EBRA-FCA for measurement of migration of the femoral component in surface arthroplasty of the hip
Studies on the migration of an implant may be the only way of monitoring the early performance of metal-on-metal prostheses. The Ein Bild Roentgen Analyse--femoral component analysis (EBRA-FCA) method was adapted to measure migration of the femoral component in a metal-on-metal surface arthroplasty...
Gespeichert in:
Veröffentlicht in: | Journal of bone and joint surgery. British volume 2005-05, Vol.87 (5), p.741-744 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 744 |
---|---|
container_issue | 5 |
container_start_page | 741 |
container_title | Journal of bone and joint surgery. British volume |
container_volume | 87 |
creator | BEAULE, P. E KRISMER, M MAYRHOFER, P WANNER, S LE DUFF, M MATTESICH, M STOECKL, B AMSTUTZ, H. C BIEDERMANN, R |
description | Studies on the migration of an implant may be the only way of monitoring the early performance of metal-on-metal prostheses. The Ein Bild Roentgen Analyse--femoral component analysis (EBRA-FCA) method was adapted to measure migration of the femoral component in a metal-on-metal surface arthroplasty of the hip using standard antero-posterior radiographs. In order to determine the accuracy and precision of this method a prosthesis was implanted into cadaver bones. Eleven series of radiographs were used to perform a zero-migration study. After adjustment of the femoral component to simulate migration of 3 mm the radiographs were repeated. All were measured independently by three different observers. The accuracy of the method was found to be +/- 1.6 mm for the x-direction and +/- 2 mm for the y-direction (95% percentile). The method was validated using 28 hips with a minimum follow-up of 3.5 years after arthroplasty. Seventeen were sound, but 11 had failed because of loosening of the femoral component. The normal (control) group had a different pattern of migration compared with that of the loose group. At 29.2 months, the control group showed a mean migration of 1.62 mm and 1.05 mm compared with 4.39 mm and 4.05 mm in the failed group, for the centre of the head and the tip of the stem, respectively (p = 0.001). In the failed group, the mean time to migration greater than 2 mm was earlier than the onset of clinical symptoms or radiological evidence of failure, 19.1 versus 32.2 months (p = 0.001) and 24.8 months (p = 0.012), respectively. EBRA-FCA is a reliable and valid tool for measuring migration of the femoral component after surface arthroplasty and can be used to predict early failure of the implant. It may be of value in determining the long-term performance of surface arthroplasty. |
doi_str_mv | 10.1302/0301-620X.87B5.15377 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_67783572</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>846355761</sourcerecordid><originalsourceid>FETCH-LOGICAL-c408t-1936da4275ac679fc30029c4781fdc78b1f4f7963c9f86b897e42a77817ce4cc3</originalsourceid><addsrcrecordid>eNpdkF1LwzAUhoMobk7_gUgR9K4zX03ay21sKgwEUfAuZFniOpqmJu3F_r2pmw68CDmH87yHwwPANYJjRCB-gASilGH4Mc75NBujjHB-AoYYUpr29SkY_iEDcBHCFkJIs4ycgwHK8ljkeAjkfPo6SRezSWKcT6yWofPa6rpNnEls-ellW7q6b9qNToy2zssqUc42ru6psk5iwkilE-nbjXdNJUO7-w1syuYSnBlZBX11-EfgfTF_mz2ly5fH59lkmSoK8zZFBWFrSTHPpGK8MIpAiAtFeY7MWvF8hQw1vGBEFSZnq7zgmmLJ45grTZUiI3C_39t499Xp0ApbBqWrStbadUGwyJKM4wje_gO3rvN1vE1gXDAeH4wQ3UPKuxC8NqLxpZV-JxAUvX_RyxW9XNH7Fz_-Y-zmsLtbWb0-hg7CI3B3AGRQsjJe1qoMR45xRlnkvgFECIyl</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>229672960</pqid></control><display><type>article</type><title>EBRA-FCA for measurement of migration of the femoral component in surface arthroplasty of the