Helical tomotherapy and two types of volumetric modulated arc therapy: dosimetric and clinical comparison for several cancer sites
Radiotherapy accelerators have undergone continuous technological developments. We investigated the differences between Radixact™ and VMAT treatment plans. Sixty patients were included in this study. Dosimetric comparison between the Radixact™ and VMAT plans was performed for six cancer sites: whole...
Gespeichert in:
Veröffentlicht in: | Radiological physics and technology 2023-06, Vol.16 (2), p.272-283 |
---|---|
Hauptverfasser: | , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Radiotherapy accelerators have undergone continuous technological developments. We investigated the differences between Radixact™ and VMAT treatment plans. Sixty patients were included in this study. Dosimetric comparison between the Radixact™ and VMAT plans was performed for six cancer sites: whole-brain, head and neck, lymphoma, lung, prostate, and rectum. The VMAT plans were generated with two Elekta linear accelerators (Synergy
®
and Versa HD™). The planning target volume (PTV) coverage, organs-at-risk dose constraints, and four dosimetric indexes were considered. The deliverability of the plans was assessed using quality assurance (gamma index evaluation) measurements; clinical judgment was included in the assessment. The mean AAPM TG218 (3%–2 mm, global normalization) gamma index values were 99.4%, 97.8%, and 96.6% for Radixact™, Versa HD™, and Synergy®, respectively. Radixact™ performed better than Versa HD™ in terms of dosimetric indexes, hippocampi
D
100%
, spinal cord
D
max
, rectum
V
38.4 Gy
, bladder
V
30 Gy
, and
V
40 Gy
. Versa HD™ saved more of the (lungs-PTV)
V
5 Gy
and (lungs-PTV)
D
mean
, heart
D
mean
, breasts
V
4 Gy
, and bowel V
45 Gy
. Regarding Synergy
®
, the head and neck Radixact™ plan saved more of the parotid gland, oral cavity, and supraglottic larynx. From a clinical point of view, for the head and neck, prostate, and rectal sites, the Radixact™ and Versa HD™ plans were similar; Radixact™ plans were preferable for the head and neck and rectum to Synergy
®
plans. The quality of linac plans has improved, and differences with tomotherapy have decreased. However, tomotherapy continues to be an essential add-on in multi-machine departments. |
---|---|
ISSN: | 1865-0333 1865-0341 |
DOI: | 10.1007/s12194-023-00716-3 |