, with correlation coefficients for all reportable absolute counts greater than 0.9. Linearity results demonstrated that the method provides accurate results throughout the reportable ranges, including clinical decision points for WBCs of 0 to 10/microL. Interassay precision and intra-assay precision for the ADVIA 120 (Bayer HealthCare, Tarrytown, NY) method were acceptable at all levels. The ADVIA 120 CSF Assay enumerates and differentiates cells via flow cytometry in a minimally diluted sample, improving the analysis of typically hypocellular CSF samples. Study results demonstrate that the automated ADVIA 120 CSF Assay is an acceptable alternative to the labor-intensive manual method.</description><subject>Autoanalysis - instrumentation</subject><subject>Biological and medical sciences</subject><subject>Cerebrospinal Fluid - cytology</subject><subject>Erythrocyte Count - methods</subject><subject>Flow Cytometry - methods</subject><subject>General aspects</subject><subject>Hematologic Tests - instrumentation</subject><subject>Hematologic Tests - methods</subject><subject>Humans</subject><subject>Leukocyte Count - methods</subject><subject>Medical sciences</subject><subject>Reference Standards</subject><subject>Reproducibility of Results</subject><subject>Tumors</subject><issn>0002-9173</issn><issn>1943-7722</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2004</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpdkE1LxDAQhoMo7rp69yS56K2aj2abeFuWrooFL4p7sqRpApG2WZMW2X9vyhYUmcN88LzDzAvAJUa3mCJxlz9v3kVOi2KzzTHbiiMwxyKlSZYRcgzmCCGSCJzRGTgL4RMhTDhKT8EMsxgEiTn4WA29a2Wva2ga9w3VPra691ZB2clmH2yAzsCqca6GSjdNgLaLhdeVd2FnIxOFg63v4Wrk-yjcaW-cb2Wn9Dk4MbIJ-mLKC_C2yV_Xj0nx8vC0XhWJwoL1CRcV57JiMZTG0iiKNeaGY0FSkQpJqOJK1lwbQxRBZJwJxgmRlZKkrugC3Bz27rz7GnToy9aG8VzZaTeEMou2MJahCKIDqOL5wWtT7rxtpd-XGJWjp-V_T6Pkato9VK2ufwWTiRG4ngAZlGyMj5_b8IdbpilDS_oDo06B0w</recordid><startdate>200405</startdate><enddate>200405</enddate><creator>AUNE, Marthe W</creator><creator>BECKER, Joanne L</creator><creator>KUNICKA, Jolanta E</creator><creator>LACKNER, Karl J</creator><creator>MINCHELLO, Elaine</creator><creator>MULLENIX, Patricia A</creator><creator>MYERS, Michael</creator><creator>PETERSEN, Annette</creator><creator>TERNSTROM, Wanda</creator><creator>WILSON, Sandra J</creator><creator>BRUGNARA, Carlo</creator><creator>CANFIELD, William</creator><creator>DORFMAN, David M</creator><creator>FIEHN, W</creator><creator>FISCHER, Gena</creator><creator>FITZPATRICK, Patricia</creator><creator>FLAMING, Timothy H</creator><creator>HENRIKSEN, Hilde-Kristin</creator><general>American Society of Clinical Pathologists</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>200405</creationdate><title>Automated flow cytometric analysis of blood cells in cerebrospinal fluid: Analytic performance</title><author>AUNE, Marthe W ; BECKER, Joanne L ; KUNICKA, Jolanta E ; LACKNER, Karl J ; MINCHELLO, Elaine ; MULLENIX, Patricia A ; MYERS, Michael ; PETERSEN, Annette ; TERNSTROM, Wanda ; WILSON, Sandra J ; BRUGNARA, Carlo ; CANFIELD, William ; DORFMAN, David M ; FIEHN, W ; FISCHER, Gena ; FITZPATRICK, Patricia ; FLAMING, Timothy H ; HENRIKSEN, Hilde-Kristin</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c195t-89b88ab5b5bce1afc31e18f81924949a23c8cad8eff2c202949a95822abca2db3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2004</creationdate><topic>Autoanalysis - instrumentation</topic><topic>Biological and medical sciences</topic><topic>Cerebrospinal Fluid - cytology</topic><topic>Erythrocyte Count - methods</topic><topic>Flow Cytometry - methods</topic><topic>General aspects</topic><topic>Hematologic Tests - instrumentation</topic><topic>Hematologic Tests - methods</topic><topic>Humans</topic><topic>Leukocyte Count - methods</topic><topic>Medical sciences</topic><topic>Reference Standards</topic><topic>Reproducibility of Results</topic><topic>Tumors</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>AUNE, Marthe W</creatorcontrib><creatorcontrib>BECKER, Joanne L</creatorcontrib><creatorcontrib>KUNICKA, Jolanta E</creatorcontrib><creatorcontrib>LACKNER, Karl J</creatorcontrib><creatorcontrib>MINCHELLO, Elaine</creatorcontrib><creatorcontrib>MULLENIX, Patricia