Femoral Head Penetration Rates and Oxidation of Highly Cross-Linked Polyethylene Hip Liners Implanted More Than 10 Years
Highly cross-linked polyethylene (HXLPE) was introduced to improve wear in total hip arthroplasty, with manufacturers implementing different thermal treatments to reduce oxidation. It is important to understand how long-term time in vivo affects the wear of these materials. The purpose of this study...
Gespeichert in:
Veröffentlicht in: | The Journal of arthroplasty 2024-09, Vol.39 (9), p.S286-S291 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | S291 |
---|---|
container_issue | 9 |
container_start_page | S286 |
container_title | The Journal of arthroplasty |
container_volume | 39 |
creator | Derr, Tabitha MacDonald, Daniel W. Malkani, Arthur L. Klein, Gregg R. Kurtz, Steven M. |
description | Highly cross-linked polyethylene (HXLPE) was introduced to improve wear in total hip arthroplasty, with manufacturers implementing different thermal treatments to reduce oxidation. It is important to understand how long-term time in vivo affects the wear of these materials. The purpose of this study was to investigate the wear and oxidative performance of first-generation HXLPE hip inserts implanted for greater than 10 years and compare annealed and remelted HXLPE formulations.
There were 49 total hip arthroplasty liners retrieved during routine revision surgery as part of an institutional review board–approved implant retrieval program. Penetration rates for the liners were calculated as the difference between the thickness of the unloaded and loaded regions divided by implantation time. Oxidation indices for the rim, locking mechanism, articulating surface, and backside regions were measured using Fourier-transform infrared spectroscopy according to American Society for Testing and Materials 2102. Mann–Whitney U tests were used to determine the statistical difference between annealed and remelted components.
The cohort had an average implantation time of 13.1 ± 2.6 years for annealed and 12.1 ± 1.7 years for remelted components. The components were revised most often for polyethylene wear, instability, and loosening. The penetration rate averaged 0.0177 ± 0.014 mm/year for annealed components and 0.015 ± 0.022 mm/year for remelted components. Penetration rates did not differ between the remelted and annealed cohorts (P = .28). Oxidation indices were found to be significantly higher in the annealed cohort for all regions of interest (P < .001).
Oxidation was found to be higher in the annealed HXLPE; however, this does not seem to be associated with greater wear as we found the average penetration rates for the cohorts were low, and the penetration rates were similar between the annealed and remelted cohorts. |
doi_str_mv | 10.1016/j.arth.2024.05.024 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_3056669031</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0883540324004571</els_id><sourcerecordid>3056669031</sourcerecordid><originalsourceid>FETCH-LOGICAL-c237t-59ec1961ad0c97d40198f503080757215c91b35a4dc48723c701605dee77fd723</originalsourceid><addsrcrecordid>eNp9kEFPGzEUhK2KqgTaP9AD8pHLbp_X6_WuxAVFpEFKRVXRQ0-Wsd8Sh911sB1E_n0dBXrsaaTRN6P3hpCvDEoGrPm2KXVI67KCqi5BlFk-kBkTvCraGpoTMoO25YWogZ-Ssxg3AIwJUX8ip7yVopbQzsjrAkcf9ECXqC39iROmoJPzE_2lE0aqJ0vvXp09er6nS_e4HvZ0HnyMxcpNT5hjfthjWu-HHM_AlmYfQ6S343bQU8rEDx-Q3q_1RBnQP6hD_Ew-9nqI-OVNz8nvxc39fFms7r7fzq9Xham4TIXo0LCuYdqC6aStgXVtL4BDC1LIignTsQcudG1N3cqKG5mXAWERpextNs7J5bF3G_zzDmNSo4sGh3wY-l1UHETTNB1wltHqiJrDcwF7tQ1u1GGvGKjD4mqjDourw-IKhMqSQxdv_buHEe2_yPvEGbg6Api_fHEYVDQOJ4PWBTRJWe_-1_8X5S-Q8A</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>3056669031</pqid></control><display><type>article</type><title>Femoral Head Penetration Rates and Oxidation of Highly Cross-Linked Polyethylene Hip Liners Implanted More Than 10 Years</title><source>Access via ScienceDirect (Elsevier)</source><creator>Derr, Tabitha ; MacDonald, Daniel W. ; Malkani, Arthur L. ; Klein, Gregg R. ; Kurtz, Steven M.</creator><creatorcontrib>Derr, Tabitha ; MacDonald, Daniel W. ; Malkani, Arthur L. ; Klein, Gregg R. ; Kurtz, Steven M.</creatorcontrib><description>Highly cross-linked polyethylene (HXLPE) was introduced to improve wear in total hip arthroplasty, with manufacturers implementing different thermal treatments to reduce oxidation. It is important to understand how long-term time in vivo affects the wear of these materials. The purpose of this study was to investigate the wear and oxidative performance of first-generation HXLPE hip inserts implanted for greater than 10 years and compare annealed and remelted HXLPE formulations.
