Detecting MLC modeling errors using radiomics‐based machine learning in patient‐specific QA with an EPID for intensity‐modulated radiation therapy

Purpose We sought to develop machine learning models to detect multileaf collimator (MLC) modeling errors with the use of radiomic features of fluence maps measured in patient‐specific quality assurance (QA) for intensity‐modulated radiation therapy (IMRT) with an electric portal imaging device (EPI...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:Medical physics (Lancaster) 2021-03, Vol.48 (3), p.991-1002
Hauptverfasser: Sakai, Madoka, Nakano, Hisashi, Kawahara, Daisuke, Tanabe, Satoshi, Takizawa, Takeshi, Narita, Akihiro, Yamada, Takumi, Sakai, Hironori, Ueda, Masataka, Sasamoto, Ryuta, Kaidu, Motoki, Aoyama, Hidefumi, Ishikawa, Hiroyuki, Utsunomiya, Satoru
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
container_end_page 1002
container_issue 3
container_start_page 991
container_title Medical physics (Lancaster)
container_volume 48
creator Sakai, Madoka
Nakano, Hisashi
Kawahara, Daisuke
Tanabe, Satoshi
Takizawa, Takeshi
Narita, Akihiro
Yamada, Takumi
Sakai, Hironori
Ueda, Masataka
Sasamoto, Ryuta
Kaidu, Motoki
Aoyama, Hidefumi
Ishikawa, Hiroyuki
Utsunomiya, Satoru
description Purpose We sought to develop machine learning models to detect multileaf collimator (MLC) modeling errors with the use of radiomic features of fluence maps measured in patient‐specific quality assurance (QA) for intensity‐modulated radiation therapy (IMRT) with an electric portal imaging device (EPID). Methods Fluence maps measured with EPID for 38 beams from 19 clinical IMRT plans were assessed. Plans with various degrees of error in MLC modeling parameters [i.e., MLC transmission factor (TF) and dosimetric leaf gap (DLG)] and plans with an MLC positional error for comparison were created. For a total of 152 error plans for each type of error, we calculated fluence difference maps for each beam by subtracting the calculated maps from the measured maps. A total of 837 radiomic features were extracted from each fluence difference map, and we determined the number of features used for the training dataset in the machine learning models by using random forest regression. Machine learning models using the five typical algorithms [decision tree, k‐nearest neighbor (kNN), support vector machine (SVM), logistic regression, and random forest] for binary classification between the error‐free plan and the plan with the corresponding error for each type of error were developed. We used part of the total dataset to perform fourfold cross‐validation to tune the models, and we used the remaining test dataset to evaluate the performance of the developed models. A gamma analysis was also performed between the measured and calculated fluence maps with the criteria of 3%/2 and 2%/2 mm for all of the types of error. Results The radiomic features and its optimal number were similar for the models for the TF and the DLG error detection, which was different from the MLC positional error. The highest sensitivity was obtained as 0.913 for the TF error with SVM and logistic regression, 0.978 for the DLG error with kNN and SVM, and 1.000 for the MLC positional error with kNN, SVM, and random forest. The highest specificity was obtained as 1.000 for the TF error with a decision tree, SVM, and logistic regression, 1.000 for the DLG error with a decision tree, logistic regression, and random forest, and 0.909 for the MLC positional error with a decision tree and logistic regression. The gamma analysis showed the poorest performance in which sensitivities were 0.737 for the TF error and the DLG error and 0.882 for the MLC positional error for 3%/2 mm. The addition of another type of er
doi_str_mv 10.1002/mp.14699
format Article
