Occasional Single Beat Regurgitation Observed with the Medtronic Advantage Bileaflet Heart Valve
The purpose of this clinical study was to obtain further evidence of the underlying mechanism causing the echocardiographically detected phenomenon of single beat regurgitation in a new bileaflet heart valve. As part of a prospective multicenter trial at our institution, 63 patients received the Adv...
Gespeichert in:
Veröffentlicht in: | The Annals of thoracic surgery 2006-08, Vol.82 (2), p.537-541 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 541 |
---|---|
container_issue | 2 |
container_start_page | 537 |
container_title | The Annals of thoracic surgery |
container_volume | 82 |
creator | Eichinger, Walter B. Wagner, Ina M. Bleiziffer, Sabine von Canal, Friederike Günzinger, Ralf Ruzicka, Daniel J. Busch, Ulrich Bauernschmitt, Robert Lange, Ruediger |
description | The purpose of this clinical study was to obtain further evidence of the underlying mechanism causing the echocardiographically detected phenomenon of single beat regurgitation in a new bileaflet heart valve. As part of a prospective multicenter trial at our institution, 63 patients received the Advantage bileaflet mechanical heart valve (Medtronic, Minneapolis, Minnesota) in aortic position. During routine follow-up performed at discharge and annually after the operation, intermittent moderate transvalvular regurgitation was detected by echocardiography in 5 patients.
Fluoroscopy of leaflet motion (n = 4), invasive blood pressure measurements in the ascending aorta (n = 3) and digital phonocardiography (n = 5) was obtained in the patients showing an intermittent regurgitation during echocardiography.
Valve thrombosis, sutures, or pannus ingrowth impairing valve closure was not detected. Fluoroscopy of leaflet motion showed intermittent incomplete closure of either one of the two leaflets in the same prosthesis. This could be correlated with a distinct diastolic blood pressure drop in the same cardiac cycle. Digital phonocardiography showed pathologic closure sounds in those cycles in which echocardiographically the intermittent regurgitation was observed.
Some patients with the Medtronic Advantage prosthesis in the aortic position show an intermittent inability of complete valve closure that leads to a single beat transvalvular regurgitation. As thrombotic or other material that might cause a disturbance of leaflet motion could not be detected, and the patients seem not to be exposed to any risk except for some chronic regurgitant volume, we decided not to replace the prostheses. |
doi_str_mv | 10.1016/j.athoracsur.2006.03.043 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_68672019</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0003497506005893</els_id><sourcerecordid>68672019</sourcerecordid><originalsourceid>FETCH-LOGICAL-c407t-2bd361c608dd21e0357a19eac25236931a6556d198c93420ac2a23a0431559433</originalsourceid><addsrcrecordid>eNqFkM1u2zAQhImiReKmeYWCp9ykLkmRso5J0CQFUhjo35Vdk2ubhiwlJOWgb1-mNuBjT4vFfLuDGca4gFqAMJ-2NebNGNGlKdYSwNSgamjUGzYTWsvKSN29ZTMAUFXTtfqcvU9pW1ZZ5DN2LszcqFa3M_Z74RymMA7Y8-9hWPfEbwgz_0brKa5Dxlw0vlgminvy_CXkDc8b4l_J5zgOwfFrv8ch47ochp5w1VPmD4Qx81_Y7-kDe7fCPtHlcV6wn3eff9w-VI-L-y-314-Va6DNlVx6ZYQzMPdeCgKlWxQdoZNaKtMpgUZr40U3d51qJBQBpcKSuSTqGqUu2NXh71McnydK2e5CctT3ONA4JVsStxJEV8D5AXRxTCnSyj7FsMP4xwqwr-3arT21a1_btaAs_PP4ePSYljvyp8NjnQW4OQBUku4DRZtcoMGRD5Fctn4M_3f5C5MBkD0</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>68672019</pqid></control><display><type>article</type><title>Occasional Single Beat Regurgitation Observed with the Medtronic Advantage Bileaflet Heart Valve</title><source>MEDLINE</source><source>Alma/SFX Local Collection</source><source>EZB Electronic Journals Library</source><creator>Eichinger, Walter B. ; Wagner, Ina M. ; Bleiziffer, Sabine ; von Canal, Friederike ; Günzinger, Ralf ; Ruzicka, Daniel J. ; Busch, Ulrich ; Bauernschmitt, Robert ; Lange, Ruediger</creator><creatorcontrib>Eichinger, Walter B. ; Wagner, Ina M. ; Bleiziffer, Sabine ; von Canal, Friederike ; Günzinger, Ralf ; Ruzicka, Daniel J. ; Busch, Ulrich ; Bauernschmitt, Robert ; Lange, Ruediger</creatorcontrib><description>The purpose of this clinical study was to obtain further evidence of the underlying mechanism causing the echocardiographically detected phenomenon of single beat regurgitation in a new bileaflet heart valve. As part of a prospective multicenter trial at our institution, 63 patients received the Advantage bileaflet mechanical heart valve (Medtronic, Minneapolis, Minnesota) in aortic position. During routine follow-up performed at discharge and annually after the operation, intermittent moderate transvalvular regurgitation was detected by echocardiography in 5 patients.