There were 49 total hip arthroplasty liners retrieved during routine revision surgery as part of an institutional review board–approved implant retrieval program. Penetration rates for the liners were calculated as the difference between the thickness of the unloaded and loaded regions divided by implantation time. Oxidation indices for the rim, locking mechanism, articulating surface, and backside regions were measured using Fourier-transform infrared spectroscopy according to American Society for Testing and Materials 2102. Mann–Whitney U tests were used to determine the statistical difference between annealed and remelted components.
The cohort had an average implantation time of 13.1 ± 2.6 years for annealed and 12.1 ± 1.7 years for remelted components. The components were revised most often for polyethylene wear, instability, and loosening. The penetration rate averaged 0.0177 ± 0.014 mm/year for annealed components and 0.015 ± 0.022 mm/year for remelted components. Penetration rates did not differ between the remelted and annealed cohorts (P = .28). Oxidation indices were found to be significantly higher in the annealed cohort for all regions of interest (P < .001).
Oxidation was found to be higher in the annealed HXLPE; however, this does not seem to be associated with greater wear as we found the average penetration rates for the cohorts were low, and the penetration rates were similar between the annealed and remelted cohorts.</description><identifier>ISSN: 0883-5403</identifier><identifier>ISSN: 1532-8406</identifier><identifier>EISSN: 1532-8406</identifier><identifier>DOI: 10.1016/j.arth.2024.05.024</identifier><identifier>PMID: 38754708</identifier><language>eng</language><publisher>United States: Elsevier Inc</publisher><subject>HXLPE ; polyethylene ; retrieval studies ; revision reasons ; total hip arthroplasty</subject><ispartof>The Journal of arthroplasty, 2024-09, Vol.39 (9), p.S286-S291</ispartof><rights>2024 The Authors</rights><rights>Copyright © 2024 The Authors. Published by Elsevier Inc. All rights reserved.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c237t-59ec1961ad0c97d40198f503080757215c91b35a4dc48723c701605dee77fd723</cites><orcidid>0000-0001-8811-4106 ; 0000-0003-0967-1006</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/j.arth.2024.05.024$$EHTML$$P50$$Gelsevier$$Hfree_for_read</linktohtml><link.rule.ids>314,780,784,3550,27924,27925,45995</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/38754708$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Derr, Tabitha</creatorcontrib><creatorcontrib>MacDonald, Daniel W.</creatorcontrib><creatorcontrib>Malkani, Arthur L.</creatorcontrib><creatorcontrib>Klein, Gregg R.</creatorcontrib><creatorcontrib>Kurtz, Steven M.</creatorcontrib><title>Femoral Head Penetration Rates and Oxidation of Highly Cross-Linked Polyethylene Hip Liners Implanted More Than 10 Years</title><title>The Journal of arthroplasty</title><addtitle>J Arthroplasty</addtitle><description>Highly cross-linked polyethylene (HXLPE) was introduced to improve wear in total hip arthroplasty, with manufacturers implementing different thermal treatments to reduce oxidation. It is important to understand how long-term time in vivo affects the wear of these materials. The purpose of this study was to investigate the wear and oxidative performance of first-generation HXLPE hip inserts implanted for greater than 10 years and compare annealed and remelted HXLPE formulations.
