A multicenter study on the effects of different methods of disinfecting medical external-use ultrasound probes

Microbial contamination of external-use ultrasound probes is a serious and overlooked issue. We assessed the effects of different methods of disinfecting medical external-use ultrasound probes. On-site disinfection experiments were conducted at 10 hospitals; the tips and sides of external-use ultras...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:American journal of infection control 2024-02, Vol.52 (2), p.167-175
Hauptverfasser: Chen, Lingling, Zhang, Hongmei, Chang, Fan, Yu, Hua, Lu, Man, Zhao, Jing, Ma, Xiaojuan, Zhou, Hong, Zhou, Liuying, Tan, Jing, Wang, Jingyu, Luo, Dingqiang, Chen, Min, Yin, Lixue
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 175
container_issue 2
container_start_page 167
container_title American journal of infection control
container_volume 52
creator Chen, Lingling
Zhang, Hongmei
Chang, Fan
Yu, Hua
Lu, Man
Zhao, Jing
Ma, Xiaojuan
Zhou, Hong
Zhou, Liuying
Tan, Jing
Wang, Jingyu
Luo, Dingqiang
Chen, Min
Yin, Lixue
description Microbial contamination of external-use ultrasound probes is a serious and overlooked issue. We assessed the effects of different methods of disinfecting medical external-use ultrasound probes. On-site disinfection experiments were conducted at 10 hospitals; the tips and sides of external-use ultrasound probes were sampled before and after disinfection, and 3 disinfection methods were assessed (use of a new ultraviolet (UV) ultrasound probe disinfector, wiping with ordinary paper towels or cleaning with disinfectant wipes). For the new UV probe disinfector, the median microbial death rates for the tips and sides of the external-use ultrasound probe were 93.67% and 97.50%, respectively, which were higher than those for wiping with paper towels and cleaning with disinfectant wipes (12.50% and 10.00% and 20.00% and 21.42%, respectively); the rates of microorganisms exceeding the standard were 15.0% and 13.3%, respectively, which were lower than those for wiping with paper towels and cleaning with disinfectant wipes (53.3% and 60.0% and 46.7% and 38.3%, respectively). The death rates of different species of microorganisms were high, ranging from 87.5% to 100%. The new UV ultrasound probe disinfector significantly reduced the risk of potential nosocomial infections according to the low microbial death rate for conventional disinfection methods. •Microbial contamination of external-use ultrasound probes is serious and overlooked.•Wiping with paper towels and cleaning with disinfectant wipes are two commonly used methods for cleaning and disinfecting external-use ultrasound probes.•The new ultraviolet ultrasound probe disinfector uses 280-nm ultraviolet rays to destroy the DNA or RNA of pathogens on the probe and block their replication and reproduction.•The new ultraviolet ultrasound probe disinfector could significantly reduce the risk of potential nosocomial infections due to the low microbial death rate for conventional disinfection methods.
doi_str_mv 10.1016/j.ajic.2023.06.007
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2824691482</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><els_id>S0196655323004108</els_id><sourcerecordid>2824691482</sourcerecordid><originalsourceid>FETCH-LOGICAL-c307t-3ffe7d7f64c2c84f277fee8d8e42d38caf197e24140af51894e2e4d1d6d1785e3</originalsourceid><addsrcrecordid>eNp9kE1LxDAQhoMoun78AQ-So5fWfDVpwYuIXyB40XPIJhM3SzfVpBX335tlV4-eZhieeYd5EDqnpKaEyqtlbZbB1owwXhNZE6L20Iw2TFWcdXIfzQjtZCWbhh-h45yXhJCOy-YQHXHFukaqbobiDV5N_RgsxBESzuPk1niIeFwABu_BjhkPHrtQ-lQYvIJxMbjdMIe4QUJ8L3MXrOkxfJecaPpqyoBLcjJ5mKLDH2mYQz5FB970Gc529QS93d-93j5Wzy8PT7c3z5XlRI0VL9eUU14Ky2wrPFPKA7SuBcEcb63xtFPABBXE-Ia2nQAGwlEnHVVtA_wEXW5zy9nPCfKoVyFb6HsTYZiyZi0TsqOiZQVlW9SmIecEXn-ksDJprSnRG896qTee9cazJlIXz2XpYpc_zcvnfyu_YgtwvQWgfPkVIOlsA0RbLKViTLsh_Jf_A7fWkP4</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2824691482</pqid></control><display><type>article</type><title>A multicenter study on the effects of different methods of disinfecting medical external-use ultrasound probes</title><source>MEDLINE</source><source>ScienceDirect Journals (5 years ago - present)</source><creator>Chen, Lingling ; Zhang, Hongmei ; Chang, Fan ; Yu, Hua ; Lu, Man ; Zhao, Jing ; Ma, Xiaojuan ; Zhou, Hong ; Zhou, Liuying ; Tan, Jing ; Wang, Jingyu ; Luo, Dingqiang ; Chen, Min ; Yin, Lixue</creator><creatorcontrib>Chen, Lingling ; Zhang, Hongmei ; Chang, Fan ; Yu, Hua ; Lu, Man ; Zhao, Jing ; Ma, Xiaojuan ; Zhou, Hong ; Zhou, Liuying ; Tan, Jing ; Wang, Jingyu ; Luo, Dingqiang ; Chen, Min ; Yin, Lixue</creatorcontrib><description>Microbial contamination of external-use ultrasound probes is a serious and overlooked issue. We assessed the effects of different methods of disinfecting medical external-use ultrasound probes. On-site disinfection experiments were conducted at 10 hospitals; the tips and sides of external-use ultrasound probes were sampled before and after disinfection, and 3 disinfection methods were assessed (use of a new ultraviolet (UV) ultrasound probe disinfector, wiping with ordinary paper towels or cleaning with disinfectant wipes). For the new UV probe disinfector, the median microbial death rates for the tips and sides of the external-use ultrasound probe were 93.67% and 97.50%, respectively, which were higher than those for wiping with paper towels and cleaning with disinfectant wipes (12.50% and 10.00% and 20.00% and 21.42%, respectively); the rates of microorganisms exceeding the standard were 15.0% and 13.3%, respectively, which were lower than those for wiping with paper towels and cleaning with disinfectant wipes (53.3% and 60.0% and 46.7% and 38.3%, respectively). The death rates of different species of microorganisms were high, ranging from 87.5% to 100%. The new UV ultrasound probe disinfector significantly reduced the risk of potential nosocomial infections according to the low microbial death rate for conventional disinfection methods. •Microbial contamination of external-use ultrasound probes is serious and overlooked.•Wiping with paper towels and cleaning with disinfectant wipes are two commonly used methods for cleaning and disinfecting external-use ultrasound probes.•The new ultraviolet ultrasound probe disinfector uses 280-nm ultraviolet rays to destroy the DNA or RNA of pathogens on the probe and block their replication and reproduction.•The new ultraviolet ultrasound probe disinfector could significantly reduce the risk of potential nosocomial infections due to the low microbial death rate for conventional disinfection methods.</description><identifier>ISSN: 0196-6553</identifier><identifier>ISSN: 1527-3296</identifier><identifier>EISSN: 1527-3296</identifier><identifier>DOI: 10.1016/j.ajic.2023.06.007</identifier><identifier>PMID: 37295679</identifier><language>eng</language><publisher>United States: Elsevier Inc</publisher><subject>Bacterial contamination ; Cross Infection - prevention &amp; control ; Disinfectants - pharmacology ; disinfection ; Disinfection - methods ; Hospitals ; Humans ; Medical external-use ultrasound probe ; Ultrasonography - methods</subject><ispartof>American journal of infection control, 2024-02, Vol.52 (2), p.167-175</ispartof><rights>2023</rights><rights>Copyright © 2023 Association for Professionals in Infection Control and Epidemiology, Inc. Published by Elsevier Inc. All rights reserved.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c307t-3ffe7d7f64c2c84f277fee8d8e42d38caf197e24140af51894e2e4d1d6d1785e3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,27923,27924</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/37295679$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Chen, Lingling</creatorcontrib><creatorcontrib>Zhang, Hongmei</creatorcontrib><creatorcontrib>Chang, Fan</creatorcontrib><creatorcontrib>Yu, Hua</creatorcontrib><creatorcontrib>Lu, Man</creatorcontrib><creatorcontrib>Zhao, Jing</creatorcontrib><creatorcontrib>Ma, Xiaojuan</creatorcontrib><creatorcontrib>Zhou, Hong</creatorcontrib><creatorcontrib>Zhou, Liuying</creatorcontrib><creatorcontrib>Tan, Jing</creatorcontrib><creatorcontrib>Wang, Jingyu</creatorcontrib><creatorcontrib>Luo, Dingqiang</creatorcontrib><creatorcontrib>Chen, Min</creatorcontrib><creatorcontrib>Yin, Lixue</creatorcontrib><title>A multicenter study on the effects of different methods of disinfecting medical external-use ultrasound probes</title><title>American journal of infection control</title><addtitle>Am J Infect Control</addtitle><description>Microbial contamination of external-use ultrasound probes is a serious and overlooked issue. We assessed the effects of different methods of disinfecting medical external-use ultrasound probes. On-site disinfection experiments were conducted at 10 hospitals; the tips and sides of external-use ultrasound probes were sampled before and after disinfection, and 3 disinfection methods were assessed (use