Assessment of endotracheal cuff pressure by continuous monitoring: a pilot study

Endotracheal tube cuff pressure must be maintained within a narrow therapeutic range to prevent complications. Cuff pressure is measured and adjusted intermittently. To assess the accuracy and feasibility of continuous monitoring of cuff pressure, describe changes in cuff pressure over time, and ide...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of critical care 2009-03, Vol.18 (2), p.133-143
Hauptverfasser: Sole, Mary Lou, Penoyer, Daleen Aragon, Su, XioaGang, Jimenez, Edgar, Kalita, Samar J, Poalillo, Elizabeth, Byers, Jacqueline F, Bennett, Melody, Ludy, Jeffery E
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 143
container_issue 2
container_start_page 133
container_title American journal of critical care
container_volume 18
creator Sole, Mary Lou
Penoyer, Daleen Aragon
Su, XioaGang
Jimenez, Edgar
Kalita, Samar J
Poalillo, Elizabeth
Byers, Jacqueline F
Bennett, Melody
Ludy, Jeffery E
description Endotracheal tube cuff pressure must be maintained within a narrow therapeutic range to prevent complications. Cuff pressure is measured and adjusted intermittently. To assess the accuracy and feasibility of continuous monitoring of cuff pressure, describe changes in cuff pressure over time, and identify clinical factors that influence cuff pressure. In a pilot study, data were collected for a mean of 9.3 hours on 10 patients who were orally intubated and receiving mechanical ventilation. Sixty percent of the patients were white, mean age was 55 years, and mean intubation time was 2.8 days. The initial cuff pressure was adjusted to a minimum of 20 cm H2O. The pilot balloon of the endotracheal tube was connected to a transducer and a pressure monitor. Cuff pressure was recorded every 0.008 seconds during a typical 12-hour shift and was reduced to 1-minute means. Patient care activities and interventions were recorded on a personal digital assistant. Values obtained with the cufflator-manometer and the transducer were congruent. Only 54% of cuff pressure measurements were within the recommended range of 20 to 30 cm H2O. The cuff pressure was high in 16% of measurements and low in 30%. No statistically significant changes over time were noted. Endotracheal suctioning, coughing, and positioning affected cuff pressure. Continuous monitoring of cuff pressure is feasible, accurate, and safe. Cuff pressures vary widely among patients.
doi_str_mv 10.4037/ajcc2009441
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_66979087</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>66979087</sourcerecordid><originalsourceid>FETCH-LOGICAL-c318t-7b79d78afbc1c32cf65db16dedce87f9c23fc9a8ca45d25eda9959e0b90392a03</originalsourceid><addsrcrecordid>eNpNkEtLxDAUhYMozji6ci9ZuZFqHm3TuJPBFwzoQsFdSW9utEPbjEm6mH9vZQZ0dQ_cj8PhI-Scs-ucSXVj1gCCMZ3n_IDMuZYqU5x9HE6ZlSKTosxn5CTGNWM8r5Q6JjOuRVFwJufk9S5GjLHHIVHvKA7Wp2DgC01HYXSObsL0HgPSZkvBD6kdRj9G2vuhTT60w-ctNXTTdj7RmEa7PSVHznQRz_Z3Qd4f7t-WT9nq5fF5ebfKQPIqZapR2qrKuAY4SAGuLGzDS4sWsFJOg5AOtKnA5IUVBVqjdaGRNZpJLQyTC3K5690E_z1iTHXfRsCuMwNOA-uy1EqzSk3g1Q6E4GMM6OpNaHsTtjVn9a_A-p_Aib7Y145Nj_aP3RuTP7A1bgc</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>66979087</pqid></control><display><type>article</type><title>Assessment of endotracheal cuff pressure by continuous monitoring: a pilot study</title><source>MEDLINE</source><source>Journals@Ovid Complete</source><creator>Sole, Mary Lou ; Penoyer, Daleen Aragon ; Su, XioaGang ; Jimenez, Edgar ; Kalita, Samar J ; Poalillo, Elizabeth ; Byers, Jacqueline F ; Bennett, Melody ; Ludy, Jeffery