Acoustic Properties of Axial and Centrifugal Flow Left Ventricular Assist Devices and Prediction of Pump Thrombosis

To characterize the properties of the audible tones produced by current left ventricular assist device (LVAD) pumps approved for use, and to ascertain if changes in those may be present in the setting of pump thrombosis. From August 31, 2016, to January 16, 2020, LVAD recipients consented to have su...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Mayo Clinic proceedings 2021-04, Vol.96 (4), p.887-900
Hauptverfasser: Boilson, Barry A, Bechtum, Elizabeth L, Behnken, Amy L, Loga, Laura A, Luckhardt, Angela J, Schettle, Sarah D, Clavell, Alfredo L, Dragomir-Daescu, Dan, Stulak, John M
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 900
container_issue 4
container_start_page 887
container_title Mayo Clinic proceedings
container_volume 96
creator Boilson, Barry A
Bechtum, Elizabeth L
Behnken, Amy L
Loga, Laura A
Luckhardt, Angela J
Schettle, Sarah D
Clavell, Alfredo L
Dragomir-Daescu, Dan
Stulak, John M
description To characterize the properties of the audible tones produced by current left ventricular assist device (LVAD) pumps approved for use, and to ascertain if changes in those may be present in the setting of pump thrombosis. From August 31, 2016, to January 16, 2020, LVAD recipients consented to have surface recordings obtained using a high-fidelity digital stethoscope. Audio data were analyzed using digital recording and editing software to produce an acoustic spectrogram by Fast Fourier transformation. Recordings were obtained in 53 patient encounters (27 HeartMate II, 19 HeartWare and 7 HeartMate 3). In 12 patients (9 HeartMate II, 3 HeartWare) there was a clinical concern for pump thrombosis. In all patients and pump models, a fundamental frequency was noted, and the second and third harmonics were also clearly detectable. Where thrombosis occurred in the HeartMate II pump, the absolute (normal -46.9 [-57.5,-42.9] dB vs thrombosis -41.4 [-49.8,-26.8] dB; P=.08) and relative (normal 0.72 [0.62, 0.92] vs thrombosis 0.95 [0.86, 1.24]; P=.01) third harmonic frequencies were increased in amplitude. Where paired data were available, an increase in the absolute and relative third harmonic frequencies was observed in all patients. In the case of the HeartWare device, a consistent difference in harmonic amplitudes in the setting of thrombosis could not be identified. A consistent pattern of fundamental and harmonic frequencies is common to all LVADs currently approved for use. Alterations in the amplitude of higher order harmonics may signal the onset of pump thrombosis in axial flow LVADs.
doi_str_mv 10.1016/j.mayocp.2020.10.043
format Article
fullrecord <record><control><sourceid>gale_proqu</sourceid><recordid>TN_cdi_proquest_miscellaneous_2508888456</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><galeid>A659290021</galeid><sourcerecordid>A659290021</sourcerecordid><originalsourceid>FETCH-LOGICAL-c408t-806e4980c0415a72a6b4967058775d4782537f0d1bd24f993c3024c04ff1caf63</originalsourceid><addsrcrecordid>eNp1kcGP1CAYxYnRuOPqf2AMJ-Ol9YNCW47N6OomkziH1SthKOwwaUsFqrv_vdRZN9mDcCC8_N6XvO8h9JZASYDUH0_lqO69nksKdJVKYNUztCGC0YJzVj9HGwDKi5qI-gK9ivEEAI0Q7CW6qKqWMBBkg2Kn_RKT03gf_GxCciZib3F359SA1dTjrZlScHa5zf-rwf_GO2MT_vFX1cugAu5idDHhT-aX09m9mvbB9E4n56d12H4ZZ3xzDH48-Iy-Ri-sGqJ58_Beou9Xn2-2X4vdty_X225XaAZtKlqoDRMtaGCEq4aq-sBE3QBvm4b3rGkprxoLPTn0lFkhKl0BZZm2lmhl6-oSfTjPnYP_uZiY5OiiNsOgJpNDS8qhzYfxFS3PaE5ppJusT0HpfHszOu0nY13Wu5oLKvJWSTa8f2LITDJ36VYtMcr_g0ejhnSMfljW3cSnIDuDOvgYg7FyDm5U4V4SkGvj8iTPjcu18VXNjWfbu4eQy2E0_aPpX8XVH-K8pwQ</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2508888456</pqid></control><display><type>article</type><title>Acoustic Properties of Axial and Centrifugal Flow Left Ventricular Assist Devices and Prediction of Pump Thrombosis</title><source>MEDLINE</source><source>Alma/SFX Local Collection</source><creator>Boilson, Barry A ; Bechtum, Elizabeth L ; Behnken, Amy L ; Loga, Laura A ; Luckhardt, Angela J ; Schettle, Sarah D ; Clavell, Alfredo L ; Dragomir-Daescu, Dan ; Stulak, John