Construction and evaluation of a flow injection micro-analyser based on urethane–acrylate resin

A flow injection micro-analyser with an integrated injection device and photometric detection is described. Channels measuring 205–295 μm depth by 265–290 μm maximum width were manufactured by deep UV lithography on two layers of urethane–acrylate oligomers-based photoresist. Hypodermic syringe need...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytica chimica acta 2007-11, Vol.603 (2), p.159-166
Hauptverfasser: Fonseca, Alexandre, Raimundo, Ivo M., Rohwedder, Jarbas J.R., Ferreira, Luiz Otávio S.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 166
container_issue 2
container_start_page 159
container_title Analytica chimica acta
container_volume 603
creator Fonseca, Alexandre
Raimundo, Ivo M.
Rohwedder, Jarbas J.R.
Ferreira, Luiz Otávio S.
description A flow injection micro-analyser with an integrated injection device and photometric detection is described. Channels measuring 205–295 μm depth by 265–290 μm maximum width were manufactured by deep UV lithography on two layers of urethane–acrylate oligomers-based photoresist. Hypodermic syringe needles (450 μm diameter) were connected to the channels for introduction of solutions into the system. Plastic optical fibres were connected to the ends of a 5.0 mm long channel, in order to conduct the light from and to a homemade photometer. The device has a total volume of 7.0 μL and three different sample volumes (0.09, 0.22 and 0.30 μL) can be inserted into the system by choosing the appropriate loop of the hydrodynamic injection approach. The micro-analyser, designed as a single line manifold, was evaluated by determining chloride in waters (mercuric thiocyanate method), and chromium (VI) in wastewater and total chromium in metallic alloys (diphenylcarbazide method). For chloride determination two micro-pumps were employed to impel the solutions, while for chromium determination this task was performed by a conventional peristaltic pump. The results obtained in all determinations did not differ significantly from the reference methods at a confidence level of 95%. In the chloride determination, a flow rate of 50 μL min −1 was used, providing a sample frequency of 45 injection h −1, generating ca. 0.7 mg of Hg(II) after an 8-h working day (ca. 20 mL of solution). This result suggests the potential of the micro-analyser towards the reduction of waste, following the philosophy of Green Chemistry.
doi_str_mv 10.1016/j.aca.2007.09.051
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_68444761</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0003267007016121</els_id><sourcerecordid>14864613</sourcerecordid><originalsourceid>FETCH-LOGICAL-c443t-d1cdca75f0a86083c8db3728382beaf747b0b58f6f0da143428222be14312a773</originalsourceid><addsrcrecordid>eNqFkc1u1DAQxy0EotvCA3BBucAtYfwR2xEntIKCVIkLnK2JP4RXWafYSdHe-g68IU-Cl12pN3qyx_7NX6P5EfKKQkeByne7Di12DEB1MHTQ0ydkQ7XireBMPCUbAOAtkwouyGUpu1oyCuI5uaBqkFxzuSG4nVNZ8mqXOKcGk2v8HU4r_ivn0GATpvlXE9POn5B9tHluMeF0KD43Ixbvmvq-Zr_8wOT_3P9Gmw8TLr7JvsT0gjwLOBX_8nxeke-fPn7bfm5vvl5_2X64aa0QfGkdtc6i6gOglqC51W7kimmu2egxKKFGGHsdZACHVHDBNGP1q14pQ6X4FXl7yr3N88_Vl8XsY7F-mupQ81qM1EIIJemjIIehklI8ClKhpaiBFaQnsK6mlOyDuc1xj_lgKJijKbMz1ZQ5mjIwmGqq9rw-h6_j3ruHjrOaCrw5A1gsTiFjsrE8cAOjfc-PQe9PnK_LvYs-m2KjT9a7mKsz4-b4nzH-AtaHsbY</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>14864613</pqid></control><display><type>article</type><title>Construction and evaluation of a flow injection micro-analyser based on urethane–acrylate resin</title><source>Elsevier ScienceDirect Journals</source><creator>Fonseca, Alexandre ; Raimundo, Ivo M. ; Rohwedder, Jarbas J.R. ; Ferreira, Luiz Otávio S.</creator><creatorcontrib>Fonseca, Alexandre ; Raimundo, Ivo M. ; Rohwedder, Jarbas J.R. ; Ferreira, Luiz Otávio S.