Progress in the Development of the DexAide Right Ventricular Assist Device

The DexAide right ventricular assist device (RVAD) is an implantable centrifugal pump modified from the CorAide left ventricular assist device. As previously published, in vitro performance testing of the DexAide RVAD has met design criteria, and the nominal operating condition of 4 l/min and 20 mm...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:ASAIO journal (1992) 2006-11, Vol.52 (6), p.630-633
Hauptverfasser: Fukamachi, Kiyotaka, Ootaki, Yoshio, Horvath, David J, Massiello, Alex L, Kamohara, Keiji, Akiyama, Masatoshi, Cingoz, Faruk, Ootaki, Chiyo, Kopcak, Michael W, Dessoffy, Raymond, Liu, Jenny, Chen, Ji-Feng, Zhou, Qun, Benefit, Stephen, Golding, Leonard A.R, Alfini, Sue, Rowe, David T
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 633
container_issue 6
container_start_page 630
container_title ASAIO journal (1992)
container_volume 52
creator Fukamachi, Kiyotaka
Ootaki, Yoshio
Horvath, David J
Massiello, Alex L
Kamohara, Keiji
Akiyama, Masatoshi
Cingoz, Faruk
Ootaki, Chiyo
Kopcak, Michael W
Dessoffy, Raymond
Liu, Jenny
Chen, Ji-Feng
Zhou, Qun
Benefit, Stephen
Golding, Leonard A.R
Alfini, Sue
Rowe, David T
description The DexAide right ventricular assist device (RVAD) is an implantable centrifugal pump modified from the CorAide left ventricular assist device. As previously published, in vitro performance testing of the DexAide RVAD has met design criteria, and the nominal operating condition of 4 l/min and 20 mm Hg pressure rise was achieved at 2,000 rpm, with a power consumption of 1.9 watts. In vivo studies in 14 calves have demonstrated acceptable hemodynamic characteristics. The calf inflow cannula design is still evolving to minimize depositions on the cannula observed in most experiments. Fitting studies were performed in 5 cadavers and 2 patients to reconfigure the cannulae for use in humans. The design and development of external electronics have been completed for the stand-alone RVAD system, and verification tests are under way in preparation for preclinical tests. Work on the external electronics design for the biventricular assist system is ongoing.In conclusion, the initial in vitro and in vivo studies have demonstrated acceptable hemodynamic characteristics of the DexAide RVAD. The design and development of the external electronic components for the stand-alone RVAD system have been completed. The calf inflow cannula is being redesigned, and chronic in vivo tests are under way.
doi_str_mv 10.1097/01.mat.0000240700.03478.d0
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_68167042</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>68167042</sourcerecordid><originalsourceid>FETCH-LOGICAL-c4144-729a9fc02705c6a139cf63c5eaadffcd34f60e31269274e64f9bd1fc007ecb983</originalsourceid><addsrcrecordid>eNpFkFtP3DAQhS1Exa38BRT1gbeEmdhrx31bcUdIrRBFvFleZ8ymTTaLnRT67-tlI-EXe2bOOWN9jH1DKBC0OgMsOjsUkE4pQAEUwIWqihp22AHOeJVrwZ930xtmVV5qlPvsMMbfAGnIcY_to0JUMMMDdvcz9C-BYsyaVTYsKbugv9T2645WQ9b7qfU-b2rKHpqX5ZA9pUlo3NjakM1jbOKw8TSOvrIv3raRjqf7iP26unw8v8nvf1zfns_vcydQiFyV2mrvoEwfcNIi185L7mZkbe29q7nwEohjKXWpBEnh9aLGZABFbqErfsROt7nr0L-OFAfTNdFR29oV9WM0skKpQJRJ-H0rdKGPMZA369B0NvwzCGZD0gCaRNJ8kjQfJE0NyXwybRkXHdWf1gldEoit4K1vBwrxTzu-UTBLsu2wnCIryEsAiZiqfNMS_D9_xn-_</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>68167042</pqid></control><display><type>article</type><title>Progress in the Development of the DexAide Right Ventricular Assist Device</title><source>MEDLINE</source><source>Journals@Ovid LWW Legacy Archive</source><source>Journals@Ovid Complete</source><source>EZB-FREE-00999 freely available EZB journals</source><creator>Fukamachi, Kiyotaka ; Ootaki, Yoshio ; Horvath, David J ; Massiello, Alex L ; Kamohara, Keiji ; Akiyama, Masatoshi ; Cingoz, Faruk ; Ootaki, Chiyo ; Kopcak, Michael W ; Dessoffy, Raymond ; Liu, Jenny ; Chen, Ji-Feng ; Zhou, Qun ; Benefit, Stephen ; Golding, Leonard A.R ; Alfini, Sue ; Rowe, David T</creator><creatorcontrib>Fukamachi, Kiyotaka ; Ootaki, Yoshio ; Horvath, David J ; Massiello, Alex