Implant efficiency and clinical performance of Aveir™ VR and Micra™ VR leadless pacemaker: A multicenter comparative analysis of 67 patients
Introduction Leadless pacemaker (LP) is a novel pacemaker that has been proven to be effective and safe; however, the majority of LPs in previous reports were the Medtronic Micra™ VR LP. We aim to evaluate the implant efficiency and clinical performance of the Aveir™ VR LP compared to the Micra™ VR...
Gespeichert in:
Veröffentlicht in: | Pacing and clinical electrophysiology 2023-08, Vol.46 (8), p.827-832 |
---|---|
Hauptverfasser: | , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Introduction
Leadless pacemaker (LP) is a novel pacemaker that has been proven to be effective and safe; however, the majority of LPs in previous reports were the Medtronic Micra™ VR LP. We aim to evaluate the implant efficiency and clinical performance of the Aveir™ VR LP compared to the Micra™ VR LP.
Method
We performed a retrospective analysis in two healthcare systems (Sparrow Hospital and Ascension Health System, Michigan) in patients implanted with LPs between January 1, 2018, and April 1, 2022. The parameters were collected at implantation, 3 months and 6 months.
Results
A total of 67 patients were included in the study. The Micra™ VR group had shorter time in the electrophysiology lab (41 ± 12 vs. 55 ± 11.5 min, p = .008) and shorter fluoroscopic time (6.5 ± 2.2 vs. 11.5 ± 4.5 min, p |
---|---|