Optimum Design Parameters for a Therapist-Constructed Positive-Expiratory-Pressure Therapy Bottle Device

Positive-expiratory-pressure (PEP) therapy uses positive airway pressure generated by a either a fixed-orifice resistor or a threshold resistor. We hypothesized that tubing diameter and length, and the diameter of the PEP bottle's air-escape orifice would impact the PEP pressure delivered to th...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Respiratory care 2009-04, Vol.54 (4), p.504-508
Hauptverfasser: GEMERASCA MESTRINER, Régis, FERNANDES, Rafael Oliveira, STEFFEN, Luis Carlos, FAGUNDES DONADIO, Marcio Vinicius
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 508
container_issue 4
container_start_page 504
container_title Respiratory care
container_volume 54
creator GEMERASCA MESTRINER, Régis
FERNANDES, Rafael Oliveira
STEFFEN, Luis Carlos
FAGUNDES DONADIO, Marcio Vinicius
description Positive-expiratory-pressure (PEP) therapy uses positive airway pressure generated by a either a fixed-orifice resistor or a threshold resistor. We hypothesized that tubing diameter and length, and the diameter of the PEP bottle's air-escape orifice would impact the PEP pressure delivered to the airway and determine whether the PEP bottle acts as a threshold resistor or a fixed-orifice resistor. We designed a model composed of a bottle partially filled with water, a compressed air source, a pneumotachometer, and a manometer, to evaluate the effects of various tubing diameters (range 2-25 mm inner diameter) and lengths (range 20-80 cm long). In the first set of experiments, the PEP bottle had an open top, so there was no pressure other than the atmospheric pressure against the air escaping from the immersed tubing. The distal tip of the tube was 10 cm below the surface of the water (ie, a water-column pressure of 10 cm H(2)O), and we tested flows of 1, 5, 10, 15, 20, and 25 L/min. In the second set of experiments we tested a PEP bottle, the top of which was closed except for an air-escape orifice (4, 6, 8, 9, or 10 mm). With tubing of 2-6 mm inner diameter, the length of the tubing and the flow significantly affected the PEP pressure (ie, the system was not a threshold resistor). With tubing > or = 8 mm inner diameter there were no significant PEP-pressure differences with any of the tubing lengths or flows tested, which indicates a threshold-resistor system. The 4-mm and 6-mm air-escape orifices significantly increased the PEP pressure, whereas the 8 mm air-escape orifice did not increase the PEP pressure. To obtain a threshold-resistor PEP bottle system (ie, the PEP pressure is generated only by the water-column pressure), the tubing must be > or = 8 mm inner diameter, and the air-escape orifice must be > or = 8 mm.
format Article
fullrecord <record><control><sourceid>gale_proqu</sourceid><recordid>TN_cdi_proquest_miscellaneous_67086046</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A198171920</galeid><sourcerecordid>A198171920</sourcerecordid><originalsourceid>FETCH-LOGICAL-g337t-7e2e0754a7e95ef7c66c8abddf26cbceb39fa58b2a473c3eb935a60683953b0c3</originalsourceid><addsrcrecordid>eNpt0d9LwzAQB_Aiis7pvyAF0bdK2rRJ86jzJwy2h_lcrul1i6RNTVJx_70RpyhIHo6Ez_c4LnvRJBU5TSgr8v1oQkhGkpRm-VF07NxLuLK8EIfRUSpoxtOST6LNYvCqG7v4Fp1a9_ESLHTo0bq4NTaGeLVBC4NyPpmZ3nk7So9NvDROefWGyd37oCx4Y7fJ0qJzo8VdZBvfGO81hs5vSuJJdNCCdni6q9Po-f5uNXtM5ouHp9n1PFlTyn3CMUPCixw4igJbLhmTJdRN02ZM1hJrKlooyjqDnFNJsRa0AEZYSUVBayLpNLr86jtY8zqi81WnnEStoUczuopxUjKSswDPv-AaNFaqb423ID9xdZ2KMuWpyEhQV_-ocBrslDQ9tiq8_wlc_gpsELTfOKNHr8L6_sKz3aBj3WFTDVZ1YLfV998EcLED4CTo1kIvlftxWUpFsIJ-ABXqmGg</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>67086046</pqid></control><display><type>article</type><title>Optimum Design Parameters for a Therapist-Constructed Positive-Expiratory-Pressure Therapy Bottle Device</title><source>MEDLINE</source><source>EZB Electronic Journals Library</source><creator>GEMERASCA MESTRINER, Régis ; FERNANDES, Rafael Oliveira ; STEFFEN, Luis Carlos ; FAGUNDES DONADIO, Marcio Vinicius</creator><creatorcontrib>GEMERASCA MESTRINER, Régis ; FERNANDES, Rafael Oliveira ; STEFFEN, Luis Carlos ; FAGUNDES DONADIO, Marcio