The Arthur Vick Award: kinematics of the metacarpophalangeal joint after surface replacement arthroplasty

Prosthetic replacement in the hand must address such unique challenges as preservation of the collateral ligaments, tendon balancing,and Stability. Surface replacement arthroplasty can be an alternative to other current implants. The purpose of this study was to evaluate the metacarpophalangeal join...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Zeitschrift für Orthopädie und Unfallchirurgie 2007-03, Vol.145 (2), p.199-206
Hauptverfasser: Fayaz, H C, Beckenbaugh, R D, An, K N, Klawitter, J J, Jerosch, J, Rehart, S, Cooney, 3rd, W P
Format: Artikel
Sprache:ger
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 206
container_issue 2
container_start_page 199
container_title Zeitschrift für Orthopädie und Unfallchirurgie
container_volume 145
creator Fayaz, H C
Beckenbaugh, R D
An, K N
Klawitter, J J
Jerosch, J
Rehart, S
Cooney, 3rd, W P
description Prosthetic replacement in the hand must address such unique challenges as preservation of the collateral ligaments, tendon balancing,and Stability. Surface replacement arthroplasty can be an alternative to other current implants. The purpose of this study was to evaluate the metacarpophalangeal joint kinematics after surface replacement arthroplasty. The kinematics of pyrolytic carbon as a surface replacement implant for the metacarpophalangeal joint (MCP) was compared with the intact MCP joint in eight fresh cadaver long fingers by means of an electromagnetic tracking system (Polhemus, Colchester, VT). The eight human cadaver MCP joints were tested before implantation, after implantation, after collateral ligaments resection, and after collateral ligaments reconstruction. The kinematics of the MCP joint is reproduced by the joint surface replacement arthroplasty when normal ligament tension was present. The maximum angular displacement of the pyrocarbon implant was 378 for lateral deviation and 338 for rotation during the passive flexion and extension motion. The instantaneus center of rotation (ICR) after implant insertion was nearly identical to the center of rotation of the normal joint. The results also indicated that the collateral ligaments provide the primary stability of the MCP joint. No significant differences in lateral and rotational stability after surface replacement arthroplasty were noted. While collateral ligaments resection significantly affected the stability of the MCP joint. The ICR of the pyrocarbon implant most closely matched that of the intact MCP joint. The pyrocarbon implant provides suitable stability to radio-ulnar deviation and rotational stresses as a resurfacing implant and it simulates the kinematics of the intact MCP joint. By using new materials and taking the anatomical and biomechanical requirements into consideration, the endoprosthesis of the finger joints has created an option to achieve good long-term results. The inadequate results of earlier and current prostheses are a consequence of their mechanical construction and their materials. The success of the new implants could be proven by preferably long-term, controlled studies.
doi_str_mv 10.1055/s-2007-965173
format Article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_proquest_miscellaneous_70486191</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>70486191</sourcerecordid><originalsourceid>FETCH-LOGICAL-p139t-597c75b5404be7ed225b361d2c88e1cd1d47992b20b2e82ae0aacaf255376ba3</originalsourceid><addsrcrecordid>eNo1kD1PwzAYhD2AaCmMrMgTW8B2Yjtmqyq-pEosFWv02nlD3SZNsB2h_nuCKNPpdI9OpyPkhrN7zqR8iJlgTGdGSa7zMzLnpSoypYyekcsYd4wpLg27IDOuCyOk4nPiN1uky5C2Y6Af3u3p8htC_Uj3_oAdJO8i7RuaJqjDBA7C0A9baOHwidDSXe8PiUKTMNA4hgYc0oBDO2mHv8lUHPrJx3S8IucNtBGvT7ogm-enzeo1W7-_vK2W62zguUmZNNppaWXBCosaayGkzRWvhStL5K7mdaGNEVYwK7AUgAymWY2QMtfKQr4gd3-1Q-i_Royp6nx02E6TsR9jpVlRKm74BN6ewNF2WFdD8B2EY_X_Tf4DJu9lyg</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>70486191</pqid></control><display><type>article</type><title>The Arthur Vick Award: kinematics of the metacarpophalangeal joint after surface replacement arthroplasty</title><source>MEDLINE</source><source>Thieme Connect Journals</source><creator>Fayaz, H C ; Beckenbaugh, R D ; An, K N ; Klawitter, J J ; Jerosch, J ; Rehart, S ; Cooney, 3rd, W P</creator><creatorcontrib>Fayaz, H C ; Beckenbaugh, R D ; An, K N ; Klawitter, J J ; Jerosch, J ; Rehart, S ; Cooney, 3rd, W P</creatorcontrib><description>Prosthetic replacement in the hand must address such unique challenges as preservation of the collateral ligaments, tendon balancing,and