The dynamic calibration of an electrical capacitance tomography sensor applied to the fluidized bed drying of pharmaceutical granule

Electrical capacitance tomographic data collected in a lab-scale fluidized bed used for the drying of pharmaceutical granule have been corrected for the influence of moisture on the permittivity of the drying material. The correction is based on a linear least-squares fit to measurements of capacita...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Measurement science & technology 2005-06, Vol.16 (6), p.1281-1290
Hauptverfasser: Chaplin, Gareth, Pugsley, Todd, van der Lee, Loni, Kantzas, Apostolos, Winters, Conrad
Format: Artikel
Sprache:eng
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 1290
container_issue 6
container_start_page 1281
container_title Measurement science & technology
container_volume 16
creator Chaplin, Gareth
Pugsley, Todd
van der Lee, Loni
Kantzas, Apostolos
Winters, Conrad
description Electrical capacitance tomographic data collected in a lab-scale fluidized bed used for the drying of pharmaceutical granule have been corrected for the influence of moisture on the permittivity of the drying material. The correction is based on a linear least-squares fit to measurements of capacitance in a packed bed of granule at various moisture contents. X-ray tomography has been used to independently verify this correction procedure. The influence of permittivity models and number of iterations used for the reconstruction of tomograms have also been examined. It has been determined that the Bottcher permittivity model performs best at bed moistures above approximately 5 wt% while the parallel model is superior at bed moisture below this value. The reconstruction technique based on iterative linear back-projection utilized for the reconstruction of ECT data required approximately 50 iterations to successfully reproduce the density behaviour seen in the x-ray tomographs. Instability in the reconstruction technique at higher numbers of iterations indicates that a linear least-squares fit does not completely capture the response of the sensor to moisture changes. For future applications, changes in bed voidage associated with the drying of Pharmaceuticals must be addressed and included in this calibration procedure in order to implement this calibration technique throughout the drying process. Nevertheless, the viability of this technique for on-line calibration of an ECT sensor applied to the drying process has been demonstrated.
doi_str_mv 10.1088/0957-0233/16/6/007
format Article
fullrecord <record><control><sourceid>proquest_iop_p</sourceid><recordid>TN_cdi_proquest_miscellaneous_28600356</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>28600356</sourcerecordid><originalsourceid>FETCH-LOGICAL-c447t-83c487d9907c1ea86528b9398e7591457037b366694e2af1b81f0898825626893</originalsourceid><addsrcrecordid>eNqNkb9PxCAYhonRxPPHP-DEZOJQC6WlMJqLv5JLXM6ZUErvMLQgtEOd_cPlPONygw5fSL73-Z7hBYArjG4xYixHvKozVBCSY5rTHKH6CCwwoTijFcLHYPELnIKzGN9QIhDnC_C53mrYzoPsjYJKWtMEORo3QNdBOUBttRqDSUEKvVRmlIPScHS92wTptzOMeoguQOm9NbpNCRyTsbOTac1HWjRp2jCbYbNT-q0MvVR6Gr-dyTFMVl-Ak07aqC9_3nPw-nC_Xj5lq5fH5-XdKlNlWY8ZI6pkdcs5qhXWktGqYA0nnOm64risakTqhlBKeakL2eGG4Q4xzlhR0YIyTs7B9d7rg3ufdBxFb6LS1spBuymKgtOKMcT-BhlFiFQ0gcUeVMHFGHQnfDC9DLPASOx-RuyKF7viBaaCitR7Osr2R8b5__E3h_whJ3zbkS-KEpwg</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>28600356</pqid></control><display><type>article</type><title>The dynamic calibration of an electrical capacitance tomography sensor applied to the fluidized bed drying of pharmaceutical granule</title><source>IOP Publishing Journals</source><source>Institute of Physics (IOP) Journals - HEAL-Link</source><creator>Chaplin, Gareth ; Pugsley, Todd ; van der Lee, Loni ; Kantzas, Apostolos ; Winters, Conrad</creator><creatorcontrib>Chaplin, Gareth ; Pugsley, Todd ; van der Lee, Loni ; Kantzas, Apostolos ; Winters, Conrad</creatorcontrib><description>Electrical capacitance tomographic data collected in a lab-scale fluidized bed used for the drying of pharmaceutical granule have been corrected for the influence of moisture on the permittivity of the drying material. The correction is based on a linear least-squares fit to measurements of capacitance in a packed bed of granule at various moisture contents. X-ray tomography has been used to independently verify this correction procedure. The influence of permittivity models and number of iterations used for the reconstruction of tomograms have also been examined. It has been determined that the Bottcher permittivity model