Fracture of ProFile nickel-titanium rotary instruments: a laboratory simulation assessment

Aim  To determine the incidence of ProFile nickel–titanium rotary instrument fracture in an endodontic laboratory simulation. Methodology  Three hundred and sixty dental students used 2880 ProFile nickel–titanium rotary instruments to prepare 1440 simulated root canals in 720 plastic teeth, and anot...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International endodontic journal 2006-06, Vol.39 (6), p.502-509
Hauptverfasser: Di Fiore, P. M., Genov, K. I., Komaroff, E., Dasanayake, A. P., Lin, L.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 509
container_issue 6
container_start_page 502
container_title International endodontic journal
container_volume 39
creator Di Fiore, P. M.
Genov, K. I.
Komaroff, E.
Dasanayake, A. P.
Lin, L.
description Aim  To determine the incidence of ProFile nickel–titanium rotary instrument fracture in an endodontic laboratory simulation. Methodology  Three hundred and sixty dental students used 2880 ProFile nickel–titanium rotary instruments to prepare 1440 simulated root canals in 720 plastic teeth, and another 2880 ProFile nickel–titanium rotary instruments to prepare 1440 natural root canals in 720 extracted teeth. A standardized crown‐down rotary instrumentation technique was used, which included measures to prevent fracture. Rotary instrument fracture was monitored during and after completion of the laboratory simulation exercises to determine the incidence of fracture. When fracture occurred, data were collected concerning the size of the instrument, the length of the instrument fragment, the location of the fragment in the canal and the curvature of the canal in which the instrument fractured. Results  The incidence of instrument fracture was 0.41% in plastic simulated canals and 0.31% in natural root canals. The overall incidence of instrument fracture was 0.36%. Of the instruments that fractured 67% were size 25, 0.04 taper; and 81% of the fragments were located in the apical third of the canal. The mean, median and mode of the fragment lengths were all 3 mm. Conclusions  The low overall incidence of fracture in this study suggests that ProFile rotary instruments are safe for use by dental students in laboratory simulations and that if preventive measures are taken the incidence of instrument fracture can be minimized.
doi_str_mv 10.1111/j.1365-2591.2006.01123.x
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_67933343</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>67933343</sourcerecordid><originalsourceid>FETCH-LOGICAL-c4053-baf8988d2cc8ac3e4d9a2b5beef289112e9309cc5c54a7601793ed7072c7917c3</originalsourceid><addsrcrecordid>eNqNkMFO3DAQhi1EBQvlFZBPvSW14ziOkThUW5YFrVokipC4WI53InlJYrAddXn7OuwKrvXFlub_ZjwfQpiSnKbzfZNTVvGs4JLmBSFVTigtWL49QLOPwiGaEVqyrKhrfoxOQtgQQjhh9Agd06oSpSirGXpaeG3i6AG7Ft95t7Ad4MGaZ-iyaKMe7Nhj76L2b9gOIfqxhyGGC6xxpxvndXSpEmw_djpaN2AdAoQwhb6iL63uApzt71P0sLj6M19mq9_XN_Mfq8yUhLOs0W0t63pdGFNrw6BcS100vAFoi1qmvUAyIo3hhpdaVIQKyWAtiCiMkFQYdoq-7fq-ePc6Qoiqt8FA1-kB3BhUlQDGSpaC9S5ovAvBQ6tevO3TZooSNXlVGzXpU5M-NXlV717VNqHn-xlj08P6E9yLTIHLXeBvEvj2343VzdXt9Ep8tuNtiLD94LV_Tv9ngqvHX9fqaUXny-XPeyXZPzeXlz8</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>67933343</pqid></control><display><type>article</type><title>Fracture of ProFile nickel-titanium rotary instruments: a laboratory simulation assessment</title><source>MEDLINE</source><source>Wiley Online Library Journals Frontfile Complete</source><creator>Di Fiore, P. M. ; Genov, K. I. ; Komaroff, E. ; Dasanayake, A. P. ; Lin, L.</creator><creatorcontrib>Di Fiore, P. M. ; Genov, K. I. ; Komaroff, E. ; Dasanayake, A. P. ; Lin, L.