Stability. Surface replacement arthroplasty can be an alternative to other current implants. The purpose of this study was to evaluate the metacarpophalangeal joint kinematics after surface replacement arthroplasty. The kinematics of pyrolytic carbon as a surface replacement implant for the metacarpophalangeal joint (MCP) was compared with the intact MCP joint in eight fresh cadaver long fingers by means of an electromagnetic tracking system (Polhemus, Colchester, VT). The eight human cadaver MCP joints were tested before implantation, after implantation, after collateral ligaments resection, and after collateral ligaments reconstruction. The kinematics of the MCP joint is reproduced by the joint surface replacement arthroplasty when normal ligament tension was present. The maximum angular displacement of the pyrocarbon implant was 378 for lateral deviation and 338 for rotation during the passive flexion and extension motion. The instantaneus center of rotation (ICR) after implant insertion was nearly identical to the center of rotation of the normal joint. The results also indicated that the collateral ligaments provide the primary stability of the MCP joint. No significant differences in lateral and rotational stability after surface replacement arthroplasty were noted. While collateral ligaments resection significantly affected the stability of the MCP joint. The ICR of the pyrocarbon implant most closely matched that of the intact MCP joint. The pyrocarbon implant provides suitable stability to radio-ulnar deviation and rotational stresses as a resurfacing implant and it simulates the kinematics of the intact MCP joint. By using new materials and taking the anatomical and biomechanical requirements into consideration, the endoprosthesis of the finger joints has created an option to achieve good long-term results. The inadequate results of earlier and current prostheses are a consequence of their mechanical construction and their materials. The success of the new implants could be proven by preferably long-term, controlled studies.</description><identifier>ISSN: 1864-6697</identifier><identifier>DOI: 10.1055/s-2007-965173</identifier><identifier>PMID: 17492561</identifier><language>ger</language><publisher>Germany</publisher><subject>Arthroplasty, Replacement, Finger - instrumentation ; Arthroplasty, Replacement, Finger - methods ; Awards and Prizes ; Biomechanical Phenomena - history ; Cadaver ; Elasticity ; Equipment Design ; Equipment Failure Analysis ; Fayaz ; Germany ; History, 21st Century ; Humans ; In Vitro Techniques ; Joint Instability - etiology ; Joint Instability - physiopathology ; Metacarpophalangeal Joint - physiopathology ; Metacarpophalangeal Joint - surgery ; Minnesota ; Orthopedics - history ; Range of Motion, Articular ; Stress, Mechanical ; Treatment Outcome</subject><ispartof>Zeitschrift für Orthopädie und Unfallchirurgie, 2007-03, Vol.145 (2), p.199-206</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/17492561$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Fayaz, H C</creatorcontrib><creatorcontrib>Beckenbaugh, R D</creatorcontrib><creatorcontrib>An, K N</creatorcontrib><creatorcontrib>Klawitter, J J</creatorcontrib><creatorcontrib>Jerosch, J</creatorcontrib><creatorcontrib>Rehart, S</creatorcontrib><creatorcontrib>Cooney, 3rd, W P</creatorcontrib><title>The Arthur Vick Award: kinematics of the metacarpophalangeal joint after surface replacement arthroplasty</title><title>Zeitschrift für Orthopädie und Unfallchirurgie</title><addtitle>Z Orthop Unfall</addtitle><description>Prosthetic replacement in the hand must address such unique challenges as preservation of the collateral ligaments, tendon balancing,and Stability. Surface replacement arthroplasty can be an alternative to other current implants. The purpose of this study was to evaluate the metacarpophalangeal joint kinematics after surface replacement arthroplasty. The kinematics of pyrolytic carbon as a surface replacement implant for the metacarpophalangeal joint (MCP) was compared with the intact MCP joint in eight fresh cadaver long fingers by means of an electromagnetic tracking system (Polhemus, Colchester, VT). The eight human cadaver MCP joints were tested before implantation, after implantation, after collateral ligaments resection, and after collateral ligaments reconstruction. The kinematics of the MCP joint is reproduced by the joint surface replacement arthroplasty when normal ligament tension was present. The maximum angular displacement of the pyrocarbon implant was 378 for lateral deviation and 338 for rotation during