Characteristics of microSilicon diode detector for electron beam dosimetry

A microSilicon™ (PTW type 60023), a new unshielded diode detector succeeding Diode E (model 60017, PTW), was characterized for electron beam dosimetry and compared with other detectors. Electron beams generated from a TrueBeam linear accelerator were measured using the microSilicon, Diode E, and mic...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Journal of radiation research 2021-11, Vol.62 (6), p.1130
Hauptverfasser: Akino, Yuichi, Das, Indra J, Fujiwara, Masateru, Kaneko, Akari, Masutani, Takashi, Mizuno, Hirokazu, Isohashi, Fumiaki, Suzuki, Osamu, Seo, Yuji, Tamari, Keisuke, Ogawa, Kazuhiko
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page
container_issue 6
container_start_page 1130
container_title Journal of radiation research
container_volume 62
creator Akino, Yuichi
Das, Indra J
Fujiwara, Masateru
Kaneko, Akari
Masutani, Takashi
Mizuno, Hirokazu
Isohashi, Fumiaki
Suzuki, Osamu
Seo, Yuji
Tamari, Keisuke
Ogawa, Kazuhiko
description A microSilicon™ (PTW type 60023), a new unshielded diode detector succeeding Diode E (model 60017, PTW), was characterized for electron beam dosimetry and compared with other detectors. Electron beams generated from a TrueBeam linear accelerator were measured using the microSilicon, Diode E, and microDiamond synthetic single-crystal diamond detector. Positional accuracy of microSilicon was measured by data collected in air and water. The percent depth dose (PDD), off-center ratio (OCR), dose-response linearity, dose rate dependence, and cone factors were evaluated. The PDDs were compared with data measured using a PPC40 plane-parallel ionization chamber. The maximum variations of depth of 50% and 90% of the maximum dose, and practical depth among all detectors and energies were 0.9 mm. The maximum variations of the bremsstrahlung dose among all detectors and energies were within 0.3%. OCR showed good agreement within 1% for the flat and tail regions. The microSilicon detector showed a penumbra width similar to microDiamond, whereas Diode E showed the steepest penumbra shape. All detectors showed good dose-response linearity and stability against the dose rate; only Diode E demonstrated logarithmic dose rate dependency. The cone factor measured with microSilicon was within ±1% for all energies and cone sizes. We demonstrated that the characteristics of microSilicon is suitable for electron beam dosimetry. The microSilicon detector can be a good alternative for electron beam dosimetry in terms of providing an appropriate PDD curve without corrections, high spatial resolution for OCR measurements and cone factors.
doi_str_mv 10.1093/jrr/rrab085
format Article
fullrecord <record><control><sourceid>gale_proqu</sourceid><recordid>TN_cdi_proquest_miscellaneous_2576653761</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A769112259</galeid><sourcerecordid>A769112259</sourcerecordid><originalsourceid>FETCH-LOGICAL-c413t-d02c8f791f4035c15d0384933f7490c361845bdbb95de8ae70b67f8a08f421df3</originalsourceid><addsrcrecordid>eNptkU1LBDEMhosoun6cvMuAF0FW02k7nR5l8RPBg3ounTbVysxW29nD_nu77CoIEkJC8qQ0eQk5pnBBQbHLj5QuUzIdtGKLTCjjaqqokNtkArzkDBrYI_s5fwDUEgTskj3GhVCtlBPyMHs3ydgRU8hjsLmKvhqCTfE59MHGeeVCdFg5HNGOMVW-OPYlT6XXoRkqF3MYcEzLQ7LjTZ_xaBMPyOvN9cvsbvr4dHs_u3qcWk7ZOHVQ29ZLRT0HJiwVDljLFWNecgWWNbTlonNdp4TD1qCErpG-NdB6XlPn2QE5W7_7meLXAvOoh5At9r2ZY1xkXQvZNILJhhb0dI2-mR51mPs4lmVXuL6SjaK0roUq1MU_VDGHw-oG6EOp_xk4Xw-UO-Wc0OvPFAaTlpqCXmmiiyZ6o0mhTzb_XXQDul_2RwT2Db4vh3k</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2576653761</pqid></control><display><type>article</type><title>Characteristics of microSilicon diode detector for electron beam dosimetry</title><source>DOAJ Directory of Open Access Journals</source><source>Access via Oxford University Press (Open Access Collection)</source><source>EZB-FREE-00999 freely available EZB