Using simulated noise to define optimal QT intervals for computer analysis of ambulatory ECG
The ambulatory electrocardiogram (ECG) is an important medical tool, not only for diagnosis of adverse cardiac events, but also to predict the risk of such events occurring. The 24-hour ambulatory ECG has certain problems and drawbacks because the signal is corrupted by noise from various sources an...
Gespeichert in:
Veröffentlicht in: | Medical engineering & physics 1999-01, Vol.21 (1), p.15-25 |
---|---|
Hauptverfasser: | , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 25 |
---|---|
container_issue | 1 |
container_start_page | 15 |
container_title | Medical engineering & physics |
container_volume | 21 |
creator | Tikkanen, P.E. Sellin, L.C. Kinnunen, H.O. Huikuri, H.V. |
description | The ambulatory electrocardiogram (ECG) is an important medical tool, not only for diagnosis of adverse cardiac events, but also to predict the risk of such events occurring. The 24-hour ambulatory ECG has certain problems and drawbacks because the signal is corrupted by noise from various sources and also several other conditions which may alter the ECG morphology. We have developed a Windows based program for the computer analysis of ambulatory ECG which attempts to address these problems. The software includes options for importing ECG data, different methods of waveform analysis, data-viewing, and exporting the extracted time series. In addition, the modular structure allows for flexible maintenance and expansion of the software. The ECG was recorded using a Holter device and oversampled to enhance the fidelity of the low sampling rate of the ambulatory ECG. The influence of different sampling rates on the interval variability were studied. The noise sensitivity of the implemented algorithm was tested with several types of simulated noise and the precision of the interval measurement was reported with
sd values. Our simulations showed that, in most of the cases, defining the end of QT interval at the maximum of the T wave gave the most precise measurement. The definition of the onset of the ventricular repolarization duration is most precisely made on the maximum or descending maximal slope of the R wave. We also analyzed some examples of time series from patients using power spectrum estimates in order to validate the low level QT interval variability. |
doi_str_mv | 10.1016/S1350-4533(99)00018-1 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_69715218</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S1350453399000181</els_id><sourcerecordid>394362</sourcerecordid><originalsourceid>FETCH-LOGICAL-c421t-9a9b4430fcc3eaf467d9566fb27e519f1bc7bd8fdad2493b1327479f499e6a43</originalsourceid><addsrcrecordid>eNqF0E1rHCEYwHEpDU2a5CO0eCihPUzqo87MeiplyRsEQujmFhBHH4tlZtzqTGC_fdzshvaWkyI_H-VPyCdg58Cg-f4LRM0qWQvxValvjDFYVPCOHMGiFZVkgr0v-1dySD7m_KcgKRvxgRwC45yBEEfk8SGH8TfNYZh7M6GjYwwZ6RSpQx9GpHE9hcH09H5FwzhhejJ9pj4mauOwnssBNaPpNzlkGj01Q7edE9OGXiyvTsiBLxxP9-sxWV1erJbX1e3d1c3y521lJYepUkZ1UgrmrRVovGxap-qm8R1vsQblobNt5xbeGcelEh0I3spWeakUNkaKY3K2G7tO8e-MedJDyBb73owY56wb1ULNYfEm5CBUCaQKrHfQpphzQq_XqVRIGw1Mb_Prl_x621YrpV_yayj3Pu8fmLsB3X-3dr0L-LIHJlvT-2RGG_I_19ZlFC_sx45hqfYUMOlsA44WXUhoJ-1ieOMnz8dIoVM</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>21394639</pqid></control><display><type>article</type><title>Using simulated noise to define optimal QT intervals for computer analysis of ambulatory ECG</title><source>MEDLINE</source><source>Elsevier ScienceDirect Journals</source><creator>Tikkanen, P.E. ; Sellin, L.C. ; Kinnunen, H.O. ; Huikuri, H.V.</creator><creatorcontrib>Tikkanen, P.E. ; Sellin, L.C. ; Kinnunen, H.O. ; Huikuri, H.V.</creatorcontrib><description>The ambulatory electrocardiogram (ECG) is an important medical tool, not only for diagnosis of adverse cardiac events, but also to predict the risk of such events occurring. The 24-hour ambulatory ECG has certain problems and drawbacks because the signal is corrupted by noise from various sources and also several other conditions which may alter the ECG morphology. We have developed a Windows based program for the computer analysis of ambulatory ECG which attempts to address these problems. The software includes options for importing ECG data, different methods of waveform analysis, data-viewing, and exporting the extracted time series. In addition, the modular structure allows for flexible maintenance and expansion of the software. The ECG was recorded using a Holter device and oversampled to enhance the fidelity of the low sampling rate of the ambulatory ECG. The influence of different sampling rates on the interval variability were studied. The noise sensitivity of the implemented algorithm was tested with several types of simulated noise and the precision of the interval measurement was reported with
