Phantom-Less Nonlinear Magnetic Resonance Imaging Calibration With Multiple Input Blood Flow Model
Previous work used phantoms to calibrate the nonlinear relationship between the gadolinium contrast concentration and the intensity of the magnetic resonance imaging signal. This work proposes a new nonlinear calibration procedure without phantoms and considers the variation of contrast agent mass m...
Gespeichert in:
Veröffentlicht in: | Topics in magnetic resonance imaging 2023-02, Vol.32 (1), p.5-13 |
---|---|
Hauptverfasser: | , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
container_end_page | 13 |
---|---|
container_issue | 1 |
container_start_page | 5 |
container_title | Topics in magnetic resonance imaging |
container_volume | 32 |
creator | Saka, Tomoki Gotoh, Toshiyuki Kagei, Seiichiro Iwasawa, Tae Takimoto, Rogerio Y. Tsuzuki, Marcos S. G. |
description | Previous work used phantoms to calibrate the nonlinear relationship between the gadolinium contrast concentration and the intensity of the magnetic resonance imaging signal. This work proposes a new nonlinear calibration procedure without phantoms and considers the variation of contrast agent mass minimum combined with the multiple input blood flow system. This also proposes a new single-input method with meaningful variables that is not influenced by reperfusion or noise generated by aliasing. The reperfusion in the lung is usually neglected and is not considered by the indicator dilution method. However, in cases of lung cancer, reperfusion cannot be neglected. A new multiple input method is formulated, and the contribution of the pulmonary artery and bronchial artery to lung perfusion can be considered and evaluated separately.
The calibration procedure applies the minimum variation of contrast agent mass in 3 different regions: (1) pulmonary artery, (2) left atrium, and (3) aorta. It was compared with four dimensional computerized tomography with iodine, which has a very high proportional relationship between contrast agent concentration and signal intensity.
Nonlinear calibration was performed without phantoms, and it is in the range of phantom calibration. It successfully separated the contributions of the pulmonary and bronchial arteries. The proposed multiple input method was verified in 6 subjects with lung cancer, and perfusion from the bronchial artery, rich in oxygen, was identified as very high in the cancer region.
Nonlinear calibration of the contrast agent without phantoms is possible. Separate contributions of the pulmonary artery and aorta can be determined. |
doi_str_mv | 10.1097/RMR.0000000000000302 |
format | Article |
fullrecord | <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2773114830</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2773114830</sourcerecordid><originalsourceid>FETCH-LOGICAL-c2622-4a5953b8de493e715d8e86e4e7abaca1a11a03085900e3f9e3b6281248102b933</originalsourceid><addsrcrecordid>eNpdkMlOwzAQhi0EYn8DhHzkEvCW7QgVm9QCqkAco0kybQ2OXWxHFW9P2BFzmUX__KP5CDng7JizMj-ZTqbH7G9IJtbINk9llnDG1PqfeovshPDEGFd5WWySLZnlMs2E3Cb13QJsdF0yxhDojbNGWwRPJzC3GHVDpxicBdsgve5gru2cjsDo2kPUztJHHRd00puol2ZQ2GUf6ZlxrqUXxq3oxLVo9sjGDEzA_a-8Sx4uzu9HV8n49vJ6dDpOGpEJkShIy1TWRYuqlJjztC2wyFBhDjU0wIFzGH4s0pIxlLMSZZ2JggtVcCbqUspdcvTpu_TupccQq06HBo0Bi64PlchzybkqJBuk6lPaeBeCx1m19LoD_1pxVr3TrQa61X-6w9rh14W-7rD9WfrG-eu7ciaiD8-mX6GvFggmLj78BFciEYMXE0OXfIzkG24xg5s</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2773114830</pqid></control><display><type>article</type><title>Phantom-Less Nonlinear Magnetic Resonance Imaging Calibration With Multiple Input Blood Flow Model</title><source>MEDLINE</source><source>Journals@Ovid LWW Legacy Archive</source><source>EZB-FREE-00999 freely available EZB journals</source><creator>Saka, Tomoki ; Gotoh, Toshiyuki ; Kagei, Seiichiro ; Iwasawa, Tae ; Takimoto, Rogerio Y. ; Tsuzuki, Marcos S. G.</creator><creatorcontrib>Saka, Tomoki ; Gotoh, Toshiyuki ; Kagei, Seiichiro ; Iwasawa, Tae ; Takimoto, Rogerio Y. ; Tsuzuki, Marcos S. G.</creatorcontrib><description>Previous work used phantoms to calibrate the nonlinear relationship between the gadolinium contrast concentration and the intensity of the magnetic resonance imaging signal. This work proposes a new nonlinear calibration procedure without phantoms and considers the variation of contrast agent mass minimum combined with the multiple input blood flow system. This also proposes a new single-input method with meaningful variables that is not influenced by reperfusion or noise generated by aliasing. The reperfusion in the lung is usually neglected and is not considered by the indicator dilution method. However, in cases of lung cancer, reperfusion cannot be neglected. A new multiple input method is formulated, and the contribution of the pulmonary artery and bronchial artery to lung perfusion can be considered and evaluated separately.
