A Complication of Pacemaker Lead Extraction: Paradoxical Embolism of a Lead Fragment in a Leg Artery
Extraction of three chronically implanted pacemaker leads was performed via the implant vein in a 71‐year‐old man. One of the leads broke in the subclavian scar tissue. The lead fragment migrated into the left peroneotibial trunk artery. Transesophageal echocardiography showed patent foramen ovale a...
Gespeichert in:
Veröffentlicht in: | Pacing and clinical electrophysiology 1998-12, Vol.21 (12), p.2699-2700 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Extraction of three chronically implanted pacemaker leads was performed via the implant vein in a 71‐year‐old man. One of the leads broke in the subclavian scar tissue. The lead fragment migrated into the left peroneotibial trunk artery. Transesophageal echocardiography showed patent foramen ovale associated with right‐to‐left atrial shunt. This complication of lead extraction is discussed along with the role of echocardiography prior to lead removal. |
---|---|
ISSN: | 0147-8389 1540-8159 |
DOI: | 10.1111/j.1540-8159.1998.tb00053.x |