fullrecord <record><control><sourceid>proquest_cross</sourceid><recordid>TN_cdi_proquest_miscellaneous_2474465705</recordid><sourceformat>XML</sourceformat><sourcesystem>PC</sourcesystem><sourcerecordid>2474465705</sourcerecordid><originalsourceid>FETCH-LOGICAL-c3219-e99229996fb9b63ed76d5a2969813e93a60133c5f02154c9de364a9bbb72042b3</originalsourceid><addsrcrecordid>eNp1kctKxDAUhoMoOl7AJ5As3VTTJE3NUsYrjKig65Kmp06kTWuSMszOR3Dp8_kkpo6XlavDge98P_wHof2UHKWE0OO2P0q5kHINTSjPWcIpketoQojkCeUk20Lb3j8TQgTLyCbaYoydUC7yCXo_gwA6GPuEb2ZT3HYVNOMCznXO48GPi1OV6Vqj_cfrW6k8VLhVem4s4AaUsyNiLO5VMGBDZHwP2tRG4_tTvDBhjpXF53fXZ7juXCQDWG_CMoIxbmhUiMIxIt53Foc5ONUvd9FGrRoPe99zBz1enD9Mr5LZ7eX19HSWaEZTmYCUlEopRV3KUjCoclFlikohT1IGkilBUsZ0VhOaZlzLCpjgSpZlmVPCacl20OHK27vuZQAfitZ4DU2jLHSDL2KfnIssJ9kfql3nvYO66J1plVsWKSnGPxRtX3z9IaIH39ahbKH6BX-Kj0CyAhamgeW_ouLmbiX8BHgdlYI</addsrcrecordid><sourcetype>Aggregation Database</sourcetype><iscdi>true</iscdi><recordtype>article</recordtype><pqid>2474465705</pqid></control><display><type>article</type><title>Detecting MLC modeling errors using radiomics‐based machine learning in patient‐specific QA with an EPID for intensity‐modulated radiation therapy</title><source>MEDLINE</source><source>Wiley Online Library Journals Frontfile Complete</source><source>Alma/SFX Local Collection</source><creator>Sakai, Madoka ; Nakano, Hisashi ; Kawahara, Daisuke ; Tanabe, Satoshi ; Takizawa, Takeshi ; Narita, Akihiro ; Yamada, Takumi ; Sakai, Hironori ; Ueda, Masataka ; Sasamoto, Ryuta ; Kaidu, Motoki ; Aoyama, Hidefumi ; Ishikawa, Hiroyuki ; Utsunomiya, Satoru</creator><creatorcontrib>Sakai, Madoka ; Nakano, Hisashi ; Kawahara, Daisuke ; Tanabe, Satoshi ; Takizawa, Takeshi ; Narita, Akihiro ; Yamada, Takumi ; Sakai, Hironori ; Ueda, Masataka ; Sasamoto, Ryuta ; Kaidu, Motoki ; Aoyama, Hidefumi ; Ishikawa, Hiroyuki ; Utsunomiya, Satoru</creatorcontrib><description>Purpose We sought to develop machine learning models to detect multileaf collimator (MLC) modeling errors with the use of radiomic features of fluence maps measured in patient‐specific quality assurance (QA) for intensity‐modulated radiation therapy (IMRT) with an electric portal imaging device (EPID). Methods Fluence maps measured with EPID for 38 beams from 19 clinical IMRT plans were assessed. Plans with various degrees of error in MLC modeling parameters [i.e., MLC transmission factor (TF) and dosimetric leaf gap (DLG)] and plans with an MLC positional error for comparison were created. For a total of 152 error plans for each type of error, we calculated fluence difference maps for each beam by subtracting the calculated maps from the measured maps. A total of 837 radiomic features were extracted from each fluence difference map, and we determined the number of features used for the training dataset in the machine learning models by using random forest regression. Machine learning models using the five typical algorithms [decision tree, k‐nearest neighbor (kNN), support vector machine (SVM), logistic regression, and random forest] for binary classification between the error‐free plan and the plan with the corresponding error for each type of error were developed. We used part of the total dataset to perform fourfold cross‐validation to tune the models, and we used the remaining test dataset to evaluate the performance of the developed models. A gamma analysis was also performed between the measured and calculated fluence maps with the criteria of 3%/2 and 2%/2 mm for all of the types of error. Results The radiomic features and its optimal number were similar for the models for the TF and the DLG error detection, which was different from the MLC positional error. The highest sensitivity was obtained as 0.913 for the TF error with SVM and logistic regression, 0.978 for the DLG error with kNN and SVM, and 1.000 for the MLC positional error with kNN, SVM, and random forest. The highest specificity was obtained as 1.000 for the TF error with a decision tree, SVM, and logistic regression, 1.000 for the DLG error with a decision tree, logistic regression, and random forest, and 0.909 for the MLC positional error with a decision tree and logistic regression. The