sd values. Our simulations showed that, in most of the cases, defining the end of QT interval at the maximum of the T wave gave the most precise measurement. The definition of the onset of the ventricular repolarization duration is most precisely made on the maximum or descending maximal slope of the R wave. We also analyzed some examples of time series from patients using power spectrum estimates in order to validate the low level QT interval variability.</description><identifier>ISSN: 1350-4533</identifier><identifier>EISSN: 1873-4030</identifier><identifier>DOI: 10.1016/S1350-4533(99)00018-1</identifier><identifier>PMID: 10220133</identifier><language>eng</language><publisher>Oxford: Elsevier Ltd</publisher><subject>Algorithms ; automatic ECG analysis ; Biological and medical sciences ; Cardiovascular system ; Computer aided diagnosis ; Computer Simulation ; ECG waveform detection ; Electrocardiography, Ambulatory ; Electrocardiography. Vectocardiography ; Electrodiagnosis. Electric activity recording ; Electrophysiology ; Humans ; Investigative techniques, diagnostic techniques (general aspects) ; Long QT Syndrome - diagnosis ; Medical sciences ; Models, Theoretical ; oversampling ; QT interval variability ; Reference Values ; Reproducibility of Results ; Sensitivity and Specificity ; Signal Processing, Computer-Assisted ; simulated noise ; Software ; Software Design ; Spurious signal noise ; timing precision ; Waveform analysis</subject><ispartof>Medical engineering & physics, 1999-01, Vol.21 (1), p.15-25</ispartof><rights>1999 IPEM</rights><rights>1999 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c421t-9a9b4430fcc3eaf467d9566fb27e519f1bc7bd8fdad2493b1327479f499e6a43</citedby><cites>FETCH-LOGICAL-c421t-9a9b4430fcc3eaf467d9566fb27e519f1bc7bd8fdad2493b1327479f499e6a43</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://www.sciencedirect.com/science/article/pii/S1350453399000181$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,776,780,3537,27901,27902,65306</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&idt=1750002$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/10220133$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Tikkanen, P.E.</creatorcontrib><creatorcontrib>Sellin, L.C.</creatorcontrib><creatorcontrib>Kinnunen, H.O.</creatorcontrib><creatorcontrib>Huikuri, H.V.</creatorcontrib><title>Using simulated noise to define optimal QT intervals for computer analysis of ambulatory ECG</title><title>Medical engineering & physics</title><addtitle>Med Eng Phys</addtitle><description>The ambulatory electrocardiogram (ECG) is an important medical tool, not only for diagnosis of adverse cardiac events, but also to predict the risk of such events occurring. The 24-hour ambulatory ECG has certain problems and drawbacks because the signal is corrupted by noise from various sources and also several other conditions which may alter the ECG morphology. We have developed a Windows based program for the computer analysis of ambulatory ECG which attempts to address these problems. The software includes options for importing ECG data, different methods of waveform analysis, data-viewing, and exporting the extracted time series. In addition, the modular structure allows for flexible maintenance and expansion of the software. The ECG was recorded using a Holter device and oversampled to enhance the fidelity of the low sampling rate of the ambulatory ECG. The influence of different sampling rates on the interval variability were studied. The noise sensitivity of the implemented algorithm was tested with several types of simulated noise and the precision of the interval measurement was reported with