</creatorcontrib><description>Aim  To determine the incidence of ProFile nickel–titanium rotary instrument fracture in an endodontic laboratory simulation. Methodology  Three hundred and sixty dental students used 2880 ProFile nickel–titanium rotary instruments to prepare 1440 simulated root canals in 720 plastic teeth, and another 2880 ProFile nickel–titanium rotary instruments to prepare 1440 natural root canals in 720 extracted teeth. A standardized crown‐down rotary instrumentation technique was used, which included measures to prevent fracture. Rotary instrument fracture was monitored during and after completion of the laboratory simulation exercises to determine the incidence of fracture. When fracture occurred, data were collected concerning the size of the instrument, the length of the instrument fragment, the location of the fragment in the canal and the curvature of the canal in which the instrument fractured. Results  The incidence of instrument fracture was 0.41% in plastic simulated canals and 0.31% in natural root canals. The overall incidence of instrument fracture was 0.36%. Of the instruments that fractured 67% were size 25, 0.04 taper; and 81% of the fragments were located in the apical third of the canal. The mean, median and mode of the fragment lengths were all 3 mm. Conclusions  The low overall incidence of fracture in this study suggests that ProFile rotary instruments are safe for use by dental students in laboratory simulations and that if preventive measures are taken the incidence of instrument fracture can be minimized.</description><identifier>ISSN: 0143-2885</identifier><identifier>EISSN: 1365-2591</identifier><identifier>DOI: 10.1111/j.1365-2591.2006.01123.x</identifier><identifier>PMID: 16674746</identifier><language>eng</language><publisher>Oxford, UK: Blackwell Publishing Ltd</publisher><subject>Dental Alloys - chemistry ; Dental Pulp Cavity - anatomy &amp; histology ; Dentistry ; Equipment Failure ; Equipment Failure Analysis ; Equipment Safety ; fracture ; Humans ; instruments ; Materials Testing ; Models, Anatomic ; Nickel - chemistry ; nickel-titanium ; Plastics ; Root Canal Preparation - instrumentation ; Root Canal Preparation - methods ; rotary ; Rotation ; Surface Properties ; Titanium - chemistry</subject><ispartof>International endodontic journal, 2006-06, Vol.39 (6), p.502-509</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c4053-baf8988d2cc8ac3e4d9a2b5beef289112e9309cc5c54a7601793ed7072c7917c3</citedby><cites>FETCH-LOGICAL-c4053-baf8988d2cc8ac3e4d9a2b5beef289112e9309cc5c54a7601793ed7072c7917c3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1111%2Fj.1365-2591.2006.01123.x$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1111%2Fj.1365-2591.2006.01123.x$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>314,778,782,1414,27907,27908,45557,45558</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/16674746$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Di Fiore, P. M.</creatorcontrib><creatorcontrib>Genov, K. I.</creatorcontrib><creatorcontrib>Komaroff, E.</creatorcontrib><creatorcontrib>Dasanayake, A. P.</creatorcontrib><creatorcontrib>Lin, L.</creatorcontrib><title>Fracture of ProFile nickel-titanium rotary instruments: a laboratory simulation assessment</title><title>International endodontic journal</title><addtitle>Int Endod J</addtitle><description>Aim  To determine the incidence of ProFile nickel–titanium rotary instrument fracture in an endodontic laboratory simulation. Methodology  Three hundred and sixty dental students used 2880 ProFile nickel–titanium rotary instruments to prepare 1440 simulated root canals in 720 plastic teeth, and another 2880 ProFile nickel–titanium rotary instruments to prepare 1440 natural root canals in 720 extracted teeth. A standardized crown‐down rotary instrumentation technique was used, which included measures to prevent fracture. Rotary instrument fracture was monitored during and after completion of the laboratory simulation exercises