Vinicius</creatorcontrib><description>Positive-expiratory-pressure (PEP) therapy uses positive airway pressure generated by a either a fixed-orifice resistor or a threshold resistor. We hypothesized that tubing diameter and length, and the diameter of the PEP bottle's air-escape orifice would impact the PEP pressure delivered to the airway and determine whether the PEP bottle acts as a threshold resistor or a fixed-orifice resistor. We designed a model composed of a bottle partially filled with water, a compressed air source, a pneumotachometer, and a manometer, to evaluate the effects of various tubing diameters (range 2-25 mm inner diameter) and lengths (range 20-80 cm long). In the first set of experiments, the PEP bottle had an open top, so there was no pressure other than the atmospheric pressure against the air escaping from the immersed tubing. The distal tip of the tube was 10 cm below the surface of the water (ie, a water-column pressure of 10 cm H(2)O), and we tested flows of 1, 5, 10, 15, 20, and 25 L/min. In the second set of experiments we tested a PEP bottle, the top of which was closed except for an air-escape orifice (4, 6, 8, 9, or 10 mm). With tubing of 2-6 mm inner diameter, the length of the tubing and the flow significantly affected the PEP pressure (ie, the system was not a threshold resistor). With tubing &gt; or = 8 mm inner diameter there were no significant PEP-pressure differences with any of the tubing lengths or flows tested, which indicates a threshold-resistor system. The 4-mm and 6-mm air-escape orifices significantly increased the PEP pressure, whereas the 8 mm air-escape orifice did not increase the PEP pressure. To obtain a threshold-resistor PEP bottle system (ie, the PEP pressure is generated only by the water-column pressure), the tubing must be &gt; or = 8 mm inner diameter, and the air-escape orifice must be &gt; or = 8 mm.</description><identifier>ISSN: 0020-1324</identifier><identifier>EISSN: 1943-3654</identifier><identifier>PMID: 19327187</identifier><identifier>CODEN: RECACP</identifier><language>eng</language><publisher>Irving, TX: Daedalus</publisher><subject>Airway (Medicine) ; Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy ; Biological and medical sciences ; Care and treatment ; Emergency and intensive respiratory care ; Equipment and supplies ; Equipment Design ; Humans ; Intensive care medicine ; Materials Testing ; Medical sciences ; Positive-Pressure Respiration - instrumentation ; Respiratory therapy ; Respiratory Therapy - instrumentation</subject><ispartof>Respiratory care, 2009-04, Vol.54 (4), p.504-508</ispartof><rights>2009 INIST-CNRS</rights><rights>COPYRIGHT 2009 Daedalus Enterprises, Inc.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784</link.rule.ids><backlink>$$Uhttp://pascal-francis.inist.fr/vibad/index.php?action=getRecordDetail&amp;idt=21391939$$DView record in Pascal Francis$$Hfree_for_read</backlink><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/19327187$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>GEMERASCA MESTRINER, Régis</creatorcontrib><creatorcontrib>FERNANDES, Rafael Oliveira</creatorcontrib><creatorcontrib>STEFFEN, Luis Carlos</creatorcontrib><creatorcontrib>FAGUNDES DONADIO, Marcio Vinicius</creatorcontrib><title>Optimum Design Parameters for a Therapist-Constructed Positive-Expiratory-Pressure Therapy Bottle Device</title><title>Respiratory care</title><addtitle>Respir Care</addtitle><description>Positive-expiratory-pressure (PEP) therapy uses positive airway pressure generated by a either a fixed-orifice resistor or a threshold resistor. We hypothesized that tubing diameter and length, and the diameter of the PEP bottle's air-escape orifice would impact the PEP pressure delivered to the airway and determine whether the PEP bottle acts as a threshold resistor or a fixed-orifice resistor. We designed a model composed of a bottle partially filled with water, a compressed air source, a pneumotachometer, and a manometer, to evaluate the effects of various tubing diameters (range 2-25 mm inner diameter) and lengths (range 20-80 cm long). In the first set of experiments, the PEP bottle had an open top, so there