Retrieval analysis of an explanted Mobi-C cervical disc replacement: A case study

•This is the first reported explant analysis of a Mobi-C cervical disc.•It was shown that migration of the insert occurred, leading to compression of the spinal cord, pain and eventually a revision operation.•While they may provide more normal biomechanics, non-constrained designs, such as the Mobi-...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical engineering & physics 2021-04, Vol.90, p.54-65
Hauptverfasser: Kandemir, Göksu, Pitsika, Marina, Nissen, Justin J., Bowey, Andrew, Joyce, Thomas J.