</creatorcontrib><description>A flow injection micro-analyser with an integrated injection device and photometric detection is described. Channels measuring 205–295 μm depth by 265–290 μm maximum width were manufactured by deep UV lithography on two layers of urethane–acrylate oligomers-based photoresist. Hypodermic syringe needles (450 μm diameter) were connected to the channels for introduction of solutions into the system. Plastic optical fibres were connected to the ends of a 5.0 mm long channel, in order to conduct the light from and to a homemade photometer. The device has a total volume of 7.0 μL and three different sample volumes (0.09, 0.22 and 0.30 μL) can be inserted into the system by choosing the appropriate loop of the hydrodynamic injection approach. The micro-analyser, designed as a single line manifold, was evaluated by determining chloride in waters (mercuric thiocyanate method), and chromium (VI) in wastewater and total chromium in metallic alloys (diphenylcarbazide method). For chloride determination two micro-pumps were employed to impel the solutions, while for chromium determination this task was performed by a conventional peristaltic pump. The results obtained in all determinations did not differ significantly from the reference methods at a confidence level of 95%. In the chloride determination, a flow rate of 50 μL min −1 was used, providing a sample frequency of 45 injection h −1, generating ca. 0.7 mg of Hg(II) after an 8-h working day (ca. 20 mL of solution). This result suggests the potential of the micro-analyser towards the reduction of waste, following the philosophy of Green Chemistry.</description><identifier>ISSN: 0003-2670</identifier><identifier>EISSN: 1873-4324</identifier><identifier>DOI: 10.1016/j.aca.2007.09.051</identifier><identifier>PMID: 17963836</identifier><identifier>CODEN: ACACAM</identifier><language>eng</language><publisher>Amsterdam: Elsevier B.V</publisher><subject>Analysis methods ; Analytical chemistry ; Applied sciences ; Chemistry ; Deep ultraviolet lithography ; Exact sciences and technology ; Flow analysis ; General, instrumentation ; Lab-on-a-chip ; Micro-FIA ; Microfluidic devices ; Natural water pollution ; Pollution ; Water treatment and pollution</subject><ispartof>Analytica chimica acta, 2007-11, Vol.603 (2), p.159-166</ispartof><rights>2007 Elsevier B.V.</rights><rights>2008 INIST-CNRS</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c443t-d1cdca75f0a86083c8db3728382beaf747b0b58f6f0da143428222be14312a773</citedby><cites>FETCH-LOGICAL-c443t-d1cdca75f0a86083c8db3728382beaf747b0b58f6f0da143428222be14312a773</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/j.aca.2007.09.051$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>314,778,782,3539,27907,27908,45978</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&amp;idt=19215531$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/17963836$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Fonseca, Alexandre</creatorcontrib><creatorcontrib>Raimundo, Ivo M.</creatorcontrib><creatorcontrib>Rohwedder, Jarbas J.R.</creatorcontrib><creatorcontrib>Ferreira, Luiz Otávio S.</creatorcontrib><title>Construction and evaluation of a flow injection micro-analyser based on urethane–acrylate resin</title><title>Analytica chimica acta</title><addtitle>Anal Chim Acta</addtitle><description>A flow injection micro-analyser with an integrated injection device and photometric detection is described. Channels measuring 205–295 μm depth by 265–290 μm maximum width were manufactured by deep UV lithography on two layers of urethane–acrylate oligomers-based photoresist. Hypodermic syringe needles (450 μm diameter) were connected to the channels for introduction of solutions into the system. Plastic optical fibres were connected to the ends of a 5.0 mm long channel, in order to conduct the light from and to a homemade photometer. The device has a total volume of 7.0 μL and three different sample volumes (0.09, 0.22 and 0.30 μL) can be inserted into the system by choosing the appropriate loop