M</creator><creatorcontrib>Boilson, Barry A ; Bechtum, Elizabeth L ; Behnken, Amy L ; Loga, Laura A ; Luckhardt, Angela J ; Schettle, Sarah D ; Clavell, Alfredo L ; Dragomir-Daescu, Dan ; Stulak, John M</creatorcontrib><description>To characterize the properties of the audible tones produced by current left ventricular assist device (LVAD) pumps approved for use, and to ascertain if changes in those may be present in the setting of pump thrombosis. From August 31, 2016, to January 16, 2020, LVAD recipients consented to have surface recordings obtained using a high-fidelity digital stethoscope. Audio data were analyzed using digital recording and editing software to produce an acoustic spectrogram by Fast Fourier transformation. Recordings were obtained in 53 patient encounters (27 HeartMate II, 19 HeartWare and 7 HeartMate 3). In 12 patients (9 HeartMate II, 3 HeartWare) there was a clinical concern for pump thrombosis. In all patients and pump models, a fundamental frequency was noted, and the second and third harmonics were also clearly detectable. Where thrombosis occurred in the HeartMate II pump, the absolute (normal -46.9 [-57.5,-42.9] dB vs thrombosis -41.4 [-49.8,-26.8] dB; P=.08) and relative (normal 0.72 [0.62, 0.92] vs thrombosis 0.95 [0.86, 1.24]; P=.01) third harmonic frequencies were increased in amplitude. Where paired data were available, an increase in the absolute and relative third harmonic frequencies was observed in all patients. In the case of the HeartWare device, a consistent difference in harmonic amplitudes in the setting of thrombosis could not be identified. A consistent pattern of fundamental and harmonic frequencies is common to all LVADs currently approved for use. Alterations in the amplitude of higher order harmonics may signal the onset of pump thrombosis in axial flow LVADs.</description><identifier>ISSN: 0025-6196</identifier><identifier>EISSN: 1942-5546</identifier><identifier>DOI: 10.1016/j.mayocp.2020.10.043</identifier><identifier>PMID: 33814091</identifier><language>eng</language><publisher>England: Elsevier, Inc</publisher><subject>Acoustics ; Aged ; Blood Circulation - physiology ; Blood clot ; Female ; Heart Failure - surgery ; Heart-Assist Devices ; Humans ; Male ; Middle Aged ; Retrospective Studies ; Risk factors ; Thrombosis ; Thrombosis - prevention &amp; control</subject><ispartof>Mayo Clinic proceedings, 2021-04, Vol.96 (4), p.887-900</ispartof><rights>Copyright © 2020 Mayo Foundation for Medical Education and Research. Published by Elsevier Inc. All rights reserved.</rights><rights>COPYRIGHT 2021 Elsevier, Inc.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c408t-806e4980c0415a72a6b4967058775d4782537f0d1bd24f993c3024c04ff1caf63</citedby><cites>FETCH-LOGICAL-c408t-806e4980c0415a72a6b4967058775d4782537f0d1bd24f993c3024c04ff1caf63</cites><orcidid>0000-0003-2043-4204</orcidid></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><link.rule.ids>314,776,780,27901,27902</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/33814091$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Boilson, Barry A</creatorcontrib><creatorcontrib>Bechtum, Elizabeth L</creatorcontrib><creatorcontrib>Behnken, Amy L</creatorcontrib><creatorcontrib>Loga, Laura A</creatorcontrib><creatorcontrib>Luckhardt, Angela J</creatorcontrib><creatorcontrib>Schettle, Sarah D</creatorcontrib><creatorcontrib>Clavell, Alfredo L</creatorcontrib><creatorcontrib>Dragomir-Daescu, Dan</creatorcontrib><creatorcontrib>Stulak, John M</creatorcontrib><title>Acoustic Properties of Axial and Centrifugal Flow Left Ventricular Assist Devices and Prediction of Pump Thrombosis</title><title>Mayo Clinic proceedings</title><addtitle>Mayo Clin Proc</addtitle><description>To characterize the properties of the audible tones produced by current left ventricular assist device (LVAD) pumps approved for use, and to ascertain if changes in those may be present in the setting of pump thrombosis. From August 31, 2016, to January 16, 2020, LVAD recipients consented to have surface recordings obtained using a high-fidelity digital stethoscope. Audio data