Liquid-Crystal Thermography: Illumination Spectral Effects. Part 1—Experiments
Experiments have been performed to examine the spectral effects of the illumination source on the hue-temperature characteristics of thermochromic liquid crystals (TLCs) used in a liquid-crystal thermography system. Five illumination sources were compared in this study. It was found that “full spect...
Gespeichert in:
Veröffentlicht in: | Journal of heat transfer 2005-06, Vol.127 (6), p.581-587 |
---|---|
Hauptverfasser: | , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Experiments have been performed to examine the spectral effects of
the illumination source on the hue-temperature characteristics of thermochromic
liquid crystals (TLCs) used in a liquid-crystal thermography system. Five
illumination sources were compared in this study. It was found that “full
spectrum” sources, which have a relatively uniform radiant intensity across the
visible spectrum, tend to have the lowest temperature uncertainties and the
broadest useful ranges, which are desirable calibration attributes. Radiation in
the infrared, which leads to (usually undesirable) heating of a test surface,
and in the ultraviolet, which can damage TLCs, are discussed for the various
light sources. Experimental observations of the effect that UV damage has on
liquid crystal calibrations are also provided. The use of a new method called
background subtraction and the use of white balancing are investigated as
methods of improving the calibration characteristics of TLCs. The uncertainty in
temperature associated with different illumination sources and both background
subtraction and white balancing is determined and discussed. It is shown that
these methods can reduce the uncertainty in some cases. |
---|---|