L ; Kamohara, Keiji ; Akiyama, Masatoshi ; Cingoz, Faruk ; Ootaki, Chiyo ; Kopcak, Michael W ; Dessoffy, Raymond ; Liu, Jenny ; Chen, Ji-Feng ; Zhou, Qun ; Benefit, Stephen ; Golding, Leonard A.R ; Alfini, Sue ; Rowe, David T</creatorcontrib><description>The DexAide right ventricular assist device (RVAD) is an implantable centrifugal pump modified from the CorAide left ventricular assist device. As previously published, in vitro performance testing of the DexAide RVAD has met design criteria, and the nominal operating condition of 4 l/min and 20 mm Hg pressure rise was achieved at 2,000 rpm, with a power consumption of 1.9 watts. In vivo studies in 14 calves have demonstrated acceptable hemodynamic characteristics. The calf inflow cannula design is still evolving to minimize depositions on the cannula observed in most experiments. Fitting studies were performed in 5 cadavers and 2 patients to reconfigure the cannulae for use in humans. The design and development of external electronics have been completed for the stand-alone RVAD system, and verification tests are under way in preparation for preclinical tests. Work on the external electronics design for the biventricular assist system is ongoing.In conclusion, the initial in vitro and in vivo studies have demonstrated acceptable hemodynamic characteristics of the DexAide RVAD. The design and development of the external electronic components for the stand-alone RVAD system have been completed. The calf inflow cannula is being redesigned, and chronic in vivo tests are under way.</description><identifier>ISSN: 1058-2916</identifier><identifier>EISSN: 1538-943X</identifier><identifier>DOI: 10.1097/01.mat.0000240700.03478.d0</identifier><identifier>PMID: 17117051</identifier><language>eng</language><publisher>United States: Copyright by the American Society for Artificial Internal Organs</publisher><subject>Animals ; Biomedical Engineering ; Blood Pressure ; Cattle ; Heart Failure - surgery ; Heart-Assist Devices - trends ; Humans ; Prosthesis Design ; Pulsatile Flow</subject><ispartof>ASAIO journal (1992), 2006-11, Vol.52 (6), p.630-633</ispartof><rights>Copyright © 2006 by the American Society for Artificial Internal Organs</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c4144-729a9fc02705c6a139cf63c5eaadffcd34f60e31269274e64f9bd1fc007ecb983</citedby><cites>FETCH-LOGICAL-c4144-729a9fc02705c6a139cf63c5eaadffcd34f60e31269274e64f9bd1fc007ecb983</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf><![CDATA[$$Uhttp://ovidsp.ovid.com/ovidweb.cgi?T=JS&CSC=Y&NEWS=N&PAGE=fulltext&PDF=y&D=ovft&AN=00002480-200611000-00004$$EPDF$$P50$$Gwolterskluwer$$H]]></linktopdf><linktohtml>$$Uhttp://ovidsp.ovid.com/ovidweb.cgi?T=JS&amp;NEWS=n&amp;CSC=Y&amp;PAGE=fulltext&amp;D=ovft&amp;AN=00002480-200611000-00004$$EHTML$$P50$$Gwolterskluwer$$H</linktohtml><link.rule.ids>314,780,784,4609,27924,27925,64666,65461</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/17117051$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Fukamachi, Kiyotaka</creatorcontrib><creatorcontrib>Ootaki, Yoshio</creatorcontrib><creatorcontrib>Horvath, David J</creatorcontrib><creatorcontrib>Massiello, Alex L</creatorcontrib><creatorcontrib>Kamohara, Keiji</creatorcontrib><creatorcontrib>Akiyama, Masatoshi</creatorcontrib><creatorcontrib>Cingoz, Faruk</creatorcontrib><creatorcontrib>Ootaki, Chiyo</creatorcontrib><creatorcontrib>Kopcak, Michael W</creatorcontrib><creatorcontrib>Dessoffy, Raymond</creatorcontrib><creatorcontrib>Liu, Jenny</creatorcontrib><creatorcontrib>Chen, Ji-Feng</creatorcontrib><creatorcontrib>Zhou, Qun</creatorcontrib><creatorcontrib>Benefit, Stephen</creatorcontrib><creatorcontrib>Golding, Leonard A.R</creatorcontrib><creatorcontrib>Alfini, Sue</creatorcontrib><creatorcontrib>Rowe, David T</creatorcontrib><title>Progress in the Development of the DexAide Right Ventricular Assist Device</title><title>ASAIO journal (1992)</title><addtitle>ASAIO J</addtitle><description>The DexAide right ventricular assist device (RVAD) is an implantable centrifugal pump modified from the CorAide