Bioresorbable Poly (L-Lactic Acid) Flow Diverter Versus Cobalt-Chromium Flow Diverter: In Vitro and In Vivo Analysis

Permanent metallic flow diverter (FD) implantation for treatment of intracranial aneurysms requires antiplatelet therapy for an unclear duration and restricts postprocedural endovascular access. Bioresorbable FDs are being developed as a solution to these issues, but the biological reactions and phe...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Stroke (1970) 2023-06, Vol.54 (6), p.1627-1635
Hauptverfasser: Sasaki, Natsuhi, Ishii, Akira, Yagi, Shinichi, Nishi, Hidehisa, Akiyama, Ryo, Okawa, Masakazu, Abekura, Yu, Tsuji, Hirofumi, Sakurai, Shinichi, Miyamoto, Susumu
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 1635
container_issue 6
container_start_page 1627
container_title Stroke (1970)
container_volume 54
creator Sasaki, Natsuhi
Ishii, Akira
Yagi, Shinichi
Nishi, Hidehisa
Akiyama, Ryo
Okawa, Masakazu
Abekura, Yu
Tsuji, Hirofumi
Sakurai, Shinichi
Miyamoto, Susumu
description Permanent metallic flow diverter (FD) implantation for treatment of intracranial aneurysms requires antiplatelet therapy for an unclear duration and restricts postprocedural endovascular access. Bioresorbable FDs are being developed as a solution to these issues, but the biological reactions and phenomena induced by bioresorbable FDs have not been compared with those of metallic FDs. We have developed a bioresorbable poly (L-lactic acid) FD (PLLA-FD) and compared it with an FD composed of cobalt-chromium and platinum-tungsten (CoCr-FD). FD mechanical performance and in vitro degradation of the PLLA-FD were evaluated. For in vivo testing in a rabbit aneurysm model, FDs were implanted at the aneurysm site and the abdominal aorta in the PLLA-FD group (n=21) and CoCr-FD group (n=15). Aneurysm occlusion rate, branch patency, and thrombus formation within the FD were evaluated at 3, 6, and 12 months. Local inflammation and neointima structure were also evaluated. Mean strut, porosity, and pore density for the PLLA-FD were 41.7 μm, 60%, and 20 pores per mm , respectively. The proportion of aneurysms exhibiting a neck remnant or complete occlusion did not significantly differ between the groups; however, the complete occlusion rate was significantly higher in the PLLA-FD group (48% versus 13%; =0.0399). Branch occlusion and thrombus formation within the FD were not observed in either group. In the PLLA-FD group, CD68 immunoreactivity was significantly higher, but neointimal thickness decreased over time and did not significantly differ from that of the CoCr-FD at 12 months. Collagen fibers significantly predominated over elastic fibers in the neointima in the PLLA-FD group. The opposite was observed in the CoCr-FD group. The PLLA-FD was as effective as the CoCr-FD in this study and is feasible for aneurysm treatment. No morphological or pathological problems were observed with PLLA-FD over a 1-year period.