performs best at bed moistures above approximately 5 wt% while the parallel model is superior at bed moisture below this value. The reconstruction technique based on iterative linear back-projection utilized for the reconstruction of ECT data required approximately 50 iterations to successfully reproduce the density behaviour seen in the x-ray tomographs. Instability in the reconstruction technique at higher numbers of iterations indicates that a linear least-squares fit does not completely capture the response of the sensor to moisture changes. For future applications, changes in bed voidage associated with the drying of Pharmaceuticals must be addressed and included in this calibration procedure in order to implement this calibration technique throughout the drying process. Nevertheless, the viability of this technique for on-line calibration of an ECT sensor applied to the drying process has been demonstrated.</description><identifier>ISSN: 0957-0233</identifier><identifier>EISSN: 1361-6501</identifier><identifier>DOI: 10.1088/0957-0233/16/6/007</identifier><language>eng</language><publisher>IOP Publishing</publisher><ispartof>Measurement science &amp; technology, 2005-06, Vol.16 (6), p.1281-1290</ispartof><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c447t-83c487d9907c1ea86528b9398e7591457037b366694e2af1b81f0898825626893</citedby><cites>FETCH-LOGICAL-c447t-83c487d9907c1ea86528b9398e7591457037b366694e2af1b81f0898825626893</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://iopscience.iop.org/article/10.1088/0957-0233/16/6/007/pdf$$EPDF$$P50$$Giop$$H</linktopdf><link.rule.ids>314,776,780,27901,27902,53805,53885</link.rule.ids></links><search><creatorcontrib>Chaplin, Gareth</creatorcontrib><creatorcontrib>Pugsley, Todd</creatorcontrib><creatorcontrib>van der Lee, Loni</creatorcontrib><creatorcontrib>Kantzas, Apostolos</creatorcontrib><creatorcontrib>Winters, Conrad</creatorcontrib><title>The dynamic calibration of an electrical capacitance tomography sensor applied to the fluidized bed drying of pharmaceutical granule</title><title>Measurement science &amp; technology</title><description>Electrical capacitance tomographic data collected in a lab-scale fluidized bed used for the drying of pharmaceutical granule have been corrected for the influence of moisture on the permittivity of the drying material. The correction is based on a linear least-squares fit to measurements of capacitance in a packed bed of granule at various moisture contents. X-ray tomography has been used to independently verify this correction procedure. The influence of permittivity models and number of iterations used for the reconstruction of tomograms have also been examined. It has been determined that the Bottcher permittivity model performs best at bed moistures above approximately 5 wt% while the parallel model is superior at bed moisture below this value. The reconstruction technique based on iterative linear back-projection utilized for the reconstruction of ECT data required approximately 50 iterations to successfully reproduce the density behaviour seen in the x-ray tomographs. Instability in the reconstruction technique at higher numbers of iterations indicates that a linear least-squares fit does not completely capture the response of the sensor to moisture changes. For future applications, changes in bed voidage associated with the drying of Pharmaceuticals must be addressed and included in this calibration procedure in order to implement this calibration technique throughout the drying process. Nevertheless, the viability of this technique for on-line calibration of an ECT sensor applied to the drying process has been demonstrated.</description><issn>0957-0233</issn><issn>1361-6501</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2005</creationdate><recordtype>article</recordtype><recordid>eNqNkb9PxCAYhonRxPPHP-DEZOJQC6WlMJqLv5JLXM6ZUErvMLQgtEOd_cPlPONygw5fSL73-Z7hBYArjG4xYixHvKozVBCSY5rTHKH6CCwwoTijFcLHYPELnIKzGN9QIhDnC_C53mrYzoPsjYJKWtMEORo3QNdBOUBttRqDSUEKvVRmlIPScHS92wTptzOMeoguQOm9NbpNCRyTsbOTac1HWjRp2jCbYbNT-q0MvVR6Gr-dyTFMVl-Ak07aqC9_3nPw-nC_Xj5lq5fH5-XdKlNlWY8ZI6pkdcs5qhXWktGqYA0nnOm64risakTqhlBKeakL2eGG4Q4xzlhR0YIyTs7B9d7rg3ufdBxFb6LS1spBuymKgtOKMcT-BhlFiFQ0gcUeVMHFGHQnfDC9DLPASOx-RuyKF7viBaaCitR7Osr2R8b5__E3h_whJ3zbkS-KEpwg</recordid><startdate>20050601</startdate><enddate>20050601</enddate><creator>Chaplin, Gareth</creator><creator>Pugsley, Todd</creator><creator>van der Lee, Loni</creator><creator>Kantzas, Apostolos</creator><creator>Winters, Conrad</creator><general>IOP