to determine the incidence of fracture. When fracture occurred, data were collected concerning the size of the instrument, the length of the instrument fragment, the location of the fragment in the canal and the curvature of the canal in which the instrument fractured. Results  The incidence of instrument fracture was 0.41% in plastic simulated canals and 0.31% in natural root canals. The overall incidence of instrument fracture was 0.36%. Of the instruments that fractured 67% were size 25, 0.04 taper; and 81% of the fragments were located in the apical third of the canal. The mean, median and mode of the fragment lengths were all 3 mm. Conclusions  The low overall incidence of fracture in this study suggests that ProFile rotary instruments are safe for use by dental students in laboratory simulations and that if preventive measures are taken the incidence of instrument fracture can be minimized.</description><subject>Dental Alloys - chemistry</subject><subject>Dental Pulp Cavity - anatomy &amp; histology</subject><subject>Dentistry</subject><subject>Equipment Failure</subject><subject>Equipment Failure Analysis</subject><subject>Equipment Safety</subject><subject>fracture</subject><subject>Humans</subject><subject>instruments</subject><subject>Materials Testing</subject><subject>Models, Anatomic</subject><subject>Nickel - chemistry</subject><subject>nickel-titanium</subject><subject>Plastics</subject><subject>Root Canal Preparation - instrumentation</subject><subject>Root Canal Preparation - methods</subject><subject>rotary</subject><subject>Rotation</subject><subject>Surface Properties</subject><subject>Titanium - chemistry</subject><issn>0143-2885</issn><issn>1365-2591</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2006</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNqNkMFO3DAQhi1EBQvlFZBPvSW14ziOkThUW5YFrVokipC4WI53InlJYrAddXn7OuwKrvXFlub_ZjwfQpiSnKbzfZNTVvGs4JLmBSFVTigtWL49QLOPwiGaEVqyrKhrfoxOQtgQQjhh9Agd06oSpSirGXpaeG3i6AG7Ft95t7Ad4MGaZ-iyaKMe7Nhj76L2b9gOIfqxhyGGC6xxpxvndXSpEmw_djpaN2AdAoQwhb6iL63uApzt71P0sLj6M19mq9_XN_Mfq8yUhLOs0W0t63pdGFNrw6BcS100vAFoi1qmvUAyIo3hhpdaVIQKyWAtiCiMkFQYdoq-7fq-ePc6Qoiqt8FA1-kB3BhUlQDGSpaC9S5ovAvBQ6tevO3TZooSNXlVGzXpU5M-NXlV717VNqHn-xlj08P6E9yLTIHLXeBvEvj2343VzdXt9Ep8tuNtiLD94LV_Tv9ngqvHX9fqaUXny-XPeyXZPzeXlz8</recordid><startdate>200606</startdate><enddate>200606</enddate><creator>Di Fiore, P. M.</creator><creator>Genov, K. I.</creator><creator>Komaroff, E.</creator><creator>Dasanayake, A. P.</creator><creator>Lin, L.</creator><general>Blackwell Publishing Ltd</general><scope>BSCLL</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>200606</creationdate><title>Fracture of ProFile nickel-titanium rotary instruments: a laboratory simulation assessment</title><author>Di Fiore, P. M. ; Genov, K. I. ; Komaroff, E. ; Dasanayake, A. P. ; Lin, L.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c4053-baf8988d2cc8ac3e4d9a2b5beef289112e9309cc5c54a7601793ed7072c7917c3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2006</creationdate><topic>Dental Alloys - chemistry</topic><topic>Dental Pulp Cavity - anatomy &amp; histology</topic><topic>Dentistry</topic><topic>Equipment Failure</topic><topic>Equipment Failure Analysis</topic><topic>Equipment Safety</topic><topic>fracture</topic><topic>Humans</topic><topic>instruments</topic><topic>Materials Testing</topic><topic>Models, Anatomic</topic><topic>Nickel - chemistry</topic><topic>nickel-titanium</topic><topic>Plastics</topic><topic>Root Canal Preparation - instrumentation</topic><topic>Root Canal Preparation - methods</topic><topic>rotary</topic><topic>Rotation</topic><topic>Surface Properties</topic><topic>Titanium - chemistry</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Di Fiore, P. M.</creatorcontrib><creatorcontrib>Genov, K. I.</creatorcontrib><creatorcontrib>Komaroff, E.