Real-Time Optical Coherence Tomography of the Anterior Segment at 1310 nm

BACKGROUND Recent advances in high-speed scanning technology have enabled a new generation of optical coherence tomographic (OCT) systems to perform imaging at video rate. Here, a handheld OCT probe capable of imaging the anterior segment of the eye at high frame rates is demonstrated for the first...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Archives of ophthalmology (1960) 2001-08, Vol.119 (8), p.1179-1185
Hauptverfasser: Radhakrishnan, Sunita, Rollins, Andrew M, Roth, Jonathan E, Yazdanfar, Siavash, Westphal, Volker, Bardenstein, David S, Izatt, Joseph A
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:BACKGROUND Recent advances in high-speed scanning technology have enabled a new generation of optical coherence tomographic (OCT) systems to perform imaging at video rate. Here, a handheld OCT probe capable of imaging the anterior segment of the eye at high frame rates is demonstrated for the first time. OBJECTIVE To demonstrate real-time OCT imaging of anterior segment structures. DESIGN Survey of anterior segment structures in normal human subjects. SETTING Laboratory. MAIN OUTCOME MEASURES Achieving real-time imaging of the anterior segment, satisfactory image quality, and convenience of a handheld probe. RESULTS Optical coherence tomographic imaging of the anterior segment of the eyes of human subjects was performed using 1310-nm wavelength light with an image rate of 8 frames per second. Imaging trials demonstrated clear resolution of corneal epithelium and stroma, sclerocorneal junction, sclera, iris pigment epithelium and stroma, and anterior lens capsule. The anterior chamber angle was clearly visualized. Limited imaging of the ciliary body was performed. Real-time imaging of pupillary constriction in response to light stimulus was also performed. CONCLUSION High-speed OCT at 1310-nm wavelength is a potentially useful technique for noninvasive assessment of anterior segment structures. CLINICAL RELEVANCE Our results suggest that real-time OCT has potential applications in glaucoma evaluation and refractive surgery.Arch Ophthalmol. 2001;119:1179-1185-->
ISSN:0003-9950
2168-6165
1538-3601
2168-6173
DOI:10.1001/archopht.119.8.1179