An image-guided hybrid robot system for dental implant surgery
Purpose Dental implant surgery is an effective method for remediating the loss of teeth. Robot is expected to increase the accuracy of dental implant surgery. However, most of them are industrial serial robot, with low stiffness and non-unique inverse kinematic solution, which may reduce the success...
Gespeichert in:
Veröffentlicht in: | International journal for computer assisted radiology and surgery 2022, Vol.17 (1), p.15-26 |
---|---|
Hauptverfasser: | , , , , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|
Zusammenfassung: | Purpose
Dental implant surgery is an effective method for remediating the loss of teeth. Robot is expected to increase the accuracy of dental implant surgery. However, most of them are industrial serial robot, with low stiffness and non-unique inverse kinematic solution, which may reduce the success rate and safety of robotic surgery. Compared to serial robot, parallel robot is more stiffness and has unique inverse kinematic. However, its workspace is small, which may not meet surgical requirements. Therefore, a novel hybrid robot dedicated to dental implant is proposed.
Methods
The hybrid robot is composed of three translation joints, two revolute joints, and Stewart parallel manipulator. Stewart is used for performing surgical operation, while the joints are used for enlarging the workspace of Stewart. In order to ensure the safety of robot motion, physical human–robot interaction based on a variable admittance controller is applied in the robotic system. In addition, considering the small workspace of Stewart, an optimal model is proposed to minimize the joint movement of Stewart in adjusting the orientation of drill bit.
Results
Phantom experiments were carried out based on the prototype robot. In the experiments, the optimal model could be solved after 20 iterations, finding an ideal joint configuration. The proposed variable admittance controller could enhance comfort level effectively. The accuracy of robot is evaluated by angle, entry and exit deviation, which are 0.74 ± 0.25°, 0.93 ± 0.28 mm, and 0.96 ± 0.23 mm, respectively.
Conclusion
The phantom experiments validate the functionality of the proposed hybrid robot. The satisfactory performance makes it more widely used in the practical dental implant surgery in the future. |
---|---|
ISSN: | 1861-6410 1861-6429 |
DOI: | 10.1007/s11548-021-02484-0 |