were analyzed using digital recording and editing software to produce an acoustic spectrogram by Fast Fourier transformation. Recordings were obtained in 53 patient encounters (27 HeartMate II, 19 HeartWare and 7 HeartMate 3). In 12 patients (9 HeartMate II, 3 HeartWare) there was a clinical concern for pump thrombosis. In all patients and pump models, a fundamental frequency was noted, and the second and third harmonics were also clearly detectable. Where thrombosis occurred in the HeartMate II pump, the absolute (normal -46.9 [-57.5,-42.9] dB vs thrombosis -41.4 [-49.8,-26.8] dB; P=.08) and relative (normal 0.72 [0.62, 0.92] vs thrombosis 0.95 [0.86, 1.24]; P=.01) third harmonic frequencies were increased in amplitude. Where paired data were available, an increase in the absolute and relative third harmonic frequencies was observed in all patients. In the case of the HeartWare device, a consistent difference in harmonic amplitudes in the setting of thrombosis could not be identified. A consistent pattern of fundamental and harmonic frequencies is common to all LVADs currently approved for use. Alterations in the amplitude of higher order harmonics may signal the onset of pump thrombosis in axial flow LVADs.</description><subject>Acoustics</subject><subject>Aged</subject><subject>Blood Circulation - physiology</subject><subject>Blood clot</subject><subject>Female</subject><subject>Heart Failure - surgery</subject><subject>Heart-Assist Devices</subject><subject>Humans</subject><subject>Male</subject><subject>Middle Aged</subject><subject>Retrospective Studies</subject><subject>Risk factors</subject><subject>Thrombosis</subject><subject>Thrombosis - prevention &amp; control</subject><issn>0025-6196</issn><issn>1942-5546</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp1kcGP1CAYxYnRuOPqf2AMJ-Ol9YNCW47N6OomkziH1SthKOwwaUsFqrv_vdRZN9mDcCC8_N6XvO8h9JZASYDUH0_lqO69nksKdJVKYNUztCGC0YJzVj9HGwDKi5qI-gK9ivEEAI0Q7CW6qKqWMBBkg2Kn_RKT03gf_GxCciZib3F359SA1dTjrZlScHa5zf-rwf_GO2MT_vFX1cugAu5idDHhT-aX09m9mvbB9E4n56d12H4ZZ3xzDH48-Iy-Ri-sGqJ58_Beou9Xn2-2X4vdty_X225XaAZtKlqoDRMtaGCEq4aq-sBE3QBvm4b3rGkprxoLPTn0lFkhKl0BZZm2lmhl6-oSfTjPnYP_uZiY5OiiNsOgJpNDS8qhzYfxFS3PaE5ppJusT0HpfHszOu0nY13Wu5oLKvJWSTa8f2LITDJ36VYtMcr_g0ejhnSMfljW3cSnIDuDOvgYg7FyDm5U4V4SkGvj8iTPjcu18VXNjWfbu4eQy2E0_aPpX8XVH-K8pwQ</recordid><startdate>202104</startdate><enddate>202104</enddate><creator>Boilson, Barry A</creator><creator>Bechtum, Elizabeth L</creator><creator>Behnken, Amy L</creator><creator>Loga, Laura A</creator><creator>Luckhardt, Angela J</creator><creator>Schettle, Sarah D</creator><creator>Clavell, Alfredo L</creator><creator>Dragomir-Daescu, Dan</creator><creator>Stulak, John M</creator><general>Elsevier, Inc</general><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope><orcidid>https://orcid.org/0000-0003-2043-4204</orcidid></search><sort><creationdate>202104</creationdate><title>Acoustic Properties of Axial and Centrifugal Flow Left Ventricular Assist Devices and Prediction of Pump Thrombosis</title><author>Boilson, Barry A ; Bechtum, Elizabeth L ; Behnken, Amy L ; Loga, Laura A ; Luckhardt, Angela J ; Schettle, Sarah D ; Clavell, Alfredo L ; Dragomir-Daescu, Dan ; Stulak, John M</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c408t-806e4980c0415a72a6b4967058775d4782537f0d1bd24f993c3024c04ff1caf63</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><topic>Acoustics</topic><topic>Aged</topic><topic>Blood Circulation - physiology</topic><topic>Blood clot</topic><topic>Female</topic><topic>Heart Failure - surgery</topic><topic>Heart-Assist Devices</topic><topic>Humans</topic><topic>Male</topic><topic>Middle Aged</topic><topic>Retrospective Studies</topic><topic>Risk factors</topic><topic>Thrombosis</topic><topic>Thrombosis - prevention &amp; control</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Boilson, Barry A</creatorcontrib><creatorcontrib>Bechtum, Elizabeth L</creatorcontrib><creatorcontrib>Behnken, Amy L</creatorcontrib><creatorcontrib>Loga, Laura A</creatorcontrib><creatorcontrib>Luckhardt, Angela J</creatorcontrib><creatorcontrib>Schettle, Sarah