doi_str_mv 10.1161/STROKEAHA.122.042043
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2809542419</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2809542419</sourcerecordid><originalsourceid>FETCH-LOGICAL-c4642-528bbd0fce0637df891dde9a08ee3916aebab792a35d7de1144fcb3bdaa3ba213</originalsourceid><addsrcrecordid>eNpVkUtvFDEQhC0EIkvgHyDkYzh449c8zG1YEhKxUhCEXC173KM1eNaJPZPV_nscbRLEqdVSVXXrK4TeM7pkrGanP69_XH076y66JeN8SSWnUrxAC1ZxSWTN25doQalQhEuljtCbnH9TSrloq9foSDRMqJa1CzR99jFBjskaGwB_j2GPT9ZkbfrJ97jrvfuIz0Pc4S_-HtIECd9AynPGq2hNmMhqk-Lo5_F_0Sd8ucU3fkoRm607LPcRd1sT9tnnt-jVYEKGd4_zGP06P7teXZD11dfLVbcmvawlJxVvrXV06IHWonFDq5hzoAxtAYRitYHydKO4EZVrHDAm5dBbYZ0xwhrOxDE6OeTepng3Q5706HMPIZgtxDlr3lJVSS6ZKlJ5kPYp5pxg0LfJjybtNaP6gbd-5q0Lb33gXWwfHi_MdgT3bHoC_C93F0MBk_-EeQdJb6DA2-jSCG3qhhJemqF12chDSVz8Bfx6jHA</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2809542419</pqid></control><display><type>article</type><title>Bioresorbable Poly (L-Lactic Acid) Flow Diverter Versus Cobalt-Chromium Flow Diverter: In Vitro and In Vivo Analysis</title><source>MEDLINE</source><source>American Heart Association Journals</source><source>Journals@Ovid Ovid Autoload</source><source>EZB-FREE-00999 freely available EZB journals</source><source>Alma/SFX Local Collection</source><creator>Sasaki, Natsuhi ; Ishii, Akira ; Yagi, Shinichi ; Nishi, Hidehisa ; Akiyama, Ryo ; Okawa, Masakazu ; Abekura, Yu ; Tsuji, Hirofumi ; Sakurai, Shinichi ; Miyamoto, Susumu</creator><creatorcontrib>Sasaki, Natsuhi ; Ishii, Akira ; Yagi, Shinichi ; Nishi, Hidehisa ; Akiyama, Ryo ; Okawa, Masakazu ; Abekura, Yu ; Tsuji, Hirofumi ; Sakurai, Shinichi ; Miyamoto, Susumu</creatorcontrib><description>Permanent metallic flow diverter (FD) implantation for treatment of intracranial aneurysms requires antiplatelet therapy for an unclear duration and restricts postprocedural endovascular access. Bioresorbable FDs are being developed as a solution to these issues, but the biological reactions and phenomena induced by bioresorbable FDs have not been compared with those of metallic FDs. We have developed a bioresorbable poly (L-lactic acid) FD (PLLA-FD) and compared it with an FD composed of cobalt-chromium and platinum-tungsten (CoCr-FD). FD mechanical performance and in vitro degradation of the PLLA-FD were evaluated. For in vivo testing in a rabbit aneurysm model, FDs were implanted at the aneurysm site and the abdominal aorta in the PLLA-FD group (n=21) and CoCr-FD group (n=15). Aneurysm occlusion rate, branch patency, and thrombus formation within the FD were evaluated at 3, 6, and 12 months. Local inflammation and neointima structure were also evaluated. Mean strut, porosity, and pore density for the PLLA-FD were 41.7 μm, 60%, and 20 pores per mm , respectively. The proportion of aneurysms exhibiting a neck remnant or complete occlusion did not significantly differ between the groups; however, the complete occlusion rate was significantly higher in the PLLA-FD group (48% versus 13%; =0.0399). Branch occlusion and thrombus formation within the FD were not observed in either group. In the PLLA-FD group, CD68 immunoreactivity was significantly higher, but neointimal thickness decreased over time and did not significantly differ from that of the CoCr-FD at 12 months. Collagen fibers significantly predominated over elastic fibers in the neointima in the PLLA-FD group. The opposite was observed in the CoCr-FD group. The PLLA-FD was as effective as the CoCr-FD in this study and is feasible for aneurysm treatment. No morphological or pathological problems were observed with PLLA-FD over a 1-year period.