Goldmann Applanation Tonometry in the Conscious Rat

To determine whether the Goldmann applanation tonometer can be modified to measure intraocular pressure (IOP) in the conscious rat. In anesthetized rats Goldmann tonometers were tested that had reduced biprism angles in the applanating tips and reduced weights in the tonometer body from those used i...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Investigative ophthalmology & visual science 2001-02, Vol.42 (2), p.340-342
Hauptverfasser: Cohan, Bruce E, Bohr, David F
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 342
container_issue 2
container_start_page 340
container_title Investigative ophthalmology & visual science
container_volume 42
creator Cohan, Bruce E
Bohr, David F
description To determine whether the Goldmann applanation tonometer can be modified to measure intraocular pressure (IOP) in the conscious rat. In anesthetized rats Goldmann tonometers were tested that had reduced biprism angles in the applanating tips and reduced weights in the tonometer body from those used in humans and species with similar size eyes. Tonometers with tips with biprism angles of 48 degrees and an applied weight of 25 mg per Goldmann scale division (2 g full scale) were calibrated for the rat against manometrically measured IOP. Tonometers, thus modified, were then used in conscious, unsedated rats. In conscious rats the measured mean Goldmann value was 15.5 +/- 0.6 mm Hg (confidence interval = 14.1, 16.6 mm Hg). This was the plateau level reached after the repeated applanations (approximately 10) required to eliminate an artifactual decline in initial Goldmann readings, which was larger than that in humans. The Goldmann applanation tonometer was modified to measure IOP in the conscious, unsedated rat. This instrument, the standard for measuring this key physiological parameter in the human eye, can now be applied to the laboratory rat. This may advance the use of this important animal as a model in IOP and glaucoma research.
format Article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_proquest_miscellaneous_70620673</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>70620673</sourcerecordid><originalsourceid>FETCH-LOGICAL-h180t-8db27a89663737de8750e47b829c52a29f6e84a1b45b87fb1a6bbb973abdf4aa3</originalsourceid><addsrcrecordid>eNo9z11LwzAYBeAgipvTvyAF0btCvpNejqFTGAgyr8ObNrWRNqlNR9m_t7Dp1bl5OJxzgZZECJoLpdklWmLCZY455gt0k9I3xpQQiq_RghAyE8mXiG1jW3UQQrbu-xYCjD6GbB9D7Nw4HDMfsrFx2SaGVPp4SNkHjLfoqoY2ubtzrtDny_N-85rv3rdvm_Uub4jGY64rSxXoQkqmmKqcVgI7rqymRSko0KKWTnMglgurVW0JSGttoRjYquYAbIWeTr39EH8OLo2m86l07TzTzVOMwpJiqdgM78_wYDtXmX7wHQxH83dzBg9nAKmEth4glD79u4JyUehZPZ5U47-ayQ_OpA7adi4lZpomTg01jGP2C4HhaA0</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>70620673</pqid></control><display><type>article</type><title>Goldmann Applanation Tonometry in the Conscious Rat</title><source>MEDLINE</source><source>EZB-FREE-00999 freely available EZB journals</source><creator>Cohan, Bruce E ; Bohr, David F</creator><creatorcontrib>Cohan, Bruce E ; Bohr, David F</creatorcontrib><description>To determine whether the Goldmann applanation tonometer can be modified to measure intraocular pressure (IOP) in the conscious rat. In anesthetized rats Goldmann tonometers were tested that had reduced biprism angles in the applanating tips and reduced weights in the tonometer body from those used in humans and species with similar size eyes. Tonometers with tips with biprism angles of 48 degrees and an applied weight of 25 mg per Goldmann scale division (2 g full scale) were calibrated for the rat against manometrically measured IOP. Tonometers, thus modified, were then used in conscious, unsedated rats. In conscious rats the measured mean Goldmann value was 15.5 +/- 0.6 mm Hg (confidence interval = 14.1, 16.6 mm Hg). This was the plateau level reached after the repeated applanations (approximately 10) required to eliminate an artifactual decline in initial Goldmann readings, which was larger than that in humans. The Goldmann applanation tonometer was modified to measure IOP in the conscious, unsedated rat. This instrument, the standard for measuring this key physiological parameter in the human eye, can now be applied to the laboratory rat. This may advance the use of this important animal as a model in IOP and glaucoma research.