</creatorcontrib><creatorcontrib>Dasanayake, A. P.</creatorcontrib><creatorcontrib>Lin, L.</creatorcontrib><collection>Istex</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>International endodontic journal</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Di Fiore, P. M.</au><au>Genov, K. I.</au><au>Komaroff, E.</au><au>Dasanayake, A. P.</au><au>Lin, L.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Fracture of ProFile nickel-titanium rotary instruments: a laboratory simulation assessment</atitle><jtitle>International endodontic journal</jtitle><addtitle>Int Endod J</addtitle><date>2006-06</date><risdate>2006</risdate><volume>39</volume><issue>6</issue><spage>502</spage><epage>509</epage><pages>502-509</pages><issn>0143-2885</issn><eissn>1365-2591</eissn><abstract>Aim  To determine the incidence of ProFile nickel–titanium rotary instrument fracture in an endodontic laboratory simulation. Methodology  Three hundred and sixty dental students used 2880 ProFile nickel–titanium rotary instruments to prepare 1440 simulated root canals in 720 plastic teeth, and another 2880 ProFile nickel–titanium rotary instruments to prepare 1440 natural root canals in 720 extracted teeth. A standardized crown‐down rotary instrumentation technique was used, which included measures to prevent fracture. Rotary instrument fracture was monitored during and after completion of the laboratory simulation exercises to determine the incidence of fracture. When fracture occurred, data were collected concerning the size of the instrument, the length of the instrument fragment, the location of the fragment in the canal and the curvature of the canal in which the instrument fractured. Results  The incidence of instrument fracture was 0.41% in plastic simulated canals and 0.31% in natural root canals. The overall incidence of instrument fracture was 0.36%. Of the instruments that fractured 67% were size 25, 0.04 taper; and 81% of the fragments were located in the apical third of the canal. The mean, median and mode of the fragment lengths were all 3 mm. Conclusions  The low overall incidence of fracture in this study suggests that ProFile rotary instruments are safe for use by dental students in laboratory simulations and that if preventive measures are taken the incidence of instrument fracture can be minimized.</abstract><cop>Oxford, UK</cop><pub>Blackwell Publishing Ltd</pub><pmid>16674746</pmid><doi>10.1111/j.1365-2591.2006.01123.x</doi><tpages>8</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0143-2885
ispartof International endodontic journal, 2006-06, Vol.39 (6), p.502-509
issn 0143-2885
1365-2591
language eng
recordid cdi_proquest_miscellaneous_67933343
source MEDLINE; Wiley Online Library Journals Frontfile Complete
subjects Dental Alloys - chemistry
Dental Pulp Cavity - anatomy & histology
Dentistry
Equipment Failure
Equipment Failure Analysis
Equipment Safety
fracture
Humans
instruments
Materials Testing
Models, Anatomic
Nickel - chemistry
nickel-titanium
Plastics
Root Canal Preparation - instrumentation
Root Canal Preparation - methods
rotary
Rotation
Surface Properties
Titanium - chemistry
title Fracture of ProFile nickel-titanium rotary instruments: a laboratory simulation assessment
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-17T08%3A05%3A22IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Fracture%20of%20ProFile%20nickel-titanium%20rotary%20instruments:%20a%20laboratory%20simulation%20assessment&rft.jtitle=International%20endodontic%20journal&rft.au=Di%20Fiore,%20P.%20M.&rft.date=2006-06&rft.volume=39&rft.issue=6&rft.spage=502&rft.epage=509&rft.pages=502-509&rft.issn=0143-2885&rft.eissn=1365-2591&rft_id=info:doi/10.1111/j.1365-2591.2006.01123.x&rft_dat=%3Cproquest_cross%3E67933343%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=67933343&rft_id=info:pmid/16674746&rfr_iscdi=true