Measuring CSF shunt flow with MRI using flow enhancement of signal intensity (FENSI)
To develop and validate a noninvasive imaging technique for accurately assessing very slow CSF flow within shunt tubes in pediatric patients with hydrocephalus, aiming to identify obstructions that might impede CSF drainage. A simulation of shunt flow enhancement of signal intensity (shunt-FENSI) si...
Gespeichert in:
Veröffentlicht in: | Magnetic resonance in medicine 2024-08, Vol.92 (2), p.807-819 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | To develop and validate a noninvasive imaging technique for accurately assessing very slow CSF flow within shunt tubes in pediatric patients with hydrocephalus, aiming to identify obstructions that might impede CSF drainage.
A simulation of shunt flow enhancement of signal intensity (shunt-FENSI) signal is used to establish the relationship between signal change and flow rate. The quantification of flow enhancement of signal intensity data involves normalization, curve fitting, and calibration to match simulated data. Additionally, a phase sweep method is introduced to accommodate the impact of magnetic field inhomogeneity on the flow measurement. The method is tested in flow phantoms, healthy adults, intensive care unit patients with external ventricular drains (EVD), and shunt patients. EVDs enable shunt-flow measurements to be acquired with a ground truth measure of CSF drainage.
The flow-rate-to-signal simulation establishes signal-flow relationships and takes into account the T
of draining fluid. The phase sweep method accurately accounts for phase accumulation due to frequency offsets at the shunt. Results in phantom and healthy human participants reveal reliable quantification of flow rates using controlled flows and agreement with the flow simulation. EVD patients display reliable measures of flow rates. Shunt patient results demonstrate feasibility of the method and consistent flow rates for functional shunts.
The results demonstrate the technique's applicability, accuracy, and potential for diagnosing and noninvasively monitoring hydrocephalus. Limitations of the current approach include a high sensitivity to motion and strict requirement of imaging slice prescription. |
---|---|
ISSN: | 0740-3194 1522-2594 1522-2594 |