Development of a new automated IL-6 immunoassay
Quantitative detection of interleukin-6 (IL-6) in serum and plasma can help monitor immune responses and the development of acute inflammation to guide patient management. We developed an IL-6 immunoassay for use with the automated ARCHITECT system for detecting an increase in the inflammatory respo...
Gespeichert in:
Veröffentlicht in: | Journal of immunological methods 2022-05, Vol.504, p.113262-113262, Article 113262 |
---|---|
Hauptverfasser: | , , , , , , , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Quantitative detection of interleukin-6 (IL-6) in serum and plasma can help monitor immune responses and the development of acute inflammation to guide patient management. We developed an IL-6 immunoassay for use with the automated ARCHITECT system for detecting an increase in the inflammatory response.
Immunized mouse sera were tested and selected B-cells were harvested for fusion with myeloma cells. A panel of monoclonal antibodies were produced, from which capture and detection monoclonal antibodies for the prototype IL-6 immunoassay were selected and screened on the ARCHITECT instrument. The antibody pair that most effectively captured and detected IL-6 was selected to develop a prototype IL-6 immunoassay. Calibrator and panel preparations using an internal recombinant IL-6 standard were compared to serum panels prepared with the IL-6 International Standard 89/548. Assay specificity and spike recovery were determined, and assay sensitivity was compared with the Roche EUA Elecsys IL-6 assay run on the cobas analyzer.
Twenty-one antibodies in 441 antibody pairs were screened. The prototype IL-6 assay showed high sensitivity with an estimated limit of detection of 0.317 pg/mL and limit of quantitation of |
---|---|
ISSN: | 0022-1759 1872-7905 |