Instrument and methods for surface dilatational rheology measurements

We describe an instrument combining the advantages of two methods, axisymmetric drop shape analysis for well-deformed drops and capillary pressure tensiometry for spherical drops, both used for measuring the interfacial tension and interfacial rheological parameters. The rheological parameters are t...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Review of scientific instruments 2008-10, Vol.79 (10), p.104102-104102-10
Hauptverfasser: Russev, Stoyan C., Alexandrov, Nikola, Marinova, Krastanka G., Danov, Krassimir D., Denkov, Nikolai D., Lyutov, Lyudmil, Vulchev, Vassil, Bilke-Krause, Christine
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 104102-10
container_issue 10
container_start_page 104102
container_title Review of scientific instruments
container_volume 79
creator Russev, Stoyan C.
Alexandrov, Nikola
Marinova, Krastanka G.
Danov, Krassimir D.
Denkov, Nikolai D.
Lyutov, Lyudmil
Vulchev, Vassil
Bilke-Krause, Christine
description We describe an instrument combining the advantages of two methods, axisymmetric drop shape analysis for well-deformed drops and capillary pressure tensiometry for spherical drops, both used for measuring the interfacial tension and interfacial rheological parameters. The rheological parameters are the complex interfacial elasticity, and the surface elasticity and viscosity of Kelvin-Voigt and Maxwell rheological models. The instrument is applicable for investigation of the effect of different types of surfactants (nonionic, ionic, proteins, and polymers) on the interfacial rheological properties both of air/water and oil/water interfaces, and of interfaces between liquids with equal mass densities. A piezodriven system and a specially designed interface unit, implemented in the instrument, ensure precise control for standard periodic waveforms of surface deformation (sine, square, triangle, and sawtooth) at a fixed frequency, or produce surface deformation at constant rate. The interface unit ensures accurate synchronization between the pressure measurement and the surface control, which is used for real-time data processing and feedback control of drop area in some of the applications.
doi_str_mv 10.1063/1.3000569
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_69856949</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>69856949</sourcerecordid><originalsourceid>FETCH-LOGICAL-c375t-1fa6706d8b76eff967d2c5e0dab58795f901abd37d4b4a9fa21dbf9616b7ab553</originalsourceid><addsrcrecordid>eNp10EFPwyAYxnFiNG5OD34B05OJh04oFMrFxCxTlyzxomcCBVxNWybQw769zDbZydN7-b3P4Q_ALYJLBCl-REsMISwpPwNzBCueM1rgczCHEJOcMlLNwFUI3_CIELoEM8QhIQwXc7De9CH6oTN9zGSvs87EndMhs85nYfBW1ibTTSujjI3rZZv5nXGt-zokKRMwx89wDS6sbIO5me4CfL6sP1Zv-fb9dbN63uY1ZmXMkZWUQaorxaixllOmi7o0UEtVVoyXlkMklcZME0Ukt7JAWiWGqGKJlHgB7sfdvXc_gwlRdE2oTdvK3rghCMqrVIHwBB9GWHsXgjdW7H3TSX8QCIpjM4HE1CzZu2l0UJ3RJzlFSuBpBKFuxg7_r51yipRT_OXEv2PsfT0</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>69856949</pqid></control><display><type>article</type><title>Instrument and methods for surface dilatational rheology measurements</title><source>AIP Journals Complete</source><source>AIP Digital Archive</source><source>Alma/SFX Local Collection</source><creator>Russev, Stoyan C. ; Alexandrov, Nikola ; Marinova, Krastanka G. ; Danov, Krassimir D. ; Denkov, Nikolai D. ; Lyutov, Lyudmil ; Vulchev, Vassil ; Bilke-Krause, Christine</creator><creatorcontrib>Russev, Stoyan C. ; Alexandrov, Nikola ; Marinova, Krastanka G. ; Danov, Krassimir D. ; Denkov, Nikolai D. ; Lyutov, Lyudmil ; Vulchev, Vassil ; Bilke-Krause, Christine</creatorcontrib><description>We describe an instrument combining the advantages of two methods, axisymmetric drop shape analysis for well-deformed drops and capillary pressure tensiometry for spherical drops, both used for measuring the interfacial tension and interfacial rheological parameters. The rheological parameters are the complex interfacial elasticity, and the surface elasticity and viscosity of Kelvin-Voigt and Maxwell rheological models. The instrument is applicable for investigation of the effect of different types of surfactants (nonionic, ionic, proteins, and polymers) on the interfacial rheological properties both of air/water and oil/water interfaces, and of interfaces between liquids with equal mass densities. A piezodriven system and a specially designed interface unit, implemented in the instrument, ensure precise control for standard periodic waveforms of surface deformation (sine, square, triangle, and sawtooth) at a fixed frequency, or produce surface deformation at constant rate. The interface unit ensures accurate synchronization between the pressure measurement and the surface control, which is used for real-time data processing and feedback control of drop area in some of the applications.