D</creatorcontrib><creatorcontrib>Clavell, Alfredo L</creatorcontrib><creatorcontrib>Dragomir-Daescu, Dan</creatorcontrib><creatorcontrib>Stulak, John M</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Mayo Clinic proceedings</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Boilson, Barry A</au><au>Bechtum, Elizabeth L</au><au>Behnken, Amy L</au><au>Loga, Laura A</au><au>Luckhardt, Angela J</au><au>Schettle, Sarah D</au><au>Clavell, Alfredo L</au><au>Dragomir-Daescu, Dan</au><au>Stulak, John M</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Acoustic Properties of Axial and Centrifugal Flow Left Ventricular Assist Devices and Prediction of Pump Thrombosis</atitle><jtitle>Mayo Clinic proceedings</jtitle><addtitle>Mayo Clin Proc</addtitle><date>2021-04</date><risdate>2021</risdate><volume>96</volume><issue>4</issue><spage>887</spage><epage>900</epage><pages>887-900</pages><issn>0025-6196</issn><eissn>1942-5546</eissn><abstract>To characterize the properties of the audible tones produced by current left ventricular assist device (LVAD) pumps approved for use, and to ascertain if changes in those may be present in the setting of pump thrombosis. From August 31, 2016, to January 16, 2020, LVAD recipients consented to have surface recordings obtained using a high-fidelity digital stethoscope. Audio data were analyzed using digital recording and editing software to produce an acoustic spectrogram by Fast Fourier transformation. Recordings were obtained in 53 patient encounters (27 HeartMate II, 19 HeartWare and 7 HeartMate 3). In 12 patients (9 HeartMate II, 3 HeartWare) there was a clinical concern for pump thrombosis. In all patients and pump models, a fundamental frequency was noted, and the second and third harmonics were also clearly detectable. Where thrombosis occurred in the HeartMate II pump, the absolute (normal -46.9 [-57.5,-42.9] dB vs thrombosis -41.4 [-49.8,-26.8] dB; P=.08) and relative (normal 0.72 [0.62, 0.92] vs thrombosis 0.95 [0.86, 1.24]; P=.01) third harmonic frequencies were increased in amplitude. Where paired data were available, an increase in the absolute and relative third harmonic frequencies was observed in all patients. In the case of the HeartWare device, a consistent difference in harmonic amplitudes in the setting of thrombosis could not be identified. A consistent pattern of fundamental and harmonic frequencies is common to all LVADs currently approved for use. Alterations in the amplitude of higher order harmonics may signal the onset of pump thrombosis in axial flow LVADs.</abstract><cop>England</cop><pub>Elsevier, Inc</pub><pmid>33814091</pmid><doi>10.1016/j.mayocp.2020.10.043</doi><tpages>14</tpages><orcidid>https://orcid.org/0000-0003-2043-4204</orcidid></addata></record>
fulltext fulltext
identifier ISSN: 0025-6196
ispartof Mayo Clinic proceedings, 2021-04, Vol.96 (4), p.887-900
issn 0025-6196
1942-5546
language eng
recordid cdi_proquest_miscellaneous_2508888456
source MEDLINE; Alma/SFX Local Collection
subjects Acoustics
Aged
Blood Circulation - physiology
Blood clot
Female
Heart Failure - surgery
Heart-Assist Devices
Humans
Male
Middle Aged
Retrospective Studies
Risk factors
Thrombosis
Thrombosis - prevention & control
title Acoustic Properties of Axial and Centrifugal Flow Left Ventricular Assist Devices and Prediction of Pump Thrombosis
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-02-01T03%3A18%3A37IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-gale_proqu&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Acoustic%20Properties%20of%20Axial%20and%20Centrifugal%20Flow%20Left%20Ventricular%20Assist%20Devices%20and%20Prediction%20of%20Pump%20Thrombosis&rft.jtitle=Mayo%20Clinic%20proceedings&rft.au=Boilson,%20Barry%20A&rft.date=2021-04&rft.volume=96&rft.issue=4&rft.spage=887&rft.epage=900&rft.pages=887-900&rft.issn=0025-6196&rft.eissn=1942-5546&rft_id=info:doi/10.1016/j.mayocp.2020.10.043&rft_dat=%3Cgale_proqu%3EA659290021%3C/gale_proqu%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2508888456&rft_id=info:pmid/33814091&rft_galeid=A659290021&rfr_iscdi=true