Retrieval analysis of an explanted Mobi-C cervical disc replacement: A case study
•This is the first reported explant analysis of a Mobi-C cervical disc.•It was shown that migration of the insert occurred, leading to compression of the spinal cord, pain and eventually a revision operation.•While they may provide more normal biomechanics, non-constrained designs, such as the Mobi-...
Gespeichert in:
Veröffentlicht in: | Medical engineering & physics 2021-04, Vol.90, p.54-65 |
---|---|
Hauptverfasser: | , , , , |
Format: | Artikel |
Sprache: | eng |
Schlagworte: | |
Online-Zugang: | Volltext |
Tags: |
Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
|