sd values. Our simulations showed that, in most of the cases, defining the end of QT interval at the maximum of the T wave gave the most precise measurement. The definition of the onset of the ventricular repolarization duration is most precisely made on the maximum or descending maximal slope of the R wave. We also analyzed some examples of time series from patients using power spectrum estimates in order to validate the low level QT interval variability.</description><subject>Algorithms</subject><subject>automatic ECG analysis</subject><subject>Biological and medical sciences</subject><subject>Cardiovascular system</subject><subject>Computer aided diagnosis</subject><subject>Computer Simulation</subject><subject>ECG waveform detection</subject><subject>Electrocardiography, Ambulatory</subject><subject>Electrocardiography. Vectocardiography</subject><subject>Electrodiagnosis. Electric activity recording</subject><subject>Electrophysiology</subject><subject>Humans</subject><subject>Investigative techniques, diagnostic techniques (general aspects)</subject><subject>Long QT Syndrome - diagnosis</subject><subject>Medical sciences</subject><subject>Models, Theoretical</subject><subject>oversampling</subject><subject>QT interval variability</subject><subject>Reference Values</subject><subject>Reproducibility of Results</subject><subject>Sensitivity and Specificity</subject><subject>Signal Processing, Computer-Assisted</subject><subject>simulated noise</subject><subject>Software</subject><subject>Software Design</subject><subject>Spurious signal noise</subject><subject>timing precision</subject><subject>Waveform analysis</subject><issn>1350-4533</issn><issn>1873-4030</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>1999</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqF0E1rHCEYwHEpDU2a5CO0eCihPUzqo87MeiplyRsEQujmFhBHH4tlZtzqTGC_fdzshvaWkyI_H-VPyCdg58Cg-f4LRM0qWQvxValvjDFYVPCOHMGiFZVkgr0v-1dySD7m_KcgKRvxgRwC45yBEEfk8SGH8TfNYZh7M6GjYwwZ6RSpQx9GpHE9hcH09H5FwzhhejJ9pj4mauOwnssBNaPpNzlkGj01Q7edE9OGXiyvTsiBLxxP9-sxWV1erJbX1e3d1c3y521lJYepUkZ1UgrmrRVovGxap-qm8R1vsQblobNt5xbeGcelEh0I3spWeakUNkaKY3K2G7tO8e-MedJDyBb73owY56wb1ULNYfEm5CBUCaQKrHfQpphzQq_XqVRIGw1Mb_Prl_x621YrpV_yayj3Pu8fmLsB3X-3dr0L-LIHJlvT-2RGG_I_19ZlFC_sx45hqfYUMOlsA44WXUhoJ-1ieOMnz8dIoVM</recordid><startdate>19990101</startdate><enddate>19990101</enddate><creator>Tikkanen, P.E.</creator><creator>Sellin, L.C.</creator><creator>Kinnunen, H.O.</creator><creator>Huikuri, H.V.</creator><general>Elsevier Ltd</general><general>Elsevier Science</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>19990101</creationdate><title>Using simulated noise to define optimal QT intervals for computer analysis of ambulatory ECG</title><author>Tikkanen, P.E. ; Sellin, L.C. ; Kinnunen, H.O. ; Huikuri, H.V.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c421t-9a9b4430fcc3eaf467d9566fb27e519f1bc7bd8fdad2493b1327479f499e6a43</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>1999</creationdate><topic>Algorithms</topic><topic>automatic ECG analysis</topic><topic>Biological and medical sciences</topic><topic>Cardiovascular system</topic><topic>Computer aided diagnosis</topic><topic>Computer Simulation</topic><topic>ECG waveform detection</topic><topic>Electrocardiography, Ambulatory</topic><topic>Electrocardiography. Vectocardiography</topic><topic>Electrodiagnosis. Electric activity recording</topic><topic>Electrophysiology</topic><topic>Humans</topic><topic>Investigative techniques, diagnostic techniques (general aspects)</topic><topic>Long QT Syndrome - diagnosis</topic><topic>Medical sciences</topic><topic>Models, Theoretical</topic><topic>oversampling</topic><topic>QT interval variability</topic><topic>Reference Values</topic><topic>Reproducibility of Results</topic><topic>Sensitivity and Specificity</topic><topic>Signal Processing, Computer-Assisted</topic><topic>simulated noise</topic><topic>Software</topic><topic>Software Design</topic><topic>Spurious signal noise</topic><topic>timing precision</topic><topic>Waveform analysis</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Tikkanen, P.E.