</description><identifier>ISSN: 0146-0404</identifier><identifier>EISSN: 1552-5783</identifier><identifier>PMID: 11157864</identifier><identifier>CODEN: IOVSDA</identifier><language>eng</language><publisher>Rockville, MD: ARVO</publisher><subject>Animals ; Biological and medical sciences ; Calibration ; Consciousness ; Glaucoma and intraocular pressure ; Intraocular Pressure ; Medical sciences ; Ophthalmology ; Rats ; Rats, Inbred BN ; Tonometry, Ocular - methods</subject><ispartof>Investigative ophthalmology &amp; visual science, 2001-02, Vol.42 (2), p.340-342</ispartof><rights>2001 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&amp;idt=924598$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/11157864$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Cohan, Bruce E</creatorcontrib><creatorcontrib>Bohr, David F</creatorcontrib><title>Goldmann Applanation Tonometry in the Conscious Rat</title><title>Investigative ophthalmology &amp; visual science</title><addtitle>Invest Ophthalmol Vis Sci</addtitle><description>To determine whether the Goldmann applanation tonometer can be modified to measure intraocular pressure (IOP) in the conscious rat. In anesthetized rats Goldmann tonometers were tested that had reduced biprism angles in the applanating tips and reduced weights in the tonometer body from those used in humans and species with similar size eyes. Tonometers with tips with biprism angles of 48 degrees and an applied weight of 25 mg per Goldmann scale division (2 g full scale) were calibrated for the rat against manometrically measured IOP. Tonometers, thus modified, were then used in conscious, unsedated rats. In conscious rats the measured mean Goldmann value was 15.5 +/- 0.6 mm Hg (confidence interval = 14.1, 16.6 mm Hg). This was the plateau level reached after the repeated applanations (approximately 10) required to eliminate an artifactual decline in initial Goldmann readings, which was larger than that in humans. The Goldmann applanation tonometer was modified to measure IOP in the conscious, unsedated rat. This instrument, the standard for measuring this key physiological parameter in the human eye, can now be applied to the laboratory rat. This may advance the use of this important animal as a model in IOP and glaucoma research.</description><subject>Animals</subject><subject>Biological and medical sciences</subject><subject>Calibration</subject><subject>Consciousness</subject><subject>Glaucoma and intraocular pressure</subject><subject>Intraocular Pressure</subject><subject>Medical sciences</subject><subject>Ophthalmology</subject><subject>Rats</subject><subject>Rats, Inbred BN</subject><subject>Tonometry, Ocular - methods</subject><issn>0146-0404</issn><issn>1552-5783</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2001</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNo9z11LwzAYBeAgipvTvyAF0btCvpNejqFTGAgyr8ObNrWRNqlNR9m_t7Dp1bl5OJxzgZZECJoLpdklWmLCZY455gt0k9I3xpQQiq_RghAyE8mXiG1jW3UQQrbu-xYCjD6GbB9D7Nw4HDMfsrFx2SaGVPp4SNkHjLfoqoY2ubtzrtDny_N-85rv3rdvm_Uub4jGY64rSxXoQkqmmKqcVgI7rqymRSko0KKWTnMglgurVW0JSGttoRjYquYAbIWeTr39EH8OLo2m86l07TzTzVOMwpJiqdgM78_wYDtXmX7wHQxH83dzBg9nAKmEth4glD79u4JyUehZPZ5U47-ayQ_OpA7adi4lZpomTg01jGP2C4HhaA0</recordid><startdate>200102</startdate><enddate>200102</enddate><creator>Cohan, Bruce E</creator><creator>Bohr, David F</creator><general>ARVO</general><general>Association for Research in Vision and Ophtalmology</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>7X8</scope></search><sort><creationdate>200102</creationdate><title>Goldmann