E</creator><creatorcontrib>Sole, Mary Lou ; Penoyer, Daleen Aragon ; Su, XioaGang ; Jimenez, Edgar ; Kalita, Samar J ; Poalillo, Elizabeth ; Byers, Jacqueline F ; Bennett, Melody ; Ludy, Jeffery E</creatorcontrib><description>Endotracheal tube cuff pressure must be maintained within a narrow therapeutic range to prevent complications. Cuff pressure is measured and adjusted intermittently. To assess the accuracy and feasibility of continuous monitoring of cuff pressure, describe changes in cuff pressure over time, and identify clinical factors that influence cuff pressure. In a pilot study, data were collected for a mean of 9.3 hours on 10 patients who were orally intubated and receiving mechanical ventilation. Sixty percent of the patients were white, mean age was 55 years, and mean intubation time was 2.8 days. The initial cuff pressure was adjusted to a minimum of 20 cm H2O. The pilot balloon of the endotracheal tube was connected to a transducer and a pressure monitor. Cuff pressure was recorded every 0.008 seconds during a typical 12-hour shift and was reduced to 1-minute means. Patient care activities and interventions were recorded on a personal digital assistant. Values obtained with the cufflator-manometer and the transducer were congruent. Only 54% of cuff pressure measurements were within the recommended range of 20 to 30 cm H2O. The cuff pressure was high in 16% of measurements and low in 30%. No statistically significant changes over time were noted. Endotracheal suctioning, coughing, and positioning affected cuff pressure. Continuous monitoring of cuff pressure is feasible, accurate, and safe. Cuff pressures vary widely among patients.</description><identifier>ISSN: 1062-3264</identifier><identifier>EISSN: 1937-710X</identifier><identifier>DOI: 10.4037/ajcc2009441</identifier><identifier>PMID: 19255103</identifier><language>eng</language><publisher>United States</publisher><subject>Critical Care - methods ; Female ; Humans ; Intubation, Intratracheal ; Male ; Middle Aged ; Monitoring, Physiologic - methods ; Nursing ; Pilot Projects ; Pressure ; Reproducibility of Results ; Respiration, Artificial</subject><ispartof>American journal of critical care, 2009-03, Vol.18 (2), p.133-143</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c318t-7b79d78afbc1c32cf65db16dedce87f9c23fc9a8ca45d25eda9959e0b90392a03</citedby></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,777,781,27905,27906</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/19255103$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Sole, Mary Lou</creatorcontrib><creatorcontrib>Penoyer, Daleen Aragon</creatorcontrib><creatorcontrib>Su, XioaGang</creatorcontrib><creatorcontrib>Jimenez, Edgar</creatorcontrib><creatorcontrib>Kalita, Samar J</creatorcontrib><creatorcontrib>Poalillo, Elizabeth</creatorcontrib><creatorcontrib>Byers, Jacqueline F</creatorcontrib><creatorcontrib>Bennett, Melody</creatorcontrib><creatorcontrib>Ludy, Jeffery E</creatorcontrib><title>Assessment of endotracheal cuff pressure by continuous monitoring: a pilot study</title><title>American journal of critical care</title><addtitle>Am J Crit Care</addtitle><description>Endotracheal tube cuff pressure must be maintained within a narrow therapeutic range to prevent complications. Cuff pressure is measured and adjusted intermittently. To assess the accuracy and feasibility of continuous monitoring of cuff pressure, describe changes in cuff pressure over time, and identify clinical factors that influence cuff pressure. In a pilot study, data were collected for a mean of 9.3 hours on 10 patients who