</description><identifier>ISSN: 0039-2499</identifier><identifier>EISSN: 1524-4628</identifier><identifier>DOI: 10.1161/STROKEAHA.122.042043</identifier><identifier>PMID: 37139818</identifier><language>eng</language><publisher>United States: Lippincott Williams &amp; Wilkins</publisher><subject>Absorbable Implants ; Animals ; Chromium ; Cobalt ; Endovascular Procedures ; Intracranial Aneurysm - diagnostic imaging ; Intracranial Aneurysm - surgery ; Lactic Acid ; Neointima ; Rabbits ; Stents ; Treatment Outcome ; Vascular Diseases</subject><ispartof>Stroke (1970), 2023-06, Vol.54 (6), p.1627-1635</ispartof><rights>Lippincott Williams &amp; Wilkins</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c4642-528bbd0fce0637df891dde9a08ee3916aebab792a35d7de1144fcb3bdaa3ba213</citedby><cites>FETCH-LOGICAL-c4642-528bbd0fce0637df891dde9a08ee3916aebab792a35d7de1144fcb3bdaa3ba213</cites><orcidid>0000-0002-5579-179X ; 0000-0002-8700-3209 ; 0000-0002-5414-1293 ; 0000-0002-3648-3572 ; 0000-0003-1763-2517 ; 0000-0001-8947-8024 ; 0000-0001-8206-3352 ; 0000-0003-4234-2749</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,3687,27924,27925</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/37139818$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Sasaki, Natsuhi</creatorcontrib><creatorcontrib>Ishii, Akira</creatorcontrib><creatorcontrib>Yagi, Shinichi</creatorcontrib><creatorcontrib>Nishi, Hidehisa</creatorcontrib><creatorcontrib>Akiyama, Ryo</creatorcontrib><creatorcontrib>Okawa, Masakazu</creatorcontrib><creatorcontrib>Abekura, Yu</creatorcontrib><creatorcontrib>Tsuji, Hirofumi</creatorcontrib><creatorcontrib>Sakurai, Shinichi</creatorcontrib><creatorcontrib>Miyamoto, Susumu</creatorcontrib><title>Bioresorbable Poly (L-Lactic Acid) Flow Diverter Versus Cobalt-Chromium Flow Diverter: In Vitro and In Vivo Analysis</title><title>Stroke (1970)</title><addtitle>Stroke</addtitle><description>Permanent metallic flow diverter (FD) implantation for treatment of intracranial aneurysms requires antiplatelet therapy for an unclear duration and restricts postprocedural endovascular access. Bioresorbable FDs are being developed as a solution to these issues, but the biological reactions and phenomena induced by bioresorbable FDs have not been compared with those of metallic FDs. We have developed a bioresorbable poly (L-lactic acid) FD (PLLA-FD) and compared it with an FD composed of cobalt-chromium and platinum-tungsten (CoCr-FD). FD mechanical performance and in vitro degradation of the PLLA-FD were evaluated. For in vivo testing in a rabbit aneurysm model, FDs were implanted at the aneurysm site and the abdominal aorta in the PLLA-FD group (n=21) and CoCr-FD group (n=15). Aneurysm occlusion rate, branch patency, and thrombus formation within the FD were evaluated at 3, 6, and 12 months. Local inflammation and neointima structure were also evaluated. Mean strut, porosity, and pore density for the PLLA-FD were 41.7 μm, 60%, and 20 pores per mm , respectively. The proportion of aneurysms exhibiting a neck remnant or complete occlusion did not significantly differ between the groups; however, the complete occlusion rate was significantly higher in the PLLA-FD group (48% versus 13%; =0.0399). Branch occlusion and thrombus formation within the FD were not observed in either group. In the PLLA-FD group, CD68 immunoreactivity was significantly higher, but neointimal thickness decreased over time and did not significantly differ from that of the CoCr-FD at 12 months. Collagen fibers significantly predominated over elastic fibers in the neointima in the PLLA-FD group. The opposite was observed in the CoCr-FD group. The PLLA-FD was as effective as the CoCr-FD in this study and is feasible for aneurysm treatment. No morphological or pathological problems were observed with PLLA-FD over a 1-year period.