Quality Control of Portal Imaging with PTW EPID QC PHANTOM
Purpose: Quality assurance (QA) and quality control (QC) of different electronic portal imaging devices (EPID) and portal images with the PTW EPID QC PHANTOM ® . Material and Methods: Characteristic properties of images of different file formats were measured on Siemens OptiVue500aSi ® , Siemens Bea...
Gespeichert in:
Veröffentlicht in: | Strahlentherapie und Onkologie 2009, Vol.185 (1), p.56-60 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose:
Quality assurance (QA) and quality control (QC) of different electronic portal imaging devices (EPID) and portal images with the PTW EPID QC PHANTOM
®
.
Material and Methods:
Characteristic properties of images of different file formats were measured on Siemens OptiVue500aSi
®
, Siemens BeamView Plus
®
, Elekta iView
®
, and Varian PortalVision™ and analyzed with the epidSoft
®
2.0 program in four radiation therapy centers. The portal images were taken with Kodak X-OMAT V
®
and the Kodak Portal Localisation ReadyPack
®
films and evaluated with the same program.
Results:
The optimal exposition both for EPIDs and portal films of different kind was determined. For double exposition, the 2+1 MU values can be recommended in the case of Siemens OptiVue500aSi
®
, Elekta iView
®
and Kodak Portal Localisation ReadyPack
®
films, while for Siemens BeamView Plus
®
, Varian PortalVision™ and Kodak X-OMAT V
®
film 7+7 MU is recommended.
Conclusion:
The PTW EPID QC PHANTOM
®
can be used not only for amorphous silicon EPIDs but also for images taken with a video-based system or by using an ionization chamber matrix or for portal film. For analysis of QC tests, a standardized format (used at the acceptance test) should be applied, as the results are dependent on the file format used. |
---|---|
ISSN: | 0179-7158 1439-099X |
DOI: | 10.1007/s00066-009-1905-z |