journals</source><source>PubMed Central</source><creator>Akino, Yuichi ; Das, Indra J ; Fujiwara, Masateru ; Kaneko, Akari ; Masutani, Takashi ; Mizuno, Hirokazu ; Isohashi, Fumiaki ; Suzuki, Osamu ; Seo, Yuji ; Tamari, Keisuke ; Ogawa, Kazuhiko</creator><creatorcontrib>Akino, Yuichi ; Das, Indra J ; Fujiwara, Masateru ; Kaneko, Akari ; Masutani, Takashi ; Mizuno, Hirokazu ; Isohashi, Fumiaki ; Suzuki, Osamu ; Seo, Yuji ; Tamari, Keisuke ; Ogawa, Kazuhiko</creatorcontrib><description>A microSilicon™ (PTW type 60023), a new unshielded diode detector succeeding Diode E (model 60017, PTW), was characterized for electron beam dosimetry and compared with other detectors. Electron beams generated from a TrueBeam linear accelerator were measured using the microSilicon, Diode E, and microDiamond synthetic single-crystal diamond detector. Positional accuracy of microSilicon was measured by data collected in air and water. The percent depth dose (PDD), off-center ratio (OCR), dose-response linearity, dose rate dependence, and cone factors were evaluated. The PDDs were compared with data measured using a PPC40 plane-parallel ionization chamber. The maximum variations of depth of 50% and 90% of the maximum dose, and practical depth among all detectors and energies were 0.9 mm. The maximum variations of the bremsstrahlung dose among all detectors and energies were within 0.3%. OCR showed good agreement within 1% for the flat and tail regions. The microSilicon detector showed a penumbra width similar to microDiamond, whereas Diode E showed the steepest penumbra shape. All detectors showed good dose-response linearity and stability against the dose rate; only Diode E demonstrated logarithmic dose rate dependency. The cone factor measured with microSilicon was within ±1% for all energies and cone sizes. We demonstrated that the characteristics of microSilicon is suitable for electron beam dosimetry. The microSilicon detector can be a good alternative for electron beam dosimetry in terms of providing an appropriate PDD curve without corrections, high spatial resolution for OCR measurements and cone factors.</description><identifier>ISSN: 0449-3060</identifier><identifier>EISSN: 1349-9157</identifier><identifier>DOI: 10.1093/jrr/rrab085</identifier><identifier>PMID: 34559877</identifier><language>eng</language><publisher>England: Oxford University Press</publisher><subject>Detectors ; Electrons ; Ionization ; Nuclear energy ; Radiotherapy</subject><ispartof>Journal of radiation research, 2021-11, Vol.62 (6), p.1130</ispartof><rights>The Author(s) 2021. Published by Oxford University Press on behalf of The Japanese Radiation Research Society and Japanese Society for Radiation Oncology.</rights><rights>COPYRIGHT 2021 Oxford University Press</rights><lds50>peer_reviewed</lds50><oa>free_for_read</oa><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c413t-d02c8f791f4035c15d0384933f7490c361845bdbb95de8ae70b67f8a08f421df3</citedby><cites>FETCH-LOGICAL-c413t-d02c8f791f4035c15d0384933f7490c361845bdbb95de8ae70b67f8a08f421df3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,780,784,864,27924,27925</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/34559877$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Akino, Yuichi</creatorcontrib><creatorcontrib>Das, Indra J</creatorcontrib><creatorcontrib>Fujiwara, Masateru</creatorcontrib><creatorcontrib>Kaneko, Akari</creatorcontrib><creatorcontrib>Masutani, Takashi</creatorcontrib><creatorcontrib>Mizuno, Hirokazu</creatorcontrib><creatorcontrib>Isohashi, Fumiaki</creatorcontrib><creatorcontrib>Suzuki, Osamu</creatorcontrib><creatorcontrib>Seo, Yuji</creatorcontrib><creatorcontrib>Tamari, Keisuke</creatorcontrib><creatorcontrib>Ogawa, Kazuhiko</creatorcontrib><title>Characteristics of microSilicon diode detector for electron beam dosimetry</title><title>Journal of radiation research</title><addtitle>J Radiat