Direct monitoring of toxic compounds in air using a portable mass spectrometer

A portable tandem mass spectrometer, capable of performing atmospheric pressure chemical ionization (APCI) using a direct atmospheric inlet, is applied to the real-time monitoring of toxic compounds in air. Analytes of interest include dimethyl methylphosphonate, arsine, benzene, toluene, pyridine a...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Analyst (London) 2006-01, Vol.131 (4), p.556-567
Hauptverfasser: MULLIGAN, Christopher C, JUSTES, Dina R, NOLL, Robert J, SANDERS, Nathaniel L, LAUGHLIN, Brian C, COOKS, R. Graham
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A portable tandem mass spectrometer, capable of performing atmospheric pressure chemical ionization (APCI) using a direct atmospheric inlet, is applied to the real-time monitoring of toxic compounds in air. Analytes of interest include dimethyl methylphosphonate, arsine, benzene, toluene, pyridine and vinyl acetate. The detection, identification and quantification of organic and inorganic compounds in air is demonstrated using short analysis times (