left ventricular assist device. As previously published, in vitro performance testing of the DexAide RVAD has met design criteria, and the nominal operating condition of 4 l/min and 20 mm Hg pressure rise was achieved at 2,000 rpm, with a power consumption of 1.9 watts. In vivo studies in 14 calves have demonstrated acceptable hemodynamic characteristics. The calf inflow cannula design is still evolving to minimize depositions on the cannula observed in most experiments. Fitting studies were performed in 5 cadavers and 2 patients to reconfigure the cannulae for use in humans. The design and development of external electronics have been completed for the stand-alone RVAD system, and verification tests are under way in preparation for preclinical tests. Work on the external electronics design for the biventricular assist system is ongoing.In conclusion, the initial in vitro and in vivo studies have demonstrated acceptable hemodynamic characteristics of the DexAide RVAD. The design and development of the external electronic components for the stand-alone RVAD system have been completed. The calf inflow cannula is being redesigned, and chronic in vivo tests are under way.</description><subject>Animals</subject><subject>Biomedical Engineering</subject><subject>Blood Pressure</subject><subject>Cattle</subject><subject>Heart Failure - surgery</subject><subject>Heart-Assist Devices - trends</subject><subject>Humans</subject><subject>Prosthesis Design</subject><subject>Pulsatile Flow</subject><issn>1058-2916</issn><issn>1538-943X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2006</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpFkFtP3DAQhS1Exa38BRT1gbeEmdhrx31bcUdIrRBFvFleZ8ymTTaLnRT67-tlI-EXe2bOOWN9jH1DKBC0OgMsOjsUkE4pQAEUwIWqihp22AHOeJVrwZ930xtmVV5qlPvsMMbfAGnIcY_to0JUMMMDdvcz9C-BYsyaVTYsKbugv9T2645WQ9b7qfU-b2rKHpqX5ZA9pUlo3NjakM1jbOKw8TSOvrIv3raRjqf7iP26unw8v8nvf1zfns_vcydQiFyV2mrvoEwfcNIi185L7mZkbe29q7nwEohjKXWpBEnh9aLGZABFbqErfsROt7nr0L-OFAfTNdFR29oV9WM0skKpQJRJ-H0rdKGPMZA369B0NvwzCGZD0gCaRNJ8kjQfJE0NyXwybRkXHdWf1gldEoit4K1vBwrxTzu-UTBLsu2wnCIryEsAiZiqfNMS_D9_xn-_</recordid><startdate>200611</startdate><enddate>200611</enddate><creator>Fukamachi, Kiyotaka</creator><creator>Ootaki, Yoshio</creator><creator>Horvath, David J</creator><creator>Massiello, Alex L</creator><creator>Kamohara, Keiji</creator><creator>Akiyama, Masatoshi</creator><creator>Cingoz, Faruk</creator><creator>Ootaki, Chiyo</creator><creator>Kopcak, Michael W</creator><creator>Dessoffy, Raymond</creator><creator>Liu, Jenny</creator><creator>Chen, Ji-Feng</creator><creator>Zhou, Qun</creator><creator>Benefit, Stephen</creator><creator>Golding, Leonard A.R</creator><creator>Alfini, Sue</creator><creator>Rowe, David T</creator><general>Copyright by the American Society for Artificial Internal Organs</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>200611</creationdate><title>Progress in the Development of the DexAide Right Ventricular Assist Device</title><author>Fukamachi, Kiyotaka ; Ootaki, Yoshio ; Horvath, David J ; Massiello, Alex L ; Kamohara, Keiji ; Akiyama, Masatoshi ; Cingoz, Faruk ; Ootaki, Chiyo ; Kopcak, Michael W ; Dessoffy, Raymond ; Liu, Jenny ; Chen, Ji-Feng ; Zhou, Qun ; Benefit, Stephen ; Golding, Leonard A.R ; Alfini, Sue ; Rowe, David T</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c4144-729a9fc02705c6a139cf63c5eaadffcd34f60e31269274e64f9bd1fc007ecb983</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2006</creationdate><topic>Animals</topic><topic>Biomedical Engineering</topic><topic>Blood Pressure</topic><topic>Cattle</topic><topic>Heart Failure - surgery</topic><topic>Heart-Assist Devices - trends</topic><topic>Humans</topic><topic>Prosthesis Design</topic><topic>Pulsatile Flow</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Fukamachi, Kiyotaka</creatorcontrib><creatorcontrib>Ootaki, Yoshio</creatorcontrib><creatorcontrib>Horvath, David J</creatorcontrib><creatorcontrib>Massiello, Alex