were orally intubated and receiving mechanical ventilation. Sixty percent of the patients were white, mean age was 55 years, and mean intubation time was 2.8 days. The initial cuff pressure was adjusted to a minimum of 20 cm H2O. The pilot balloon of the endotracheal tube was connected to a transducer and a pressure monitor. Cuff pressure was recorded every 0.008 seconds during a typical 12-hour shift and was reduced to 1-minute means. Patient care activities and interventions were recorded on a personal digital assistant. Values obtained with the cufflator-manometer and the transducer were congruent. Only 54% of cuff pressure measurements were within the recommended range of 20 to 30 cm H2O. The cuff pressure was high in 16% of measurements and low in 30%. No statistically significant changes over time were noted. Endotracheal suctioning, coughing, and positioning affected cuff pressure. Continuous monitoring of cuff pressure is feasible, accurate, and safe. Cuff pressures vary widely among patients.</description><subject>Critical Care - methods</subject><subject>Female</subject><subject>Humans</subject><subject>Intubation, Intratracheal</subject><subject>Male</subject><subject>Middle Aged</subject><subject>Monitoring, Physiologic - methods</subject><subject>Nursing</subject><subject>Pilot Projects</subject><subject>Pressure</subject><subject>Reproducibility of Results</subject><subject>Respiration, Artificial</subject><issn>1062-3264</issn><issn>1937-710X</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2009</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpNkEtLxDAUhYMozji6ci9ZuZFqHm3TuJPBFwzoQsFdSW9utEPbjEm6mH9vZQZ0dQ_cj8PhI-Scs-ucSXVj1gCCMZ3n_IDMuZYqU5x9HE6ZlSKTosxn5CTGNWM8r5Q6JjOuRVFwJufk9S5GjLHHIVHvKA7Wp2DgC01HYXSObsL0HgPSZkvBD6kdRj9G2vuhTT60w-ctNXTTdj7RmEa7PSVHznQRz_Z3Qd4f7t-WT9nq5fF5ebfKQPIqZapR2qrKuAY4SAGuLGzDS4sWsFJOg5AOtKnA5IUVBVqjdaGRNZpJLQyTC3K5690E_z1iTHXfRsCuMwNOA-uy1EqzSk3g1Q6E4GMM6OpNaHsTtjVn9a_A-p_Aib7Y145Nj_aP3RuTP7A1bgc</recordid><startdate>200903</startdate><enddate>200903</enddate><creator>Sole, Mary Lou</creator><creator>Penoyer, Daleen Aragon</creator><creator>Su, XioaGang</creator><creator>Jimenez, Edgar</creator><creator>Kalita, Samar J</creator><creator>Poalillo, Elizabeth</creator><creator>Byers, Jacqueline F</creator><creator>Bennett, Melody</creator><creator>Ludy, Jeffery E</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>200903</creationdate><title>Assessment of endotracheal cuff pressure by continuous monitoring: a pilot study</title><author>Sole, Mary Lou ; Penoyer, Daleen Aragon ; Su, XioaGang ; Jimenez, Edgar ; Kalita, Samar J ; Poalillo, Elizabeth ; Byers, Jacqueline F ; Bennett, Melody ; Ludy, Jeffery E</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c318t-7b79d78afbc1c32cf65db16dedce87f9c23fc9a8ca45d25eda9959e0b90392a03</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2009</creationdate><topic>Critical Care - methods</topic><topic>Female</topic><topic>Humans</topic><topic>Intubation, Intratracheal</topic><topic>Male</topic><topic>Middle Aged</topic><topic>Monitoring, Physiologic - methods</topic><topic>Nursing</topic><topic>Pilot Projects</topic><topic>Pressure</topic><topic>Reproducibility of Results</topic><topic>Respiration, Artificial</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Sole, Mary Lou</creatorcontrib><creatorcontrib>Penoyer, Daleen Aragon</creatorcontrib><creatorcontrib>Su, XioaGang</creatorcontrib><creatorcontrib>Jimenez, Edgar</creatorcontrib><creatorcontrib>Kalita, Samar J</creatorcontrib><creatorcontrib>Poalillo, Elizabeth</creatorcontrib><creatorcontrib>Byers, Jacqueline