Validation of a RP-LC method for the simultaneous determination of isoniazid, pyrazinamide and rifampicin in a pharmaceutical formulation
A simple and accurate liquid chromatographic method was developed and validated for estimation of isoniazid (ISN), pyrazinamide (PYR) and rifampicin (RIF) in combined dosage forms. Drugs were chromatographed on a reverse phase C18 column using a mobile phase gradient and monitored at the correspondi...
Gespeichert in:
Veröffentlicht in: | Journal of pharmaceutical and biomedical analysis 2002-08, Vol.29 (6), p.1089-1096 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A simple and accurate liquid chromatographic method was developed and validated for estimation of isoniazid (ISN), pyrazinamide (PYR) and rifampicin (RIF) in combined dosage forms. Drugs were chromatographed on a reverse phase C18 column using a mobile phase gradient and monitored at the corresponding maximum of each compounds. Peaks were identified with retention time as compared with standards and confirmed with characteristic spectra using diode-array detector. Solution concentrations were measured on a weight basis to avoid the use of an internal standard. The method does not require any specific sample preparation except the use of a guard column. The method is linear (
r
2>0.999), precise (RSD%: 0.50% for ISN, 0.12% for PYR and 0.98% for RIF), accurate (overall average recovery yields: 98.55% for ISN, 98.51 for PYR and 98.56% for RIF) and selective. Due to its simplicity and accuracy the method is suitable for routine quality control analysis of antitubercolosis combination dosage form. |
---|---|
ISSN: | 0731-7085 1873-264X |
DOI: | 10.1016/S0731-7085(02)00150-4 |