gamma analysis showed the poorest performance in which sensitivities were 0.737 for the TF error and the DLG error and 0.882 for the MLC positional error for 3%/2 mm. The addition of another type of error to fluence maps significantly reduced the sensitivity for the TF and the DLG error, whereas no effect was observed for the MLC positional error detection. Conclusions Compared to the conventional gamma analysis, the radiomics‐based machine learning models showed higher sensitivity and specificity in detecting a single type of the MLC modeling error and the MLC positional error. Although the developed models need further improvement for detecting multiple types of error, radiomics‐based IMRT QA was shown to be a promising approach for detecting the MLC modeling error.</description><identifier>ISSN: 0094-2405</identifier><identifier>EISSN: 2473-4209</identifier><identifier>DOI: 10.1002/mp.14699</identifier><identifier>PMID: 33382467</identifier><language>eng</language><publisher>United States</publisher><subject>Gamma Rays ; Humans ; IMRT QA ; Machine Learning ; quality assurance ; Radiometry ; radiomics ; Radiotherapy Dosage ; Radiotherapy Planning, Computer-Assisted ; Radiotherapy, Intensity-Modulated</subject><ispartof>Medical physics (Lancaster), 2021-03, Vol.48 (3), p.991-1002</ispartof><rights>2020 American Association of Physicists in Medicine</rights><rights>2020 American Association of Physicists in Medicine.</rights><lds50>peer_reviewed</lds50><woscitedreferencessubscribed>false</woscitedreferencessubscribed><citedby>FETCH-LOGICAL-c3219-e99229996fb9b63ed76d5a2969813e93a60133c5f02154c9de364a9bbb72042b3</citedby><cites>FETCH-LOGICAL-c3219-e99229996fb9b63ed76d5a2969813e93a60133c5f02154c9de364a9bbb72042b3</cites></display><links><openurl>$$Topenurl_article</openurl><openurlfulltext>$$Topenurlfull_article</openurlfulltext><thumbnail>$$Tsyndetics_thumb_exl</thumbnail><linktopdf>$$Uhttps://onlinelibrary.wiley.com/doi/pdf/10.1002%2Fmp.14699$$EPDF$$P50$$Gwiley$$H</linktopdf><linktohtml>$$Uhttps://onlinelibrary.wiley.com/doi/full/10.1002%2Fmp.14699$$EHTML$$P50$$Gwiley$$H</linktohtml><link.rule.ids>314,776,780,1411,27903,27904,45553,45554</link.rule.ids><backlink>$$Uhttps://www.ncbi.nlm.nih.gov/pubmed/33382467$$D View this record in MEDLINE/PubMed$$Hfree_for_read</backlink></links><search><creatorcontrib>Sakai, Madoka</creatorcontrib><creatorcontrib>Nakano, Hisashi</creatorcontrib><creatorcontrib>Kawahara, Daisuke</creatorcontrib><creatorcontrib>Tanabe, Satoshi</creatorcontrib><creatorcontrib>Takizawa, Takeshi</creatorcontrib><creatorcontrib>Narita, Akihiro</creatorcontrib><creatorcontrib>Yamada, Takumi</creatorcontrib><creatorcontrib>Sakai, Hironori</creatorcontrib><creatorcontrib>Ueda, Masataka</creatorcontrib><creatorcontrib>Sasamoto, Ryuta</creatorcontrib><creatorcontrib>Kaidu, Motoki</creatorcontrib><creatorcontrib>Aoyama, Hidefumi</creatorcontrib><creatorcontrib>Ishikawa, Hiroyuki</creatorcontrib><creatorcontrib>Utsunomiya, Satoru</creatorcontrib><title>Detecting MLC modeling errors using radiomics‐based machine learning in patient‐specific QA with an EPID for intensity‐modulated radiation therapy</title><title>Medical physics (Lancaster)</title><addtitle>Med Phys</addtitle><description>Purpose We sought to develop machine learning models to detect multileaf collimator (MLC) modeling errors with the use of radiomic features of fluence maps measured in patient‐specific quality assurance (QA) for intensity‐modulated radiation therapy (IMRT) with an electric portal imaging device (EPID). Methods Fluence maps measured with EPID for 38 beams from 19 clinical IMRT plans were assessed. Plans with various degrees of error in MLC modeling parameters [i.e., MLC transmission factor (TF) and dosimetric leaf gap (DLG)] and plans with an MLC positional error for comparison were created. For a total of 152 error plans for each type of error, we calculated fluence difference maps for each beam by subtracting the calculated maps