A</creatorcontrib><creatorcontrib>MYERS, Michael</creatorcontrib><creatorcontrib>PETERSEN, Annette</creatorcontrib><creatorcontrib>TERNSTROM, Wanda</creatorcontrib><creatorcontrib>WILSON, Sandra J</creatorcontrib><creatorcontrib>BRUGNARA, Carlo</creatorcontrib><creatorcontrib>CANFIELD, William</creatorcontrib><creatorcontrib>DORFMAN, David M</creatorcontrib><creatorcontrib>FIEHN, W</creatorcontrib><creatorcontrib>FISCHER, Gena</creatorcontrib><creatorcontrib>FITZPATRICK, Patricia</creatorcontrib><creatorcontrib>FLAMING, Timothy H</creatorcontrib><creatorcontrib>HENRIKSEN, Hilde-Kristin</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>American journal of clinical pathology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>AUNE, Marthe W</au><au>BECKER, Joanne L</au><au>KUNICKA, Jolanta E</au><au>LACKNER, Karl J</au><au>MINCHELLO, Elaine</au><au>MULLENIX, Patricia A</au><au>MYERS, Michael</au><au>PETERSEN, Annette</au><au>TERNSTROM, Wanda</au><au>WILSON, Sandra J</au><au>BRUGNARA, Carlo</au><au>CANFIELD, William</au><au>DORFMAN, David M</au><au>FIEHN, W</au><au>FISCHER, Gena</au><au>FITZPATRICK, Patricia</au><au>FLAMING, Timothy H</au><au>HENRIKSEN, Hilde-Kristin</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Automated flow cytometric analysis of blood cells in cerebrospinal fluid: Analytic performance</atitle><jtitle>American journal of clinical pathology</jtitle><addtitle>Am J Clin Pathol</addtitle><date>2004-05</date><risdate>2004</risdate><volume>121</volume><issue>5</issue><spage>690</spage><epage>700</epage><pages>690-700</pages><issn>0002-9173</issn><eissn>1943-7722</eissn><coden>AJCPAI</coden><abstract>We compared the performance of an automated method for obtaining RBC and WBC counts and WBC differential counts in cerebrospinal fluid (CSF) samples with the reference manual method. Results from 325 samples from 10 worldwide clinical sites were used to demonstrate the accuracy, precision, and linearity of the method. Accuracy statistics for absolute cell counts showed a high correlation between methods, with correlation coefficients for all reportable absolute counts greater than 0.9. Linearity results demonstrated that the method provides accurate results throughout the reportable ranges, including clinical decision points for WBCs of 0 to 10/microL. Interassay precision and intra-assay precision for the ADVIA 120 (Bayer HealthCare, Tarrytown, NY) method were acceptable at all levels. The ADVIA 120 CSF Assay enumerates and differentiates cells via flow cytometry in a minimally diluted sample, improving the analysis of typically hypocellular CSF samples. Study results demonstrate that the automated ADVIA 120 CSF Assay is an acceptable alternative to the labor-intensive manual method.</abstract><cop>Chicago, IL</cop><pub>American Society of Clinical Pathologists</pub><pmid>15151209</pmid><doi>10.1309/EKFW9E3LLFXE15X9</doi><tpages>11</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0002-9173
ispartof American journal of clinical pathology, 2004-05, Vol.121 (5), p.690-700
issn 0002-9173
1943-7722
language eng
recordid cdi_proquest_miscellaneous_71945570
source MEDLINE; EZB Electronic Journals Library
subjects Autoanalysis - instrumentation
Biological and medical sciences
Cerebrospinal Fluid - cytology
Erythrocyte Count - methods
Flow Cytometry - methods
General aspects
Hematologic Tests - instrumentation
Hematologic Tests - methods
Humans
Leukocyte Count - methods
Medical sciences
Reference Standards
Reproducibility of Results
Tumors
title Automated flow cytometric analysis of blood cells in cerebrospinal fluid: Analytic performance
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-11-29T14%3A20%3A30IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Automated%20flow%20cytometric%20analysis%20of%20blood%20cells%20in%20cerebrospinal%20fluid:%20Analytic%20performance&rft.jtitle=American%20journal%20of%20clinical%20pathology&rft.au=AUNE,%20Marthe%20W&rft.date=2004-05&rft.volume=121&rft.issue=5&rft.spage=690&rft.epage=700&rft.pages=690-700&rft.issn=0002-9173&rft.eissn=1943-7722&rft.coden=AJCPAI&rft_id=info:doi/10.1309/EKFW9E3LLFXE15X9&rft_dat=%3Cproquest_cross%3E71945570%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=71945570&rft_id=info:pmid/15151209&rfr_iscdi=true