was no pressure other than the atmospheric pressure against the air escaping from the immersed tubing. The distal tip of the tube was 10 cm below the surface of the water (ie, a water-column pressure of 10 cm H(2)O), and we tested flows of 1, 5, 10, 15, 20, and 25 L/min. In the second set of experiments we tested a PEP bottle, the top of which was closed except for an air-escape orifice (4, 6, 8, 9, or 10 mm). With tubing of 2-6 mm inner diameter, the length of the tubing and the flow significantly affected the PEP pressure (ie, the system was not a threshold resistor). With tubing &gt; or = 8 mm inner diameter there were no significant PEP-pressure differences with any of the tubing lengths or flows tested, which indicates a threshold-resistor system. The 4-mm and 6-mm air-escape orifices significantly increased the PEP pressure, whereas the 8 mm air-escape orifice did not increase the PEP pressure. To obtain a threshold-resistor PEP bottle system (ie, the PEP pressure is generated only by the water-column pressure), the tubing must be &gt; or = 8 mm inner diameter, and the air-escape orifice must be &gt; or = 8 mm.</description><subject>Airway (Medicine)</subject><subject>Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy</subject><subject>Biological and medical sciences</subject><subject>Care and treatment</subject><subject>Emergency and intensive respiratory care</subject><subject>Equipment and supplies</subject><subject>Equipment Design</subject><subject>Humans</subject><subject>Intensive care medicine</subject><subject>Materials Testing</subject><subject>Medical sciences</subject><subject>Positive-Pressure Respiration - instrumentation</subject><subject>Respiratory therapy</subject><subject>Respiratory Therapy - instrumentation</subject><issn>0020-1324</issn><issn>1943-3654</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2009</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpt0d9LwzAQB_Aiis7pvyAF0bdK2rRJ86jzJwy2h_lcrul1i6RNTVJx_70RpyhIHo6Ez_c4LnvRJBU5TSgr8v1oQkhGkpRm-VF07NxLuLK8EIfRUSpoxtOST6LNYvCqG7v4Fp1a9_ESLHTo0bq4NTaGeLVBC4NyPpmZ3nk7So9NvDROefWGyd37oCx4Y7fJ0qJzo8VdZBvfGO81hs5vSuJJdNCCdni6q9Po-f5uNXtM5ouHp9n1PFlTyn3CMUPCixw4igJbLhmTJdRN02ZM1hJrKlooyjqDnFNJsRa0AEZYSUVBayLpNLr86jtY8zqi81WnnEStoUczuopxUjKSswDPv-AaNFaqb423ID9xdZ2KMuWpyEhQV_-ocBrslDQ9tiq8_wlc_gpsELTfOKNHr8L6_sKz3aBj3WFTDVZ1YLfV998EcLED4CTo1kIvlftxWUpFsIJ-ABXqmGg</recordid><startdate>20090401</startdate><enddate>20090401</enddate><creator>GEMERASCA MESTRINER, Régis</creator><creator>FERNANDES, Rafael Oliveira</creator><creator>STEFFEN, Luis Carlos</creator><creator>FAGUNDES DONADIO, Marcio Vinicius</creator><general>Daedalus</general><general>Daedalus Enterprises, Inc</general><scope>IQODW</scope><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>7X8</scope></search><sort><creationdate>20090401</creationdate><title>Optimum Design Parameters for a Therapist-Constructed Positive-Expiratory-Pressure Therapy Bottle Device</title><author>GEMERASCA MESTRINER, Régis ; FERNANDES, Rafael Oliveira ; STEFFEN, Luis Carlos ; FAGUNDES DONADIO, Marcio Vinicius</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-g337t-7e2e0754a7e95ef7c66c8abddf26cbceb39fa58b2a473c3eb935a60683953b0c3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2009</creationdate><topic>Airway (Medicine)</topic><topic>Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy</topic><topic>Biological and medical sciences</topic><topic>Care and treatment</topic><topic>Emergency and intensive respiratory care</topic><topic>Equipment and supplies</topic><topic>Equipment Design</topic><topic>Humans</topic><topic>Intensive care medicine</topic><topic>Materials Testing</topic><topic>Medical sciences</topic><topic>Positive-Pressure Respiration - instrumentation</topic><topic>Respiratory therapy</topic><topic>Respiratory Therapy - instrumentation</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>GEMERASCA MESTRINER, Régis</creatorcontrib><creatorcontrib>FERNANDES, Rafael Oliveira</creatorcontrib><creatorcontrib>STEFFEN, Luis Carlos</creatorcontrib><creatorcontrib>FAGUNDES DONADIO, Marcio Vinicius</creatorcontrib><collection>Pascal-Francis</collection><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>MEDLINE - Academic</collection><jtitle>Respiratory care</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>GEMERASCA MESTRINER, Régis</au><au>FERNANDES, Rafael Oliveira</au><au>STEFFEN, Luis Carlos</au><au>FAGUNDES DONADIO, Marcio Vinicius</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Optimum Design Parameters for a Therapist-Constructed Positive-Expiratory-Pressure Therapy Bottle Device</atitle><jtitle>Respiratory care</jtitle><addtitle>Respir Care</addtitle><date>2009-04-01</date><risdate>2009</risdate><volume>54</volume><issue>4</issue><spage>504</spage><epage>508</epage><pages>504-508</pages><issn>0020-1324</issn><eissn>1943-3654</eissn><coden>RECACP</coden><abstract>Positive-expiratory-pressure (PEP) therapy uses positive airway pressure generated by a either a fixed-orifice resistor or a threshold resistor. We hypothesized that tubing diameter and length, and the diameter of the PEP bottle's air-escape orifice would impact the PEP pressure delivered to the airway and determine whether the PEP bottle acts as a threshold resistor or a fixed-orifice resistor. We designed a model composed of a bottle partially filled with water, a compressed air source, a pneumotachometer, and a manometer, to evaluate the effects of various tubing diameters (range 2-25 mm inner diameter) and lengths (range 20-80 cm long). In the first set of experiments, the PEP bottle had an open top, so there was no pressure other than the atmospheric pressure against the air escaping from the immersed tubing. The distal tip of the tube was 10 cm below the surface of the water (ie, a water-column pressure of 10 cm H(2)O), and we tested flows of 1, 5, 10, 15, 20, and 25 L/min. In the second set of experiments we tested a PEP bottle, the top of which was closed except for an air-escape orifice (4, 6, 8, 9, or 10 mm). With tubing of 2-6 mm inner diameter, the length of the tubing and the flow significantly affected the PEP pressure (ie, the system was not a threshold resistor). With tubing &gt; or = 8 mm inner diameter there were no significant PEP-pressure differences with any of the tubing lengths or flows tested, which indicates a threshold-resistor system. The 4-mm and 6-mm air-escape orifices significantly increased the PEP pressure, whereas the 8 mm air-escape orifice did not increase the PEP pressure. To obtain a threshold-resistor PEP bottle system (ie, the PEP pressure is generated only by the water-column pressure), the tubing must be &gt; or = 8 mm inner diameter, and the air-escape orifice must be &gt; or = 8 mm.</abstract><cop>Irving, TX</cop><pub>Daedalus</pub><pmid>19327187</pmid><tpages>5</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0020-1324
ispartof Respiratory care, 2009-04, Vol.54 (4), p.504-508
issn 0020-1324
1943-3654
language eng
recordid cdi_proquest_miscellaneous_67086046
source MEDLINE; EZB Electronic Journals Library
subjects Airway (Medicine)
Anesthesia. Intensive care medicine. Transfusions. Cell therapy and gene therapy
Biological and medical sciences
Care and treatment
Emergency and intensive respiratory care
Equipment and supplies
Equipment Design
Humans
Intensive care medicine
Materials Testing
Medical sciences
Positive-Pressure Respiration - instrumentation
Respiratory therapy
Respiratory Therapy - instrumentation
title Optimum Design Parameters for a Therapist-Constructed Positive-Expiratory-Pressure Therapy Bottle Device
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-07T17%3A59%3A55IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_proqu&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Optimum%20Design%20Parameters%20for%20a%20Therapist-Constructed%20Positive-Expiratory-Pressure%20Therapy%20Bottle%20Device&rft.jtitle=Respiratory%20care&rft.au=GEMERASCA%20MESTRINER,%20R%C3%A9gis&rft.date=2009-04-01&rft.volume=54&rft.issue=4&rft.spage=504&rft.epage=508&rft.pages=504-508&rft.issn=0020-1324&rft.eissn=1943-3654&rft.coden=RECACP&rft_id=info:doi/&rft_dat=%3Cgale_proqu%3EA198171920%3C/gale_proqu%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=67086046&rft_id=info:pmid/19327187&rft_galeid=A198171920&rfr_iscdi=true