the passive flexion and extension motion. The instantaneus center of rotation (ICR) after implant insertion was nearly identical to the center of rotation of the normal joint. The results also indicated that the collateral ligaments provide the primary stability of the MCP joint. No significant differences in lateral and rotational stability after surface replacement arthroplasty were noted. While collateral ligaments resection significantly affected the stability of the MCP joint. The ICR of the pyrocarbon implant most closely matched that of the intact MCP joint. The pyrocarbon implant provides suitable stability to radio-ulnar deviation and rotational stresses as a resurfacing implant and it simulates the kinematics of the intact MCP joint. By using new materials and taking the anatomical and biomechanical requirements into consideration, the endoprosthesis of the finger joints has created an option to achieve good long-term results. The inadequate results of earlier and current prostheses are a consequence of their mechanical construction and their materials. The success of the new implants could be proven by preferably long-term, controlled studies.</description><subject>Arthroplasty, Replacement, Finger - instrumentation</subject><subject>Arthroplasty, Replacement, Finger - methods</subject><subject>Awards and Prizes</subject><subject>Biomechanical Phenomena - history</subject><subject>Cadaver</subject><subject>Elasticity</subject><subject>Equipment Design</subject><subject>Equipment Failure Analysis</subject><subject>Fayaz</subject><subject>Germany</subject><subject>History, 21st Century</subject><subject>Humans</subject><subject>In Vitro Techniques</subject><subject>Joint Instability - etiology</subject><subject>Joint Instability - physiopathology</subject><subject>Metacarpophalangeal Joint - physiopathology</subject><subject>Metacarpophalangeal Joint - surgery</subject><subject>Minnesota</subject><subject>Orthopedics - history</subject><subject>Range of Motion, Articular</subject><subject>Stress, Mechanical</subject><subject>Treatment Outcome</subject><issn>1864-6697</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2007</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNo1kD1PwzAYhD2AaCmMrMgTW8B2Yjtmqyq-pEosFWv02nlD3SZNsB2h_nuCKNPpdI9OpyPkhrN7zqR8iJlgTGdGSa7zMzLnpSoypYyekcsYd4wpLg27IDOuCyOk4nPiN1uky5C2Y6Af3u3p8htC_Uj3_oAdJO8i7RuaJqjDBA7C0A9baOHwidDSXe8PiUKTMNA4hgYc0oBDO2mHv8lUHPrJx3S8IucNtBGvT7ogm-enzeo1W7-_vK2W62zguUmZNNppaWXBCosaayGkzRWvhStL5K7mdaGNEVYwK7AUgAymWY2QMtfKQr4gd3-1Q-i_Royp6nx02E6TsR9jpVlRKm74BN6ewNF2WFdD8B2EY_X_Tf4DJu9lyg</recordid><startdate>200703</startdate><enddate>200703</enddate><creator>Fayaz, H C</creator><creator>Beckenbaugh, R D</creator><creator>An, K N</creator><creator>Klawitter, J J</creator><creator>Jerosch, J</creator><creator>Rehart, S</creator><creator>Cooney, 3rd, W P</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>7X8</scope></search><sort><creationdate>200703</creationdate><title>The Arthur Vick Award: kinematics of the metacarpophalangeal joint after surface replacement arthroplasty</title><author>Fayaz, H C ; Beckenbaugh, R D ; An, K N ; Klawitter, J J ; Jerosch, J ; Rehart, S ; Cooney, 3rd, W P</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-p139t-597c75b5404be7ed225b361d2c88e1cd1d47992b20b2e82ae0aacaf255376ba3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>ger</language><creationdate>2007</creationdate><topic>Arthroplasty, Replacement, Finger - instrumentation</topic><topic>Arthroplasty, Replacement, Finger - methods</topic><topic>Awards and Prizes</topic><topic>Biomechanical Phenomena - history</topic><topic>Cadaver</topic><topic>Elasticity</topic><topic>Equipment Design</topic><topic>Equipment Failure Analysis</topic><topic>Fayaz</topic><topic>Germany</topic><topic>History, 21st Century</topic><topic>Humans</topic><topic>In Vitro Techniques</topic><topic>Joint Instability - etiology</topic><topic>Joint Instability - physiopathology</topic><topic>Metacarpophalangeal Joint - physiopathology</topic><topic>Metacarpophalangeal Joint - surgery</topic><topic>Minnesota</topic><topic>Orthopedics - history</topic><topic>Range of Motion, Articular</topic><topic>Stress, Mechanical</topic><topic>Treatment Outcome</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Fayaz, H C</creatorcontrib><creatorcontrib>Beckenbaugh, R D</creatorcontrib><creatorcontrib>An, K N</creatorcontrib><creatorcontrib>Klawitter, J J</creatorcontrib><creatorcontrib>Jerosch, J</creatorcontrib><creatorcontrib>Rehart, S</creatorcontrib><creatorcontrib>Cooney, 3rd, W