Procedural Performance of Ultrathin, Biodegradable Polymer-Coated Stents Versus Durable Polymer-Coated Stents Based on Intracoronary Imaging
Thinner stent struts might lead to a higher risk of recoil and subsequently a smaller minimal stent area (MSA), which is known to be the strongest predictor of stent failure. We compared procedural performance between an ultrathin-strut biodegradable-polymer sirolimus-eluting stent (BP-SES) and a du...
Gespeichert in:
Veröffentlicht in: | The Journal of invasive cardiology 2022-11, Vol.34 (11), p.E811 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Thinner stent struts might lead to a higher risk of recoil and subsequently a smaller minimal stent area (MSA), which is known to be the strongest predictor of stent failure. We compared procedural performance between an ultrathin-strut biodegradable-polymer sirolimus-eluting stent (BP-SES) and a durable-polymer zotarolimus-eluting stent (DP-ZES) using intracoronary imaging.
A consecutive cohort of patients underwent percutaneous coronary intervention (PCI) with either BP-SES or DP-ZES in a pseudorandomized fashion between July 2018 and October 2019. In the present subanalysis, we included cases in which post-PCI imaging with intravascular ultrasound (IVUS) or optical coherence tomography (OCT) was performed. The primary endpoint of the study was MSA. Secondary endpoints included percentage stent expansion and presence of residual edge disease, malapposition, tissue protrusion, submedial edge dissections, or edge hematoma.
A total of 141 treated lesions (78 BP-SES and 63 DP-ZES) in 127 patients were analyzed. Median age was 69.3 years (interquartile range [IQR], 57.3-75.6) and 74.0% of patients were male. All baseline and procedural characteristics were comparable between both groups. Median MSA was 5.80 mm² (IQR, 4.40-7.24) for BP-SES and 6.35 mm² (IQR, 4.76-8.31) for DP-ZES (P=.15). No significant differences in stent expansion, residual edge disease and presence of malapposition, tissue protrusion, submedial edge dissections, or edge hematomas were found. Stent diameter and stent length were found to be independent predictors of MSA.
No significant differences in MSA were found between lesions treated with BP-SES vs DP-ZES. BP-SES and DP-ZES were comparable in terms of procedural performance. |
---|---|
ISSN: | 1557-2501 1557-2501 |
DOI: | 10.25270/jic/22.00174 |