Fluid balance error in continuous renal replacement therapy: a technical note
The reliability and safety of continuous renal replacement therapy (CRRT) machines have improved, yet there still remains the potential for fluid balance errors to occur during treatment. In vitro testing of two Kimal Hygieia CRRT machines (Plus and Ultima) was performed. Normal saline to simulate t...
Gespeichert in:
Veröffentlicht in: | International journal of artificial organs 2007-05, Vol.30 (5), p.434-440 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 440 |
---|---|
container_issue | 5 |
container_start_page | 434 |
container_title | International journal of artificial organs |
container_volume | 30 |
creator | Bagshaw, S M Baldwin, I Fealy, N Bellomo, R |
description | The reliability and safety of continuous renal replacement therapy (CRRT) machines have improved, yet there still remains the potential for fluid balance errors to occur during treatment.
In vitro testing of two Kimal Hygieia CRRT machines (Plus and Ultima) was performed. Normal saline to simulate the blood circuit and standard bicarbonate-based fluid for replacement were used. All tests were performed in CVVH mode at four ultrafiltration (UF) rates. The testing was based on creation of a voluntary fluid balance error by clamping the line that fills the replacement fluid chamber to stop flow to the (simulated) patient. The time to alarms and fluid balance errors were recorded. The alarms were overridden and the accumulated fluid balance error allowed by the machine was determined.
The alarm occurred approximately 1 minute after the replacement fluid line was clamped at all UF rates. There was no limit to the number of times the alarm could be overridden and the accumulated negative fluid balance was proportional to the prescribed UF rate. After the replacement fluid chamber was allowed to re-fill, the machine attempted to correct the fluid deficit and consistently delivered excess fluid to generate a positive fluid balance error.
The Hygieia machines appear designed with appropriate alarm and safety features. However, simulated fluid balance errors raise caution for operators. Clinicians and nurses need to understand the clinical implications of alarm overrides. Fluid balance errors caused by failure to acknowledge and correct replacement fluid failure alarms may cause harm to patients. |
doi_str_mv | 10.1177/039139880703000511 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_70574602</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>20305560</sourcerecordid><originalsourceid>FETCH-LOGICAL-c446t-8443af1392a46da3392576f00f8b14c374690af12f80bc54df252cd1a0aa9b793</originalsourceid><addsrcrecordid>eNqFkD9PwzAQxS0EoqXwBRiQJ7bA-V-csKGKAlIRC8yR49hqUOIE2xn67XHUSgwMLHc3_N7Tu4fQNYE7QqS8B1YSVhYFSGAAIAg5QUsiKc9y4HCKljOQzcQCXYTwBUByzsU5WhApBClBLtHbppvaBteqU04bbLwfPG4d1oOLrZuGKWBvnOrSHDulTW9cxHFnvBr3D1jhaPTOtToBbojmEp1Z1QVzddwr9Ll5-li_ZNv359f14zbTnOcxKzhnyqbsVPG8USwdQuYWwBY14ZpJnpeQAGoLqLXgjaWC6oYoUKqsZclW6PbgO_rhezIhVn0btOnSEyZFriSI5AH0X5Cm5oTIIYH0AGo_hOCNrUbf9srvKwLV3Hb1t-0kujm6T3Vvml_JsV72A92-eRY</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>20305560</pqid></control><display><type>article</type><title>Fluid balance error in continuous renal replacement therapy: a technical note</title><source>MEDLINE</source><source>SAGE Complete A-Z List</source><creator>Bagshaw, S M ; Baldwin, I ; Fealy, N ; Bellomo, R</creator><creatorcontrib>Bagshaw, S M ; Baldwin, I ; Fealy, N ; Bellomo, R</creatorcontrib><description>The reliability and safety of continuous renal replacement therapy (CRRT) machines have improved, yet there still remains the potential for fluid balance errors to occur during treatment.
In vitro testing of two Kimal Hygieia CRRT machines (Plus and Ultima) was performed. Normal saline to simulate the blood circuit and standard bicarbonate-based fluid for replacement were used. All tests were performed in CVVH mode at four ultrafiltration (UF) rates. The testing was based on creation of a voluntary fluid balance error by clamping the line that fills the replacement fluid chamber to stop flow to the (simulated) patient. The time to alarms and fluid balance errors were recorded. The alarms were overridden and the accumulated fluid balance error allowed by the machine was determined.
The alarm occurred approximately 1 minute after the replacement fluid line was clamped at all UF rates. There was no limit to the number of times the alarm could be overridden and the accumulated negative fluid balance was proportional to the prescribed UF rate. After the replacement fluid chamber was allowed to re-fill, the machine attempted to correct the fluid deficit and consistently delivered excess fluid to generate a positive fluid balance error.
