Dosimetric Studies of a mMLC for Intensity Modulated Radiation Surgery for Irregular Lesions

In this work we present the commissioning of a micro Multileaf Collimator (mMLC) of the NovalisTM Intensity Modulated Radiation Surgery (IMRS) system. This system can deliver conventional stereotactic radiosurgery, conformal arcs and beams, and IMRS. The last modalities are performed using the mMLC....

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Hauptverfasser: Garcia-Garduno, Olivia Amanda, Rodriguez-Villafuerte, Mercedes, Larraga-Gutierrez, Jose Manuel, Celis-Lopez, Miguel Angel
Format: Tagungsbericht
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!