Performance assessment of the surface-guided radiation therapy system: Varian Identify

Image-guided radiotherapy (IGRT) systems using ionizing radiation may increase the risk of secondary cancer and normal tissue toxicity due to additional radiation exposure caused by large field sizes or repeated scans during X-ray imaging. As an alternative to these modalities, surface-guided radiot...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical dosimetry : official journal of the American Association of Medical Dosimetrists 2024, Vol.49 (3), p.222-228
Hauptverfasser: Ayrancıoğlu, Oğuzhan, Ayrancıoğlu, Ceren, Arıkan, Şerife Ceren, Alıcıkuş, Lütfiye Zümre Arıcan
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 228
container_issue 3
container_start_page 222
container_title Medical dosimetry : official journal of the American Association of Medical Dosimetrists
container_volume 49
creator Ayrancıoğlu, Oğuzhan
Ayrancıoğlu, Ceren
Arıkan, Şerife Ceren
Alıcıkuş, Lütfiye Zümre Arıcan
description Image-guided radiotherapy (IGRT) systems using ionizing radiation may increase the risk of secondary cancer and normal tissue toxicity due to additional radiation exposure caused by large field sizes or repeated scans during X-ray imaging. As an alternative to these modalities, surface-guided radiotherapy (SGRT) systems which do not employ ionizing radiation have been developed. This study presents a comprehensive performance evaluation of the Varian Identify SGRT system by using an anthropomorphic Alderson Rando phantom in three different aspects: (a) the accuracy and reproducibility of the system in different regions of interest (ROI) for varying couch displacements, (b) the setup accuracy of the system for patient positioning based on different computed tomography (CT) slice thicknesses, and (c) the potential influence of obstructing SGRT cameras by the gantry on the system's overall accuracy and reproducibility. The accuracy and reproducibility of the SGRT system fell within 1 mm and 1°. Nevertheless, in certain situations, these values were observed to exceed prescribed limits. Consequently, concerning SGRT tolerance limits for treatment applications, careful consideration of ROIs and offset values of the system is crucial. We also recommend that patients should ideally be set up during 0° gantry rotation, and the on-board imaging (OBI) system should be retracted to prevent obstruction of the cameras. Additionally, reference CT images with a slice thickness of under 3 mm are recommended for this purpose.
doi_str_mv 10.1016/j.meddos.2024.01.001
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2923326585</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0958394724000013</els_id><sourcerecordid>2923326585</sourcerecordid><originalsourceid>FETCH-LOGICAL-c311t-8a5cb9b78e66971c9e0437d43dc18e97e2306d3600e2f6983e6ab1ff40c1947e3</originalsourceid><addsrcrecordid>eNp9kEtv2zAQhImiReO4_QdFoGMvUpcPSVQPAYogLyBAe0h8JWhy2dCwJJcrBfC_Dw07Pea0h5nZwXyMfeNQceDNj03Vo_cjVQKEqoBXAPwDW3DdylKBEB_ZArpal7JT7Rk7J9oAQK1AfmZnUksBWqsFW_3BFMbU28FhYYmQqMdhKsZQTM9Y0JyCdVj-naNHXyTro53iOBzEZHf7gvY0Yf-zWNkU7VDc-xyOYf-FfQp2S_j1dJfs6eb68equfPh9e3_166F0kvOp1LZ2627damyaruWuQ1Cy9Up6xzV2LQoJjZcNAIrQdFpiY9c8BAWO51kol-z78e8ujf9mpMn0kRxut3bAcSYjOiGlaGpdZ6s6Wl0aiRIGs0uxt2lvOJgDUbMxR6LmQNQAN5lojl2cGuZ1lv-H3hBmw-XRgHnnS8RkyEXMOH1M6Cbjx_h-wytFjIl-</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2923326585</pqid></control><display><type>article</type><title>Performance assessment of the surface-guided radiation therapy system: Varian Identify</title><source>Access via ScienceDirect (Elsevier)</source><creator>Ayrancıoğlu, Oğuzhan ; Ayrancıoğlu, Ceren ; Arıkan, Şerife Ceren ; Alıcıkuş, Lütfiye Zümre Arıcan</creator><creatorcontrib>Ayrancıoğlu, Oğuzhan ; Ayrancıoğlu, Ceren ; Arıkan, Şerife Ceren ; Alıcıkuş, Lütfiye Zümre Arıcan</creatorcontrib><description>Image-guided radiotherapy (IGRT) systems using ionizing radiation may increase the risk of secondary cancer and normal tissue toxicity due to additional radiation exposure caused by large field sizes or repeated scans during X-ray imaging. As an alternative to these modalities, surface-guided radiotherapy (SGRT) systems which do not employ ionizing radiation have been developed. This study presents a comprehensive performance evaluation of the Varian Identify SGRT system by using an anthropomorphic Alderson Rando phantom in three different aspects: (a) the accuracy and reproducibility of the system in different regions of interest (ROI) for varying couch displacements, (b) the setup accuracy of the system for patient positioning based on different computed tomography (CT) slice thicknesses, and (c) the potential influence of obstructing SGRT cameras by the gantry on the system's overall accuracy and reproducibility. The accuracy and reproducibility of the SGRT system fell within 1 mm and 1°. Nevertheless, in certain situations, these values were observed to exceed prescribed limits. Consequently, concerning SGRT tolerance limits for treatment applications, careful consideration of ROIs and offset values of the system is crucial. We also recommend that patients should ideally be set up during 0° gantry rotation, and the on-board imaging (OBI) system should be retracted to prevent obstruction of the cameras. Additionally, reference CT images with a slice thickness of under 3 mm are recommended for this purpose.</description><identifier>ISSN: 0958-3947</identifier><identifier>ISSN: 1873-4022</identifier><identifier>EISSN: 1873-4022</identifier><identifier>DOI: 10.1016/j.meddos.2024.01.001</identifier><identifier>PMID: 38320884</identifier><language>eng</language><publisher>United States: Elsevier Inc</publisher><subject>Offsets ; Performance assessment ; Setup accuracy ; Surface-guided radiation therapy</subject><ispartof>Medical dosimetry : official journal of the American Association of Medical Dosimetrists, 2024, Vol.49 (3), p.222-228</ispartof><rights>2024 American Association of Medical Dosimetrists</rights><rights>Copyright © 2024 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c311t-8a5cb9b78e66971c9e0437d43dc18e97e2306d3600e2f6983e6ab1ff40c1947e3</cites><orcidid>0000-0002-0124-6534 ; 0000-0002-5445-6797 ; 0009-0000-6084-182X ; 0000-0002-2484-1096</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttps://dx.doi.org/10.1016/j.meddos.2024.01.001$$EHTML$$P50$$Gelsevier$$H</linktohtml><link.rule.ids>315,781,785,3551,4025,27928,27929,27930,46000</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/38320884$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Ayrancıoğlu, Oğuzhan</creatorcontrib><creatorcontrib>Ayrancıoğlu, Ceren</creatorcontrib><creatorcontrib>Arıkan, Şerife Ceren</creatorcontrib><creatorcontrib>Alıcıkuş, Lütfiye Zümre Arıcan</creatorcontrib><title>Performance assessment of the surface-guided radiation therapy system: Varian Identify</title><title>Medical dosimetry : official journal of the American Association of Medical Dosimetrists</title><addtitle>Med Dosim</addtitle><description>Image-guided radiotherapy (IGRT) systems using ionizing radiation may increase the risk of secondary cancer and normal tissue toxicity due to additional radiation exposure caused by large field sizes or repeated scans during X-ray imaging. As an alternative to these modalities, surface-guided