The calibration procedure applies the minimum variation of contrast agent mass in 3 different regions: (1) pulmonary artery, (2) left atrium, and (3) aorta. It was compared with four dimensional computerized tomography with iodine, which has a very high proportional relationship between contrast agent concentration and signal intensity.
Nonlinear calibration was performed without phantoms, and it is in the range of phantom calibration. It successfully separated the contributions of the pulmonary and bronchial arteries. The proposed multiple input method was verified in 6 subjects with lung cancer, and perfusion from the bronchial artery, rich in oxygen, was identified as very high in the cancer region.
Nonlinear calibration of the contrast agent without phantoms is possible. Separate contributions of the pulmonary artery and aorta can be determined.</description><identifier>ISSN: 1536-1004</identifier><identifier>EISSN: 1536-1004</identifier><identifier>DOI: 10.1097/RMR.0000000000000302</identifier><identifier>PMID: 36735623</identifier><language>eng</language><publisher>United States: Wolters Kluwer</publisher><subject>Calibration ; Contrast Media ; Humans ; Lung Neoplasms - diagnostic imaging ; Magnetic Resonance Imaging - methods ; Tomography, X-Ray Computed</subject><ispartof>Topics in magnetic resonance imaging, 2023-02, Vol.32 (1), p.5-13</ispartof><rights>Wolters Kluwer</rights><rights>Copyright © 2023 The Author(s). Published by Wolters Kluwer Health, Inc.</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><cites>FETCH-LOGICAL-c2622-4a5953b8de493e715d8e86e4e7abaca1a11a03085900e3f9e3b6281248102b933</cites><orcidid>0000-0002-8495-2337</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktohtml>$$Uhttp://ovidsp.ovid.com/ovidweb.cgi?T=JS&NEWS=n&CSC=Y&PAGE=fulltext&D=ovft&AN=00002142-202302000-00002$$EHTML$$P50$$Gwolterskluwer$$H</linktohtml><link.rule.ids>314,776,780,4595,27901,27902,65206</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/36735623$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Saka, Tomoki</creatorcontrib><creatorcontrib>Gotoh, Toshiyuki</creatorcontrib><creatorcontrib>Kagei, Seiichiro</creatorcontrib><creatorcontrib>Iwasawa, Tae</creatorcontrib><creatorcontrib>Takimoto, Rogerio Y.</creatorcontrib><creatorcontrib>Tsuzuki, Marcos S. G.</creatorcontrib><title>Phantom-Less Nonlinear Magnetic Resonance Imaging Calibration With Multiple Input Blood Flow Model</title><title>Topics in magnetic resonance imaging</title><addtitle>Top Magn Reson Imaging</addtitle><description>Previous work used phantoms to calibrate the nonlinear relationship between the gadolinium contrast concentration and the intensity of the magnetic resonance imaging signal. This work proposes a new nonlinear calibration procedure without phantoms and considers the variation of contrast agent mass minimum combined with the multiple input blood flow system. This also proposes a new single-input method with meaningful variables that is not influenced by reperfusion or noise generated by aliasing. The reperfusion in the lung is usually neglected and is not considered by the indicator dilution method. However, in cases of lung cancer, reperfusion cannot be neglected. A new multiple input method is formulated, and the contribution of the pulmonary artery and bronchial artery to lung perfusion can be considered and evaluated separately.
