Two data acquisition and processing systems for a compact gamma-camera

A compact gamma-camera is constructed using a 3-mm-thick single NaI(Tl) crystal viewed by a 5-in-diameter position-sensitive photomultiplier tube. In order to use this system in applications where the counting rate is high and flexibility of data handling is desired, a faster data acquisition and pr...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:IEEE transactions on nuclear science 1993-08, Vol.40 (4), p.1165-1168
Hauptverfasser: He, Z., Ramsden, D., Holder, S.M., Duncan, S.H.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext bestellen
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A compact gamma-camera is constructed using a 3-mm-thick single NaI(Tl) crystal viewed by a 5-in-diameter position-sensitive photomultiplier tube. In order to use this system in applications where the counting rate is high and flexibility of data handling is desired, a faster data acquisition and processing system is required. The design and operation of two data collection and processing systems capable of handling events within 30 mu s are presented. One of the systems is based on the use of a commercial Vector 32C/8500 processor board, while the other makes use of multiple T805 transputers. A comparison of the two systems is provided.< >
ISSN:0018-9499
1558-1578