High-Resolution Magnetic Resonance Imaging of Trabecular Bone in the Wrist at 3 Tesla: Initial Results
Purpose: To evaluate the feasibility of high-resolution magnetic resonance imaging (MRI) of trabecular bone of the wrist at 3 Tesla (3T) in vivo and to assess the potential benefit of the increased resolution for clinical assessment of structural changes in spongy bone. Material and Methods: High-re...
Gespeichert in:
Veröffentlicht in: | Acta radiologica (1987) 2005-05, Vol.46 (3), p.306-309 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose: To evaluate the feasibility of high-resolution magnetic resonance imaging (MRI) of trabecular bone of the wrist at 3 Tesla (3T) in vivo and to assess the potential benefit of the increased resolution for clinical assessment of structural changes in spongy bone.
Material and Methods: High-resolution MRI of the wrist was performed with a whole-body 3T MR scanner using a dedicated circularly polarized transmit-receive wrist-coil. Two 3D-FISP sequences with a spatial resolution of 300×300×300 µm3 in a measuring time of TA = 7:51 min, and 200×200×200 µm3 in TA = 9:33 min were applied. Seven young healthy volunteers and three elderly subjects with suspected osteoporosis were examined. The signal-to-noise ratio (SNR) in the optimized setup at 3T was compared to measurements at 1.5T.
Results: The images at 3T allow microscopic analysis of the bone structure at an isotropic spatial resolution of 200 µm in examination times of |
---|---|
ISSN: | 0284-1851 1600-0455 |
DOI: | 10.1080/02841850510021076 |