The calibration procedure applies the minimum variation of contrast agent mass in 3 different regions: (1) pulmonary artery, (2) left atrium, and (3) aorta. It was compared with four dimensional computerized tomography with iodine, which has a very high proportional relationship between contrast agent concentration and signal intensity.
Nonlinear calibration was performed without phantoms, and it is in the range of phantom calibration. It successfully separated the contributions of the pulmonary and bronchial arteries. The proposed multiple input method was verified in 6 subjects with lung cancer, and perfusion from the bronchial artery, rich in oxygen, was identified as very high in the cancer region.
Nonlinear calibration of the contrast agent without phantoms is possible. Separate contributions of the pulmonary artery and aorta can be determined.</description><subject>Calibration</subject><subject>Contrast Media</subject><subject>Humans</subject><subject>Lung Neoplasms - diagnostic imaging</subject><subject>Magnetic Resonance Imaging - methods</subject><subject>Tomography, X-Ray Computed</subject><issn>1536-1004</issn><issn>1536-1004</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2023</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNpdkMlOwzAQhi0EYn8DhHzkEvCW7QgVm9QCqkAco0kybQ2OXWxHFW9P2BFzmUX__KP5CDng7JizMj-ZTqbH7G9IJtbINk9llnDG1PqfeovshPDEGFd5WWySLZnlMs2E3Cb13QJsdF0yxhDojbNGWwRPJzC3GHVDpxicBdsgve5gru2cjsDo2kPUztJHHRd00puol2ZQ2GUf6ZlxrqUXxq3oxLVo9sjGDEzA_a-8Sx4uzu9HV8n49vJ6dDpOGpEJkShIy1TWRYuqlJjztC2wyFBhDjU0wIFzGH4s0pIxlLMSZZ2JggtVcCbqUspdcvTpu_TupccQq06HBo0Bi64PlchzybkqJBuk6lPaeBeCx1m19LoD_1pxVr3TrQa61X-6w9rh14W-7rD9WfrG-eu7ciaiD8-mX6GvFggmLj78BFciEYMXE0OXfIzkG24xg5s</recordid><startdate>20230201</startdate><enddate>20230201</enddate><creator>Saka, Tomoki</creator><creator>Gotoh, Toshiyuki</creator><creator>Kagei, Seiichiro</creator><creator>Iwasawa, Tae</creator><creator>Takimoto, Rogerio Y.</creator><creator>Tsuzuki, Marcos S. G.</creator><general>Wolters Kluwer</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><orcidid>https://orcid.org/0000-0002-8495-2337</orcidid></search><sort><creationdate>20230201</creationdate><title>Phantom-Less Nonlinear Magnetic Resonance Imaging Calibration With Multiple Input Blood Flow Model</title><author>Saka, Tomoki ; Gotoh, Toshiyuki ; Kagei, Seiichiro ; Iwasawa, Tae ; Takimoto, Rogerio Y. ; Tsuzuki, Marcos S. G.</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c2622-4a5953b8de493e715d8e86e4e7abaca1a11a03085900e3f9e3b6281248102b933</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2023</creationdate><topic>Calibration</topic><topic>Contrast Media</topic><topic>Humans</topic><topic>Lung Neoplasms - diagnostic imaging</topic><topic>Magnetic Resonance Imaging - methods</topic><topic>Tomography, X-Ray Computed</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Saka, Tomoki</creatorcontrib><creatorcontrib>Gotoh, Toshiyuki</creatorcontrib><creatorcontrib>Kagei, Seiichiro</creatorcontrib><creatorcontrib>Iwasawa, Tae</creatorcontrib><creatorcontrib>Takimoto, Rogerio Y.</creatorcontrib><creatorcontrib>Tsuzuki, Marcos S. G.</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Topics in magnetic resonance imaging</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Saka, Tomoki</au><au>Gotoh, Toshiyuki</au><au>Kagei, Seiichiro</au><au>Iwasawa, Tae</au><au>Takimoto, Rogerio Y.</au><au>Tsuzuki, Marcos S. G.