from the measured maps. A total of 837 radiomic features were extracted from each fluence difference map, and we determined the number of features used for the training dataset in the machine learning models by using random forest regression. Machine learning models using the five typical algorithms [decision tree, k‐nearest neighbor (kNN), support vector machine (SVM), logistic regression, and random forest] for binary classification between the error‐free plan and the plan with the corresponding error for each type of error were developed. We used part of the total dataset to perform fourfold cross‐validation to tune the models, and we used the remaining test dataset to evaluate the performance of the developed models. A gamma analysis was also performed between the measured and calculated fluence maps with the criteria of 3%/2 and 2%/2 mm for all of the types of error. Results The radiomic features and its optimal number were similar for the models for the TF and the DLG error detection, which was different from the MLC positional error. The highest sensitivity was obtained as 0.913 for the TF error with SVM and logistic regression, 0.978 for the DLG error with kNN and SVM, and 1.000 for the MLC positional error with kNN, SVM, and random forest. The highest specificity was obtained as 1.000 for the TF error with a decision tree, SVM, and logistic regression, 1.000 for the DLG error with a decision tree, logistic regression, and random forest, and 0.909 for the MLC positional error with a decision tree and logistic regression. The gamma analysis showed the poorest performance in which sensitivities were 0.737 for the TF error and the DLG error and 0.882 for the MLC positional error for 3%/2 mm. The addition of another type of error to fluence maps significantly reduced the sensitivity for the TF and the DLG error, whereas no effect was observed for the MLC positional error detection. Conclusions Compared to the conventional gamma analysis, the radiomics‐based machine learning models showed higher sensitivity and specificity in detecting a single type of the MLC modeling error and the MLC positional error. Although the developed models need further improvement for detecting multiple types of error, radiomics‐based IMRT QA was shown to be a promising approach for detecting the MLC modeling error.</description><subject>Gamma Rays</subject><subject>Humans</subject><subject>IMRT QA</subject><subject>Machine Learning</subject><subject>quality assurance</subject><subject>Radiometry</subject><subject>radiomics</subject><subject>Radiotherapy Dosage</subject><subject>Radiotherapy Planning, Computer-Assisted</subject><subject>Radiotherapy, Intensity-Modulated</subject><issn>0094-2405</issn><issn>2473-4209</issn><fulltext>true</fulltext><rsrctype>article</rsrctype><creationdate>2021</creationdate><recordtype>article</recordtype><sourceid>EIF</sourceid><recordid>eNp1kctKxDAUhoMoOl7AJ5As3VTTJE3NUsYrjKig65Kmp06kTWuSMszOR3Dp8_kkpo6XlavDge98P_wHof2UHKWE0OO2P0q5kHINTSjPWcIpketoQojkCeUk20Lb3j8TQgTLyCbaYoydUC7yCXo_gwA6GPuEb2ZT3HYVNOMCznXO48GPi1OV6Vqj_cfrW6k8VLhVem4s4AaUsyNiLO5VMGBDZHwP2tRG4_tTvDBhjpXF53fXZ7juXCQDWG_CMoIxbmhUiMIxIt53Foc5ONUvd9FGrRoPe99zBz1enD9Mr5LZ7eX19HSWaEZTmYCUlEopRV3KUjCoclFlikohT1IGkilBUsZ0VhOaZlzLCpjgSpZlmVPCacl20OHK27vuZQAfitZ4DU2jLHSDL2KfnIssJ9kfql3nvYO66J1plVsWKSnGPxRtX3z9IaIH39ahbKH6BX-Kj0CyAhamgeW_ouLmbiX8BHgdlYI</recordid><startdate>202103</startdate><enddate>202103</enddate><creator>Sakai, Madoka</creator><creator>Nakano, Hisashi</creator><creator>Kawahara, Daisuke</creator><creator>Tanabe, Satoshi</creator><creator>Takizawa, Takeshi</creator><creator>Narita, Akihiro</creator><creator>Yamada, Takumi</creator><creator>Sakai, Hironori</creator><creator>Ueda, Masataka</creator><creator>Sasamoto, Ryuta</creator><creator>Kaidu, Motoki</creator><creator>Aoyama, Hidefumi</creator><creator>Ishikawa, Hiroyuki</creator><creator>Utsunomiya, Satoru</creator><scope>CGR</scope><scope>CUY</scope><scope>CVF</scope><scope>ECM</scope><scope>EIF</scope><scope>NPM</scope><scope>AAYXX</scope><scope>CITATION</scope><scope>7X8</scope></search><sort><creationdate>202103</creationdate><title>Detecting