Validation of a high-performance liquid chromatographic method with fluorescence detection for the simultaneous determination of tetracyclines residues in bovine milk

A high-performance liquid chromatography-fluorescence detection method was optimized and validated to determine tetracyclines residues in bovine milk. Post-column derivatization using metal complexation in non-aqueous reagent increased the fluorescence of chelates by a factor up to 2.54 compared to...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analytica chimica acta 2007-01, Vol.581 (1), p.108-117
Hauptverfasser: Spisso, Bernardete Ferraz, de Oliveira e Jesus, Ana Luiza, de Araújo Júnior, Marcus Antônio Gonçalves, Monteiro, Mychelle Alves
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A high-performance liquid chromatography-fluorescence detection method was optimized and validated to determine tetracyclines residues in bovine milk. Post-column derivatization using metal complexation in non-aqueous reagent increased the fluorescence of chelates by a factor up to 2.54 compared to water (signal-to-noise ratio enhancement). Overall recoveries ranged from 61 to 115%, with RSD r from 5 to 15% ( n = 54). Detection limits ranged from 5 to 35 μg kg −1. Limits of quantification were established at 50 μg kg −1. Decision limits (CCα) were 109, 108 and 124 μg kg −1 and detection capabilities (CCβ) 119, 117 and 161 μg kg −1 for oxytetracycline, tetracycline and chlortetracycline, respectively. The method was applied successfully in a national monitoring program.
ISSN:0003-2670
1873-4324