Validation of the NucliSens Extractor in combination with the hepatitis C virus Cobas Amplicor 2.0 assay in four laboratories in the Netherlands utilizing nucleic acid amplification technology for blood screening

Background and Objectives Since July 1 1999, four laboratories in the Netherlands have been routinely screening plasma minipools for the release of labile blood components utilizing hepatitis C virus nucleic acid amplification technology (HCV NAT). This report describes the performance evaluation of...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Vox sanguinis 2001-07, Vol.81 (1), p.12-20
Hauptverfasser: Cuijpers, H. T. M., Molijn, M. H. J., Bos, H. J., Peeters, A. P. W., Van Der Poel, C. L., Lelie, P. N.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Background and Objectives Since July 1 1999, four laboratories in the Netherlands have been routinely screening plasma minipools for the release of labile blood components utilizing hepatitis C virus nucleic acid amplification technology (HCV NAT). This report describes the performance evaluation of the HCV NAT method and the quality control results obtained during 6 months of routine screening. Materials and Methods Plasma minipools of 48 donations were prepared on a Tecan Genesis robot. HCV RNA was isolated from 2 ml of plasma by using the NucliSens Extractor and amplified and detected with the Cobas HCV Amplicor 2.0 test system. For validation of the test system the laboratories used viral quality control (VQC) reagents of CLB. Results Initial robustness experiments demonstrated consistent detection of PeliSpy HCV RNA samples of 140 genome equivalents/ml (geq/ml) in each station of the installed Nuclisens Extractors. Further ‘stress’ tests with a highly viraemic sample of ≈ 5·106 geq/ml did not contaminate negative samples processed on all Extractor stations in subsequent runs. In the validation period prior to July 1999, 1021 pools were tested with the following performance characteristics: 0·1%, initially false reactive; 0·89%, failure of internal control detection; 0·97%, no eluate generated by the Extractor; and 100% reactivity of the PeliSpy 140 geq/ml control in 176 Extractor runs and a 98% reactivity rate of the PeliSpy 38 geq/ml control in 102 test runs. By testing the PeliCheck HCV RNA genotype 1 dilution panels 49 times, an overall 95% detection limit of 30 geq/ml (≈ 8 IU/ml) and a 50% detection limit of 5 geq/ml was found by the four laboratories. In the first 6 months of routine screening, the minimum requirement for invalid results (2%) was exceeded with some batches of silica and NucliSens Extractor cartridges. From November 1999 to February 2000, the manufacturer (Organon Teknika) improved the protocol for silica absorption of the Nuclisens Extractor – the cartridge design as well as the software of the Extractor. During the next 6 months of observation in 2000, the percentages of false initial reactives and invalids were 0·05% and 1·4%, respectively, in 8962 pools tested. Of these invalid results, 0·74% and 0·66% were caused by Extractor failure and negative internal control signals, respectively. The PeliSpy HCV RNA ‘stop or go’ run control of 140 geq/ml was 100% reactive, but invalid in 16/1375 (1·2%) of cases. The PeliSpy run control
ISSN:0042-9007
1423-0410
DOI:10.1046/j.1423-0410.2001.00055.x