of the hydrodynamic injection approach. The micro-analyser, designed as a single line manifold, was evaluated by determining chloride in waters (mercuric thiocyanate method), and chromium (VI) in wastewater and total chromium in metallic alloys (diphenylcarbazide method). For chloride determination two micro-pumps were employed to impel the solutions, while for chromium determination this task was performed by a conventional peristaltic pump. The results obtained in all determinations did not differ significantly from the reference methods at a confidence level of 95%. In the chloride determination, a flow rate of 50 μL min −1 was used, providing a sample frequency of 45 injection h −1, generating ca. 0.7 mg of Hg(II) after an 8-h working day (ca. 20 mL of solution). This result suggests the potential of the micro-analyser towards the reduction of waste, following the philosophy of Green Chemistry.</description><subject>Analysis methods</subject><subject>Analytical chemistry</subject><subject>Applied sciences</subject><subject>Chemistry</subject><subject>Deep ultraviolet lithography</subject><subject>Exact sciences and technology</subject><subject>Flow analysis</subject><subject>General, instrumentation</subject><subject>Lab-on-a-chip</subject><subject>Micro-FIA</subject><subject>Microfluidic devices</subject><subject>Natural water pollution</subject><subject>Pollution</subject><subject>Water treatment and pollution</subject><issn>0003-2670</issn><issn>1873-4324</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2007</creationdate><recordtype>article</recordtype><recordid>eNqFkc1u1DAQxy0EotvCA3BBucAtYfwR2xEntIKCVIkLnK2JP4RXWafYSdHe-g68IU-Cl12pN3qyx_7NX6P5EfKKQkeByne7Di12DEB1MHTQ0ydkQ7XireBMPCUbAOAtkwouyGUpu1oyCuI5uaBqkFxzuSG4nVNZ8mqXOKcGk2v8HU4r_ivn0GATpvlXE9POn5B9tHluMeF0KD43Ixbvmvq-Zr_8wOT_3P9Gmw8TLr7JvsT0gjwLOBX_8nxeke-fPn7bfm5vvl5_2X64aa0QfGkdtc6i6gOglqC51W7kimmu2egxKKFGGHsdZACHVHDBNGP1q14pQ6X4FXl7yr3N88_Vl8XsY7F-mupQ81qM1EIIJemjIIehklI8ClKhpaiBFaQnsK6mlOyDuc1xj_lgKJijKbMz1ZQ5mjIwmGqq9rw-h6_j3ruHjrOaCrw5A1gsTiFjsrE8cAOjfc-PQe9PnK_LvYs-m2KjT9a7mKsz4-b4nzH-AtaHsbY</recordid><startdate>20071112</startdate><enddate>20071112</enddate><creator>Fonseca, Alexandre</creator><creator>Raimundo, Ivo M.</creator><creator>Rohwedder, Jarbas J.R.</creator><creator>Ferreira, Luiz Otávio S.</creator><general>Elsevier B.V</general><general>Elsevier</general><scope>IQODW</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7ST</scope><scope>C1K</scope><scope>SOI</scope><scope>7U5</scope><scope>8FD</scope><scope>L7M</scope><scope>7X8</scope></search><sort><creationdate>20071112</creationdate><title>Construction and evaluation of a flow injection micro-analyser based on urethane–acrylate resin</title><author>Fonseca, Alexandre ; Raimundo, Ivo M. ; Rohwedder, Jarbas J.R. ; Ferreira, Luiz Otávio S.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c443t-d1cdca75f0a86083c8db3728382beaf747b0b58f6f0da143428222be14312a773</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2007</creationdate><topic>Analysis methods</topic><topic>Analytical chemistry</topic><topic>Applied sciences</topic><topic>Chemistry</topic><topic>Deep ultraviolet lithography</topic><topic>Exact sciences and technology</topic><topic>Flow analysis</topic><topic>General, instrumentation</topic><topic>Lab-on-a-chip</topic><topic>Micro-FIA</topic><topic>Microfluidic devices</topic><topic>Natural water pollution</topic><topic>Pollution</topic><topic>Water treatment and pollution</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Fonseca, Alexandre</creatorcontrib><creatorcontrib>Raimundo, Ivo M.</creatorcontrib><creatorcontrib>Rohwedder, Jarbas J.R.</creatorcontrib><creatorcontrib>Ferreira, Luiz Otávio S.