</description><subject>Absorbable Implants</subject><subject>Animals</subject><subject>Chromium</subject><subject>Cobalt</subject><subject>Endovascular Procedures</subject><subject>Intracranial Aneurysm - diagnostic imaging</subject><subject>Intracranial Aneurysm - surgery</subject><subject>Lactic Acid</subject><subject>Neointima</subject><subject>Rabbits</subject><subject>Stents</subject><subject>Treatment Outcome</subject><subject>Vascular Diseases</subject><issn>0039-2499</issn><issn>1524-4628</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2023</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpVkUtvFDEQhC0EIkvgHyDkYzh449c8zG1YEhKxUhCEXC173KM1eNaJPZPV_nscbRLEqdVSVXXrK4TeM7pkrGanP69_XH076y66JeN8SSWnUrxAC1ZxSWTN25doQalQhEuljtCbnH9TSrloq9foSDRMqJa1CzR99jFBjskaGwB_j2GPT9ZkbfrJ97jrvfuIz0Pc4S_-HtIECd9AynPGq2hNmMhqk-Lo5_F_0Sd8ucU3fkoRm607LPcRd1sT9tnnt-jVYEKGd4_zGP06P7teXZD11dfLVbcmvawlJxVvrXV06IHWonFDq5hzoAxtAYRitYHydKO4EZVrHDAm5dBbYZ0xwhrOxDE6OeTepng3Q5706HMPIZgtxDlr3lJVSS6ZKlJ5kPYp5pxg0LfJjybtNaP6gbd-5q0Lb33gXWwfHi_MdgT3bHoC_C93F0MBk_-EeQdJb6DA2-jSCG3qhhJemqF12chDSVz8Bfx6jHA</recordid><startdate>20230601</startdate><enddate>20230601</enddate><creator>Sasaki, Natsuhi</creator><creator>Ishii, Akira</creator><creator>Yagi, Shinichi</creator><creator>Nishi, Hidehisa</creator><creator>Akiyama, Ryo</creator><creator>Okawa, Masakazu</creator><creator>Abekura, Yu</creator><creator>Tsuji, Hirofumi</creator><creator>Sakurai, Shinichi</creator><creator>Miyamoto, Susumu</creator><general>Lippincott Williams &amp; Wilkins</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><orcidid>https://orcid.org/0000-0002-5579-179X</orcidid><orcidid>https://orcid.org/0000-0002-8700-3209</orcidid><orcidid>https://orcid.org/0000-0002-5414-1293</orcidid><orcidid>https://orcid.org/0000-0002-3648-3572</orcidid><orcidid>https://orcid.org/0000-0003-1763-2517</orcidid><orcidid>https://orcid.org/0000-0001-8947-8024</orcidid><orcidid>https://orcid.org/0000-0001-8206-3352</orcidid><orcidid>https://orcid.org/0000-0003-4234-2749</orcidid></search><sort><creationdate>20230601</creationdate><title>Bioresorbable Poly (L-Lactic Acid) Flow Diverter Versus Cobalt-Chromium Flow Diverter: In Vitro and In Vivo Analysis</title><author>Sasaki, Natsuhi ; Ishii, Akira ; Yagi, Shinichi ; Nishi, Hidehisa ; Akiyama, Ryo ; Okawa, Masakazu ; Abekura, Yu ; Tsuji, Hirofumi ; Sakurai, Shinichi ; Miyamoto, Susumu</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c4642-528bbd0fce0637df891dde9a08ee3916aebab792a35d7de1144fcb3bdaa3ba213</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2023</creationdate><topic>Absorbable Implants</topic><topic>Animals</topic><topic>Chromium</topic><topic>Cobalt</topic><topic>Endovascular Procedures</topic><topic>Intracranial Aneurysm - diagnostic imaging</topic><topic>Intracranial Aneurysm - surgery</topic><topic>Lactic Acid</topic><topic>Neointima</topic><topic>Rabbits</topic><topic>Stents</topic><topic>Treatment Outcome</topic><topic>Vascular Diseases</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Sasaki, Natsuhi</creatorcontrib><creatorcontrib>Ishii, Akira</creatorcontrib><creatorcontrib>Yagi, Shinichi</creatorcontrib><creatorcontrib>Nishi, Hidehisa</creatorcontrib><creatorcontrib>Akiyama, Ryo</creatorcontrib><creatorcontrib>Okawa, Masakazu</creatorcontrib><creatorcontrib>Abekura, Yu</creatorcontrib><creatorcontrib>Tsuji, Hirofumi</creatorcontrib><creatorcontrib>Sakurai, Shinichi</creatorcontrib><creatorcontrib>Miyamoto, Susumu</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Stroke (1970)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Sasaki, Natsuhi</au><au>Ishii, Akira</au><au>Yagi, Shinichi</au><au>Nishi, Hidehisa</au><au>Akiyama, Ryo</au><au>Okawa, Masakazu</au><au>Abekura, Yu</au><au>Tsuji, Hirofumi</au><au>Sakurai, Shinichi</au><au>Miyamoto, Susumu</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Bioresorbable Poly (L-Lactic Acid) Flow Diverter Versus Cobalt-Chromium Flow Diverter: In Vitro and In Vivo Analysis</atitle><jtitle>Stroke (1970)</jtitle><addtitle>Stroke</addtitle><date>2023-06-01</date><risdate>2023</risdate><volume>54</volume><issue>6</issue><spage>1627</spage><epage>1635</epage><pages>1627-1635</pages><issn>0039-2499</issn><eissn>1524-4628</eissn><abstract>Permanent