A new architecture for non-linear analogue-digital conversion

A modified architecture is proposed for the non-linear analogue-digital conversion of a transducer response. It is proved that by making use of a dual-slope A/D converter and by changing the clock frequency fed to the A/D converter during the deintegration period and blocking some of the pulses duri...

Ausführliche Beschreibung

Gespeichert in:
Bibliographische Detailangaben
Veröffentlicht in:International journal of electronics 1985-03, Vol.58 (3), p.497-502
1. Verfasser: SUNDARSINGH, V. P.
Format: Artikel
Sprache:eng
Schlagworte:
Online-Zugang:Volltext
Tags: Tag hinzufügen
Keine Tags, Fügen Sie den ersten Tag hinzu!
Beschreibung
Zusammenfassung:A modified architecture is proposed for the non-linear analogue-digital conversion of a transducer response. It is proved that by making use of a dual-slope A/D converter and by changing the clock frequency fed to the A/D converter during the deintegration period and blocking some of the pulses during this period, a nonlinear A/D conversion can be achieved. This method was used to linearize a Pt-PtRh thermocouple response and excellent results were obtained. Further, it is proved that the frequency change depends only on the maximum slope of the transducer response that is to be corrected.
ISSN:0020-7217
1362-3060
DOI:10.1080/00207218508939046