radiotherapy (SGRT) systems which do not employ ionizing radiation have been developed. This study presents a comprehensive performance evaluation of the Varian Identify SGRT system by using an anthropomorphic Alderson Rando phantom in three different aspects: (a) the accuracy and reproducibility of the system in different regions of interest (ROI) for varying couch displacements, (b) the setup accuracy of the system for patient positioning based on different computed tomography (CT) slice thicknesses, and (c) the potential influence of obstructing SGRT cameras by the gantry on the system's overall accuracy and reproducibility. The accuracy and reproducibility of the SGRT system fell within 1 mm and 1°. Nevertheless, in certain situations, these values were observed to exceed prescribed limits. Consequently, concerning SGRT tolerance limits for treatment applications, careful consideration of ROIs and offset values of the system is crucial. We also recommend that patients should ideally be set up during 0° gantry rotation, and the on-board imaging (OBI) system should be retracted to prevent obstruction of the cameras. Additionally, reference CT images with a slice thickness of under 3 mm are recommended for this purpose.</description><subject>Offsets</subject><subject>Performance assessment</subject><subject>Setup accuracy</subject><subject>Surface-guided radiation therapy</subject><issn>0958-3947</issn><issn>1873-4022</issn><issn>1873-4022</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2024</creationdate><recordtype>article</recordtype><recordid>eNp9kEtv2zAQhImiReO4_QdFoGMvUpcPSVQPAYogLyBAe0h8JWhy2dCwJJcrBfC_Dw07Pea0h5nZwXyMfeNQceDNj03Vo_cjVQKEqoBXAPwDW3DdylKBEB_ZArpal7JT7Rk7J9oAQK1AfmZnUksBWqsFW_3BFMbU28FhYYmQqMdhKsZQTM9Y0JyCdVj-naNHXyTro53iOBzEZHf7gvY0Yf-zWNkU7VDc-xyOYf-FfQp2S_j1dJfs6eb68equfPh9e3_166F0kvOp1LZ2627damyaruWuQ1Cy9Up6xzV2LQoJjZcNAIrQdFpiY9c8BAWO51kol-z78e8ujf9mpMn0kRxut3bAcSYjOiGlaGpdZ6s6Wl0aiRIGs0uxt2lvOJgDUbMxR6LmQNQAN5lojl2cGuZ1lv-H3hBmw-XRgHnnS8RkyEXMOH1M6Cbjx_h-wytFjIl-</recordid><startdate>2024</startdate><enddate>2024</enddate><creator>Ayrancıoğlu, Oğuzhan</creator><creator>Ayrancıoğlu, Ceren</creator><creator>Arıkan, Şerife Ceren</creator><creator>Alıcıkuş, Lütfiye Zümre Arıcan</creator><general>Elsevier Inc</general><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><orcidid>https://orcid.org/0000-0002-0124-6534</orcidid><orcidid>https://orcid.org/0000-0002-5445-6797</orcidid><orcidid>https://orcid.org/0009-0000-6084-182X</orcidid><orcidid>https://orcid.org/0000-0002-2484-1096</orcidid></search><sort><creationdate>2024</creationdate><title>Performance assessment of the surface-guided radiation therapy system: Varian Identify</title><author>Ayrancıoğlu, Oğuzhan ; Ayrancıoğlu, Ceren ; Arıkan, Şerife Ceren ; Alıcıkuş, Lütfiye Zümre Arıcan</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c311t-8a5cb9b78e66971c9e0437d43dc18e97e2306d3600e2f6983e6ab1ff40c1947e3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2024</creationdate><topic>Offsets</topic><topic>Performance assessment</topic><topic>Setup accuracy</topic><topic>Surface-guided radiation therapy</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Ayrancıoğlu, Oğuzhan</creatorcontrib><creatorcontrib>Ayrancıoğlu, Ceren</creatorcontrib><creatorcontrib>Arıkan, Şerife Ceren</creatorcontrib><creatorcontrib>Alıcıkuş, Lütfiye Zümre Arıcan</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Medical dosimetry : official journal of the American Association of Medical Dosimetrists</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Ayrancıoğlu, Oğuzhan</au><au>Ayrancıoğlu, Ceren</au><au>Arıkan, Şerife