hip</title><source>MEDLINE</source><source>Alma/SFX Local Collection</source><creator>BEAULE, P. E ; KRISMER, M ; MAYRHOFER, P ; WANNER, S ; LE DUFF, M ; MATTESICH, M ; STOECKL, B ; AMSTUTZ, H. C ; BIEDERMANN, R</creator><creatorcontrib>BEAULE, P. E ; KRISMER, M ; MAYRHOFER, P ; WANNER, S ; LE DUFF, M ; MATTESICH, M ; STOECKL, B ; AMSTUTZ, H. C ; BIEDERMANN, R</creatorcontrib><description>Studies on the migration of an implant may be the only way of monitoring the early performance of metal-on-metal prostheses. The Ein Bild Roentgen Analyse--femoral component analysis (EBRA-FCA) method was adapted to measure migration of the femoral component in a metal-on-metal surface arthroplasty of the hip using standard antero-posterior radiographs. In order to determine the accuracy and precision of this method a prosthesis was implanted into cadaver bones. Eleven series of radiographs were used to perform a zero-migration study. After adjustment of the femoral component to simulate migration of 3 mm the radiographs were repeated. All were measured independently by three different observers. The accuracy of the method was found to be +/- 1.6 mm for the x-direction and +/- 2 mm for the y-direction (95% percentile). The method was validated using 28 hips with a minimum follow-up of 3.5 years after arthroplasty. Seventeen were sound, but 11 had failed because of loosening of the femoral component. The normal (control) group had a different pattern of migration compared with that of the loose group. At 29.2 months, the control group showed a mean migration of 1.62 mm and 1.05 mm compared with 4.39 mm and 4.05 mm in the failed group, for the centre of the head and the tip of the stem, respectively (p = 0.001). In the failed group, the mean time to migration greater than 2 mm was earlier than the onset of clinical symptoms or radiological evidence of failure, 19.1 versus 32.2 months (p = 0.001) and 24.8 months (p = 0.012), respectively. EBRA-FCA is a reliable and valid tool for measuring migration of the femoral component after surface arthroplasty and can be used to predict early failure of the implant. It may be of value in determining the long-term performance of surface arthroplasty.</description><edition>British volume</edition><identifier>ISSN: 0301-620X</identifier><identifier>ISSN: 2049-4394</identifier><identifier>EISSN: 2044-5377</identifier><identifier>EISSN: 2049-4408</identifier><identifier>DOI: 10.1302/0301-620X.87B5.15377</identifier><identifier>PMID: 15855382</identifier><identifier>CODEN: JBSUAK</identifier><language>eng</language><publisher>London: British Editorial Society of Bone and Joint Surgery</publisher><subject>Adult ; Aged ; Arthroplasty, Replacement, Hip - methods ; Biological and medical sciences ; Bone Diseases - surgery ; Cadaver ; Diseases of the osteoarticular system ; Equipment Failure Analysis - methods ; Female ; Femur - surgery ; Hip Joint - surgery ; Hip Prosthesis ; Humans ; Joint Diseases - surgery ; Male ; Medical sciences ; Middle Aged ; Orthopedic surgery ; Prosthesis Design ; Prosthesis Failure ; Reproducibility of Results ; Software ; Surgery (general aspects). Transplantations, organ and tissue grafts. Graft diseases</subject><ispartof>Journal of bone and joint surgery. British volume, 2005-05, Vol.87 (5), p.741-744</ispartof><rights>2005 INIST-CNRS</rights><rights>Copyright British Editorial Society of Bone & Joint Surgery May 