Lead stuck (frozen) in header: salvage by bone cutter versus other techniques. jfisher@montefiore.org

It is occasionally difficult to disconnect leads from headers at the time of pulse generator replacement without injuring the fragile leads. Over a 2.5-year period we encountered this problem in six cases (1.7% of pulse generator replacements). The posterior portion of the header was clipped off usi...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Pacing and clinical electrophysiology 2004-08, Vol.27 (8), p.1136-1143
Hauptverfasser: Fisher, John D, Lapman, Peter, Kim, Soo G, Ferrick, Kevin J, Gross, Jay N, Palma, Eugen C, Delvecchio, Alexander
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 1143
container_issue 8
container_start_page 1136
container_title Pacing and clinical electrophysiology
container_volume 27
creator Fisher, John D
Lapman, Peter
Kim, Soo G
Ferrick, Kevin J
Gross, Jay N
Palma, Eugen C
Delvecchio, Alexander
description It is occasionally difficult to disconnect leads from headers at the time of pulse generator replacement without injuring the fragile leads. Over a 2.5-year period we encountered this problem in six cases (1.7% of pulse generator replacements). The posterior portion of the header was clipped off using an orthopedic bone cutter in four cases. The cut was aligned with the deep end of the lead socket in the header. A metal rod was then used to push the lead out of the socket. Bench testing of alternative methods was done on previously explanted pulse generators that were firmly held in a vice. Motorized microtools were used to drill holes from the end of the header to the deep end of the socket; or with a rotary saw attachment to slice off the back of the header, allowing a retained lead to be pushed out. The latter was also done with a hand held razor saw, and attempts were made with a scalpel. Lead removal in the clinical cases was accomplished quickly in the four cases using the bone-cutter, without trauma to the lead. Bench testing results varied. The bone cutter was the most efficient method for most brands, but was ineffective on one. The motorized tool was difficult to position, produced sprays of plastic particles, and would have been risky in a clinical setting. The razor saw was difficult to use safely, or efficiently, except in some headers that resisted the bone cutter. The scalpel failed except in one "soft header" pacemaker. An orthopedic bone cutter is a useful tool for removing a retained lead from a pulse generator header. Different header designs and materials necessitate knowledge of several lead detachment methods.
format Article
fullrecord <record><control><sourceid>proquest_pubme</sourceid><recordid>TN_cdi_proquest_miscellaneous_66792196</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>66792196</sourcerecordid><originalsourceid>FETCH-LOGICAL-p544-25c29e6d29f94f51268692f780539bb3863a93a651267547ebbf61622daab30f3</originalsourceid><addsrcrecordid>eNo1UDtPwzAY9ACipfAXkCcEQyrHr8RMoKo8pEgs3SM7-dykJHGxnUrl15OKMp3uodPpLtCcpDxLcparGboOYUcIkYSLKzRLBSNCST5HUICucYhj9YUfrHc_MDzidsDNJIN_wkF3B70FbI7YuAFwNcYIHh_AhzFgF5uJRKiaof0eISzxzrZh0p57N0SwrfOwdH57gy6t7gLcnnGBNq_rzeo9KT7fPlYvRbIXnCdUVFSBrKmyiluRUplLRW2WE8GUMSyXTCum5cnJBM_AGCtTSWmttWHEsgW6_6vde3eaE8u-DRV0nR7AjaGUMlM0VXIK3p2Do-mhLve-7bU_lv-_sF9thV3g</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>66792196</pqid></control><display><type>article</type><title>Lead stuck (frozen) in header: salvage by bone cutter versus other techniques. jfisher@montefiore.org</title><source>MEDLINE</source><source>Wiley Online Library Journals Frontfile Complete</source><creator>Fisher, John D ; Lapman, Peter ; Kim, Soo G ; Ferrick, Kevin J ; Gross, Jay N ; Palma, Eugen C ; Delvecchio, Alexander</creator><creatorcontrib>Fisher, John D ; Lapman, Peter ; Kim, Soo G ; Ferrick, Kevin J ; Gross, Jay N ; Palma, Eugen C ; Delvecchio, Alexander</creatorcontrib><description>It is occasionally difficult to disconnect leads from headers at the time of pulse generator replacement without injuring the fragile leads. Over a 2.5-year period we encountered this problem in six cases (1.7% of pulse generator replacements). The posterior portion of the header was clipped off using an orthopedic bone cutter in four cases. The cut was aligned with the deep end of the lead socket in the header. A metal rod was then used to push the lead out of the socket. Bench testing of alternative methods was done on previously explanted pulse generators that were firmly held in a vice. Motorized microtools were used to drill holes from the end of the header to the deep end of the socket; or with a rotary saw