Publishing</general><scope>AAYXX</scope><scope>CITATION</scope><scope>8FD</scope><scope>F28</scope><scope>FR3</scope></search><sort><creationdate>20050601</creationdate><title>The dynamic calibration of an electrical capacitance tomography sensor applied to the fluidized bed drying of pharmaceutical granule</title><author>Chaplin, Gareth ; Pugsley, Todd ; van der Lee, Loni ; Kantzas, Apostolos ; Winters, Conrad</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c447t-83c487d9907c1ea86528b9398e7591457037b366694e2af1b81f0898825626893</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2005</creationdate><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Chaplin, Gareth</creatorcontrib><creatorcontrib>Pugsley, Todd</creatorcontrib><creatorcontrib>van der Lee, Loni</creatorcontrib><creatorcontrib>Kantzas, Apostolos</creatorcontrib><creatorcontrib>Winters, Conrad</creatorcontrib><collection>CrossRef</collection><collection>Technology Research Database</collection><collection>ANTE: Abstracts in New Technology &amp; Engineering</collection><collection>Engineering Research Database</collection><jtitle>Measurement science &amp; technology</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Chaplin, Gareth</au><au>Pugsley, Todd</au><au>van der Lee, Loni</au><au>Kantzas, Apostolos</au><au>Winters, Conrad</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>The dynamic calibration of an electrical capacitance tomography sensor applied to the fluidized bed drying of pharmaceutical granule</atitle><jtitle>Measurement science &amp; technology</jtitle><date>2005-06-01</date><risdate>2005</risdate><volume>16</volume><issue>6</issue><spage>1281</spage><epage>1290</epage><pages>1281-1290</pages><issn>0957-0233</issn><eissn>1361-6501</eissn><abstract>Electrical capacitance tomographic data collected in a lab-scale fluidized bed used for the drying of pharmaceutical granule have been corrected for the influence of moisture on the permittivity of the drying material. The correction is based on a linear least-squares fit to measurements of capacitance in a packed bed of granule at various moisture contents. X-ray tomography has been used to independently verify this correction procedure. The influence of permittivity models and number of iterations used for the reconstruction of tomograms have also been examined. It has been determined that the Bottcher permittivity model performs best at bed moistures above approximately 5 wt% while the parallel model is superior at bed moisture below this value. The reconstruction technique based on iterative linear back-projection utilized for the reconstruction of ECT data required approximately 50 iterations to successfully reproduce the density behaviour seen in the x-ray tomographs. Instability in the reconstruction technique at higher numbers of iterations indicates that a linear least-squares fit does not completely capture the response of the sensor to moisture changes. For future applications, changes in bed voidage associated with the drying of Pharmaceuticals must be addressed and included in this calibration procedure in order to implement this calibration technique throughout the drying process. Nevertheless, the viability of this technique for on-line calibration of an ECT sensor applied to the drying process has been demonstrated.</abstract><pub>IOP Publishing</pub><doi>10.1088/0957-0233/16/6/007</doi><tpages>10</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0957-0233
ispartof Measurement science & technology, 2005-06, Vol.16 (6), p.1281-1290
issn 0957-0233
1361-6501
language eng
recordid cdi_proquest_miscellaneous_28600356
source IOP Publishing Journals; Institute of Physics (IOP) Journals - HEAL-Link
title The dynamic calibration of an electrical capacitance tomography sensor applied to the fluidized bed drying of pharmaceutical granule
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-02T13%3A30%3A35IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_iop_p&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=The%20dynamic%20calibration%20of%20an%20electrical%20capacitance%20tomography%20sensor%20applied%20to%20the%20fluidized%20bed%20drying%20of%20pharmaceutical%20granule&rft.jtitle=Measurement%20science%20&%20technology&rft.au=Chaplin,%20Gareth&rft.date=2005-06-01&rft.volume=16&rft.issue=6&rft.spage=1281&rft.epage=1290&rft.pages=1281-1290&rft.issn=0957-0233&rft.eissn=1361-6501&rft_id=info:doi/10.1088/0957-0233/16/6/007&rft_dat=%3Cproquest_iop_p%3E28600356%3C/proquest_iop_p%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=28600356&rft_id=info:pmid/&rfr_iscdi=true