2005</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c408t-1936da4275ac679fc30029c4781fdc78b1f4f7963c9f86b897e42a77817ce4cc3</citedby><cites>FETCH-LOGICAL-c408t-1936da4275ac679fc30029c4781fdc78b1f4f7963c9f86b897e42a77817ce4cc3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,27901,27902</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&idt=16764682$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/15855382$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>BEAULE, P. E</creatorcontrib><creatorcontrib>KRISMER, M</creatorcontrib><creatorcontrib>MAYRHOFER, P</creatorcontrib><creatorcontrib>WANNER, S</creatorcontrib><creatorcontrib>LE DUFF, M</creatorcontrib><creatorcontrib>MATTESICH, M</creatorcontrib><creatorcontrib>STOECKL, B</creatorcontrib><creatorcontrib>AMSTUTZ, H. C</creatorcontrib><creatorcontrib>BIEDERMANN, R</creatorcontrib><title>EBRA-FCA for measurement of migration of the femoral component in surface arthroplasty of the hip</title><title>Journal of bone and joint surgery. British volume</title><addtitle>J Bone Joint Surg Br</addtitle><description>Studies on the migration of an implant may be the only way of monitoring the early performance of metal-on-metal prostheses. The Ein Bild Roentgen Analyse--femoral component analysis (EBRA-FCA) method was adapted to measure migration of the femoral component in a metal-on-metal surface arthroplasty of the hip using standard antero-posterior radiographs. In order to determine the accuracy and precision of this method a prosthesis was implanted into cadaver bones. Eleven series of radiographs were used to perform a zero-migration study. After adjustment of the femoral component to simulate migration of 3 mm the radiographs were repeated. All were measured independently by three different observers. The accuracy of the method was found to be +/- 1.6 mm for the x-direction and +/- 2 mm for the y-direction (95% percentile). The method was validated using 28 hips with a minimum follow-up of 3.5 years after arthroplasty. Seventeen were sound, but 11 had failed because of loosening of the femoral component. The normal (control) group had a different pattern of migration compared with that of the loose group. At 29.2 months, the control group showed a mean migration of 1.62 mm and 1.05 mm compared with 4.39 mm and 4.05 mm in the failed group, for the centre of the head and the tip of the stem, respectively (p = 0.001). In the failed group, the mean time to migration greater than 2 mm was earlier than the onset of clinical symptoms or radiological evidence of failure, 19.1 versus 32.2 months (p = 0.001) and 24.8 months (p = 0.012), respectively. EBRA-FCA is a reliable and valid tool for measuring migration of the femoral component after surface arthroplasty and can be used to predict early failure of the implant. It may be of value in determining the long-term performance of surface arthroplasty.</description><subject>Adult</subject><subject>Aged</subject><subject>Arthroplasty, Replacement, Hip - methods</subject><subject>Biological and medical sciences</subject><subject>Bone Diseases - surgery</subject><subject>Cadaver</subject><subject>Diseases of the osteoarticular system</subject><subject>Equipment Failure Analysis - methods</subject><subject>Female</subject><subject>Femur - surgery</subject><subject>Hip Joint - surgery</subject><subject>Hip Prosthesis</subject><subject>Humans</subject><subject>Joint Diseases - surgery</subject><subject>Male</subject><subject>Medical sciences</subject><subject>Middle Aged</subject><subject>Orthopedic surgery</subject><subject>Prosthesis Design</subject><subject>Prosthesis Failure</subject><subject>Reproducibility of Results</subject><subject>Software</subject><subject>Surgery (general aspects). Transplantations, organ and tissue grafts. Graft