of a new ultraviolet (UV) ultrasound probe disinfector, wiping with ordinary paper towels or cleaning with disinfectant wipes). For the new UV probe disinfector, the median microbial death rates for the tips and sides of the external-use ultrasound probe were 93.67% and 97.50%, respectively, which were higher than those for wiping with paper towels and cleaning with disinfectant wipes (12.50% and 10.00% and 20.00% and 21.42%, respectively); the rates of microorganisms exceeding the standard were 15.0% and 13.3%, respectively, which were lower than those for wiping with paper towels and cleaning with disinfectant wipes (53.3% and 60.0% and 46.7% and 38.3%, respectively). The death rates of different species of microorganisms were high, ranging from 87.5% to 100%. The new UV ultrasound probe disinfector significantly reduced the risk of potential nosocomial infections according to the low microbial death rate for conventional disinfection methods. •Microbial contamination of external-use ultrasound probes is serious and overlooked.•Wiping with paper towels and cleaning with disinfectant wipes are two commonly used methods for cleaning and disinfecting external-use ultrasound probes.•The new ultraviolet ultrasound probe disinfector uses 280-nm ultraviolet rays to destroy the DNA or RNA of pathogens on the probe and block their replication and reproduction.•The new ultraviolet ultrasound probe disinfector could significantly reduce the risk of potential nosocomial infections due to the low microbial death rate for conventional disinfection methods.</description><subject>Bacterial contamination</subject><subject>Cross Infection - prevention &amp; control</subject><subject>Disinfectants - pharmacology</subject><subject>disinfection</subject><subject>Disinfection - methods</subject><subject>Hospitals</subject><subject>Humans</subject><subject>Medical external-use ultrasound probe</subject><subject>Ultrasonography - methods</subject><issn>0196-6553</issn><issn>1527-3296</issn><issn>1527-3296</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2024</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp9kE1LxDAQhoMoun78AQ-So5fWfDVpwYuIXyB40XPIJhM3SzfVpBX335tlV4-eZhieeYd5EDqnpKaEyqtlbZbB1owwXhNZE6L20Iw2TFWcdXIfzQjtZCWbhh-h45yXhJCOy-YQHXHFukaqbobiDV5N_RgsxBESzuPk1niIeFwABu_BjhkPHrtQ-lQYvIJxMbjdMIe4QUJ8L3MXrOkxfJecaPpqyoBLcjJ5mKLDH2mYQz5FB970Gc529QS93d-93j5Wzy8PT7c3z5XlRI0VL9eUU14Ky2wrPFPKA7SuBcEcb63xtFPABBXE-Ia2nQAGwlEnHVVtA_wEXW5zy9nPCfKoVyFb6HsTYZiyZi0TsqOiZQVlW9SmIecEXn-ksDJprSnRG896qTee9cazJlIXz2XpYpc_zcvnfyu_YgtwvQWgfPkVIOlsA0RbLKViTLsh_Jf_A7fWkP4</recordid><startdate>202402</startdate><enddate>202402</enddate><creator>Chen, Lingling</creator><creator>Zhang, Hongmei</creator><creator>Chang, Fan</creator><creator>Yu, Hua</creator><creator>Lu, Man</creator><creator>Zhao, Jing</creator><creator>Ma, Xiaojuan</creator><creator>Zhou, Hong</creator><creator>Zhou, Liuying</creator><creator>Tan, Jing</creator><creator>Wang, Jingyu</creator><creator>Luo, Dingqiang</creator><creator>Chen, Min</creator><creator>Yin, Lixue</creator><general>Elsevier Inc</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>202402</creationdate><title>A multicenter study on the effects of different methods of disinfecting medical external-use ultrasound probes</title><author>Chen, Lingling ; Zhang, Hongmei ; Chang, Fan ; Yu, Hua ; Lu, Man ; Zhao, Jing ; Ma, Xiaojuan ; Zhou, Hong ; Zhou, Liuying ; Tan, Jing ; Wang, Jingyu ; Luo, Dingqiang ; Chen, Min ; Yin, Lixue</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c307t-3ffe7d7f64c2c84f277fee8d8e42d38caf197e24140af51894e2e4d1d6d1785e3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2024</creationdate><topic>Bacterial contamination</topic><topic>Cross Infection - prevention &amp; control</topic><topic>Disinfectants - pharmacology</topic><topic>disinfection</topic><topic>Disinfection - methods</topic><topic>Hospitals</topic><topic>Humans</topic><topic>Medical external-use ultrasound probe</topic><topic>Ultrasonography - methods</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Chen, Lingling</creatorcontrib><creatorcontrib>Zhang, Hongmei</creatorcontrib><creatorcontrib>Chang, Fan</creatorcontrib><creatorcontrib>Yu, Hua</creatorcontrib><creatorcontrib>Lu, Man</creatorcontrib><creatorcontrib>Zhao, Jing</creatorcontrib><creatorcontrib>Ma, Xiaojuan</creatorcontrib><creatorcontrib>Zhou, Hong</creatorcontrib><creatorcontrib>Zhou, Liuying</creatorcontrib><creatorcontrib>Tan, Jing</creatorcontrib><creatorcontrib>Wang, Jingyu</creatorcontrib><creatorcontrib>Luo, Dingqiang</creatorcontrib><creatorcontrib>Chen, Min</creatorcontrib><creatorcontrib>Yin, Lixue</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>American