The Hygieia machines appear designed with appropriate alarm and safety features. However, simulated fluid balance errors raise caution for operators. Clinicians and nurses need to understand the clinical implications of alarm overrides. Fluid balance errors caused by failure to acknowledge and correct replacement fluid failure alarms may cause harm to patients.</description><identifier>ISSN: 0391-3988</identifier><identifier>EISSN: 1724-6040</identifier><identifier>DOI: 10.1177/039139880703000511</identifier><identifier>PMID: 17551907</identifier><language>eng</language><publisher>United States</publisher><subject>Equipment Safety ; Hemofiltration - adverse effects ; Hemofiltration - instrumentation ; Models, Biological ; Ultrafiltration ; Water-Electrolyte Balance ; Water-Electrolyte Imbalance - diagnosis</subject><ispartof>International journal of artificial organs, 2007-05, Vol.30 (5), p.434-440</ispartof><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c446t-8443af1392a46da3392576f00f8b14c374690af12f80bc54df252cd1a0aa9b793</citedby><cites>FETCH-LOGICAL-c446t-8443af1392a46da3392576f00f8b14c374690af12f80bc54df252cd1a0aa9b793</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,777,781,27905,27906</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/17551907$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Bagshaw, S M</creatorcontrib><creatorcontrib>Baldwin, I</creatorcontrib><creatorcontrib>Fealy, N</creatorcontrib><creatorcontrib>Bellomo, R</creatorcontrib><title>Fluid balance error in continuous renal replacement therapy: a technical note</title><title>International journal of artificial organs</title><addtitle>Int J Artif Organs</addtitle><description>The reliability and safety of continuous renal replacement therapy (CRRT) machines have improved, yet there still remains the potential for fluid balance errors to occur during treatment.
In vitro testing of two Kimal Hygieia CRRT machines (Plus and Ultima) was performed. Normal saline to simulate the blood circuit and standard bicarbonate-based fluid for replacement were used. All tests were performed in CVVH mode at four ultrafiltration (UF) rates. The testing was based on creation of a voluntary fluid balance error by clamping the line that fills the replacement fluid chamber to stop flow to the (simulated) patient. The time to alarms and fluid balance errors were recorded. The alarms were overridden and the accumulated fluid balance error allowed by the machine was determined.
The alarm occurred approximately 1 minute after the replacement fluid line was clamped at all UF rates. There was no limit to the number of times the alarm could be overridden and the accumulated negative fluid balance was proportional to the prescribed UF rate. After the replacement fluid chamber was allowed to re-fill, the machine attempted to correct the fluid deficit and consistently delivered excess fluid to generate a positive fluid balance error.
The Hygieia machines appear designed with appropriate alarm and safety features. However, simulated fluid balance errors raise caution for operators. Clinicians and nurses need to understand the clinical implications of alarm overrides. Fluid balance errors caused by failure to acknowledge and correct replacement fluid failure alarms may cause harm to patients.</description><subject>Equipment Safety</subject><subject>Hemofiltration - adverse effects</subject><subject>Hemofiltration - instrumentation</subject><subject>Models, Biological</subject><subject>Ultrafiltration</subject><subject>Water-Electrolyte Balance</subject><subject>Water-Electrolyte Imbalance - diagnosis</subject><issn>0391-3988</issn><issn>1724-6040</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2007</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqFkD9PwzAQxS0EoqXwBRiQJ7bA-V-csKGKAlIRC8yR49hqUOIE2xn67XHUSgwMLHc3_N7Tu4fQNYE7QqS8B1YSVhYFSGAAIAg5QUsiKc9y4HCKljOQzcQCXYTwBUByzsU5WhApBClBLtHbppvaBteqU04bbLwfPG4d1oOLrZuGKWBvnOrSHDulTW9cxHFnvBr3D1jhaPTOtToBbojmEp1Z1QVzddwr9Ll5-li_ZNv359f14zbTnOcxKzhnyqbsVPG8USwdQuYWwBY14ZpJnpeQAGoLqLXgjaWC6oYoUKqsZclW6PbgO_rhezIhVn0btOnSEyZFriSI5AH0X5Cm5oTIIYH0AGo_hOCNrUbf9srvKwLV3Hb1t-0kujm6T3Vvml_JsV72A92-eRY</recordid><startdate>20070501</startdate><enddate>20070501</enddate><creator>Bagshaw, S M</creator><creator>Baldwin, I</creator><creator>Fealy, N</creator><creator>Bellomo, R</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7QO</scope><scope>8FD</scope><scope>FR3</scope><scope>P64</scope><scope>7X8</scope></search><sort><creationdate>20070501</creationdate><title>Fluid balance error in continuous renal replacement