Res</addtitle><description>A microSilicon™ (PTW type 60023), a new unshielded diode detector succeeding Diode E (model 60017, PTW), was characterized for electron beam dosimetry and compared with other detectors. Electron beams generated from a TrueBeam linear accelerator were measured using the microSilicon, Diode E, and microDiamond synthetic single-crystal diamond detector. Positional accuracy of microSilicon was measured by data collected in air and water. The percent depth dose (PDD), off-center ratio (OCR), dose-response linearity, dose rate dependence, and cone factors were evaluated. The PDDs were compared with data measured using a PPC40 plane-parallel ionization chamber. The maximum variations of depth of 50% and 90% of the maximum dose, and practical depth among all detectors and energies were 0.9 mm. The maximum variations of the bremsstrahlung dose among all detectors and energies were within 0.3%. OCR showed good agreement within 1% for the flat and tail regions. The microSilicon detector showed a penumbra width similar to microDiamond, whereas Diode E showed the steepest penumbra shape. All detectors showed good dose-response linearity and stability against the dose rate; only Diode E demonstrated logarithmic dose rate dependency. The cone factor measured with microSilicon was within ±1% for all energies and cone sizes. We demonstrated that the characteristics of microSilicon is suitable for electron beam dosimetry. The microSilicon detector can be a good alternative for electron beam dosimetry in terms of providing an appropriate PDD curve without corrections, high spatial resolution for OCR measurements and cone factors.</description><subject>Detectors</subject><subject>Electrons</subject><subject>Ionization</subject><subject>Nuclear energy</subject><subject>Radiotherapy</subject><issn>0449-3060</issn><issn>1349-9157</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><recordid>eNptkU1LBDEMhosoun6cvMuAF0FW02k7nR5l8RPBg3ounTbVysxW29nD_nu77CoIEkJC8qQ0eQk5pnBBQbHLj5QuUzIdtGKLTCjjaqqokNtkArzkDBrYI_s5fwDUEgTskj3GhVCtlBPyMHs3ydgRU8hjsLmKvhqCTfE59MHGeeVCdFg5HNGOMVW-OPYlT6XXoRkqF3MYcEzLQ7LjTZ_xaBMPyOvN9cvsbvr4dHs_u3qcWk7ZOHVQ29ZLRT0HJiwVDljLFWNecgWWNbTlonNdp4TD1qCErpG-NdB6XlPn2QE5W7_7meLXAvOoh5At9r2ZY1xkXQvZNILJhhb0dI2-mR51mPs4lmVXuL6SjaK0roUq1MU_VDGHw-oG6EOp_xk4Xw-UO-Wc0OvPFAaTlpqCXmmiiyZ6o0mhTzb_XXQDul_2RwT2Db4vh3k</recordid><startdate>20211101</startdate><enddate>20211101</enddate><creator>Akino, Yuichi</creator><creator>Das, Indra J</creator><creator>Fujiwara, Masateru</creator><creator>Kaneko, Akari</creator><creator>Masutani, Takashi</creator><creator>Mizuno, Hirokazu</creator><creator>Isohashi, Fumiaki</creator><creator>Suzuki, Osamu</creator><creator>Seo, Yuji</creator><creator>Tamari, Keisuke</creator><creator>Ogawa, Kazuhiko</creator><general>Oxford University Press</general><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>20211101</creationdate><title>Characteristics of microSilicon diode detector for electron beam dosimetry</title><author>Akino, Yuichi ; Das, Indra J ; Fujiwara, Masateru ; Kaneko, Akari ; Masutani, Takashi ; Mizuno, Hirokazu ; Isohashi, Fumiaki ; Suzuki, Osamu ; Seo, Yuji ; Tamari, Keisuke ; Ogawa, Kazuhiko</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c413t-d02c8f791f4035c15d0384933f7490c361845bdbb95de8ae70b67f8a08f421df3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><topic>Detectors</topic><topic>Electrons</topic><topic>Ionization</topic><topic>Nuclear energy</topic><topic>Radiotherapy</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Akino, Yuichi</creatorcontrib><creatorcontrib>Das, Indra J</creatorcontrib><creatorcontrib>Fujiwara, Masateru</creatorcontrib><creatorcontrib>Kaneko, Akari</creatorcontrib><creatorcontrib>Masutani, Takashi</creatorcontrib><creatorcontrib>Mizuno, Hirokazu</creatorcontrib><creatorcontrib>Isohashi, Fumiaki</creatorcontrib><creatorcontrib>Suzuki, Osamu</creatorcontrib><creatorcontrib>Seo, Yuji</creatorcontrib><creatorcontrib>Tamari, Keisuke</creatorcontrib><creatorcontrib>Ogawa, Kazuhiko</creatorcontrib><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Journal