Development and single-laboratory validation of a reversed-phase liquid chromatography-electrospray-tandem mass spectrometry method for identification and determination of acrylamide in foods

A rapid and accurate method using reversed-phase liquid chromatography-tandem mass spectrometry interfaced with electrospray was developed for determination of acrylamide in cooked food samples. A simplified sample treatment procedure using an extraction step with acidified water without cleanup was...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of AOAC International 2004-01, Vol.87 (1), p.107-115
Hauptverfasser: Calbiani, F, Careri, M, Elviri, L, Mangia, A, Zagnoni, I
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A rapid and accurate method using reversed-phase liquid chromatography-tandem mass spectrometry interfaced with electrospray was developed for determination of acrylamide in cooked food samples. A simplified sample treatment procedure using an extraction step with acidified water without cleanup was developed. A C18 column with an aqueous formic acid-methanol mixture as the mobile phase was used under isocratic conditions. The method was validated in-house for robustness, limits of detection (LOD) and quantitation (LOQ), linearity, recovery, and accuracy both on standard and baked-product and potato flour matrixes. Good results in the low ppb level were obtained for LOD (
ISSN:1060-3271