</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Phantom-Less Nonlinear Magnetic Resonance Imaging Calibration With Multiple Input Blood Flow Model</atitle><jtitle>Topics in magnetic resonance imaging</jtitle><addtitle>Top Magn Reson Imaging</addtitle><date>2023-02-01</date><risdate>2023</risdate><volume>32</volume><issue>1</issue><spage>5</spage><epage>13</epage><pages>5-13</pages><issn>1536-1004</issn><eissn>1536-1004</eissn><abstract>Previous work used phantoms to calibrate the nonlinear relationship between the gadolinium contrast concentration and the intensity of the magnetic resonance imaging signal. This work proposes a new nonlinear calibration procedure without phantoms and considers the variation of contrast agent mass minimum combined with the multiple input blood flow system. This also proposes a new single-input method with meaningful variables that is not influenced by reperfusion or noise generated by aliasing. The reperfusion in the lung is usually neglected and is not considered by the indicator dilution method. However, in cases of lung cancer, reperfusion cannot be neglected. A new multiple input method is formulated, and the contribution of the pulmonary artery and bronchial artery to lung perfusion can be considered and evaluated separately.
The calibration procedure applies the minimum variation of contrast agent mass in 3 different regions: (1) pulmonary artery, (2) left atrium, and (3) aorta. It was compared with four dimensional computerized tomography with iodine, which has a very high proportional relationship between contrast agent concentration and signal intensity.
Nonlinear calibration was performed without phantoms, and it is in the range of phantom calibration. It successfully separated the contributions of the pulmonary and bronchial arteries. The proposed multiple input method was verified in 6 subjects with lung cancer, and perfusion from the bronchial artery, rich in oxygen, was identified as very high in the cancer region.
Nonlinear calibration of the contrast agent without phantoms is possible. Separate contributions of the pulmonary artery and aorta can be determined.</abstract><cop>United States</cop><pub>Wolters Kluwer</pub><pmid>36735623</pmid><doi>10.1097/RMR.0000000000000302</doi><tpages>9</tpages><orcidid>https://orcid.org/0000-0002-8495-2337</orcidid><oa>free_for_read</oa></addata></record> |
fulltext | fulltext |
identifier | ISSN: 1536-1004 |
ispartof | Topics in magnetic resonance imaging, 2023-02, Vol.32 (1), p.5-13 |
issn | 1536-1004 1536-1004 |
language | eng |
recordid | cdi_proquest_miscellaneous_2773114830 |
source | MEDLINE; Journals@Ovid LWW Legacy Archive; EZB-FREE-00999 freely available EZB journals |
subjects | Calibration Contrast Media Humans Lung Neoplasms - diagnostic imaging Magnetic Resonance Imaging - methods Tomography, X-Ray Computed |
title | Phantom-Less Nonlinear Magnetic Resonance Imaging Calibration With Multiple Input Blood Flow Model |
url | https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-01T04%3A59%3A54IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Phantom-Less%20Nonlinear%20Magnetic%20Resonance%20Imaging%20Calibration%20With%20Multiple%20Input%20Blood%20Flow%20Model&rft.jtitle=Topics%20in%20magnetic%20resonance%20imaging&rft.au=Saka,%20Tomoki&rft.date=2023-02-01&rft.volume=32&rft.issue=1&rft.spage=5&rft.epage=13&rft.pages=5-13&rft.issn=1536-1004&rft.eissn=1536-1004&rft_id=info:doi/10.1097/RMR.0000000000000302&rft_dat=%3Cproquest_cross%3E2773114830%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2773114830&rft_id=info:pmid/36735623&rfr_iscdi=true |