</description><identifier>ISSN: 0034-6748</identifier><identifier>EISSN: 1089-7623</identifier><identifier>DOI: 10.1063/1.3000569</identifier><identifier>PMID: 19044732</identifier><identifier>CODEN: RSINAK</identifier><language>eng</language><publisher>United States: American Institute of Physics</publisher><ispartof>Review of scientific instruments, 2008-10, Vol.79 (10), p.104102-104102-10</ispartof><rights>2008 American Institute of Physics</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c375t-1fa6706d8b76eff967d2c5e0dab58795f901abd37d4b4a9fa21dbf9616b7ab553</citedby><cites>FETCH-LOGICAL-c375t-1fa6706d8b76eff967d2c5e0dab58795f901abd37d4b4a9fa21dbf9616b7ab553</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://pubs.aip.org/rsi/article-lookup/doi/10.1063/1.3000569$$EHTML$$P50$$Gscitation$$H</linktohtml><link.rule.ids>314,780,784,794,1559,4512,27924,27925,76384,76390</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/19044732$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Russev, Stoyan C.</creatorcontrib><creatorcontrib>Alexandrov, Nikola</creatorcontrib><creatorcontrib>Marinova, Krastanka G.</creatorcontrib><creatorcontrib>Danov, Krassimir D.</creatorcontrib><creatorcontrib>Denkov, Nikolai D.</creatorcontrib><creatorcontrib>Lyutov, Lyudmil</creatorcontrib><creatorcontrib>Vulchev, Vassil</creatorcontrib><creatorcontrib>Bilke-Krause, Christine</creatorcontrib><title>Instrument and methods for surface dilatational rheology measurements</title><title>Review of scientific instruments</title><addtitle>Rev Sci Instrum</addtitle><description>We describe an instrument combining the advantages of two methods, axisymmetric drop shape analysis for well-deformed drops and capillary pressure tensiometry for spherical drops, both used for measuring the interfacial tension and interfacial rheological parameters. The rheological parameters are the complex interfacial elasticity, and the surface elasticity and viscosity of Kelvin-Voigt and Maxwell rheological models. The instrument is applicable for investigation of the effect of different types of surfactants (nonionic, ionic, proteins, and polymers) on the interfacial rheological properties both of air/water and oil/water interfaces, and of interfaces between liquids with equal mass densities. A piezodriven system and a specially designed interface unit, implemented in the instrument, ensure precise control for standard periodic waveforms of surface deformation (sine, square, triangle, and sawtooth) at a fixed frequency, or produce surface deformation at constant rate. The interface unit ensures accurate synchronization between the pressure measurement and the surface control, which is used for real-time data processing and feedback control of drop area in some of the applications.</description><issn>0034-6748</issn><issn>1089-7623</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2008</creationdate><recordtype>article</recordtype><recordid>eNp10EFPwyAYxnFiNG5OD34B05OJh04oFMrFxCxTlyzxomcCBVxNWybQw769zDbZydN7-b3P4Q_ALYJLBCl-REsMISwpPwNzBCueM1rgczCHEJOcMlLNwFUI3_CIELoEM8QhIQwXc7De9CH6oTN9zGSvs87EndMhs85nYfBW1ibTTSujjI3rZZv5nXGt-zokKRMwx89wDS6sbIO5me4CfL6sP1Zv-fb9dbN63uY1ZmXMkZWUQaorxaixllOmi7o0UEtVVoyXlkMklcZME0Ukt7JAWiWGqGKJlHgB7sfdvXc_gwlRdE2oTdvK3rghCMqrVIHwBB9GWHsXgjdW7H3TSX8QCIpjM4HE1CzZu2l0UJ3RJzlFSuBpBKFuxg7_r51yipRT_OXEv2PsfT0</recordid><startdate>20081001</startdate><enddate>20081001</enddate><creator>Russev, Stoyan C.</creator><creator>Alexandrov, Nikola</creator><creator>Marinova, Krastanka G.</creator><creator>Danov, Krassimir D.</creator><creator>Denkov, Nikolai D.