Fluoroscopy of leaflet motion (n = 4), invasive blood pressure measurements in the ascending aorta (n = 3) and digital phonocardiography (n = 5) was obtained in the patients showing an intermittent regurgitation during echocardiography.
Valve thrombosis, sutures, or pannus ingrowth impairing valve closure was not detected. Fluoroscopy of leaflet motion showed intermittent incomplete closure of either one of the two leaflets in the same prosthesis. This could be correlated with a distinct diastolic blood pressure drop in the same cardiac cycle. Digital phonocardiography showed pathologic closure sounds in those cycles in which echocardiographically the intermittent regurgitation was observed.
Some patients with the Medtronic Advantage prosthesis in the aortic position show an intermittent inability of complete valve closure that leads to a single beat transvalvular regurgitation. As thrombotic or other material that might cause a disturbance of leaflet motion could not be detected, and the patients seem not to be exposed to any risk except for some chronic regurgitant volume, we decided not to replace the prostheses.</description><identifier>ISSN: 0003-4975</identifier><identifier>EISSN: 1552-6259</identifier><identifier>DOI: 10.1016/j.athoracsur.2006.03.043</identifier><identifier>PMID: 16863757</identifier><language>eng</language><publisher>Netherlands: Elsevier Inc</publisher><subject>Aortic Valve Insufficiency - diagnostic imaging ; Aortic Valve Insufficiency - etiology ; Blood Pressure Determination ; Fluoroscopy ; Heart Valve Prosthesis - adverse effects ; Humans ; Phonocardiography ; Prospective Studies ; Ultrasonography</subject><ispartof>The Annals of thoracic surgery, 2006-08, Vol.82 (2), p.537-541</ispartof><rights>2006 The Society of Thoracic Surgeons</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c407t-2bd361c608dd21e0357a19eac25236931a6556d198c93420ac2a23a0431559433</citedby><cites>FETCH-LOGICAL-c407t-2bd361c608dd21e0357a19eac25236931a6556d198c93420ac2a23a0431559433</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27924,27925</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/16863757$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Eichinger, Walter B.</creatorcontrib><creatorcontrib>Wagner, Ina M.</creatorcontrib><creatorcontrib>Bleiziffer, Sabine</creatorcontrib><creatorcontrib>von Canal, Friederike</creatorcontrib><creatorcontrib>Günzinger, Ralf</creatorcontrib><creatorcontrib>Ruzicka, Daniel J.</creatorcontrib><creatorcontrib>Busch, Ulrich</creatorcontrib><creatorcontrib>Bauernschmitt, Robert</creatorcontrib><creatorcontrib>Lange, Ruediger</creatorcontrib><title>Occasional Single Beat Regurgitation Observed with the Medtronic Advantage Bileaflet Heart Valve</title><title>The Annals of thoracic surgery</title><addtitle>Ann Thorac Surg</addtitle><description>The purpose of this clinical study was to obtain further evidence of the underlying mechanism causing the echocardiographically detected phenomenon of single beat regurgitation in a new bileaflet heart valve. As part of a prospective multicenter trial at our institution, 63 patients received the Advantage bileaflet mechanical heart valve (Medtronic, Minneapolis, Minnesota) in aortic position. During routine follow-up performed at discharge and annually after the operation, intermittent moderate transvalvular regurgitation was detected by echocardiography in 5 patients.