</creatorcontrib><creatorcontrib>Sellin, L.C.</creatorcontrib><creatorcontrib>Kinnunen, H.O.</creatorcontrib><creatorcontrib>Huikuri, H.V.</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Medical engineering & physics</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Tikkanen, P.E.</au><au>Sellin, L.C.</au><au>Kinnunen, H.O.</au><au>Huikuri, H.V.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Using simulated noise to define optimal QT intervals for computer analysis of ambulatory ECG</atitle><jtitle>Medical engineering & physics</jtitle><addtitle>Med Eng Phys</addtitle><date>1999-01-01</date><risdate>1999</risdate><volume>21</volume><issue>1</issue><spage>15</spage><epage>25</epage><pages>15-25</pages><issn>1350-4533</issn><eissn>1873-4030</eissn><abstract>The ambulatory electrocardiogram (ECG) is an important medical tool, not only for diagnosis of adverse cardiac events, but also to predict the risk of such events occurring. The 24-hour ambulatory ECG has certain problems and drawbacks because the signal is corrupted by noise from various sources and also several other conditions which may alter the ECG morphology. We have developed a Windows based program for the computer analysis of ambulatory ECG which attempts to address these problems. The software includes options for importing ECG data, different methods of waveform analysis, data-viewing, and exporting the extracted time series. In addition, the modular structure allows for flexible maintenance and expansion of the software. The ECG was recorded using a Holter device and oversampled to enhance the fidelity of the low sampling rate of the ambulatory ECG. The influence of different sampling rates on the interval variability were studied. The noise sensitivity of the implemented algorithm was tested with several types of simulated noise and the precision of the interval measurement was reported with
sd values. Our simulations showed that, in most of the cases, defining the end of QT interval at the maximum of the T wave gave the most precise measurement. The definition of the onset of the ventricular repolarization duration is most precisely made on the maximum or descending maximal slope of the R wave. We also analyzed some examples of time series from patients using power spectrum estimates in order to validate the low level QT interval variability.</abstract><cop>Oxford</cop><pub>Elsevier Ltd</pub><pmid>10220133</pmid><doi>10.1016/S1350-4533(99)00018-1</doi><tpages>11</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1350-4533 |
ispartof | Medical engineering & physics, 1999-01, Vol.21 (1), p.15-25 |
issn | 1350-4533 1873-4030 |
language | eng |
recordid | cdi_proquest_miscellaneous_69715218 |
source | MEDLINE; Elsevier ScienceDirect Journals |
subjects | Algorithms automatic ECG analysis Biological and medical sciences Cardiovascular system Computer aided diagnosis Computer Simulation ECG waveform detection Electrocardiography, Ambulatory Electrocardiography. Vectocardiography Electrodiagnosis. Electric activity recording Electrophysiology Humans Investigative techniques, diagnostic techniques (general aspects) Long QT Syndrome - diagnosis Medical sciences Models, Theoretical oversampling QT interval variability Reference Values Reproducibility of Results Sensitivity and Specificity Signal Processing, Computer-Assisted simulated noise Software Software Design Spurious signal noise timing precision Waveform analysis |
title | Using simulated noise to define optimal QT intervals for computer analysis of ambulatory ECG |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-31T21%3A06%3A13IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Using%20simulated%20noise%20to%20define%20optimal%20QT%20intervals%20for%20computer%20analysis%20of%20ambulatory%20ECG&rft.jtitle=Medical%20engineering%20&%20physics&rft.au=Tikkanen,%20P.E.&rft.date=1999-01-01&rft.volume=21&rft.issue=1&rft.spage=15&rft.epage=25&rft.pages=15-25&rft.issn=1350-4533&rft.eissn=1873-4030&rft_id=info:doi/10.1016/S1350-4533(99)00018-1&rft_dat=%3Cproquest_cross%3E394362%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=21394639&rft_id=info:pmid/10220133&rft_els_id=S1350453399000181&rfr_iscdi=true |