attachment to slice off the back of the header, allowing a retained lead to be pushed out. The latter was also done with a hand held razor saw, and attempts were made with a scalpel. Lead removal in the clinical cases was accomplished quickly in the four cases using the bone-cutter, without trauma to the lead. Bench testing results varied. The bone cutter was the most efficient method for most brands, but was ineffective on one. The motorized tool was difficult to position, produced sprays of plastic particles, and would have been risky in a clinical setting. The razor saw was difficult to use safely, or efficiently, except in some headers that resisted the bone cutter. The scalpel failed except in one "soft header" pacemaker. An orthopedic bone cutter is a useful tool for removing a retained lead from a pulse generator header. Different header designs and materials necessitate knowledge of several lead detachment methods.</description><identifier>ISSN: 0147-8389</identifier><identifier>PMID: 15305964</identifier><language>eng</language><publisher>United States</publisher><subject>Device Removal - methods ; Equipment Failure ; Humans ; Pacemaker, Artificial ; Surgical Instruments</subject><ispartof>Pacing and clinical electrophysiology, 2004-08, Vol.27 (8), p.1136-1143</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/15305964$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Fisher, John D</creatorcontrib><creatorcontrib>Lapman, Peter</creatorcontrib><creatorcontrib>Kim, Soo G</creatorcontrib><creatorcontrib>Ferrick, Kevin J</creatorcontrib><creatorcontrib>Gross, Jay N</creatorcontrib><creatorcontrib>Palma, Eugen C</creatorcontrib><creatorcontrib>Delvecchio, Alexander</creatorcontrib><title>Lead stuck (frozen) in header: salvage by bone cutter versus other techniques. jfisher@montefiore.org</title><title>Pacing and clinical electrophysiology</title><addtitle>Pacing Clin Electrophysiol</addtitle><description>It is occasionally difficult to disconnect leads from headers at the time of pulse generator replacement without injuring the fragile leads. Over a 2.5-year period we encountered this problem in six cases (1.7% of pulse generator replacements). The posterior portion of the header was clipped off using an orthopedic bone cutter in four cases. The cut was aligned with the deep end of the lead socket in the header. A metal rod was then used to push the lead out of the socket. Bench testing of alternative methods was done on previously explanted pulse generators that were firmly held in a vice. Motorized microtools were used to drill holes from the end of the header to the deep end of the socket; or with a rotary saw attachment to slice off the back of the header, allowing a retained lead to be pushed out. The latter was also done with a hand held razor saw, and attempts were made with a scalpel. Lead removal in the clinical cases was accomplished quickly in the four cases using the bone-cutter, without trauma to the lead. Bench testing results varied. The bone cutter was the most efficient method for most brands, but was ineffective on one. The motorized tool was difficult to position, produced sprays of plastic particles, and would have been risky in a clinical setting. The razor saw was difficult to use safely, or efficiently, except in some headers that resisted the bone cutter. The scalpel failed except in one "soft header" pacemaker. An orthopedic bone cutter is a useful tool for removing a retained lead from a pulse generator header. Different header designs and materials necessitate knowledge of several lead detachment methods.</description><subject>Device Removal - methods</subject><subject>Equipment Failure</subject><subject>Humans</subject><subject>Pacemaker, Artificial</subject><subject>Surgical Instruments</subject><issn>0147-8389</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2004</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNo1UDtPwzAY9ACipfAXkCcEQyrHr8RMoKo8pEgs3SM7-dykJHGxnUrl15OKMp3uodPpLtCcpDxLcparGboOYUcIkYSLKzRLBSNCST5HUICucYhj9YUfrHc_MDzidsDNJIN_wkF3B70FbI7YuAFwNcYIHh_AhzFgF5uJRKiaof0eISzxzrZh0p57N0SwrfOwdH57gy6t7gLcnnGBNq_rzeo9KT7fPlYvRbIXnCdUVFSBrKmyiluRUplLRW2WE8GUMSyXTCum5cnJBM_AGCtTSWmttWHEsgW6_6vde3eaE8u-DRV0nR7AjaGUMlM0VXIK3p2Do-mhLve-7bU_lv-_sF9thV3g</recordid><startdate>200408</startdate><enddate>200408</enddate><creator>Fisher, John D</creator><creator>Lapman, Peter</creator><creator>Kim, Soo G</creator><creator>Ferrick, Kevin J</creator><creator>Gross, Jay N</creator><creator>Palma, Eugen C</creator><creator>Delvecchio, Alexander</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>7X8</scope></search><sort><creationdate>200408</creationdate><title>Lead