therapy: a technical note</title><author>Bagshaw, S M ; Baldwin, I ; Fealy, N ; Bellomo, R</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c446t-8443af1392a46da3392576f00f8b14c374690af12f80bc54df252cd1a0aa9b793</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2007</creationdate><topic>Equipment Safety</topic><topic>Hemofiltration - adverse effects</topic><topic>Hemofiltration - instrumentation</topic><topic>Models, Biological</topic><topic>Ultrafiltration</topic><topic>Water-Electrolyte Balance</topic><topic>Water-Electrolyte Imbalance - diagnosis</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Bagshaw, S M</creatorcontrib><creatorcontrib>Baldwin, I</creatorcontrib><creatorcontrib>Fealy, N</creatorcontrib><creatorcontrib>Bellomo, R</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>Biotechnology Research Abstracts</collection><collection>Technology Research Database</collection><collection>Engineering Research Database</collection><collection>Biotechnology and BioEngineering Abstracts</collection><collection>MEDLINE - Academic</collection><jtitle>International journal of artificial organs</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Bagshaw, S M</au><au>Baldwin, I</au><au>Fealy, N</au><au>Bellomo, R</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Fluid balance error in continuous renal replacement therapy: a technical note</atitle><jtitle>International journal of artificial organs</jtitle><addtitle>Int J Artif Organs</addtitle><date>2007-05-01</date><risdate>2007</risdate><volume>30</volume><issue>5</issue><spage>434</spage><epage>440</epage><pages>434-440</pages><issn>0391-3988</issn><eissn>1724-6040</eissn><abstract>The reliability and safety of continuous renal replacement therapy (CRRT) machines have improved, yet there still remains the potential for fluid balance errors to occur during treatment.
In vitro testing of two Kimal Hygieia CRRT machines (Plus and Ultima) was performed. Normal saline to simulate the blood circuit and standard bicarbonate-based fluid for replacement were used. All tests were performed in CVVH mode at four ultrafiltration (UF) rates. The testing was based on creation of a voluntary fluid balance error by clamping the line that fills the replacement fluid chamber to stop flow to the (simulated) patient. The time to alarms and fluid balance errors were recorded. The alarms were overridden and the accumulated fluid balance error allowed by the machine was determined.
The alarm occurred approximately 1 minute after the replacement fluid line was clamped at all UF rates. There was no limit to the number of times the alarm could be overridden and the accumulated negative fluid balance was proportional to the prescribed UF rate. After the replacement fluid chamber was allowed to re-fill, the machine attempted to correct the fluid deficit and consistently delivered excess fluid to generate a positive fluid balance error.
The Hygieia machines appear designed with appropriate alarm and safety features. However, simulated fluid balance errors raise caution for operators. Clinicians and nurses need to understand the clinical implications of alarm overrides. Fluid balance errors caused by failure to acknowledge and correct replacement fluid failure alarms may cause harm to patients.</abstract><cop>United States</cop><pmid>17551907</pmid><doi>10.1177/039139880703000511</doi><tpages>7</tpages><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0391-3988 |
ispartof | International journal of artificial organs, 2007-05, Vol.30 (5), p.434-440 |
issn | 0391-3988 1724-6040 |
language | eng |
recordid | cdi_proquest_miscellaneous_70574602 |
source | MEDLINE; SAGE Complete A-Z List |
subjects | Equipment Safety Hemofiltration - adverse effects Hemofiltration - instrumentation Models, Biological Ultrafiltration Water-Electrolyte Balance Water-Electrolyte Imbalance - diagnosis |
title | Fluid balance error in continuous renal replacement therapy: a technical note |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-18T12%3A01%3A12IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Fluid%20balance%20error%20in%20continuous%20renal%20replacement%20therapy:%20a%20technical%20note&rft.jtitle=International%20journal%20of%20artificial%20organs&rft.au=Bagshaw,%20S%20M&rft.date=2007-05-01&rft.volume=30&rft.issue=5&rft.spage=434&rft.epage=440&rft.pages=434-440&rft.issn=0391-3988&rft.eissn=1724-6040&rft_id=info:doi/10.1177/039139880703000511&rft_dat=%3Cproquest_cross%3E20305560%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=20305560&rft_id=info:pmid/17551907&rfr_iscdi=true |