MLC modeling errors using radiomics‐based machine learning in patient‐specific QA with an EPID for intensity‐modulated radiation therapy</title><author>Sakai, Madoka ; Nakano, Hisashi ; Kawahara, Daisuke ; Tanabe, Satoshi ; Takizawa, Takeshi ; Narita, Akihiro ; Yamada, Takumi ; Sakai, Hironori ; Ueda, Masataka ; Sasamoto, Ryuta ; Kaidu, Motoki ; Aoyama, Hidefumi ; Ishikawa, Hiroyuki ; Utsunomiya, Satoru</author></sort><facets><frbrtype>5</frbrtype><frbrgroupid>cdi_FETCH-LOGICAL-c3219-e99229996fb9b63ed76d5a2969813e93a60133c5f02154c9de364a9bbb72042b3</frbrgroupid><rsrctype>articles</rsrctype><prefilter>articles</prefilter><language>eng</language><creationdate>2021</creationdate><topic>Gamma Rays</topic><topic>Humans</topic><topic>IMRT QA</topic><topic>Machine Learning</topic><topic>quality assurance</topic><topic>Radiometry</topic><topic>radiomics</topic><topic>Radiotherapy Dosage</topic><topic>Radiotherapy Planning, Computer-Assisted</topic><topic>Radiotherapy, Intensity-Modulated</topic><toplevel>peer_reviewed</toplevel><toplevel>online_resources</toplevel><creatorcontrib>Sakai, Madoka</creatorcontrib><creatorcontrib>Nakano, Hisashi</creatorcontrib><creatorcontrib>Kawahara, Daisuke</creatorcontrib><creatorcontrib>Tanabe, Satoshi</creatorcontrib><creatorcontrib>Takizawa, Takeshi</creatorcontrib><creatorcontrib>Narita, Akihiro</creatorcontrib><creatorcontrib>Yamada, Takumi</creatorcontrib><creatorcontrib>Sakai, Hironori</creatorcontrib><creatorcontrib>Ueda, Masataka</creatorcontrib><creatorcontrib>Sasamoto, Ryuta</creatorcontrib><creatorcontrib>Kaidu, Motoki</creatorcontrib><creatorcontrib>Aoyama, Hidefumi</creatorcontrib><creatorcontrib>Ishikawa, Hiroyuki</creatorcontrib><creatorcontrib>Utsunomiya, Satoru</creatorcontrib><collection>Medline</collection><collection>MEDLINE</collection><collection>MEDLINE (Ovid)</collection><collection>MEDLINE</collection><collection>MEDLINE</collection><collection>PubMed</collection><collection>CrossRef</collection><collection>MEDLINE - Academic</collection><jtitle>Medical physics (Lancaster)</jtitle></facets><delivery><delcategory>Remote Search Resource</delcategory><fulltext>fulltext</fulltext></delivery><addata><au>Sakai, Madoka</au><au>Nakano, Hisashi</au><au>Kawahara, Daisuke</au><au>Tanabe, Satoshi</au><au>Takizawa, Takeshi</au><au>Narita, Akihiro</au><au>Yamada, Takumi</au><au>Sakai, Hironori</au><au>Ueda, Masataka</au><au>Sasamoto, Ryuta</au><au>Kaidu, Motoki</au><au>Aoyama, Hidefumi</au><au>Ishikawa, Hiroyuki</au><au>Utsunomiya, Satoru</au><format>journal</format><genre>article</genre><ristype>JOUR</ristype><atitle>Detecting MLC modeling errors using radiomics‐based machine learning in patient‐specific QA with an EPID for intensity‐modulated radiation therapy</atitle><jtitle>Medical physics (Lancaster)</jtitle><addtitle>Med Phys</addtitle><date>2021-03</date><risdate>2021</risdate><volume>48</volume><issue>3</issue><spage>991</spage><epage>1002</epage><pages>991-1002</pages><issn>0094-2405</issn><eissn>2473-4209</eissn><abstract>Purpose We sought to develop machine learning models to detect multileaf collimator (MLC) modeling errors with the use of radiomic features of fluence maps measured in patient‐specific quality assurance (QA) for intensity‐modulated radiation therapy (IMRT) with an electric portal imaging device (EPID). Methods Fluence maps measured with EPID for 38 beams from 19 clinical IMRT plans were assessed. Plans with various degrees of error in MLC modeling parameters [i.e., MLC transmission factor (TF) and dosimetric leaf gap (DLG)] and plans with an MLC positional error for comparison were created. For a total of 152 error plans for each type of error, we calculated fluence difference maps for each beam by subtracting the calculated maps from the measured maps. A total of 837 radiomic features