</creatorcontrib><collection>Pascal-Francis</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>Environment Abstracts</collection><collection>Environmental Sciences and Pollution Management</collection><collection>Environment Abstracts</collection><collection>Solid State and Superconductivity Abstracts</collection><collection>Technology Research Database</collection><collection>Advanced Technologies Database with Aerospace</collection><collection>MEDLINE - Academic</collection><jtitle>Analytica chimica acta</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Fonseca, Alexandre</au><au>Raimundo, Ivo M.</au><au>Rohwedder, Jarbas J.R.</au><au>Ferreira, Luiz Otávio S.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Construction and evaluation of a flow injection micro-analyser based on urethane–acrylate resin</atitle><jtitle>Analytica chimica acta</jtitle><addtitle>Anal Chim Acta</addtitle><date>2007-11-12</date><risdate>2007</risdate><volume>603</volume><issue>2</issue><spage>159</spage><epage>166</epage><pages>159-166</pages><issn>0003-2670</issn><eissn>1873-4324</eissn><coden>ACACAM</coden><abstract>A flow injection micro-analyser with an integrated injection device and photometric detection is described. Channels measuring 205–295 μm depth by 265–290 μm maximum width were manufactured by deep UV lithography on two layers of urethane–acrylate oligomers-based photoresist. Hypodermic syringe needles (450 μm diameter) were connected to the channels for introduction of solutions into the system. Plastic optical fibres were connected to the ends of a 5.0 mm long channel, in order to conduct the light from and to a homemade photometer. The device has a total volume of 7.0 μL and three different sample volumes (0.09, 0.22 and 0.30 μL) can be inserted into the system by choosing the appropriate loop of the hydrodynamic injection approach. The micro-analyser, designed as a single line manifold, was evaluated by determining chloride in waters (mercuric thiocyanate method), and chromium (VI) in wastewater and total chromium in metallic alloys (diphenylcarbazide method). For chloride determination two micro-pumps were employed to impel the solutions, while for chromium determination this task was performed by a conventional peristaltic pump. The results obtained in all determinations did not differ significantly from the reference methods at a confidence level of 95%. In the chloride determination, a flow rate of 50 μL min −1 was used, providing a sample frequency of 45 injection h −1, generating ca. 0.7 mg of Hg(II) after an 8-h working day (ca. 20 mL of solution). This result suggests the potential of the micro-analyser towards the reduction of waste, following the philosophy of Green Chemistry.</abstract><cop>Amsterdam</cop><pub>Elsevier B.V</pub><pmid>17963836</pmid><doi>10.1016/j.aca.2007.09.051</doi><tpages>8</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0003-2670
ispartof Analytica chimica acta, 2007-11, Vol.603 (2), p.159-166
issn 0003-2670
1873-4324
language eng
recordid cdi_proquest_miscellaneous_68444761
source Elsevier ScienceDirect Journals
subjects Analysis methods
Analytical chemistry
Applied sciences
Chemistry
Deep ultraviolet lithography
Exact sciences and technology
Flow analysis
General, instrumentation
Lab-on-a-chip
Micro-FIA
Microfluidic devices
Natural water pollution
Pollution
Water treatment and pollution
title Construction and evaluation of a flow injection micro-analyser based on urethane–acrylate resin
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-17T04%3A29%3A54IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Construction%20and%20evaluation%20of%20a%20flow%20injection%20micro-analyser%20based%20on%20urethane%E2%80%93acrylate%20resin&rft.jtitle=Analytica%20chimica%20acta&rft.au=Fonseca,%20Alexandre&rft.date=2007-11-12&rft.volume=603&rft.issue=2&rft.spage=159&rft.epage=166&rft.pages=159-166&rft.issn=0003-2670&rft.eissn=1873-4324&rft.coden=ACACAM&rft_id=info:doi/10.1016/j.aca.2007.09.051&rft_dat=%3Cproquest_cross%3E14864613%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=14864613&rft_id=info:pmid/17963836&rft_els_id=S0003267007016121&rfr_iscdi=true