Ceren</au><au>Alıcıkuş, Lütfiye Zümre Arıcan</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Performance assessment of the surface-guided radiation therapy system: Varian Identify</atitle><jtitle>Medical dosimetry : official journal of the American Association of Medical Dosimetrists</jtitle><addtitle>Med Dosim</addtitle><date>2024</date><risdate>2024</risdate><volume>49</volume><issue>3</issue><spage>222</spage><epage>228</epage><pages>222-228</pages><issn>0958-3947</issn><issn>1873-4022</issn><eissn>1873-4022</eissn><abstract>Image-guided radiotherapy (IGRT) systems using ionizing radiation may increase the risk of secondary cancer and normal tissue toxicity due to additional radiation exposure caused by large field sizes or repeated scans during X-ray imaging. As an alternative to these modalities, surface-guided radiotherapy (SGRT) systems which do not employ ionizing radiation have been developed. This study presents a comprehensive performance evaluation of the Varian Identify SGRT system by using an anthropomorphic Alderson Rando phantom in three different aspects: (a) the accuracy and reproducibility of the system in different regions of interest (ROI) for varying couch displacements, (b) the setup accuracy of the system for patient positioning based on different computed tomography (CT) slice thicknesses, and (c) the potential influence of obstructing SGRT cameras by the gantry on the system's overall accuracy and reproducibility. The accuracy and reproducibility of the SGRT system fell within 1 mm and 1°. Nevertheless, in certain situations, these values were observed to exceed prescribed limits. Consequently, concerning SGRT tolerance limits for treatment applications, careful consideration of ROIs and offset values of the system is crucial. We also recommend that patients should ideally be set up during 0° gantry rotation, and the on-board imaging (OBI) system should be retracted to prevent obstruction of the cameras. Additionally, reference CT images with a slice thickness of under 3 mm are recommended for this purpose.</abstract><cop>United States</cop><pub>Elsevier Inc</pub><pmid>38320884</pmid><doi>10.1016/j.meddos.2024.01.001</doi><tpages>7</tpages><orcidid>https://orcid.org/0000-0002-0124-6534</orcidid><orcidid>https://orcid.org/0000-0002-5445-6797</orcidid><orcidid>https://orcid.org/0009-0000-6084-182X</orcidid><orcidid>https://orcid.org/0000-0002-2484-1096</orcidid></addata></record>
fulltext fulltext
identifier ISSN: 0958-3947
ispartof Medical dosimetry : official journal of the American Association of Medical Dosimetrists, 2024, Vol.49 (3), p.222-228
issn 0958-3947
1873-4022
1873-4022
language eng
recordid cdi_proquest_miscellaneous_2923326585
source Access via ScienceDirect (Elsevier)
subjects Offsets
Performance assessment
Setup accuracy
Surface-guided radiation therapy
title Performance assessment of the surface-guided radiation therapy system: Varian Identify
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-12T16%3A45%3A17IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Performance%20assessment%20of%20the%20surface-guided%20radiation%20therapy%20system:%20Varian%20Identify&rft.jtitle=Medical%20dosimetry%20:%20official%20journal%20of%20the%20American%20Association%20of%20Medical%20Dosimetrists&rft.au=Ayranc%C4%B1o%C4%9Flu,%20O%C4%9Fuzhan&rft.date=2024&rft.volume=49&rft.issue=3&rft.spage=222&rft.epage=228&rft.pages=222-228&rft.issn=0958-3947&rft.eissn=1873-4022&rft_id=info:doi/10.1016/j.meddos.2024.01.001&rft_dat=%3Cproquest_cross%3E2923326585%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2923326585&rft_id=info:pmid/38320884&rft_els_id=S0958394724000013&rfr_iscdi=true