L</creatorcontrib><creatorcontrib>Kamohara, Keiji</creatorcontrib><creatorcontrib>Akiyama, Masatoshi</creatorcontrib><creatorcontrib>Cingoz, Faruk</creatorcontrib><creatorcontrib>Ootaki, Chiyo</creatorcontrib><creatorcontrib>Kopcak, Michael W</creatorcontrib><creatorcontrib>Dessoffy, Raymond</creatorcontrib><creatorcontrib>Liu, Jenny</creatorcontrib><creatorcontrib>Chen, Ji-Feng</creatorcontrib><creatorcontrib>Zhou, Qun</creatorcontrib><creatorcontrib>Benefit, Stephen</creatorcontrib><creatorcontrib>Golding, Leonard A.R</creatorcontrib><creatorcontrib>Alfini, Sue</creatorcontrib><creatorcontrib>Rowe, David T</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>ASAIO journal (1992)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Fukamachi, Kiyotaka</au><au>Ootaki, Yoshio</au><au>Horvath, David J</au><au>Massiello, Alex L</au><au>Kamohara, Keiji</au><au>Akiyama, Masatoshi</au><au>Cingoz, Faruk</au><au>Ootaki, Chiyo</au><au>Kopcak, Michael W</au><au>Dessoffy, Raymond</au><au>Liu, Jenny</au><au>Chen, Ji-Feng</au><au>Zhou, Qun</au><au>Benefit, Stephen</au><au>Golding, Leonard A.R</au><au>Alfini, Sue</au><au>Rowe, David T</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Progress in the Development of the DexAide Right Ventricular Assist Device</atitle><jtitle>ASAIO journal (1992)</jtitle><addtitle>ASAIO J</addtitle><date>2006-11</date><risdate>2006</risdate><volume>52</volume><issue>6</issue><spage>630</spage><epage>633</epage><pages>630-633</pages><issn>1058-2916</issn><eissn>1538-943X</eissn><abstract>The DexAide right ventricular assist device (RVAD) is an implantable centrifugal pump modified from the CorAide left ventricular assist device. As previously published, in vitro performance testing of the DexAide RVAD has met design criteria, and the nominal operating condition of 4 l/min and 20 mm Hg pressure rise was achieved at 2,000 rpm, with a power consumption of 1.9 watts. In vivo studies in 14 calves have demonstrated acceptable hemodynamic characteristics. The calf inflow cannula design is still evolving to minimize depositions on the cannula observed in most experiments. Fitting studies were performed in 5 cadavers and 2 patients to reconfigure the cannulae for use in humans. The design and development of external electronics have been completed for the stand-alone RVAD system, and verification tests are under way in preparation for preclinical tests. Work on the external electronics design for the biventricular assist system is ongoing.In conclusion, the initial in vitro and in vivo studies have demonstrated acceptable hemodynamic characteristics of the DexAide RVAD. The design and development of the external electronic components for the stand-alone RVAD system have been completed. The calf inflow cannula is being redesigned, and chronic in vivo tests are under way.</abstract><cop>United States</cop><pub>Copyright by the American Society for Artificial Internal Organs</pub><pmid>17117051</pmid><doi>10.1097/01.mat.0000240700.03478.d0</doi><tpages>4</tpages><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 1058-2916
ispartof ASAIO journal (1992), 2006-11, Vol.52 (6), p.630-633
issn 1058-2916
1538-943X
language eng
recordid cdi_proquest_miscellaneous_68167042
source MEDLINE; Journals@Ovid LWW Legacy Archive; Journals@Ovid Complete; EZB-FREE-00999 freely available EZB journals
subjects Animals
Biomedical Engineering
Blood Pressure
Cattle
Heart Failure - surgery
Heart-Assist Devices - trends
Humans
Prosthesis Design
Pulsatile Flow
title Progress in the Development of the DexAide Right Ventricular Assist Device
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-02T17%3A30%3A50IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Progress%20in%20the%20Development%20of%20the%20DexAide%20Right%20Ventricular%20Assist%20Device&rft.jtitle=ASAIO%20journal%20(1992)&rft.au=Fukamachi,%20Kiyotaka&rft.date=2006-11&rft.volume=52&rft.issue=6&rft.spage=630&rft.epage=633&rft.pages=630-633&rft.issn=1058-2916&rft.eissn=1538-943X&rft_id=info:doi/10.1097/01.mat.0000240700.03478.d0&rft_dat=%3Cproquest_cross%3E68167042%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=68167042&rft_id=info:pmid/17117051&rfr_iscdi=true