F</creatorcontrib><creatorcontrib>Bennett, Melody</creatorcontrib><creatorcontrib>Ludy, Jeffery E</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>American journal of critical care</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Sole, Mary Lou</au><au>Penoyer, Daleen Aragon</au><au>Su, XioaGang</au><au>Jimenez, Edgar</au><au>Kalita, Samar J</au><au>Poalillo, Elizabeth</au><au>Byers, Jacqueline F</au><au>Bennett, Melody</au><au>Ludy, Jeffery E</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Assessment of endotracheal cuff pressure by continuous monitoring: a pilot study</atitle><jtitle>American journal of critical care</jtitle><addtitle>Am J Crit Care</addtitle><date>2009-03</date><risdate>2009</risdate><volume>18</volume><issue>2</issue><spage>133</spage><epage>143</epage><pages>133-143</pages><issn>1062-3264</issn><eissn>1937-710X</eissn><abstract>Endotracheal tube cuff pressure must be maintained within a narrow therapeutic range to prevent complications. Cuff pressure is measured and adjusted intermittently. To assess the accuracy and feasibility of continuous monitoring of cuff pressure, describe changes in cuff pressure over time, and identify clinical factors that influence cuff pressure. In a pilot study, data were collected for a mean of 9.3 hours on 10 patients who were orally intubated and receiving mechanical ventilation. Sixty percent of the patients were white, mean age was 55 years, and mean intubation time was 2.8 days. The initial cuff pressure was adjusted to a minimum of 20 cm H2O. The pilot balloon of the endotracheal tube was connected to a transducer and a pressure monitor. Cuff pressure was recorded every 0.008 seconds during a typical 12-hour shift and was reduced to 1-minute means. Patient care activities and interventions were recorded on a personal digital assistant. Values obtained with the cufflator-manometer and the transducer were congruent. Only 54% of cuff pressure measurements were within the recommended range of 20 to 30 cm H2O. The cuff pressure was high in 16% of measurements and low in 30%. No statistically significant changes over time were noted. Endotracheal suctioning, coughing, and positioning affected cuff pressure. Continuous monitoring of cuff pressure is feasible, accurate, and safe. Cuff pressures vary widely among patients.</abstract><cop>United States</cop><pmid>19255103</pmid><doi>10.4037/ajcc2009441</doi><tpages>11</tpages></addata></record>
fulltext fulltext
identifier ISSN: 1062-3264
ispartof American journal of critical care, 2009-03, Vol.18 (2), p.133-143
issn 1062-3264
1937-710X
language eng
recordid cdi_proquest_miscellaneous_66979087
source MEDLINE; Journals@Ovid Complete
subjects Critical Care - methods
Female
Humans
Intubation, Intratracheal
Male
Middle Aged
Monitoring, Physiologic - methods
Nursing
Pilot Projects
Pressure
Reproducibility of Results
Respiration, Artificial
title Assessment of endotracheal cuff pressure by continuous monitoring: a pilot study
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-19T18%3A38%3A36IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Assessment%20of%20endotracheal%20cuff%20pressure%20by%20continuous%20monitoring:%20a%20pilot%20study&rft.jtitle=American%20journal%20of%20critical%20care&rft.au=Sole,%20Mary%20Lou&rft.date=2009-03&rft.volume=18&rft.issue=2&rft.spage=133&rft.epage=143&rft.pages=133-143&rft.issn=1062-3264&rft.eissn=1937-710X&rft_id=info:doi/10.4037/ajcc2009441&rft_dat=%3Cproquest_cross%3E66979087%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=66979087&rft_id=info:pmid/19255103&rfr_iscdi=true