There were 49 total hip arthroplasty liners retrieved during routine revision surgery as part of an institutional review board–approved implant retrieval program. Penetration rates for the liners were calculated as the difference between the thickness of the unloaded and loaded regions divided by implantation time. Oxidation indices for the rim, locking mechanism, articulating surface, and backside regions were measured using Fourier-transform infrared spectroscopy according to American Society for Testing and Materials 2102. Mann–Whitney U tests were used to determine the statistical difference between annealed and remelted components.
The cohort had an average implantation time of 13.1 ± 2.6 years for annealed and 12.1 ± 1.7 years for remelted components. The components were revised most often for polyethylene wear, instability, and loosening. The penetration rate averaged 0.0177 ± 0.014 mm/year for annealed components and 0.015 ± 0.022 mm/year for remelted components. Penetration rates did not differ between the remelted and annealed cohorts (P = .28). Oxidation indices were found to be significantly higher in the annealed cohort for all regions of interest (P < .001).
Oxidation was found to be higher in the annealed HXLPE; however, this does not seem to be associated with greater wear as we found the average penetration rates for the cohorts were low, and the penetration rates were similar between the annealed and remelted cohorts.</description><subject>HXLPE</subject><subject>polyethylene</subject><subject>retrieval studies</subject><subject>revision reasons</subject><subject>total hip arthroplasty</subject><issn>0883-5403</issn><issn>1532-8406</issn><issn>1532-8406</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2024</creationdate><recordtype>article</recordtype><recordid>eNp9kEFPGzEUhK2KqgTaP9AD8pHLbp_X6_WuxAVFpEFKRVXRQ0-Wsd8Sh911sB1E_n0dBXrsaaTRN6P3hpCvDEoGrPm2KXVI67KCqi5BlFk-kBkTvCraGpoTMoO25YWogZ-Ssxg3AIwJUX8ip7yVopbQzsjrAkcf9ECXqC39iROmoJPzE_2lE0aqJ0vvXp09er6nS_e4HvZ0HnyMxcpNT5hjfthjWu-HHM_AlmYfQ6S343bQU8rEDx-Q3q_1RBnQP6hD_Ew-9nqI-OVNz8nvxc39fFms7r7fzq9Xham4TIXo0LCuYdqC6aStgXVtL4BDC1LIignTsQcudG1N3cqKG5mXAWERpextNs7J5bF3G_zzDmNSo4sGh3wY-l1UHETTNB1wltHqiJrDcwF7tQ1u1GGvGKjD4mqjDourw-IKhMqSQxdv_buHEe2_yPvEGbg6Api_fHEYVDQOJ4PWBTRJWe_-1_8X5S-Q8A</recordid><startdate>20240901</startdate><enddate>20240901</enddate><creator>Derr, Tabitha</creator><creator>MacDonald, Daniel W.</creator><creator>Malkani, Arthur L.</creator><creator>Klein, Gregg R.</creator><creator>Kurtz, Steven M.</creator><general>Elsevier Inc</general><scope>6I.</scope><scope>AAFTH</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><orcidid>https://orcid.org/0000-0001-8811-4106</orcidid><orcidid>https://orcid.org/0000-0003-0967-1006</orcidid></search><sort><creationdate>20240901</creationdate><title>Femoral Head Penetration Rates and Oxidation of Highly Cross-Linked Polyethylene Hip Liners Implanted More Than 10 Years</title><author>Derr, Tabitha ; MacDonald, Daniel W. ; Malkani, Arthur L. ; Klein, Gregg R. ; Kurtz, Steven M.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c237t-59ec1961ad0c97d40198f503080757215c91b35a4dc48723c701605dee77fd723</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2024</creationdate><topic>HXLPE</topic><topic>polyethylene</topic><topic>retrieval studies</topic><topic>revision reasons</topic><topic>total hip arthroplasty</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Derr, Tabitha</creatorcontrib><creatorcontrib>MacDonald, Daniel W.</creatorcontrib><creatorcontrib>Malkani, Arthur L.</creatorcontrib><creatorcontrib>Klein, Gregg R.</creatorcontrib><creatorcontrib>Kurtz, Steven M.</creatorcontrib><collection>ScienceDirect Open Access Titles</collection><collection>Elsevier:ScienceDirect:Open Access</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>The Journal of arthroplasty</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Derr, Tabitha</au><au>MacDonald, Daniel W.