Applanation Tonometry in the Conscious Rat</title><author>Cohan, Bruce E ; Bohr, David F</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-h180t-8db27a89663737de8750e47b829c52a29f6e84a1b45b87fb1a6bbb973abdf4aa3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2001</creationdate><topic>Animals</topic><topic>Biological and medical sciences</topic><topic>Calibration</topic><topic>Consciousness</topic><topic>Glaucoma and intraocular pressure</topic><topic>Intraocular Pressure</topic><topic>Medical sciences</topic><topic>Ophthalmology</topic><topic>Rats</topic><topic>Rats, Inbred BN</topic><topic>Tonometry, Ocular - methods</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Cohan, Bruce E</creatorcontrib><creatorcontrib>Bohr, David F</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>MEDLINE - Academic</collection><jtitle>Investigative ophthalmology &amp; visual science</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Cohan, Bruce E</au><au>Bohr, David F</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Goldmann Applanation Tonometry in the Conscious Rat</atitle><jtitle>Investigative ophthalmology &amp; visual science</jtitle><addtitle>Invest Ophthalmol Vis Sci</addtitle><date>2001-02</date><risdate>2001</risdate><volume>42</volume><issue>2</issue><spage>340</spage><epage>342</epage><pages>340-342</pages><issn>0146-0404</issn><eissn>1552-5783</eissn><coden>IOVSDA</coden><abstract>To determine whether the Goldmann applanation tonometer can be modified to measure intraocular pressure (IOP) in the conscious rat. In anesthetized rats Goldmann tonometers were tested that had reduced biprism angles in the applanating tips and reduced weights in the tonometer body from those used in humans and species with similar size eyes. Tonometers with tips with biprism angles of 48 degrees and an applied weight of 25 mg per Goldmann scale division (2 g full scale) were calibrated for the rat against manometrically measured IOP. Tonometers, thus modified, were then used in conscious, unsedated rats. In conscious rats the measured mean Goldmann value was 15.5 +/- 0.6 mm Hg (confidence interval = 14.1, 16.6 mm Hg). This was the plateau level reached after the repeated applanations (approximately 10) required to eliminate an artifactual decline in initial Goldmann readings, which was larger than that in humans. The Goldmann applanation tonometer was modified to measure IOP in the conscious, unsedated rat. This instrument, the standard for measuring this key physiological parameter in the human eye, can now be applied to the laboratory rat. This may advance the use of this important animal as a model in IOP and glaucoma research.</abstract><cop>Rockville, MD</cop><pub>ARVO</pub><pmid>11157864</pmid><tpages>3</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0146-0404
ispartof Investigative ophthalmology & visual science, 2001-02, Vol.42 (2), p.340-342
issn 0146-0404
1552-5783
language eng
recordid cdi_proquest_miscellaneous_70620673
source MEDLINE; EZB-FREE-00999 freely available EZB journals
subjects Animals
Biological and medical sciences
Calibration
Consciousness
Glaucoma and intraocular pressure
Intraocular Pressure
Medical sciences
Ophthalmology
Rats
Rats, Inbred BN
Tonometry, Ocular - methods
title Goldmann Applanation Tonometry in the Conscious Rat
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-14T22%3A02%3A33IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Goldmann%20Applanation%20Tonometry%20in%20the%20Conscious%20Rat&rft.jtitle=Investigative%20ophthalmology%20&%20visual%20science&rft.au=Cohan,%20Bruce%20E&rft.date=2001-02&rft.volume=42&rft.issue=2&rft.spage=340&rft.epage=342&rft.pages=340-342&rft.issn=0146-0404&rft.eissn=1552-5783&rft.coden=IOVSDA&rft_id=info:doi/&rft_dat=%3Cproquest_pubme%3E70620673%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=70620673&rft_id=info:pmid/11157864&rfr_iscdi=true