Correction of slab boundary artifact using histogram matching
Purpose To improve the imaging efficiency of multi‐slab three‐dimensional acquisitions by decreasing slab overlap and the number of blanked slices. Materials and Methods Application of the histogram matching technique for correction of the slab boundary artifact (SBA) due to non‐ideal radiofrequency...
Gespeichert in:
Veröffentlicht in: | Journal of magnetic resonance imaging 2002-05, Vol.15 (5), p.610-617 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
To improve the imaging efficiency of multi‐slab three‐dimensional acquisitions by decreasing slab overlap and the number of blanked slices.
Materials and Methods
Application of the histogram matching technique for correction of the slab boundary artifact (SBA) due to non‐ideal radiofrequency pulse profiles has been proposed. The technique was validated by studying its effectiveness in reducing the SBA in phantom images, as well as in proton density‐ and T2‐weighted brain images.
Results
The correction algorithm considerably decreases the disruption of image volume continuity due to the SBA, and improves slice‐to‐slice intensity continuity.
Conclusion
The proposed technique can greatly reduce the SBA when the SBA can be described as a global intensity variation along the slab. The technique may be used to improve imaging efficiency of the multi‐slab three‐dimensional fast spin‐echo sequences by decreasing the amount of required slab overlap. J. Magn. Reson. Imaging 2002;15:610–617. © 2002 Wiley‐Liss, Inc. |
---|---|