Impact of volute design features on hemodynamic performance and hemocompatibility of centrifugal blood pumps used in ECMO
Background The centrifugal blood pump volute has a significant impact on its hemodynamic performance hemocompatibility. Previous studies about the effect of volute design features on the performance of blood pumps are relatively few. Methods In the present study, the computational fluid dynamics (CF...
Gespeichert in:
Veröffentlicht in: | Artificial organs 2023-01, Vol.47 (1), p.88-104 |
---|---|
Hauptverfasser: | , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Background
The centrifugal blood pump volute has a significant impact on its hemodynamic performance hemocompatibility. Previous studies about the effect of volute design features on the performance of blood pumps are relatively few.
Methods
In the present study, the computational fluid dynamics (CFD) method was utilized to evaluate the impact of volute design factors, including spiral start position, volute tongue radius, inlet height, size, shape and diffuser pipe angle on the hemolysis index and thrombogenic potential of the centrifugal blood pump.
Results
Correlation analysis shows that flow losses affect the hemocompatibility of the blood pump by influencing shear stress and residence time. The closer the spiral start position of the volute, the better the hydraulic performance and hemocompatibility of the blood pump. Too large or too small volute inlet heights can worsen hydraulic performance and hemolysis, and higher volute inlet height can increase the thrombogenic potential. Small volute sizes exacerbate hemolysis and large volute sizes increase the thrombogenic risk, but volute size does not affect hydraulic performance. When the diffuser pipe is tangent to the base circle of the volute, the best hydraulic performance and hemolysis performance of the blood pump is achieved, but the thrombogenic potential is increased. The trapezoid volute has poor hydraulic performance and hemocompatibility. The round volute has the best hydraulic and hemolysis performance, but the thrombogenic potential is higher than that of the rectangle volute.
Conclusion
This study found that the hemolysis index shows a significant correlation with spiral start position, volute size, and diffuser pipe angle. Thrombogenic potential exhibits a good correlation with all the studied volute design features. The flow losses affect the hemocompatibility of the blood pump by influencing shear stress and residence time. The finding of this study can be used to guide the optimization of blood pump for improving the hemodynamic performance and hemocompatibility.
The volute of centrifugal blood pump has a significant impact on its hemodynamic performance hemocompatibility. Previous studies about the effect of volute design features on the performance of blood pumps are relatively few. In the present study, the Eulerian method was utilized to evaluate the impact of volute design factors, including spiral start position, volute tongue radius, inlet height, size, shape and diffuser pipe an |
---|---|