of radiation research</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Akino, Yuichi</au><au>Das, Indra J</au><au>Fujiwara, Masateru</au><au>Kaneko, Akari</au><au>Masutani, Takashi</au><au>Mizuno, Hirokazu</au><au>Isohashi, Fumiaki</au><au>Suzuki, Osamu</au><au>Seo, Yuji</au><au>Tamari, Keisuke</au><au>Ogawa, Kazuhiko</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Characteristics of microSilicon diode detector for electron beam dosimetry</atitle><jtitle>Journal of radiation research</jtitle><addtitle>J Radiat Res</addtitle><date>2021-11-01</date><risdate>2021</risdate><volume>62</volume><issue>6</issue><spage>1130</spage><pages>1130-</pages><issn>0449-3060</issn><eissn>1349-9157</eissn><abstract>A microSilicon™ (PTW type 60023), a new unshielded diode detector succeeding Diode E (model 60017, PTW), was characterized for electron beam dosimetry and compared with other detectors. Electron beams generated from a TrueBeam linear accelerator were measured using the microSilicon, Diode E, and microDiamond synthetic single-crystal diamond detector. Positional accuracy of microSilicon was measured by data collected in air and water. The percent depth dose (PDD), off-center ratio (OCR), dose-response linearity, dose rate dependence, and cone factors were evaluated. The PDDs were compared with data measured using a PPC40 plane-parallel ionization chamber. The maximum variations of depth of 50% and 90% of the maximum dose, and practical depth among all detectors and energies were 0.9 mm. The maximum variations of the bremsstrahlung dose among all detectors and energies were within 0.3%. OCR showed good agreement within 1% for the flat and tail regions. The microSilicon detector showed a penumbra width similar to microDiamond, whereas Diode E showed the steepest penumbra shape. All detectors showed good dose-response linearity and stability against the dose rate; only Diode E demonstrated logarithmic dose rate dependency. The cone factor measured with microSilicon was within ±1% for all energies and cone sizes. We demonstrated that the characteristics of microSilicon is suitable for electron beam dosimetry. The microSilicon detector can be a good alternative for electron beam dosimetry in terms of providing an appropriate PDD curve without corrections, high spatial resolution for OCR measurements and cone factors.</abstract><cop>England</cop><pub>Oxford University Press</pub><pmid>34559877</pmid><doi>10.1093/jrr/rrab085</doi><oa>free_for_read</oa></addata></record>
fulltext fulltext
identifier ISSN: 0449-3060
ispartof Journal of radiation research, 2021-11, Vol.62 (6), p.1130
issn 0449-3060
1349-9157
language eng
recordid cdi_proquest_miscellaneous_2576653761
source DOAJ Directory of Open Access Journals; Access via Oxford University Press (Open Access Collection); EZB-FREE-00999 freely available EZB journals; PubMed Central
subjects Detectors
Electrons
Ionization
Nuclear energy
Radiotherapy
title Characteristics of microSilicon diode detector for electron beam dosimetry
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2024-12-22T22%3A34%3A16IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_proqu&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Characteristics%20of%20microSilicon%20diode%20detector%20for%20electron%20beam%20dosimetry&rft.jtitle=Journal%20of%20radiation%20research&rft.au=Akino,%20Yuichi&rft.date=2021-11-01&rft.volume=62&rft.issue=6&rft.spage=1130&rft.pages=1130-&rft.issn=0449-3060&rft.eissn=1349-9157&rft_id=info:doi/10.1093/jrr/rrab085&rft_dat=%3Cgale_proqu%3EA769112259%3C/gale_proqu%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2576653761&rft_id=info:pmid/34559877&rft_galeid=A769112259&rfr_iscdi=true