journal of infection control</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Chen, Lingling</au><au>Zhang, Hongmei</au><au>Chang, Fan</au><au>Yu, Hua</au><au>Lu, Man</au><au>Zhao, Jing</au><au>Ma, Xiaojuan</au><au>Zhou, Hong</au><au>Zhou, Liuying</au><au>Tan, Jing</au><au>Wang, Jingyu</au><au>Luo, Dingqiang</au><au>Chen, Min</au><au>Yin, Lixue</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>A multicenter study on the effects of different methods of disinfecting medical external-use ultrasound probes</atitle><jtitle>American journal of infection control</jtitle><addtitle>Am J Infect Control</addtitle><date>2024-02</date><risdate>2024</risdate><volume>52</volume><issue>2</issue><spage>167</spage><epage>175</epage><pages>167-175</pages><issn>0196-6553</issn><issn>1527-3296</issn><eissn>1527-3296</eissn><abstract>Microbial contamination of external-use ultrasound probes is a serious and overlooked issue. We assessed the effects of different methods of disinfecting medical external-use ultrasound probes. On-site disinfection experiments were conducted at 10 hospitals; the tips and sides of external-use ultrasound probes were sampled before and after disinfection, and 3 disinfection methods were assessed (use of a new ultraviolet (UV) ultrasound probe disinfector, wiping with ordinary paper towels or cleaning with disinfectant wipes). For the new UV probe disinfector, the median microbial death rates for the tips and sides of the external-use ultrasound probe were 93.67% and 97.50%, respectively, which were higher than those for wiping with paper towels and cleaning with disinfectant wipes (12.50% and 10.00% and 20.00% and 21.42%, respectively); the rates of microorganisms exceeding the standard were 15.0% and 13.3%, respectively, which were lower than those for wiping with paper towels and cleaning with disinfectant wipes (53.3% and 60.0% and 46.7% and 38.3%, respectively). The death rates of different species of microorganisms were high, ranging from 87.5% to 100%. The new UV ultrasound probe disinfector significantly reduced the risk of potential nosocomial infections according to the low microbial death rate for conventional disinfection methods. •Microbial contamination of external-use ultrasound probes is serious and overlooked.•Wiping with paper towels and cleaning with disinfectant wipes are two commonly used methods for cleaning and disinfecting external-use ultrasound probes.•The new ultraviolet ultrasound probe disinfector uses 280-nm ultraviolet rays to destroy the DNA or RNA of pathogens on the probe and block their replication and reproduction.•The new ultraviolet ultrasound probe disinfector could significantly reduce the risk of potential nosocomial infections due to the low microbial death rate for conventional disinfection methods.</abstract><cop>United States</cop><pub>Elsevier Inc</pub><pmid>37295679</pmid><doi>10.1016/j.ajic.2023.06.007</doi><tpages>9</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0196-6553
ispartof American journal of infection control, 2024-02, Vol.52 (2), p.167-175
issn 0196-6553
1527-3296
1527-3296
language eng
recordid cdi_proquest_miscellaneous_2824691482
source MEDLINE; ScienceDirect Journals (5 years ago - present)
subjects Bacterial contamination
Cross Infection - prevention & control
Disinfectants - pharmacology
disinfection
Disinfection - methods
Hospitals
Humans
Medical external-use ultrasound probe
Ultrasonography - methods
title A multicenter study on the effects of different methods of disinfecting medical external-use ultrasound probes
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-11T10%3A17%3A10IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=A%20multicenter%20study%20on%20the%20effects%20of%20different%20methods%20of%20disinfecting%20medical%20external-use%20ultrasound%20probes&rft.jtitle=American%20journal%20of%20infection%20control&rft.au=Chen,%20Lingling&rft.date=2024-02&rft.volume=52&rft.issue=2&rft.spage=167&rft.epage=175&rft.pages=167-175&rft.issn=0196-6553&rft.eissn=1527-3296&rft_id=info:doi/10.1016/j.ajic.2023.06.007&rft_dat=%3Cproquest_cross%3E2824691482%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2824691482&rft_id=info:pmid/37295679&rft_els_id=S0196655323004108&rfr_iscdi=true