P</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>MEDLINE - Academic</collection><jtitle>Zeitschrift für Orthopädie und Unfallchirurgie</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Fayaz, H C</au><au>Beckenbaugh, R D</au><au>An, K N</au><au>Klawitter, J J</au><au>Jerosch, J</au><au>Rehart, S</au><au>Cooney, 3rd, W P</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>The Arthur Vick Award: kinematics of the metacarpophalangeal joint after surface replacement arthroplasty</atitle><jtitle>Zeitschrift für Orthopädie und Unfallchirurgie</jtitle><addtitle>Z Orthop Unfall</addtitle><date>2007-03</date><risdate>2007</risdate><volume>145</volume><issue>2</issue><spage>199</spage><epage>206</epage><pages>199-206</pages><issn>1864-6697</issn><abstract>Prosthetic replacement in the hand must address such unique challenges as preservation of the collateral ligaments, tendon balancing,and Stability. Surface replacement arthroplasty can be an alternative to other current implants. The purpose of this study was to evaluate the metacarpophalangeal joint kinematics after surface replacement arthroplasty. The kinematics of pyrolytic carbon as a surface replacement implant for the metacarpophalangeal joint (MCP) was compared with the intact MCP joint in eight fresh cadaver long fingers by means of an electromagnetic tracking system (Polhemus, Colchester, VT). The eight human cadaver MCP joints were tested before implantation, after implantation, after collateral ligaments resection, and after collateral ligaments reconstruction. The kinematics of the MCP joint is reproduced by the joint surface replacement arthroplasty when normal ligament tension was present. The maximum angular displacement of the pyrocarbon implant was 378 for lateral deviation and 338 for rotation during the passive flexion and extension motion. The instantaneus center of rotation (ICR) after implant insertion was nearly identical to the center of rotation of the normal joint. The results also indicated that the collateral ligaments provide the primary stability of the MCP joint. No significant differences in lateral and rotational stability after surface replacement arthroplasty were noted. While collateral ligaments resection significantly affected the stability of the MCP joint. The ICR of the pyrocarbon implant most closely matched that of the intact MCP joint. The pyrocarbon implant provides suitable stability to radio-ulnar deviation and rotational stresses as a resurfacing implant and it simulates the kinematics of the intact MCP joint. By using new materials and taking the anatomical and biomechanical requirements into consideration, the endoprosthesis of the finger joints has created an option to achieve good long-term results. The inadequate results of earlier and current prostheses are a consequence of their mechanical construction and their materials. The success of the new implants could be proven by preferably long-term, controlled studies.</abstract><cop>Germany</cop><pmid>17492561</pmid><doi>10.1055/s-2007-965173</doi><tpages>8</tpages></addata></record>
fulltext fulltext
identifier ISSN: 1864-6697
ispartof Zeitschrift für Orthopädie und Unfallchirurgie, 2007-03, Vol.145 (2), p.199-206
issn 1864-6697
language ger
recordid cdi_proquest_miscellaneous_70486191
source MEDLINE; Thieme Connect Journals
subjects Arthroplasty, Replacement, Finger - instrumentation
Arthroplasty, Replacement, Finger - methods
Awards and Prizes
Biomechanical Phenomena - history
Cadaver
Elasticity
Equipment Design
Equipment Failure Analysis
Fayaz
Germany
History, 21st Century
Humans
In Vitro Techniques
Joint Instability - etiology
Joint Instability - physiopathology
Metacarpophalangeal Joint - physiopathology
Metacarpophalangeal Joint - surgery
Minnesota
Orthopedics - history
Range of Motion, Articular
Stress, Mechanical
Treatment Outcome
title The Arthur Vick Award: kinematics of the metacarpophalangeal joint after surface replacement arthroplasty
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-03T22%3A02%3A59IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=The%20Arthur%20Vick%20Award:%20kinematics%20of%20the%20metacarpophalangeal%20joint%20after%20surface%20replacement%20arthroplasty&rft.jtitle=Zeitschrift%20f%C3%BCr%20Orthop%C3%A4die%20und%20Unfallchirurgie&rft.au=Fayaz,%20H%20C&rft.date=2007-03&rft.volume=145&rft.issue=2&rft.spage=199&rft.epage=206&rft.pages=199-206&rft.issn=1864-6697&rft_id=info:doi/10.1055/s-2007-965173&rft_dat=%3Cproquest_pubme%3E70486191%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=70486191&rft_id=info:pmid/17492561&rfr_iscdi=true