diseases</subject><issn>0301-620X</issn><issn>2049-4394</issn><issn>2044-5377</issn><issn>2049-4408</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2005</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpdkF1LwzAUhoMobk7_gUgR9K4zX03ay21sKgwEUfAuZFniOpqmJu3F_r2pmw68CDmH87yHwwPANYJjRCB-gASilGH4Mc75NBujjHB-AoYYUpr29SkY_iEDcBHCFkJIs4ycgwHK8ljkeAjkfPo6SRezSWKcT6yWofPa6rpNnEls-ellW7q6b9qNToy2zssqUc42ru6psk5iwkilE-nbjXdNJUO7-w1syuYSnBlZBX11-EfgfTF_mz2ly5fH59lkmSoK8zZFBWFrSTHPpGK8MIpAiAtFeY7MWvF8hQw1vGBEFSZnq7zgmmLJ45grTZUiI3C_39t499Xp0ApbBqWrStbadUGwyJKM4wje_gO3rvN1vE1gXDAeH4wQ3UPKuxC8NqLxpZV-JxAUvX_RyxW9XNH7Fz_-Y-zmsLtbWb0-hg7CI3B3AGRQsjJe1qoMR45xRlnkvgFECIyl</recordid><startdate>20050501</startdate><enddate>20050501</enddate><creator>BEAULE, P. E</creator><creator>KRISMER, M</creator><creator>MAYRHOFER, P</creator><creator>WANNER, S</creator><creator>LE DUFF, M</creator><creator>MATTESICH, M</creator><creator>STOECKL, B</creator><creator>AMSTUTZ, H. C</creator><creator>BIEDERMANN, R</creator><general>British Editorial Society of Bone and Joint Surgery</general><general>British Editorial Society of Bone & Joint Surgery</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>K9.</scope><scope>NAPCQ</scope><scope>7X8</scope></search><sort><creationdate>20050501</creationdate><title>EBRA-FCA for measurement of migration of the femoral component in surface arthroplasty of the hip</title><author>BEAULE, P. E ; KRISMER, M ; MAYRHOFER, P ; WANNER, S ; LE DUFF, M ; MATTESICH, M ; STOECKL, B ; AMSTUTZ, H. C ; BIEDERMANN, R</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c408t-1936da4275ac679fc30029c4781fdc78b1f4f7963c9f86b897e42a77817ce4cc3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2005</creationdate><topic>Adult</topic><topic>Aged</topic><topic>Arthroplasty, Replacement, Hip - methods</topic><topic>Biological and medical sciences</topic><topic>Bone Diseases - surgery</topic><topic>Cadaver</topic><topic>Diseases of the osteoarticular system</topic><topic>Equipment Failure Analysis - methods</topic><topic>Female</topic><topic>Femur - surgery</topic><topic>Hip Joint - surgery</topic><topic>Hip Prosthesis</topic><topic>Humans</topic><topic>Joint Diseases - surgery</topic><topic>Male</topic><topic>Medical sciences</topic><topic>Middle Aged</topic><topic>Orthopedic surgery</topic><topic>Prosthesis Design</topic><topic>Prosthesis Failure</topic><topic>Reproducibility of Results</topic><topic>Software</topic><topic>Surgery (general aspects). Transplantations, organ and tissue grafts. Graft diseases</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>BEAULE, P. E</creatorcontrib><creatorcontrib>KRISMER, M</creatorcontrib><creatorcontrib>MAYRHOFER, P</creatorcontrib><creatorcontrib>WANNER, S</creatorcontrib><creatorcontrib>LE DUFF, M</creatorcontrib><creatorcontrib>MATTESICH, M</creatorcontrib><creatorcontrib>STOECKL, B</creatorcontrib><creatorcontrib>AMSTUTZ, H. C</creatorcontrib><creatorcontrib>BIEDERMANN, R</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>ProQuest Health & Medical Complete (Alumni)</collection><collection>Nursing & Allied Health Premium</collection><collection>MEDLINE - Academic</collection><jtitle>Journal of bone and joint surgery. British volume</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>BEAULE, P. E</au><au>KRISMER, M</au><au>MAYRHOFER, P</au><au>WANNER, S</au><au>LE DUFF, M</au><au>MATTESICH, M</au><au>STOECKL, B</au><au>AMSTUTZ, H. C</au><au>BIEDERMANN, R</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>EBRA-FCA for measurement of migration of the femoral component in surface arthroplasty of the hip</atitle><jtitle>Journal of bone and joint surgery. British volume</jtitle><addtitle>J Bone Joint Surg Br</addtitle><date>2005-05-01</date><risdate>2005</risdate><volume>87</volume><issue>5</issue><spage>741</spage><epage>744</epage><pages>741-744</pages><issn>0301-620X</issn><issn>2049-4394</issn><eissn>2044-5377</eissn><eissn>2049-4408</eissn><coden>JBSUAK</coden><abstract>Studies