were extracted from each fluence difference map, and we determined the number of features used for the training dataset in the machine learning models by using random forest regression. Machine learning models using the five typical algorithms [decision tree, k‐nearest neighbor (kNN), support vector machine (SVM), logistic regression, and random forest] for binary classification between the error‐free plan and the plan with the corresponding error for each type of error were developed. We used part of the total dataset to perform fourfold cross‐validation to tune the models, and we used the remaining test dataset to evaluate the performance of the developed models. A gamma analysis was also performed between the measured and calculated fluence maps with the criteria of 3%/2 and 2%/2 mm for all of the types of error. Results The radiomic features and its optimal number were similar for the models for the TF and the DLG error detection, which was different from the MLC positional error. The highest sensitivity was obtained as 0.913 for the TF error with SVM and logistic regression, 0.978 for the DLG error with kNN and SVM, and 1.000 for the MLC positional error with kNN, SVM, and random forest. The highest specificity was obtained as 1.000 for the TF error with a decision tree, SVM, and logistic regression, 1.000 for the DLG error with a decision tree, logistic regression, and random forest, and 0.909 for the MLC positional error with a decision tree and logistic regression. The gamma analysis showed the poorest performance in which sensitivities were 0.737 for the TF error and the DLG error and 0.882 for the MLC positional error for 3%/2 mm. The addition of another type of error to fluence maps significantly reduced the sensitivity for the TF and the DLG error, whereas no effect was observed for the MLC positional error detection. Conclusions Compared to the conventional gamma analysis, the radiomics‐based machine learning models showed higher sensitivity and specificity in detecting a single type of the MLC modeling error and the MLC positional error. Although the developed models need further improvement for detecting multiple types of error, radiomics‐based IMRT QA was shown to be a promising approach for detecting the MLC modeling error.</abstract><cop>United States</cop><pmid>33382467</pmid><doi>10.1002/mp.14699</doi><tpages>12</tpages></addata></record>
fulltext fulltext
identifier ISSN: 0094-2405
ispartof Medical physics (Lancaster), 2021-03, Vol.48 (3), p.991-1002
issn 0094-2405
2473-4209
language eng
recordid cdi_proquest_miscellaneous_2474465705
source MEDLINE; Wiley Online Library Journals Frontfile Complete; Alma/SFX Local Collection
subjects Gamma Rays
Humans
IMRT QA
Machine Learning
quality assurance
Radiometry
radiomics
Radiotherapy Dosage
Radiotherapy Planning, Computer-Assisted
Radiotherapy, Intensity-Modulated
title Detecting MLC modeling errors using radiomics‐based machine learning in patient‐specific QA with an EPID for intensity‐modulated radiation therapy
url https://sfx.bib-bvb.de/sfx_tum?ctx_ver=Z39.88-2004&ctx_enc=info:ofi/enc:UTF-8&ctx_tim=2025-01-26T09%3A01%3A50IST&url_ver=Z39.88-2004&url_ctx_fmt=infofi/fmt:kev:mtx:ctx&rfr_id=info:sid/primo.exlibrisgroup.com:primo3-Article-proquest_cross&rft_val_fmt=info:ofi/fmt:kev:mtx:journal&rft.genre=article&rft.atitle=Detecting%20MLC%20modeling%20errors%20using%20radiomics%E2%80%90based%20machine%20learning%20in%20patient%E2%80%90specific%20QA%20with%20an%20EPID%20for%20intensity%E2%80%90modulated%20radiation%20therapy&rft.jtitle=Medical%20physics%20(Lancaster)&rft.au=Sakai,%20Madoka&rft.date=2021-03&rft.volume=48&rft.issue=3&rft.spage=991&rft.epage=1002&rft.pages=991-1002&rft.issn=0094-2405&rft.eissn=2473-4209&rft_id=info:doi/10.1002/mp.14699&rft_dat=%3Cproquest_cross%3E2474465705%3C/proquest_cross%3E%3Curl%3E%3C/url%3E&disable_directlink=true&sfx.directlink=off&sfx.report_link=0&rft_id=info:oai/&rft_pqid=2474465705&rft_id=info:pmid/33382467&rfr_iscdi=true