Fluoroscopy of leaflet motion (n = 4), invasive blood pressure measurements in the ascending aorta (n = 3) and digital phonocardiography (n = 5) was obtained in the patients showing an intermittent regurgitation during echocardiography.
Valve thrombosis, sutures, or pannus ingrowth impairing valve closure was not detected. Fluoroscopy of leaflet motion showed intermittent incomplete closure of either one of the two leaflets in the same prosthesis. This could be correlated with a distinct diastolic blood pressure drop in the same cardiac cycle. Digital phonocardiography showed pathologic closure sounds in those cycles in which echocardiographically the intermittent regurgitation was observed.
Some patients with the Medtronic Advantage prosthesis in the aortic position show an intermittent inability of complete valve closure that leads to a single beat transvalvular regurgitation. As thrombotic or other material that might cause a disturbance of leaflet motion could not be detected, and the patients seem not to be exposed to any risk except for some chronic regurgitant volume, we decided not to replace the prostheses.</description><subject>Aortic Valve Insufficiency - diagnostic imaging</subject><subject>Aortic Valve Insufficiency - etiology</subject><subject>Blood Pressure Determination</subject><subject>Fluoroscopy</subject><subject>Heart Valve Prosthesis - adverse effects</subject><subject>Humans</subject><subject>Phonocardiography</subject><subject>Prospective Studies</subject><subject>Ultrasonography</subject><issn>0003-4975</issn><issn>1552-6259</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2006</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqFkM1u2zAQhImiReKmeYWCp9ykLkmRso5J0CQFUhjo35Vdk2ubhiwlJOWgb1-mNuBjT4vFfLuDGca4gFqAMJ-2NebNGNGlKdYSwNSgamjUGzYTWsvKSN29ZTMAUFXTtfqcvU9pW1ZZ5DN2LszcqFa3M_Z74RymMA7Y8-9hWPfEbwgz_0brKa5Dxlw0vlgminvy_CXkDc8b4l_J5zgOwfFrv8ch47ochp5w1VPmD4Qx81_Y7-kDe7fCPtHlcV6wn3eff9w-VI-L-y-314-Va6DNlVx6ZYQzMPdeCgKlWxQdoZNaKtMpgUZr40U3d51qJBQBpcKSuSTqGqUu2NXh71McnydK2e5CctT3ONA4JVsStxJEV8D5AXRxTCnSyj7FsMP4xwqwr-3arT21a1_btaAs_PP4ePSYljvyp8NjnQW4OQBUku4DRZtcoMGRD5Fctn4M_3f5C5MBkD0</recordid><startdate>20060801</startdate><enddate>20060801</enddate><creator>Eichinger, Walter B.</creator><creator>Wagner, Ina M.</creator><creator>Bleiziffer, Sabine</creator><creator>von Canal, Friederike</creator><creator>Günzinger, Ralf</creator><creator>Ruzicka, Daniel J.</creator><creator>Busch, Ulrich</creator><creator>Bauernschmitt, Robert</creator><creator>Lange, Ruediger</creator><general>Elsevier Inc</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20060801</creationdate><title>Occasional Single Beat Regurgitation Observed with the Medtronic Advantage Bileaflet Heart Valve</title><author>Eichinger, Walter B. ; Wagner, Ina M. ; Bleiziffer, Sabine ; von Canal, Friederike ; Günzinger, Ralf ; Ruzicka, Daniel J. ; Busch, Ulrich ; Bauernschmitt, Robert ; Lange, Ruediger</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c407t-2bd361c608dd21e0357a19eac25236931a6556d198c93420ac2a23a0431559433</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2006</creationdate><topic>Aortic Valve Insufficiency - diagnostic imaging</topic><topic>Aortic Valve Insufficiency - etiology</topic><topic>Blood Pressure Determination</topic><topic>Fluoroscopy</topic><topic>Heart Valve Prosthesis - adverse effects</topic><topic>Humans</topic><topic>Phonocardiography</topic><topic>Prospective Studies</topic><topic>Ultrasonography</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Eichinger, Walter B.</creatorcontrib><creatorcontrib>Wagner, Ina M.</creatorcontrib><creatorcontrib>Bleiziffer, Sabine</creatorcontrib><creatorcontrib>von Canal, Friederike</creatorcontrib><creatorcontrib>Günzinger, Ralf</creatorcontrib><creatorcontrib>Ruzicka, Daniel J.</creatorcontrib><creatorcontrib>Busch, Ulrich</creatorcontrib><creatorcontrib>Bauernschmitt, Robert</creatorcontrib><creatorcontrib>Lange, Ruediger</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>The Annals of thoracic surgery</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Eichinger, Walter B.