Electrochemical corrosion and metal ion release from Co-Cr-Mo prosthesis with titanium plasma spray coating

The corrosion behavior of CoCrMo implants with rough titanium coatings, applied by different suppliers by either sintering or vacuum plasma spraying, has been evaluated and compared with uncoated material. The open-circuit potential, corrosion current and polarization resistance were determined by e...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Biomaterials 2005-08, Vol.26 (23), p.4747-4756
Hauptverfasser: Reclaru, Lucien, Eschler, Pierre-Yves, Lerf, Reto, Blatter, Andreas
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:The corrosion behavior of CoCrMo implants with rough titanium coatings, applied by different suppliers by either sintering or vacuum plasma spraying, has been evaluated and compared with uncoated material. The open-circuit potential, corrosion current and polarization resistance were determined by electrochemical techniques. The Co, Cr and Ti ions released from the samples into the electrolyte during a potentiostatic extraction technique were analyzed using ICP-MS. The Ti coatings from the different suppliers showed a different porous morphology, and the implants exhibited a distinct corrosion activity, underlining the importance of the coating process parameters. Among the titanium coated samples, the one with the sintered overcoat turned out to be the most resistant. Yet, on an absolute scale, they all showed a corrosion resistance inferior to that of uncoated CoCrMo or wrought titanium.
ISSN:0142-9612
1878-5905
DOI:10.1016/j.biomaterials.2005.01.004