Determination of imipenem and rifampicin in mouse plasma by high performance liquid chromatography–diode array detection
A high performance liquid chromatographic (HPLC) method for the determination of imipenem and rifampicin was developed and validated. The method involves plasma deproteinisation with methanol, gradient elution on a RP-18 column and diode array detection. Separation was carried out in 8 min using a m...
Gespeichert in:
Veröffentlicht in: | Analytica chimica acta 2008-02, Vol.608 (2), p.204-210 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | A high performance liquid chromatographic (HPLC) method for the determination of imipenem and rifampicin was developed and validated. The method involves plasma deproteinisation with methanol, gradient elution on a RP-18 column and diode array detection. Separation was carried out in 8
min using a mobile phase composed of methanol and 0.2
M borate buffer (pH 7.2). Imipenem and rifampicin were detected at 300
nm and 255
nm, respectively. A linear response was observed at plasma levels ranged between 0.3 and 30
μg
mL
−1 for imipenem and 1.5 and 20
μg
mL
−1 for rifampicin. The detection limits were 0.07
μg
mL
−1 and 0.47
μg
mL
−1 for imipenem and rifampicin, respectively. The method was applied to the determination of both compounds in mouse plasma samples. |
---|---|
ISSN: | 0003-2670 1873-4324 |
DOI: | 10.1016/j.aca.2007.12.026 |