Analysis of Factors Associated with Volumetric Data Errors in Gamma Knife Radiosurgery

Object: Gamma knife (GK) surgery is an important part of the treatment armamentarium for benign and malignant brain tumors. In general, quantitative volumetrical analysis of the tumor on neuroimaging studies is the most reliable method of assessment of the tumor’s response and is critical for accura...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Stereotactic and functional neurosurgery 2009-01, Vol.87 (1), p.1-7
Hauptverfasser: Yang, Dar-Yu, Sheehan, Jason, Liu, Yung-Sheng, ChangLai, Sheng-Pin, Pan, Hung-Chuan, Chen, Chun-Jung, Chou, Ming-Chih
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:Object: Gamma knife (GK) surgery is an important part of the treatment armamentarium for benign and malignant brain tumors. In general, quantitative volumetrical analysis of the tumor on neuroimaging studies is the most reliable method of assessment of the tumor’s response and is critical for accurate dose planning. This study evaluated various factors contributing to volumetric data error of tumors treated with GK radiosurgery. Method: Three differently shaped phantoms (spherical, rectangular, and irregular morphology) were created by immersing like shaped objects into 2% agarose gel. The volumes of phantoms were measured by laser scanning with errors