on the migration of an implant may be the only way of monitoring the early performance of metal-on-metal prostheses. The Ein Bild Roentgen Analyse--femoral component analysis (EBRA-FCA) method was adapted to measure migration of the femoral component in a metal-on-metal surface arthroplasty of the hip using standard antero-posterior radiographs. In order to determine the accuracy and precision of this method a prosthesis was implanted into cadaver bones. Eleven series of radiographs were used to perform a zero-migration study. After adjustment of the femoral component to simulate migration of 3 mm the radiographs were repeated. All were measured independently by three different observers. The accuracy of the method was found to be +/- 1.6 mm for the x-direction and +/- 2 mm for the y-direction (95% percentile). The method was validated using 28 hips with a minimum follow-up of 3.5 years after arthroplasty. Seventeen were sound, but 11 had failed because of loosening of the femoral component. The normal (control) group had a different pattern of migration compared with that of the loose group. At 29.2 months, the control group showed a mean migration of 1.62 mm and 1.05 mm compared with 4.39 mm and 4.05 mm in the failed group, for the centre of the head and the tip of the stem, respectively (p = 0.001). In the failed group, the mean time to migration greater than 2 mm was earlier than the onset of clinical symptoms or radiological evidence of failure, 19.1 versus 32.2 months (p = 0.001) and 24.8 months (p = 0.012), respectively. EBRA-FCA is a reliable and valid tool for measuring migration of the femoral component after surface arthroplasty and can be used to predict early failure of the implant. It may be of value in determining the long-term performance of surface arthroplasty.</abstract><cop>London</cop><pub>British Editorial Society of Bone and Joint Surgery</pub><pmid>15855382</pmid><doi>10.1302/0301-620X.87B5.15377</doi><tpages>4</tpages><edition>British volume</edition><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0301-620X |
ispartof | Journal of bone and joint surgery. British volume, 2005-05, Vol.87 (5), p.741-744 |
issn | 0301-620X 2049-4394 2044-5377 2049-4408 |
language | eng |
recordid | cdi_proquest_miscellaneous_67783572 |
source | MEDLINE; Alma/SFX Local Collection |
subjects | Adult Aged Arthroplasty, Replacement, Hip - methods Biological and medical sciences Bone Diseases - surgery Cadaver Diseases of the osteoarticular system Equipment Failure Analysis - methods Female Femur - surgery Hip Joint - surgery Hip Prosthesis Humans Joint Diseases - surgery Male Medical sciences Middle Aged Orthopedic surgery Prosthesis Design Prosthesis Failure Reproducibility of Results Software Surgery (general aspects). Transplantations, organ and tissue grafts. Graft diseases |
title | EBRA-FCA for measurement of migration of the femoral component in surface arthroplasty of the hip |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-18T23%3A08%3A23IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=EBRA-FCA%20for%20measurement%20of%20migration%20of%20the%20femoral%20component%20in%20surface%20arthroplasty%20of%20the%20hip&rft.jtitle=Journal%20of%20bone%20and%20joint%20surgery.%20British%20volume&rft.au=BEAULE,%20P.%20E&rft.date=2005-05-01&rft.volume=87&rft.issue=5&rft.spage=741&rft.epage=744&rft.pages=741-744&rft.issn=0301-620X&rft.eissn=2044-5377&rft.coden=JBSUAK&rft_id=info:doi/10.1302/0301-620X.87B5.15377&rft_dat=%3Cproquest_cross%3E846355761%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=229672960&rft_id=info:pmid/15855382&rfr_iscdi=true |