metallic flow diverter (FD) implantation for treatment of intracranial aneurysms requires antiplatelet therapy for an unclear duration and restricts postprocedural endovascular access. Bioresorbable FDs are being developed as a solution to these issues, but the biological reactions and phenomena induced by bioresorbable FDs have not been compared with those of metallic FDs. We have developed a bioresorbable poly (L-lactic acid) FD (PLLA-FD) and compared it with an FD composed of cobalt-chromium and platinum-tungsten (CoCr-FD). FD mechanical performance and in vitro degradation of the PLLA-FD were evaluated. For in vivo testing in a rabbit aneurysm model, FDs were implanted at the aneurysm site and the abdominal aorta in the PLLA-FD group (n=21) and CoCr-FD group (n=15). Aneurysm occlusion rate, branch patency, and thrombus formation within the FD were evaluated at 3, 6, and 12 months. Local inflammation and neointima structure were also evaluated. Mean strut, porosity, and pore density for the PLLA-FD were 41.7 μm, 60%, and 20 pores per mm , respectively. The proportion of aneurysms exhibiting a neck remnant or complete occlusion did not significantly differ between the groups; however, the complete occlusion rate was significantly higher in the PLLA-FD group (48% versus 13%; =0.0399). Branch occlusion and thrombus formation within the FD were not observed in either group. In the PLLA-FD group, CD68 immunoreactivity was significantly higher, but neointimal thickness decreased over time and did not significantly differ from that of the CoCr-FD at 12 months. Collagen fibers significantly predominated over elastic fibers in the neointima in the PLLA-FD group. The opposite was observed in the CoCr-FD group. The PLLA-FD was as effective as the CoCr-FD in this study and is feasible for aneurysm treatment. No morphological or pathological problems were observed with PLLA-FD over a 1-year period.</abstract><cop>United States</cop><pub>Lippincott Williams &amp; Wilkins</pub><pmid>37139818</pmid><doi>10.1161/STROKEAHA.122.042043</doi><tpages>9</tpages><orcidid>https://orcid.org/0000-0002-5579-179X</orcidid><orcidid>https://orcid.org/0000-0002-8700-3209</orcidid><orcidid>https://orcid.org/0000-0002-5414-1293</orcidid><orcidid>https://orcid.org/0000-0002-3648-3572</orcidid><orcidid>https://orcid.org/0000-0003-1763-2517</orcidid><orcidid>https://orcid.org/0000-0001-8947-8024</orcidid><orcidid>https://orcid.org/0000-0001-8206-3352</orcidid><orcidid>https://orcid.org/0000-0003-4234-2749</orcidid><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 0039-2499
ispartof Stroke (1970), 2023-06, Vol.54 (6), p.1627-1635
issn 0039-2499
1524-4628
language eng
recordid cdi_proquest_miscellaneous_2809542419
source MEDLINE; American Heart Association Journals; Journals@Ovid Ovid Autoload; EZB-FREE-00999 freely available EZB journals; Alma/SFX Local Collection
subjects Absorbable Implants
Animals
Chromium
Cobalt
Endovascular Procedures
Intracranial Aneurysm - diagnostic imaging
Intracranial Aneurysm - surgery
Lactic Acid
Neointima
Rabbits
Stents
Treatment Outcome
Vascular Diseases
title Bioresorbable Poly (L-Lactic Acid) Flow Diverter Versus Cobalt-Chromium Flow Diverter: In Vitro and In Vivo Analysis
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-05T19%3A42%3A41IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Bioresorbable%20Poly%20(L-Lactic%20Acid)%20Flow%20Diverter%20Versus%20Cobalt-Chromium%20Flow%20Diverter:%20In%20Vitro%20and%20In%20Vivo%20Analysis&rft.jtitle=Stroke%20(1970)&rft.au=Sasaki,%20Natsuhi&rft.date=2023-06-01&rft.volume=54&rft.issue=6&rft.spage=1627&rft.epage=1635&rft.pages=1627-1635&rft.issn=0039-2499&rft.eissn=1524-4628&rft_id=info:doi/10.1161/STROKEAHA.122.042043&rft_dat=%3Cproquest_cross%3E2809542419%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2809542419&rft_id=info:pmid/37139818&rfr_iscdi=true