</au><au>Malkani, Arthur L.</au><au>Klein, Gregg R.</au><au>Kurtz, Steven M.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Femoral Head Penetration Rates and Oxidation of Highly Cross-Linked Polyethylene Hip Liners Implanted More Than 10 Years</atitle><jtitle>The Journal of arthroplasty</jtitle><addtitle>J Arthroplasty</addtitle><date>2024-09-01</date><risdate>2024</risdate><volume>39</volume><issue>9</issue><spage>S286</spage><epage>S291</epage><pages>S286-S291</pages><issn>0883-5403</issn><issn>1532-8406</issn><eissn>1532-8406</eissn><abstract>Highly cross-linked polyethylene (HXLPE) was introduced to improve wear in total hip arthroplasty, with manufacturers implementing different thermal treatments to reduce oxidation. It is important to understand how long-term time in vivo affects the wear of these materials. The purpose of this study was to investigate the wear and oxidative performance of first-generation HXLPE hip inserts implanted for greater than 10 years and compare annealed and remelted HXLPE formulations.
There were 49 total hip arthroplasty liners retrieved during routine revision surgery as part of an institutional review board–approved implant retrieval program. Penetration rates for the liners were calculated as the difference between the thickness of the unloaded and loaded regions divided by implantation time. Oxidation indices for the rim, locking mechanism, articulating surface, and backside regions were measured using Fourier-transform infrared spectroscopy according to American Society for Testing and Materials 2102. Mann–Whitney U tests were used to determine the statistical difference between annealed and remelted components.
The cohort had an average implantation time of 13.1 ± 2.6 years for annealed and 12.1 ± 1.7 years for remelted components. The components were revised most often for polyethylene wear, instability, and loosening. The penetration rate averaged 0.0177 ± 0.014 mm/year for annealed components and 0.015 ± 0.022 mm/year for remelted components. Penetration rates did not differ between the remelted and annealed cohorts (P = .28). Oxidation indices were found to be significantly higher in the annealed cohort for all regions of interest (P < .001).
Oxidation was found to be higher in the annealed HXLPE; however, this does not seem to be associated with greater wear as we found the average penetration rates for the cohorts were low, and the penetration rates were similar between the annealed and remelted cohorts.</abstract><cop>United States</cop><pub>Elsevier Inc</pub><pmid>38754708</pmid><doi>10.1016/j.arth.2024.05.024</doi><orcidid>https://orcid.org/0000-0001-8811-4106</orcidid><orcidid>https://orcid.org/0000-0003-0967-1006</orcidid><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 0883-5403 |
ispartof | The Journal of arthroplasty, 2024-09, Vol.39 (9), p.S286-S291 |
issn | 0883-5403 1532-8406 1532-8406 |
language | eng |
recordid | cdi_proquest_miscellaneous_3056669031 |
source | Access via ScienceDirect (Elsevier) |
subjects | HXLPE polyethylene retrieval studies revision reasons total hip arthroplasty |
title | Femoral Head Penetration Rates and Oxidation of Highly Cross-Linked Polyethylene Hip Liners Implanted More Than 10 Years |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-01T21%3A06%3A37IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Femoral%20Head%20Penetration%20Rates%20and%20Oxidation%20of%20Highly%20Cross-Linked%20Polyethylene%20Hip%20Liners%20Implanted%20More%20Than%2010%20Years&rft.jtitle=The%20Journal%20of%20arthroplasty&rft.au=Derr,%20Tabitha&rft.date=2024-09-01&rft.volume=39&rft.issue=9&rft.spage=S286&rft.epage=S291&rft.pages=S286-S291&rft.issn=0883-5403&rft.eissn=1532-8406&rft_id=info:doi/10.1016/j.arth.2024.05.024&rft_dat=%3Cproquest_cross%3E3056669031%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=3056669031&rft_id=info:pmid/38754708&rft_els_id=S0883540324004571&rfr_iscdi=true |