Spatial modulation of a neutron beam by Larmor precession
We modulated the intensity of a neutron beam using Larmor precession techniques. We simulated triangular coils by using magnetised foils in electromagnets with rectangular pole shoes. Reversing the orientation of two of the magnetised foils in the spin-echo small-angle neutron scattering (SESANS)-in...
Gespeichert in:
Veröffentlicht in: | Physica. B, Condensed matter Condensed matter, 2009-09, Vol.404 (17), p.2585-2589 |
---|---|
Hauptverfasser: | , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | We modulated the intensity of a neutron beam using Larmor precession techniques. We simulated triangular coils by using magnetised foils in electromagnets with rectangular pole shoes. Reversing the orientation of two of the magnetised foils in the spin-echo small-angle neutron scattering (SESANS)-instrument in Delft and changing the field settings was sufficient to obtain a modulation with a period in the order of 1
mm. We expect to be able to go to a modulation with a period of
20
μ
m
.
This technique can be used to measure small-angle neutron scattering, analogous to the method used in SESANS, but with the advantage that all the manipulation of the neutron spin occurs before the sample. This means that the technique is insensitive to magnetic perturbations at the sample position. By positioning several of these devices after each other it will be possible to obtain a sharper modulation, or a modulation in two directions. |
---|---|
ISSN: | 0921-4526 1873-2135 |
DOI: | 10.1016/j.physb.2009.06.052 |