</au><au>Wagner, Ina M.</au><au>Bleiziffer, Sabine</au><au>von Canal, Friederike</au><au>Günzinger, Ralf</au><au>Ruzicka, Daniel J.</au><au>Busch, Ulrich</au><au>Bauernschmitt, Robert</au><au>Lange, Ruediger</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Occasional Single Beat Regurgitation Observed with the Medtronic Advantage Bileaflet Heart Valve</atitle><jtitle>The Annals of thoracic surgery</jtitle><addtitle>Ann Thorac Surg</addtitle><date>2006-08-01</date><risdate>2006</risdate><volume>82</volume><issue>2</issue><spage>537</spage><epage>541</epage><pages>537-541</pages><issn>0003-4975</issn><eissn>1552-6259</eissn><abstract>The purpose of this clinical study was to obtain further evidence of the underlying mechanism causing the echocardiographically detected phenomenon of single beat regurgitation in a new bileaflet heart valve. As part of a prospective multicenter trial at our institution, 63 patients received the Advantage bileaflet mechanical heart valve (Medtronic, Minneapolis, Minnesota) in aortic position. During routine follow-up performed at discharge and annually after the operation, intermittent moderate transvalvular regurgitation was detected by echocardiography in 5 patients.
Fluoroscopy of leaflet motion (n = 4), invasive blood pressure measurements in the ascending aorta (n = 3) and digital phonocardiography (n = 5) was obtained in the patients showing an intermittent regurgitation during echocardiography.
Valve thrombosis, sutures, or pannus ingrowth impairing valve closure was not detected. Fluoroscopy of leaflet motion showed intermittent incomplete closure of either one of the two leaflets in the same prosthesis. This could be correlated with a distinct diastolic blood pressure drop in the same cardiac cycle. Digital phonocardiography showed pathologic closure sounds in those cycles in which echocardiographically the intermittent regurgitation was observed.
Some patients with the Medtronic Advantage prosthesis in the aortic position show an intermittent inability of complete valve closure that leads to a single beat transvalvular regurgitation. As thrombotic or other material that might cause a disturbance of leaflet motion could not be detected, and the patients seem not to be exposed to any risk except for some chronic regurgitant volume, we decided not to replace the prostheses.</abstract><cop>Netherlands</cop><pub>Elsevier Inc</pub><pmid>16863757</pmid><doi>10.1016/j.athoracsur.2006.03.043</doi><tpages>5</tpages></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0003-4975 |
ispartof | The Annals of thoracic surgery, 2006-08, Vol.82 (2), p.537-541 |
issn | 0003-4975 1552-6259 |
language | eng |
recordid | cdi_proquest_miscellaneous_68672019 |
source | MEDLINE; Alma/SFX Local Collection; EZB Electronic Journals Library |
subjects | Aortic Valve Insufficiency - diagnostic imaging Aortic Valve Insufficiency - etiology Blood Pressure Determination Fluoroscopy Heart Valve Prosthesis - adverse effects Humans Phonocardiography Prospective Studies Ultrasonography |
title | Occasional Single Beat Regurgitation Observed with the Medtronic Advantage Bileaflet Heart Valve |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-03T19%3A53%3A06IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Occasional%20Single%20Beat%20Regurgitation%20Observed%20with%20the%20Medtronic%20Advantage%20Bileaflet%20Heart%20Valve&rft.jtitle=The%20Annals%20of%20thoracic%20surgery&rft.au=Eichinger,%20Walter%20B.&rft.date=2006-08-01&rft.volume=82&rft.issue=2&rft.spage=537&rft.epage=541&rft.pages=537-541&rft.issn=0003-4975&rft.eissn=1552-6259&rft_id=info:doi/10.1016/j.athoracsur.2006.03.043&rft_dat=%3Cproquest_cross%3E68672019%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=68672019&rft_id=info:pmid/16863757&rft_els_id=S0003497506005893&rfr_iscdi=true |