</creator><creator>Lyutov, Lyudmil</creator><creator>Vulchev, Vassil</creator><creator>Bilke-Krause, Christine</creator><general>American Institute of Physics</general><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20081001</creationdate><title>Instrument and methods for surface dilatational rheology measurements</title><author>Russev, Stoyan C. ; Alexandrov, Nikola ; Marinova, Krastanka G. ; Danov, Krassimir D. ; Denkov, Nikolai D. ; Lyutov, Lyudmil ; Vulchev, Vassil ; Bilke-Krause, Christine</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c375t-1fa6706d8b76eff967d2c5e0dab58795f901abd37d4b4a9fa21dbf9616b7ab553</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2008</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Russev, Stoyan C.</creatorcontrib><creatorcontrib>Alexandrov, Nikola</creatorcontrib><creatorcontrib>Marinova, Krastanka G.</creatorcontrib><creatorcontrib>Danov, Krassimir D.</creatorcontrib><creatorcontrib>Denkov, Nikolai D.</creatorcontrib><creatorcontrib>Lyutov, Lyudmil</creatorcontrib><creatorcontrib>Vulchev, Vassil</creatorcontrib><creatorcontrib>Bilke-Krause, Christine</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Review of scientific instruments</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Russev, Stoyan C.</au><au>Alexandrov, Nikola</au><au>Marinova, Krastanka G.</au><au>Danov, Krassimir D.</au><au>Denkov, Nikolai D.</au><au>Lyutov, Lyudmil</au><au>Vulchev, Vassil</au><au>Bilke-Krause, Christine</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Instrument and methods for surface dilatational rheology measurements</atitle><jtitle>Review of scientific instruments</jtitle><addtitle>Rev Sci Instrum</addtitle><date>2008-10-01</date><risdate>2008</risdate><volume>79</volume><issue>10</issue><spage>104102</spage><epage>104102-10</epage><pages>104102-104102-10</pages><issn>0034-6748</issn><eissn>1089-7623</eissn><coden>RSINAK</coden><abstract>We describe an instrument combining the advantages of two methods, axisymmetric drop shape analysis for well-deformed drops and capillary pressure tensiometry for spherical drops, both used for measuring the interfacial tension and interfacial rheological parameters. The rheological parameters are the complex interfacial elasticity, and the surface elasticity and viscosity of Kelvin-Voigt and Maxwell rheological models. The instrument is applicable for investigation of the effect of different types of surfactants (nonionic, ionic, proteins, and polymers) on the interfacial rheological properties both of air/water and oil/water interfaces, and of interfaces between liquids with equal mass densities. A piezodriven system and a specially designed interface unit, implemented in the instrument, ensure precise control for standard periodic waveforms of surface deformation (sine, square, triangle, and sawtooth) at a fixed frequency, or produce surface deformation at constant rate. The interface unit ensures accurate synchronization between the pressure measurement and the surface control, which is used for real-time data processing and feedback control of drop area in some of the applications.</abstract><cop>United States</cop><pub>American Institute of Physics</pub><pmid>19044732</pmid><doi>10.1063/1.3000569</doi><tpages>1</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0034-6748
ispartof Review of scientific instruments, 2008-10, Vol.79 (10), p.104102-104102-10
issn 0034-6748
1089-7623
language eng
recordid cdi_proquest_miscellaneous_69856949
source AIP Journals Complete; AIP Digital Archive; Alma/SFX Local Collection
title Instrument and methods for surface dilatational rheology measurements
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-26T00%3A39%3A04IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Instrument%20and%20methods%20for%20surface%20dilatational%20rheology%20measurements&rft.jtitle=Review%20of%20scientific%20instruments&rft.au=Russev,%20Stoyan%20C.&rft.date=2008-10-01&rft.volume=79&rft.issue=10&rft.spage=104102&rft.epage=104102-10&rft.pages=104102-104102-10&rft.issn=0034-6748&rft.eissn=1089-7623&rft.coden=RSINAK&rft_id=info:doi/10.1063/1.3000569&rft_dat=%3Cproquest_cross%3E69856949%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=69856949&rft_id=info:pmid/19044732&rfr_iscdi=true