stuck (frozen) in header: salvage by bone cutter versus other techniques. jfisher@montefiore.org</title><author>Fisher, John D ; Lapman, Peter ; Kim, Soo G ; Ferrick, Kevin J ; Gross, Jay N ; Palma, Eugen C ; Delvecchio, Alexander</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-p544-25c29e6d29f94f51268692f780539bb3863a93a651267547ebbf61622daab30f3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2004</creationdate><topic>Device Removal - methods</topic><topic>Equipment Failure</topic><topic>Humans</topic><topic>Pacemaker, Artificial</topic><topic>Surgical Instruments</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Fisher, John D</creatorcontrib><creatorcontrib>Lapman, Peter</creatorcontrib><creatorcontrib>Kim, Soo G</creatorcontrib><creatorcontrib>Ferrick, Kevin J</creatorcontrib><creatorcontrib>Gross, Jay N</creatorcontrib><creatorcontrib>Palma, Eugen C</creatorcontrib><creatorcontrib>Delvecchio, Alexander</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>MEDLINE - Academic</collection><jtitle>Pacing and clinical electrophysiology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Fisher, John D</au><au>Lapman, Peter</au><au>Kim, Soo G</au><au>Ferrick, Kevin J</au><au>Gross, Jay N</au><au>Palma, Eugen C</au><au>Delvecchio, Alexander</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Lead stuck (frozen) in header: salvage by bone cutter versus other techniques. jfisher@montefiore.org</atitle><jtitle>Pacing and clinical electrophysiology</jtitle><addtitle>Pacing Clin Electrophysiol</addtitle><date>2004-08</date><risdate>2004</risdate><volume>27</volume><issue>8</issue><spage>1136</spage><epage>1143</epage><pages>1136-1143</pages><issn>0147-8389</issn><abstract>It is occasionally difficult to disconnect leads from headers at the time of pulse generator replacement without injuring the fragile leads. Over a 2.5-year period we encountered this problem in six cases (1.7% of pulse generator replacements). The posterior portion of the header was clipped off using an orthopedic bone cutter in four cases. The cut was aligned with the deep end of the lead socket in the header. A metal rod was then used to push the lead out of the socket. Bench testing of alternative methods was done on previously explanted pulse generators that were firmly held in a vice. Motorized microtools were used to drill holes from the end of the header to the deep end of the socket; or with a rotary saw attachment to slice off the back of the header, allowing a retained lead to be pushed out. The latter was also done with a hand held razor saw, and attempts were made with a scalpel. Lead removal in the clinical cases was accomplished quickly in the four cases using the bone-cutter, without trauma to the lead. Bench testing results varied. The bone cutter was the most efficient method for most brands, but was ineffective on one. The motorized tool was difficult to position, produced sprays of plastic particles, and would have been risky in a clinical setting. The razor saw was difficult to use safely, or efficiently, except in some headers that resisted the bone cutter. The scalpel failed except in one "soft header" pacemaker. An orthopedic bone cutter is a useful tool for removing a retained lead from a pulse generator header. Different header designs and materials necessitate knowledge of several lead detachment methods.</abstract><cop>United States</cop><pmid>15305964</pmid><tpages>8</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0147-8389
ispartof Pacing and clinical electrophysiology, 2004-08, Vol.27 (8), p.1136-1143
issn 0147-8389
language eng
recordid cdi_proquest_miscellaneous_66792196
source MEDLINE; Wiley Online Library Journals Frontfile Complete
subjects Device Removal - methods
Equipment Failure
Humans
Pacemaker, Artificial
Surgical Instruments
title Lead stuck (frozen) in header: salvage by bone cutter versus other techniques. jfisher@montefiore.org
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-06T00%3A46%3A49IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_pubme&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Lead%20stuck%20(frozen)%20in%20header:%20salvage%20by%20bone%20cutter%20versus%20other%20techniques.%20jfisher@montefiore.org&rft.jtitle=Pacing%20and%20clinical%20electrophysiology&rft.au=Fisher,%20John%20D&rft.date=2004-08&rft.volume=27&rft.issue=8&rft.spage=1136&rft.epage=1143&rft.pages=1136-1143&rft.issn=0147-8389&rft_id=info:doi/&rft_dat=%3Cproquest_pubme%3E66792196%3C/proquest_pubme%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=66792196&rft_id=info:pmid/15305964&rfr_iscdi=true